K Number
K080049

Validate with FDA (Live)

Date Cleared
2008-01-23

(15 days)

Product Code
Regulation Number
892.1560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VascuView ultrasound system provides ultrasound imaging of peripheral vascular structures in order to provide for ultrasound guidance for placement of needles and catheters in these structures. The VascuView ultrasound system is not intended for fetal or ophthalmic applications.

Device Description

The VascuView ultrasound system provides ultrasound imaging of peripheral vascular structures in order to provide for ultrasound guidance for placement of needles and catheters in these structures. The VascuView system consists of the following components: Ultrasound Probe and Cable, Tablet PC, VascuView Software.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the VascuView Ultrasound System based on the provided K080049 510(k) summary:

This device is an ultrasound system and its performance evaluation often relies on meeting safety standards and demonstrating equivalent physical characteristics to predicate devices, rather than clinical performance metrics typical of AI/ML devices. Therefore, a direct mapping to typical AI acceptance criteria like sensitivity/specificity or human-AI reader performance is not explicitly present in the provided documentation, as this is an older submission for a traditional medical device.

Acceptance Criteria and Device Performance

The acceptance criteria for the VascuView Ultrasound System appear to be primarily focused on safety, electromagnetic compatibility, and demonstrating comparable or superior physical and performance characteristics to its predicate devices.

Acceptance CriteriaReported Device Performance
Safety Standards ComplianceCompliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, and IEC 60601-2-37 (including acoustic output limits).
Measurement Accuracy"slightly better measurement accuracy of ±2% plus 0.5% full scale" compared to predicate device's ±2% plus 1% full scale.
Software ValidationSoftware has been validated using a test protocol.
Field of View / Viewing Angle"full 60° viewing angle at a depth of up to 6.0cm (sufficient for peripheral vascular)" - claimed to provide a greater field of view than predicate.
Storage Capacity"virtually unlimited storage (tens of thousands of images on tablet pc hard drive and ability to connect external hard drives if additional storage is required)" - significantly greater than predicate device's 74 images.
Intended Use EquivalenceThe device's intended use for peripheral vascular structures for ultrasound guidance for needle/catheter placement is considered substantially equivalent to the predicate Interson SR 7.5 MHz USB Ultrasound Probe System (K070907) and the Sonosite iLook 25 Personal Imaging Tool (K021628), with some intended use limitations compared to predicates.

Study Information

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated as a "test set" in the context of clinical performance data for metrics like sensitivity/specificity. The measurement validation was done using "wire phantoms."
    • Data Provenance: The studies mentioned are non-clinical (laboratory testing using phantoms and compliance with international standards). No patient data is mentioned.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The ground truth for measurement accuracy was established by the known dimensions/properties of "wire phantoms." For safety standards, the ground truth is the standard's requirements.
  3. Adjudication method for the test set:

    • Not applicable. The non-clinical tests involved direct measurement against known standards or physical properties.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a traditional ultrasound imaging device, not an AI/ML device. Therefore, an MRMC study for AI assistance is not applicable and was not performed.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a traditional ultrasound imaging device. The "algorithm" here refers to the system's software functionality, which was validated, but not as a standalone AI diagnostic tool.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For measurement accuracy: Wire phantoms (known physical dimensions).
    • For safety and electromagnetic compatibility: International standards (IEC 60601 series).
    • For software validation: Test protocols designed to verify software functionality.
  7. The sample size for the training set:

    • Not applicable. This is not an AI/ML device, so there is no "training set" in the context of machine learning.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for an AI/ML model.

{0}------------------------------------------------

K080049
Pg. 10f 3

JAN 23 2008

Image /page/0/Picture/2 description: The image shows a logo for Escalon Vascular Access, a division of Escalon Medical Corp. The logo features the letters 'VA' in a stylized font, with the word 'ESCALON' above and 'VASCULAR ACCESS' below. To the right of the 'VA' is a symbol resembling radio waves. The text 'A Division of Escalon Medical Corp' is located below the logo.

2440 South 179th Street New Berlin, Wi 53146 Tel 800-433-8197 Fax 262-821-9927 www.escalonvascularaccess.com

510(k) Summary

Submitted By

Escalon Medical Corp 2440 South 179th Street New Berlin, WI 53146 800-433-8197 | 262-821-9182 Fax 262-821-9927

Contact Person

Mark Wallace Vice-President Quality

Date Prepared

December 27, 2007

Trade Name

VascuView Ultrasound System

Common Name

Ultrasonic Pulsed Echo Imaging System Diagnostic Ultrasound Transducer

Classification Name

Ultrasonic Pulsed Echo Imaging System FR Number: 892.1560 Product Code: 90-IYO

Diagnostic Ultrasound Transducer FR Number: 892-1570

Product Code: 90-ITX

Legally-Marketed Equivalent Devices

Interson SR 7.5 MHz USB Ultrasound Probe System (K070907)

Sonosite iLook 25 Personal Imaging Tool (K021628)

VascuView 510(k) Summary

{1}------------------------------------------------

K080049
PG-2of3

Description of the Device

The VascuView ultrasound system provides ultrasound imaging of peripheral vascular structures in order to provide for ultrasound guldance for placement of needles and catheters in these structures. The VascuView system consists of the following components:

Ultrasound Probe and Cable. The handheld ultrasound probe is the Interson SR 7.5 MHz USB ultrasound probe (the same as the predicate device listed previously). The probe consists of a single-element mechanical sector scanner that contains the ultrasound pulser and receiver within the hand piece. It has a push button to activate scanning and capturing of images and video clips. The probe is connected to the Included USB cable. The fundamental scan frequency (excitation frequency) is 7.5 MHz, allowing visualization of vessels at depths of 0.5 to 6.0 cm. The nose cone of the ultrasound probe is made of blocompatible material sufficient for its use (RT-18 TPX Polymethyl Pentene).

Tablet PC. A Kiosk tablet pc is provided which facilitates intuilive touch screen operation. The tablet pc is provided with a medical-grade power supply and medical-grade power cord. The pc provides a 5 vdc power supply to the ultrasound probe via the USB cable. The pc also provides for the operation of the software and storage of archived images and video clips.

VascuView Software. All features and controls of the system are provided for in the VascuView software which comes pre-loaded onto the tablet pc. The VascuView software uses an iconbased interface whereby the user can select any of the primary functional screens at any time by clicking on an icon located on the screen, for that function. The software enables ultrasound image capture and review, image controls for near, mid, and far gain, as well as image intensity and contrast, linear measurement calipers, image depth grid lines, zoom, display of needle guide lines, and printing.

Intended Use

The VascuView ultrasound system provides ultrasound imaging of peripheral vascular structures in order to provide for ultrasound guidance for placement of needles and catheters in these structures. The VascuView ultrasound system is not intended for fetal or ophthalmic applications.

The intended use of the VascuView system is slightly more limited than that of the Interson USB Ultrasound Probe System in that it's intended use does not include visualization of small organs (although that intended use was for the range of probes cleared under K070907). Likewise, the intended use of the VascuView system is also more limited than the Sonosite iLook 25, although the iLook 25 is marketed for vascular access.

Technological Characteristics Compared with Predicate Devices

The VascuView Ultrasound System utilizes the Interson model SR 7.5 MHz ultrasound probe, which was previously cleared under K070907 (formerly referred to as model number VC 7.5 MHz ultrasound probe - see Attachment 5A). The probe is indicated for use for peripheral vascular and small organ (intended for use in biopsy). The scope of the application for the VascuView has been limited to just peripheral vascular, primarily for assistance in vascular access (see Attachment 5B for transducer indication for use form). There are no differences or modifications made to the Interson probe.

The differences between the iLook 25 and the VascuView Ultrasound System include that the iLook is a dedicated device, while the VascuView system utilizes a tablet pc. This makes little difference in the actual use of the device since the VascuView is set up to launch immediately into the VascuView software. The VasculView does provide for a significantly larger screen, approximately 4 times greater in viewing area to aid in visualization. Additionally, while the il ook 25 is limited to storage capacity of 74 images, the VascuView system has virtually unlimited storage (tens of thousands of images on tablet pc hard drive and ability to connect external hard drives if additional storage is required).

{2}------------------------------------------------

K050049
pg. 3 of 3

The iLook 25 also provides a power Doppler mode that is intended indicate by different colors (red or blue) directionality of blood flow within vessels which, given the anatomical location of the vessel and orientation of the ultrasound probe, assists the user in determining whether the vessel is a vein or artery. This determination may also be made in either the iLook 25 or the VascuView system by observing whether the vessel is pulsing (indicating an artery) or not (indicating a vein), in addition to anatomical location.

The iLook 25 incorporates a linear array, while the VascuView system utilizes a single-element sector scan transducer. The linear array allows a deeper penetration by the il.ook (deeper than that typically required for peripheral vascular), although a narrower profile. The sector scan sweep of the VascuView provides for a full 60° viewing angle at a depth of up to 6.0cm (sufficient for peripheral vascular). Therefore, the VascuView provides a greater field of view which combined with the larger monitor offers an enhanced resolution of scan.

The stated measurement accuracy of the iLook 25 is ±2% plus 1% full scale, while the VascuView has a slightly better measurement accuracy of ±2% plus 0.5% full scale.

Non-Clinical Performance Data

The VascuView has been tested and assessed as compliant with the requirements of several international safety standards, including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, and IEC 60601-2-37 (including acoustic output limits), and others. Additionally, the measurement accuracy of the VascuView system was confirmed via a measurement validation test protocol using wire phantoms. The software has been also been validated using a test protocol.

Clinical Performance Data

None submitted.

Conclusions from Non-Clinical (and Clinical) Data

Based upon the results of the data as summarized above, the VascuView system has demonstrated that it is as safe, as effective, and performs as well as or better than the predicate devices. Furthermore, based on the comparison with the predicate devices as shown in the discussion above, as well as guidance provided by FDA 510(k) Substantial Equivalence Decision-Making Process Flowchart, the Escalon Vascular Access VascuView ultrasound system is deemed to be substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird with three horizontal lines extending from its head, representing the department's mission to protect and promote the health and well-being of Americans.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 23 2008

Escalon Medical Corp. c/o Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, N.W. BUFFALO MN 55313

Re: K080049

Trade/Device Name: VU62000 VascuView Ultrasound System Regulation Number: 21 CFR §892.1560 Regulatory Class: II Procode: IYO Dated: January 7, 2008 Received: January 8, 2008

Dear Mr. Job:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the Interson SR 7.5 MHz Mechanical Sector Probe intended for use with the VU62000 VascuView Ultrasound System, as described in your premarket notification.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions.

{4}------------------------------------------------

Page 2 -- Mr. Mark Job

Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed needicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (240) 276-0120. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (240) 276-0100. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

If you have any questions regarding the content of this letter, please contact Ms. Lauren Hefner at (240) 276-3666.

Sincerely vours,

Hori M. Zhang

So Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{5}------------------------------------------------

Appendix F

Ultrasound Device Indication For Use

510(k) Number

Device Name VU62000 VascuView Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(specify)Other(specify)
Ophthalmic
Fetal
Abdominal
Intraoperative (specify)
Intraoperative Neurological
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Transesophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral VascularN
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletal Superficial
Other (specify)

Additional Comments:

(PLEASE OO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

_
(Division Sign-Off)_

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

E-3

{6}------------------------------------------------

Ultrasound Device Indication For Use

510(k) Number

Device Name Interson SR 7.5 MHz Mechanical Sector Probe

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationABMPWD CWDColor DopplerAmplitude DopplerColor Velocity ImagingCombined (specify)Other (specify)
Ophthalmic
Fetal
Abdominal
Intraoperative (specify)
Intraoperative Neurological
Pediatric
Small Organ (specify)PNote 3
Neonatal Cephalic
Adult Cephalic
Cardiac
Transesophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral VascularPNote 3
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletal Superficial
Other (specify)

usly cleared by FDA; E= added under Appe

Additional Comments:__Note 3: Includes imaging for guidance of biopsy

(previously cleared under 510(k) 070907 - model no. VC 7.5 MHz)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)


(Division Sign-Off)

Division of Reproductive, Abdominal, and
Radiological Devices
510(k) Number [K080049](https://510k.innolitics.com/device/K080049)

F-3

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.