Intended Use

The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, illio-femoral, popliteal, and infra popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

Device Description

The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

AI/ML Overview

This document describes the AngioSculpt® PTA Scoring Balloon Catheter, a device intended for dilating lesions in various arteries and for treating obstructive lesions in arteriovenous fistulae. While it details the device's intended use, design, and comparison to predicate devices, it does not contain information about specific acceptance criteria, device performance metrics, or a study design with sample sizes, expert involvement, or adjudication methods.

This document is a 510(k) Summary, a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device. This type of submission typically focuses on demonstrating equivalence through design and functional performance similarities and verification/validation testing, rather than reporting detailed clinical study results against pre-defined acceptance criteria.

Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor details about sample sizes, ground truth establishment, or multi-reader studies.

Here's what can be extracted from the document:

  • Device Name: AngioSculpt® PTA Scoring Balloon Catheter
  • Intended Use/Indications for Use:
    • Dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, and infra popliteal arteries.
    • Treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
    • Not for use in the coronary or neuro-vasculature.
  • Regulatory Class: Class II (two)
  • Product Codes: PNO, DQY, and LIT (the first page mentions PNO, the summary mentions DQY and LIT)
  • Predicate Devices:
    • Cordis Aviator® PTA Dilatation Catheter (K013581)
    • Cordis Savvy® PTA Dilatation Catheter (K971010)
    • AngioSculpt® Scoring Balloon Catheter (K050629)
    • AngioSculpt® Scoring Balloon Catheter (K072225)
  • Device Description: A standard two-lumen catheter with a scoring balloon near the distal tip. The distal end has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force to minimize balloon slippage and assist with luminal expansion. The balloon has radiopaque markers.
  • Statement on Performance: "Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised."

Missing Information (as per your request):

  1. A table of acceptance criteria and the reported device performance: Not present in this document. The document states "adequate device performance" but provides no specific metrics or criteria.
  2. Sample sized used for the test set and the data provenance: Not present. The document mentions "design verification and validation testing" but does not detail the testing methodology, sample sizes, or data provenance.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable or not present. This type of detail is usually found in clinical study reports, which are not part of this 510(k) summary. Given the device type, "ground truth" would likely refer to physical/mechanical testing results, not expert consensus on medical images.
  4. Adjudication method for the test set: Not applicable or not present.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No, this is an angioplasty catheter, not an imaging AI device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, this is an angioplasty catheter.
  7. The type of ground truth used: For this device, ground truth would refer to engineering specifications and performance metrics (e.g., balloon burst pressure, trackability, pushability, material compatibility, dimensional accuracy). This document does not detail the specific "ground truth" used for its verification and validation.
  8. The sample size for the training set: Not applicable. This document is for a medical device (catheter), not an AI algorithm that requires a training set.
  9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a simple, clean design. The eagle is facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 31, 2016

AngioScore, Inc. c/o Ms. Karin J. Gastineau Vice President Medical, Regulatory and Quality 5055 Brandin Court Fremont, CA 94538

Re: K080151

Trade/Device Name: AngioSculpt® PTA Scoring Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: PNO Dated: March 31, 2008 Received: April 1, 2008

Dear Ms. Gastineau:

This letter corrects our substantially equivalent letter of April 25, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

Page 2 - Ms. Karin J. Gastineau

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Misti L. Malone -S

for Bram D. Zuckerman. M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: AngioSculpt® PTA Scoring Balloon Catheter

Indications for Use:

.

The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, illio-femoral, popliteal, and infra popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

------------------------------------------
(Division Sign-Off)
Division of Cardiovascular DevicesPage 1 of 1
510(k) NumberK080151
------------------------

Page 4 of 18

{3}------------------------------------------------

K080151

APR 2 5 2008

510(k) Summary for the AngioSculpt Scoring Balloon Catheter

1. Submitter's Name / Contact Person

Submitter:AngioScore, Inc.5055 Brandin CourtFremont, CA 94538
Contact Person:Karin J. GastineauVice President, Medical, Regulatory, QualityPhone: 510.933.7904Fax: 510.933.7994

Summary Preparation Date: March 31, 2008

2. General Information

Trade Name:AngioSculpt® PTA Scoring Balloon Catheter
Common / Usual Name:Angioplasty catheter
Classification Name:Catheter, angioplasty, peripheral, transluminal
Product Codes:DQY and LIT
Predicate Devices:Cordis Aviator® PTA Dilatation Catheter (K013581)
Cordis Savvy® PTA Dilatation Catheter (K971010)
AngioSculpt® Scoring Balloon Catheter (K050629)
AngioSculpt® Scoring Balloon Catheter (K072225)

3. Intended Use / Indications

The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, ilio-femoral, popliteal, and infra popliteal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

4. Device Description

The AngioSculpt catheter is a standard two-lumen catheter with a scoring balloon near the distal tip. The distal end of the catheter has a conventional nylon-blend balloon with a nitinol scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure.

{4}------------------------------------------------

5. Substantial Equivalence Comparison

The AngioSculpt catheter intended use and that of the predicate devices, the Cordis Aviator PTA Dilatation Catheter and the Cordis Savvy PTA Dilatation Catheter are substantially the same.

The indications for use of the subject device are an expanded list of indications for the 510(k) cleared AngioSculpt Scoring Balloon Catheter.

The AngioSculpt catheter shares the same fundamental design, scientific technology (operating principle), functional performance, and materials as the currently marketed AngioSculpt catheter. The subject device has similar design, operating principle, and functional performance as the Cordis Aviator and Savvy Dilitation Catheters.

Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised.

The AngioSculpt PTA Scoring Balloon Catheter is therefore, substantially equivalent to the predicate devices.

[This section intentionally left blank]

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).