K Number
K080008

Validate with FDA (Live)

Date Cleared
2009-02-23

(418 days)

Product Code
Regulation Number
866.3510
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioPlex™ 2200 ToRC IgG kit is a multiplex flow immunoassay intended for the quantitative detection of IgG antibodies to Toxoplasma gondii (T. gondii) and Rubella, and the qualitative detection of IgG antibodies to Cytomegalovirus (CMV) in human serum and EDTA or heparinized plasma. The ToRC IgG kit is intended for use with the Bio-Rad BioPlex 2200 System. This kit is intended as an aid in the determination of serological status to T. gondii, Rubella and CMV. This kit is not intended for use in screening blood or plasma donors. Performance characteristics for T. gondii and Rubella have not been evaluated in immunocompromised or immunosuppressed individuals. Performance characteristics for CMV have not been evaluated in immunosuppressed or organ transplant individuals. Performance characteristics of this kit have not been established for use in neonatal screening or for use at a point of care.

The BioPlex 2200 ToRC IgG Calibrator Set is intended for the calibration of the BioPlex 2200 ToRC IgG Reagent Pack.

The BioPlex 2200 ToRC IgG Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex 2200 Instrument and BioPlex ToRC IgG Reagent Pack in the clinical laboratory. The performance of the BioPlex 2200 ToRC IgG Control Set has not been established with any other Toxoplasma gondii, Rubella or Cytomegalovirus (CMV) IgG antibody assays.

Device Description

The BioPlex™ 2200 ToRC IgG kit uses multiplex flow immunoassay, a methodology that greatly resembles traditional EIA, but permits simultaneous detection and identification of many antibodies in a single tube. "ToRC" is an acronym for individual tests to detect antibodies to Toxoplasma gondii (T. gondii), Rubella, and Cytomegalovirus (CMV). Three (3) different populations of dyed beads are coated with cell lysates bearing T. gondii, Rubella, or CMV antigens.

The BioPlex 2200 System combines an aliquot of patient sample, sample diluent, and bead reagent into a reaction vessel; the mixture is incubated at 37°C. After a wash cycle, anti-human IgG antibody, conjugated to phycoerythrin (PE), is added to the dyed beads, and this mixture is incubated at 37°C. The excess conjugate is removed in another wash cycle, and the beads are re-suspended in wash buffer. The bead mixture then passes through the detector. The identity of the dyed beads is determined by the fluorescence of the dyes, and the amount of antibody captured by the antigen is determined by the fluorescence of the attached PE. Raw data are calculated in relative fluorescence intensity (RFI).

Three additional dyed beads, Internal Standard Bead (ISB), Serum Verification Bead (SVB) and a Reagent Blank Bead (RBB) are present in each reaction mixture to verify detector response, the addition of serum or plasma to the reaction vessel and the absence of significant non-specific binding in serum or plasma.

The instrument is calibrated using a set of six (6) distinct calibrator vials, the BioPlex 2200 ToRC IgG Calibrator Set. For T. gondii and Rubella, six (6) vials, representing six (6) different levels of antibody concentrations, are used for quantitative calibration, and results for patient samples are expressed in IU/mL. For T. gondii, results of ≤ 9 IU/mL are negative, 10 and 11 IU/mL are equivocal, and results of ≥ 12 IU/mL are reported as positive. For Rubella, results of ≤ 7 IU/mL are reported as negative, 8 and 9 IU/mL are equivocal, and ≥ 10 IU/mL are reported as positive. For CMV, four (4) vials, representing four (4) different antibody concentrations, are used for qualitative calibration. CMV results are expressed as an antibody index (Al) and results of ≤ 0.8 Al are negative, 0.9 and 1.0 Al are equivocal, and results of ≥ 1.1 Al are reported as positive.

The BioPlex 2200 ToRC IgG Control Set includes a negative control as well as two (2) multi-analyte positive controls. The BioPlex ToRC IgG Low Positive Control contains antibodies for T. gondii, Rubella and CMV and the BioPlex ToRC IgG High Positive Control contains antibodies for T. gondii and Rubella. The BioPlex ToRC IgG Positive Controls are manufactured to give positive results, with values above the cut-off for each specific analyte. The BioPlex ToRC IgG Negative Control is manufactured to give negative results, with values below the cut-off for each specific analyte. The recommended frequency for performing quality control is once every 24-hour testing period. Performing quality control is also necessary after each new assay calibration and certain service procedures.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and the study proving the device meets them:

BioPlex™ 2200 ToRC IgG Kit Performance Analysis

1. Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" but rather describes various performance characteristics evaluated against predetermined principles (e.g., CLSI standards) or by comparison to predicate devices. The "reported device performance" is then presented as the outcome of these evaluations.

Based on the provided text, the most relevant performance characteristics and their reported outcomes are summarized below:

Performance CharacteristicAcceptance Criteria (Implicit/Explicit)Reported Device Performance
ReproducibilityBased on CLSI EP5-A2 principles. Expectation of acceptable within-run and total precision (low % CV values).Internal Reproducibility:- T. gondii IgG: Within-run 3.9%-9.5%, Total 6.6%-12.3%- Rubella IgG: Within-run 3.0%-5.0%, Total 5.2%-9.9%- CMV IgG: Within-run 1.8%-5.6%, Total 3.8%-8.7%External Reproducibility:- T. gondii IgG: Within-run 3.7%-7.7%, Total 4.8%-17.6%- Rubella IgG: Within-run 4.6%-6.6%, Total 7.2%-10.3%- CMV IgG: Within-run 2.5%-5.8%, Total 4.4%-12.1%Additional Rubella Reproducibility (near cut-off): Within-run 5.2%-9.7%
LinearityBased on CLSI EP6-A principles. Expectation of acceptable dilution linearity, with all recoveries within ±20% of the predicted value.Demonstrated acceptable dilution linearity for T. gondii (3-900 IU/mL), Rubella (1-250 IU/mL), and CMV (0.2-8.0 Al). All recoveries within ±20% of the predicted value. Acceptable dilution linearity through the cut-off for T. gondii and Rubella using WHO IgG standards.
Interfering SubstancesBased on CLSI EP7-A2 principles. Expectation of no significant interference from tested substances at specified concentrations.No significant interference observed with tested substances (Hemoglobin, Bilirubin, Cholesterol, Red Blood Cells, Gamma Globulin, Triglyceride, Total Protein, Ascorbic Acid, Heparin, EDTA) at maximum levels.
Cross-ReactivityExpectation that various disease state samples do not cross-react with the three (3) assays in the BioPlex ToRC IgG kit.Evaluated specimens positive for ANA, CMV, dsDNA (SLE), EBV VCA, hCG (pregnancy), HIV, HSV-1, HSV-2, Influenza, Mumps, Myeloma, Parvovirus B19, Rheumatoid Factor, Rubella, Rubeola, T. gondii, VZV. The results demonstrated that the various disease state samples evaluated do not cross-react with the three (3) assays. Most positive results were confirmed by reference assays. Only one equivocal result was noted and not counted as discrepant.
Correlation with CDC Evaluation PanelsExpectation of high agreement with CDC Reference Methods.T. gondii IgG (N=100): BioPlex detected 70 Positives and 30 Negatives, matching CDC Reference Result (100% agreement).Rubella IgG (N=100): BioPlex detected 82 Positives and 18 Negatives across all 3 sites, matching CDC Reference Result (100% agreement).CMV IgG (N=100): BioPlex detected 60 Positives and 40 Negatives across all 3 sites, matching CDC Reference Result (100% agreement).
Performance Characteristics (vs. Predicate EIA)High positive and negative percent agreement (PPA/NPA) with predicate commercial immunoassays, with narrow 95% Confidence Intervals. Specific numerical thresholds for "high" are not explicitly stated, but the reported values demonstrate strong concordance.Rubella IgG (Pregnant Women, N=300): Exclusion of equivocal: PPA 97.2% (276/284), NPA 100% (8/8). Inclusion of equivocal: PPA 94.5% (276/292), NPA 100% (8/8).Rubella IgG (Clinical Samples, N=400): Exclusion of equivocal: PPA 96.5% (358/371), NPA 100% (12/12). Inclusion of equivocal: PPA 92.7% (358/386), NPA 85.7% (12/14).T. gondii IgG (Total, N=700): PPA 97.5% (118/121), NPA 98.8% (569/576).CMV IgG (Total, N=700): PPA 99.0% (394/398), NPA 98.7% (298/302).CMV IgG (HIV+, N=100): PPA 100% (86/86), NPA 100% (14/14).Retrospective Rubella IgG (N=280): Exclusion of equivocal: PPA 92.8% (90/97), NPA 100% (124/124). Discordant samples were mostly low positive or close to cut-off for predicate.

2. Sample Size Used for the Test Set and Data Provenance

The primary test sets for evaluating performance characteristics (agreement with predicate devices) are described as:

  • Prospective Samples:
    • Pregnant Women: 300 samples (150 from the U.S., 150 from Europe).
    • Clinical Samples submitted for ToRC IgG testing: 1200 samples (400 for Rubella IgG, 700 for T. gondii IgG, 700 for CMV IgG - likely overlapping sets for the total of 1200).
    • Immunocompromised/AIDS patients (for CMV testing): 100 samples.
    • Provenance: Samples from the U.S. and Europe, some from specific clinical populations (pregnant women, immunocompromised/AIDS patients), and others from general clinical submissions. All appear to be prospective collections relative to the study.
  • Retrospective Samples:
    • Rubella IgG: 280 samples (50 low positive, 50 high positive, 130 negative).
    • Provenance: Retrospective collection from U.S. clinical sites with known Rubella IgG status.
  • CDC Evaluation Panels:
    • T. gondii IgG: 100 samples.
    • Rubella IgG: 100 samples.
    • CMV IgG: 100 samples.
    • Provenance: Provided by the Centers for Disease Control (CDC).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

The ground truth for the test set was established primarily through comparison to predicate commercial immunoassays (e.g., bioMerieux VIDAS® TOXO IgG II, Rubella IgG, and CMV IgG kits) and CDC Reference Methods.

  • Number of Experts: Not explicitly stated. For predicate assays, the "expert" is essentially the predicate test itself, which has previously undergone its own validation. For CDC Reference Methods, the expertise would be inherent in the established CDC methodology. There is no mention of individual experts reviewing results for ground truth establishment.
  • Qualifications of Experts: Not applicable in the context of this document, as it relies on established assays and CDC reference methods rather than individual expert consensus.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method involving human review or consensus for discrepant results between the BioPlex 2200 ToRC IgG kit and the predicate devices. Instead, it reports percent agreements excluding equivocal results and including equivocal results from the BioPlex device. The discrepancies where the BioPlex 2200 disagreed with the predicate were simply reported as "Agreement" percentages. For the retrospective Rubella study, it's noted that "Most of the discordant samples were low positive or close to the cut-off for the predicate assay," providing some context, but not an adjudication process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable to this device. The BioPlex™ 2200 ToRC IgG kit is an automated in vitro diagnostic immunoassay system. It does not involve human "readers" interpreting images or clinical cases in a way that an MRMC study, typical for AI-powered diagnostic imaging, would be relevant. Its performance is evaluated against other diagnostic assays, not based on human interpretation improvement.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the studies described are all standalone performance evaluations of the BioPlex™ 2200 ToRC IgG kit. The device is an automated system designed to provide results without human intervention in the assay interpretation step. The "Total" percentages of agreement with predicate assays represent the device's standalone performance.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The ground truth was established primarily using:

  • Reference Devices/Predicate Assays: Results from other commercially available, FDA-cleared TOXO IgG, Rubella IgG, and CMV IgG immunoassays (bioMerieux VIDAS® kits) were used as the comparator "truth."
  • CDC Reference Methods: For the correlation study, CDC-provided panels were used, with their results presumably representing a highly validated reference standard.

It is not based on pathology, expert consensus (beyond the validation of the reference assays themselves), or outcomes data.

8. The Sample Size for the Training Set

The document does not explicitly state the sample size for a "training set" in the context of developing the BioPlex 2200 ToRC IgG Kit. This type of immunoassay development typically involves extensive characterization and optimization using various reagent lots, calibrators, and control materials, rather than a "training set" in the machine learning sense. The performance studies described are for validation of the finalized device.

9. How the Ground Truth for the Training Set was Established

As there is no distinct "training set" described in the context of an algorithm or machine learning for this immunoassay, information on how its ground truth was established is not provided and is likely not applicable in the conventional sense for this type of device. The ground truth for the device's calibration and controls is established by defining concentrations or expected results for these internal control materials, which are then used to standardize assay performance.

{0}------------------------------------------------

510(k) Summary for Bio-Rad Laboratories, Inc. BioPlex™ 2200 ToRC IgG

1. Applicant/Sponsor

Bio-Rad Laboratories, Inc. BioPlex Division 5500 East Second Street Benicia, CA 94510

Contact Person: Teresa Peterson Telephone: 510-741-5702 Date Prepared: February 18, 2009

2. Device Name

Proprietary Name:

Common/Usual Name:

Classification Name:

BioPlex™ 2200 ToRC IgG Kit BioPlex™ 2200 ToRC IgG Calibrator Set BioPlex™ 2200 ToRC IgG Control Set · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·

Multi-Analyte Detection System for Toxoplasma gondii IgG, Rubella IqG and Cytomegalovirus (CMV) IqG

Toxoplasma gondii serological reagents Rubella virus serological reagents Cytomegalovirus serological reagents Calibrator, multi-analyte mixture Multi-analyte controls, all kinds (assayed and unassayed)

3. Predicate Devices

• bioMerieux VIDAS® TOXO IgG IIK993319
• bioMerieux VIDAS® Rubella IgGK902925
• bioMerieux VIDAS® CMV IgGK920661

FEB 2 3 2009

{1}------------------------------------------------

4. DEVICE DESCRIPTION

The BioPlex™ 2200 ToRC IgG kit uses multiplex flow immunoassay, a methodology that greatly resembles traditional EIA, but permits simultaneous detection and identification of many antibodies in a single tube. "ToRC" is an acronym for individual tests to detect antibodies to Toxoplasma gondii (T. gondii), Rubella, and Cytomegalovirus (CMV). Three (3) different populations of dyed beads are coated with cell lysates bearing T. gondii, Rubella, or CMV antigens. and the state of the first the

The BioPlex 2200 System combines an aliquot of patient sample, sample diluent, and bead reagent into a reaction vessel; the mixture is incubated at 37°C. After a wash cycle, anti-human IgG antibody, conjugated to phycoerythrin (PE), is added to the dyed beads, and this mixture is incubated at 37°C. The excess conjugate is removed in another wash cycle, and the beads are re-suspended in wash buffer. The bead mixture then passes through the detector. The identity of the dyed beads is determined by the fluorescence of the dyes, and the amount of antibody captured by the antigen is determined by the fluorescence of the attached PE. Raw data are calculated in relative fluorescence intensity (RFI).

Three additional dyed beads, Internal Standard Bead (ISB), Serum Verification Bead (SVB) and a Reagent Blank Bead (RBB) are present in each reaction mixture to verify detector response, the addition of serum or plasma to the reaction vessel and the absence of significant non-specific binding in serum or plasma.

The instrument is calibrated using a set of six (6) distinct calibrator vials, the BioPlex 2200 ToRC IgG Calibrator Set. For T. gondii and Rubella, six (6) vials, representing six (6) different levels of antibody concentrations, are used for quantitative calibration, and results for patient samples are expressed in IU/mL. For T. gondii, results of ≤ 9 IU/mL are negative, 10 and 11 IU/mL are equivocal, and results of ≥ 12 IU/mL are reported as positive. For Rubella, results of ≤ 7 : U/mL are reported as negative, 8 and 9 /U/mL are equivocal, and ≥ 10 IU/mL are reported as positive. For CMV, four (4) vials, representing four (4) different antibody concentrations, are used for qualitative calibration. CMV results are expressed as an antibody index (Al) and results of ≤ 0.8 Al are negative, 0.9 and 1.0 Al are equivocal, and results of ≥ 1.1 Al are reported as positive.

The BioPlex 2200 ToRC IgG Control Set includes a negative control as well as two (2) multi-analyte positive controls. The BioPlex ToRC IgG Low Positive Control contains antibodies for T. gondii, Rubella and CMV and the BioPlex ToRC IgG High Positive Control contains antibodies for T. gondii and Rubella. The BioPlex ToRC IqG Positive Controls are manufactured to give positive results, with values above the cut-off for each specific analyte. The BioPlex ToRC IgG Negative Control is manufactured to give negative results, with values below the cut-off for each specific analyte. The recommended frequency for performing quality control is once every 24-hour testing

{2}------------------------------------------------

period. Performing quality control is also necessary after each new assay calibration and certain service procedures.

BioPlex™ 2200 ToRC IgG Kit Components

The BioPlex 2200 ToRC IgG kit (665-1650) contains supplies sufficient for 100 tests.

VialDescription
Bead SetOne (1) 10 mL vial containing three (3) different populationsof dyed beads coated with lysates (tachyzoites, Vero cell andnuclear) of T. gondii, Rubella and CMV; plus an InternalStandard Bead (ISB), a Serum Verification Bead (SVB) and aReagent Blank Bead (RBB); with protein stabilizers (bovine)in a saline buffer. ProClin® 300 (0.3%), sodium benzoate(0.1%) and sodium azide (<0.1%) as preservatives.
ConjugateOne (1) 5 mL vial, containing murine monoclonal anti-humanIgG/phycoerythrin conjugate, in a phosphate buffer. ProClin®300 (0.3%), sodium benzoate (0.1%) and sodium azide(<0.1%) as preservatives.
Sample DiluentOne (1) 10 mL vial, containing protein stabilizers (bovine andmurine) in a saline buffer. ProClin® 300 (0.3%), sodiumbenzoate (0.1%) and sodium azide (<0.1%) as preservatives.

Additional Required Items, Available from Bio-Rad:

Catalog #Description
663-1600BioPlex 2200 ToRC IgG Calibrator Set: Six (6) 0.5 mL vialscontaining antibodies to T. gondii, Rubella and CMV derivedfrom human disease state plasma, in a human serum matrixmade from defibrinated plasma. ProClin® 300 (0.3%),sodium benzoate (0.1%) and sodium azide (<0.1%) aspreservatives.
663-1630BioPlex 2200 ToRC IgG Control Set: Two (2) 1.5 mL vials ofLow Positive Control containing antibodies to T. gondii,Rubella and CMV; two (2) 1.5 mL vials of High PositiveControl containing antibodies to T. gondii and Rubella; andtwo (2) 1.5 mL vials of Negative Control Serum. All controlsin a human serum matrix made from defibrinated plasma,with ProClin® 300 (0.3%), sodium benzoate (0.1%) andsodium azide (<0.1%) as preservatives. All antibodies arederived from human disease state plasma.
660-0817BioPlex 2200 Sheath Fluid: Two (2) 4 L bottles containingPhosphate Buffered Saline (PBS). ProClin® 300 (0.03%) andsodium azide (<0.1%) as preservatives.
660-0818BioPlex 2200 Wash Solution: One (1) 10 L bottle containingPhosphate Buffered Saline (PBS) and Tween 20. ProClin®300 (0.03%) and sodium azide (<0.1%) as preservatives.
660-0000BioPlex 2200 Instrument and Software System.

neer a bli forto ago countrol Soll World Creative .

{3}------------------------------------------------

5. Intended Use

BioPlex™ 2200 ToRC IgG Kit

The BioPlex™ 2200 ToRC IgG kit is a multiplex flow immunoassay intended for the quantitative detection of IqG antibodies to Toxoplasma gondii (T. gondii) and Rubella, and the qualitative detection of IgG antibodies to Cytomegalovirus (CMV) in human serum and EDTA or heparinized plasma.

The ToRC IgG kit is intended for use with the Bio-Rad BioPlex 2200 System.

This kit is intended as an aid in the determination of serological status to T. gondii, Rubella and CMV. This kit is not intended for use in screening blood or plasma donors.

Performance characteristics for T. gondii and Rubella have not been evaluated in immunocompromised or immunosuppressed individuals. Performance characteristics for CMV have not been evaluated in immunosuppressed or organ transplant individuals. Performance characteristics of this kit have not been established for use in neonatal screening or for use at a point of care.

BioPlex™ 2200 ToRC IgG Calibrator Set

The BioPlex 2200 ToRC IgG Calibrator Set is intended for the calibration of the BioPlex 2200 ToRC IgG Reagent Pack.

BioPlex™ 2200 ToRC IgG Control Set

The BioPlex 2200 ToRC IgG Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex 2200 Instrument and BioPlex ToRC IgG Reagent Pack in the clinical laboratory. The performance of the BioPlex 2200 ToRC IgG Control Set has not been established with any other Toxoplasma gondii, Rubella or Cytomegalovirus (CMV) IgG antibody assays.

6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The following tables summarize the similarities and differences between the BioPlex 2200 ToRC IgG kit and the predicate devices used in comparative studies with the BioPlex 2200 ToRC IgG kit.

{4}------------------------------------------------

BioPlex™ 2200 ToRC IgG Kit vs. Predicate Devices - Similarities

:

.

BioPlex™ 2200ToRC IgG KitbioMerieux, Inc. VIDAS®TOXO IgG II (K993319)bioMerieux, Inc. VIDAS®Rubella IgG (K902925)bioMerieux, Inc. VIDAS®CMV IgG (K920661)
Intended Use- The BioPlex™ 2200 ToRCIgG kit is a multiplex flowimmunoassay intended for thequantitative detection of IgGantibodies to Toxoplasmagondii ( T. gondii ) and Rubella,and the qualitative detection ofIgG antibodies toCytomegalovirus (CMV) inhuman serum and EDTA orheparinized plasma.- The ToRC IgG kit is intendedfor use with the Bio-RadBioPlex 2200 System.- This kit is intended as an aidin the determination ofserological status to T. gondii ,Rubella and CMV. This kit isnot intended for use in thescreening of blood or plasmadonors.VIDAS® Toxo IgG II (TXG) isan automated quantitative testfor use on the VIDAS analyzerfor the measurement of anti-Toxoplasma IgG in humanserum using the ELFAtechnique. It is intended foruse as an aid in determinationof immune status.VIDAS® Rubella IgG (RBG)assay is intended for use witha VIDAS (VitekImmunoDiagnostic AssaySystem) instrument as anautomated qualitativeenzyme-linked fluorescentimmunoassay (ELFA) for thedetection of IgG antibodies torubella virus in human serum.VIDAS® Cytomegalovirus IgG(CMVG) is intended for usewith the VIDAS (VitekImmunoDiagnostic AssaySystem) instrument as a semi-quantitative automatedenzyme-linked fluorescentimmunoassay (ELFA)for thedetection of IgG antibodies toCytomegalovirus in humanserum.
AntigenPartially purified cell lysates ofT. gondii , Rubella, and CMVMembrane and cytotoxicToxoplasma antigen(RH Sabin Strain)Inactivated Rubella virus(HPV-77)Purified and inactivated CMVantigen (Strain AD 169)
Assay TypeQuantitative detection forT. gondii , Rubella, andqualitative for CMVSameSame
Analyte DetectedHuman IgG antibodiesto T. gondii , Rubella, andCMVSameSame
Signal DetectionFluorescenceSameSame
ItemBioPlex™ 2200TORC IgG KitbioMerieux, Inc. VIDAS®TOXO IgG II (K993319)bioMerieux, Inc. VIDAS®Rubella IgG (K902925)bioMerieux, Inc. VIDAS®CMV IgG (K920661)
Number of AnalytesSimultaneouslyDetectedMultiple (3)SingleSingleSingle
Enzyme ConjugatePhycoerythrin conjugatedHorseradish peroxidase conjugated
MatricesSerum and plasma (EDTAand heparin)Serum and plasma (EDTAand heparin)Serum
Assay TechnologyMultiplex flow immunoassayTwo-step enzyme immunoassay sandwich method with fluorescent detection (ELFA)
Solid PhaseAntigen-coated paramagneticmicrobead reagent.Microbeads are infused withred and infrared fluorescentdyes for bead classification.Green fluorescence from theimmunoassay label is used foranalyte measurement.Antigen-coated solid phase receptacles
Calibrator(s)Multiple CalibratorsSingle Calibrator
ControlsNegative Control and multi-analyte Positive ControlsNegative Control and PositiveControl specific for T. gondiiNegative Control and PositiveControl specific for rubellavirusNegative Control and PositiveControl specific for CMV virus

.

  1. .. ..

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

:

Bio-Rad Laboratories, Inc. 510(K)
Additional Information - K080008

510(k) Summary

,

February 18, 2009

{5}------------------------------------------------

BioPlex 2200 ToRC IgG Kit vs. Predicate Devices - Differences

·

...

.

:

. . . . . .

.

. . . . . . . . . . .

.

. .

  • ...
    . ...
    . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
    . . . . . . . . . . . . . . . . . . . . . . . . . . .
    . . . . . . . . . . . . . . . . . . . . .

: 上一篇:

... ... ........

510(k) Summary

February 18, 2009

Bio-Rad Laboratories, Inc. 510(k

{6}------------------------------------------------

7. Performance Testing

A series of studies was conducted to evaluate the performance of the BioPlex™ 2200 ToRC IgG kit. The studies included reproducibility, linearity, interfering substances, cross-reactivity, correlation with CDC evaluation panels, expected values and performance characteristics. The results of all studies demonstrated that the BioPlex 2200 ToRC IgG kit performed according to its specifications.

A. Reproducibility

Separate internal and external reproducibility studies were conducted to evaluate the reproducibility of the BioPlex 2200 ToRC IgG kit on the BioPlex 2200 instrument. Reproducibility studies were based on the principles described in Clinical and Laboratory Standards Institute (CLSI) EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods.

For the internal reproducibility study, three (3) panels made from serum and plasma (EDTA and heparinized) were assayed two (2) times in two (2) separate daily runs over 20 days (n=80). The data were analyzed for withinrun, between-run, between-day, and total precision and the standard deviation (SD) and percent coefficient of variation (% CV) were calculated. The within-run precision for positive samples, greater than their respective cut-offs. in all sample matrices ranged from 3.9% to 9.5% for T. gondii IgG, 3.0% to 5.0% for Rubella IgG, and 1.8% to 5.6% for CMV IgG. The total precision for positive samples, greater than their respective cut-offs, in all sample matrices ranged from 6.6% to 12.3% for T. gondii IgG, 5.2% to 9.9% for Rubella IgG, and 3.8% to 8.7% for CMV IgG.

An external reproducibility study was performed at each of three (3) U.S. testing facilities. Three (3) panels made from serum and plasma (EDTA and heparinized) were assayed two (2) times on two (2) separate daily runs over five (5) days at the three (3) sites (N=60). The data were analyzed for withinrun, between-run, between-day, between-site, and total precision and the standard deviation (SD) and percent coefficient of variation (% CV) were calculated. The within-run precision for positive samples, greater than their respective cut-offs, in all sample matrices ranged from 3.7% to 7.7% for T. gondii IgG, 4.6% to 6.6% for Rubella IgG, and 2.5% to 5.8% for CMV IgG. The total precision for positive samples, greater than their respective cut-offs, in all sample matrices ranged from 4.8% to 17.6% for T. gondii IgG, 7.2% to 10.3% for Rubella IgG and 4.4% to 12.1% for CMV IgG.

An additional external Rubella reproducibility study was performed at each of three (3) U.S. testing facilities. A panel of three (3) Rubella IgG samples measuring near cut-off was assayed in replicates of 40 in one run at each

{7}------------------------------------------------

The data were analyzed for within-run precision and the standard site. deviation (SD) and percent coefficient of variation (% CV) were calculated. The within-run precision ranged from 5.2% to 9.7%.

B. Linearity

Linearity of the BioPlex 2200 ToRC IgG kit was determined based on the principles described in Clinical and Laboratory Standards Institute (CLSI) EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: Statistical Approach. The assay ranges are T. gondii (3 - 900 IU/mL), Rubella (1 - 250 IU/mL) and CMV (0.2 - 8.0 Al), and were established by examining the clinical relevance of reporting high value results and the ability of the calibration curve to discriminate sample dilutions. Five (5) high positive serum samples were obtained for each analyte in the kit. The samples were selected at the high end of the assay range and serial dilutions prepared using negative serum. Linear and polynomial regression analysis of IU/mL or Al vs. sample dilution was performed to determine if the dilution curves exhibit statistically significant non-linear regression. The results demonstrated acceptable dilution linearity, as all recoveries were within ±20% of the predicted value . . . . . . . . . . . . . . . . . . . . . . . .

Linearity of BioPlex ToRC IgG T. gondii and Rubella assays was evaluated using World Health Organization (WHO) IgG standards. Dilutions of anti-Toxoplasma serum, 3rd International Standard (TOXM), and anti-Rubella, immunoglobulin, 1st International Standard (RUBI-1-94), were prepared and tested in the respective assay. Results obtained demonstrated acceptable dilution linearity through the cut-off, as all recoveries were within ±20% of the predicted value.

miemational Slandaile

{8}------------------------------------------------

C. Interfering Substances

An interfering substances study was conducted to evaluate the potential interference of specific endogenous and exogenous substances with the BioPlex 2200 ToRC IgG assays. The study was conducted based on the principles described in Clinical and Laboratory Standards Institute (CLSI) EP7-A2, Interference Testing in Clinical Chemistry. No significant interference was observed with any of the substances tested. The substances and the maximum levels tested are shown in the table below.

SubstanceConcentration
Hemoglobin500 mg/dL
Bilirubin, Unconjugated30 mg/dL
Bilirubin, Conjugated30 mg/dL
Cholesterol500 mg/dL
Red Blood Cells0.4% (v/v)
Gamma Globulin6 g/dL
Triglyceride3500 mg/dL
Total Protein (albumin)12 g/dL
Ascorbic Acid3 mg/dL
Heparin Lithium8000 units/dL
Heparin Sodium8000 units/dL
EDTA800 mg/dL

Interfering Substances

D. Cross-Reactivity miller in come the most in with and in

A cross-reactivity study was performed to determine if samples from various disease states and other potentially interfering factors interfere with test results when tested with the BioPlex 2200 ToRC IgG kit. A panel of ten (10) specimens positive for each cross reactant were evaluated for possible crossreactivity with each of the three (3) BioPlex ToRC IgG assays. Samples were also tested with reference ToRC IgG assays. The results demonstrated that the various disease state samples evaluated do not cross-react with the three (3) assays in the BioPlex ToRC IgG kit. Most of the samples evaluated were high positives for each disease state. If a specific disease state did cross react to the various assays of the BioPlex ToRC IgG kit, all samples for that particular disease would elicit a positive response. The majority of the samples that elicited a positive result were also confirmed positive by the corresponding reference ToRC IgG assay. Results are shown in the table below.

Bio-Rad Laboratories, Inc. 510(k) Additional Information - K080008 February 18, 2009

510(k) Summary 9 of 17

{9}------------------------------------------------

Cross-Reactivity

서비스 - 대부 - 대부분 - 대부분 - 대부분 - 대통령 - 대한민국 - 대한민국 -

PotentialCross-ReactantNMethodT. gondii IgGRubella IgGCMV IgG
ANA IgG10BioPlex 2200498*
10Reference4108
Discordant010
CMV IgG10BioPlex 22002910
10Reference2910
Discordant000
dsDNA (SLE clinical)10BioPlex 2200499
10Reference4*99
Discordant000
EBV VCA IgG10BioPlex 2200193
10Reference193
Discordant000
hCG (pregnancy)10BioPlex 22004910
10Reference4910
Discordant000
HIV IgG10BioPlex 22001910
10Reference1910
Discordant000
HSV-1 IgG10BioPlex 22000107
10Reference0106*
Discordant000
HSV-2 IgG10BioPlex 22003108*
10Reference3109
Discordant000
Influenza IgG10BioPlex 2200577
10Reference577
Discordant000
Mumps IgG10BioPlex 22004107
10Reference4107
Discordant000
Myeloma IgG10BioPlex 2200379
10Reference387
Discordant012
Parvovirus B19 IgG10BioPlex 22002104
10Reference2105
Discordant001
Rheumatoid FactorIgM10BioPlex 22001109
10Reference1109
Discordant000
Rubella IgG10BioPlex 22003107
10Reference3107
Discordant000
Rubeola (measles)IgG10BioPlex 22000105
10Reference0105
Discordant000
T. gondii IgG10BioPlex 2200999
10Reference999
Discordant000
VZV IgG10BioPlex 22003*107
10Reference3*107
Discordant000

*One equivocal result was not included in the count and was not considered as a false positive or negative discrepant. .

Bio-Rad Laboratories, Inc. 510(k) Additional Information - K080008

.. & R W .. .. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

: .

:

・・)

510(k) Summary 10 of 17

{10}------------------------------------------------

E. Correlation with CDC Evaluation Panels

A correlation study was performed to evaluate the characteristics of the BioPlex ToRC IgG kit with serum panels provided by the Centers for Disease Control (CDC) for T. gondii, Rubella and CMV. Three (3) panels were tested and the data were provided to the CDC for comparison against the CDC Reference Methods for the three (3) analytes. The results are presented in the tables below.

Characteristics of CDC T. gondii Reference Sera (N=100)

CDC SamplesReference ResultBioPlex Pos (+)BioPlex Neg (-)
T. gondii IgGPos (+)700
Neg (-)030

Characteristics of CDC Rubella Reference Sera (N=100)

CDC SamplesReference ResultSite 1Site 2Site 3
BioPlexPos (+)BioPlexNeg (-)BioPlexPos (+)BioPlexNeg (-)BioPlexPos (+)BioPlexNeg (-)
Rubella IgGPos (+)820820820
Neg (-)018018018

Characteristics of CDC CMV Reference Sera (N=100)

BIORINGSICALCompany of Concession of ConcessionO THE CONTRACT CARDEDANYMINIA22.25STATE AND THE COLLECTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULAT OF1And States of Children Company of ChildrenA C A C A C A P A C A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P A P AAnd Comments of the first ofCALL COLLECTIONAnd Control of ConcessionCarrely of them and any110120, 01111994A C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C C1. 10 HorAND AND OF CONSTITUTION OFA R R PCompany of the control of the control of the county of the county of the county of the county ofCar Concession100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000110001-1 475 315 31 31 31 31 31ANDARD ADDRESSS OF OLD GE GENERAL000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000of the later the propose10 T 6 7 6 2 6 1 6 1 6 1 6 1 6 1 6 4 1 6 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4
CMV laG
A 400 400 - 13.00 - 10.000 - 10.000 - 10.000

{11}------------------------------------------------

F. Expected Values/Prevalence

The observed prevalence for each of the ToRC IgG assays, individually and in dual infection, was determined using samples collected from pregnant women in the U.S. and Europe as well as in clinical samples submitted for T. gondii, Rubella, or CMV IgG testing. For T. gondii, results of ≤ 9 IU/mL are negative, 10 and 11 IU/mL are equivocal, and results of ≥ 12 IU/mL are reported as positive. For Rubella, results of ≤ 7 |U/mL are reported as negative, 8 and 9 IU/mL are equivocal, and ≥ 10 IU/mL are reported as positive. For CMV, results of ≤ 0.8 Al are negative, 0.9 and 1.0 Al are equivocal, and ≥ 1.1 Al are reported as positive. Results from all sites are shown and summarized in the tables below.

AgeT. gondii IgGUSPos/TotalT. gondii IgGUS% PrevalenceT. gondii IgGEuropePos/TotalT. gondii IgGEurope% PrevalenceRubella IgGPos/TotalRubella IgG% PrevalenceCMV IgGPos/TotalCMV IgG% Prevalence
16-255/4211.99/3129.071/7397.341/7356.2
26-355/895.633/8339.8155/17290.190/17252.3
36-451/195.316/3644.450/5590.928/5550.9
Total11/1507.358/15038.7276/30092.0159/30053.0

Prevalence of Individual Assay Positive in Pregnant Women

Note: There was 1 equivocal result for T. gondii and 6 equivocal results for Rubella

. . . . . . . . Rubena, fostils of a 7 ran

Prevalence of Dual Assay Positive in Pregnant Women

AgeT. gondii IgG /Rubella IgG UST. gondii IgG /Rubella IgG EuropeT. gondii IgG /CMV IgG UST. gondii IgG /CMV IgG EuropeRubella IgG /CMV IgG
Pos/Total% PrevalencePos/Total% PrevalencePos/Total% PrevalencePos/Total% PrevalencePos/Total% Prevalence
16-255/4211.99/3129.02/424.84/3112.939/7353.4
26-354/894.531/8337.32/892.219/8322.979/17245.9
36-451/195.316/3644.41/195.36/3616.725/5545.5
Total10/1506.756/15037.35/1503.329/15019.3143/30047.7

$\frac{1}{4}$

Bio-Rad Laboratories, Inc. 510(k) Additional Information - K080008

{12}------------------------------------------------

Prevalence of Individual Assay Positive in Samples Submitted for ToRC IgG Testing

AgeGenderT. gondii IgG% PrevalenceRubella IgG% PrevalenceCMV IgG% Prevalence
1-10Female1/911.18/988.93/933.3
Male1/812.57/887.50/80.0
11-20Female2/633.254/6385.734/6354.0
Male1/313.228/3190.316/3151.6
21-30Female30/23212.9214/23292.2123/23253.0
Male6/5610.750/5689.333/5658.9
31-40Female31/25012.4222/25088.8125/25050.0
Male19/8322.963/8375.954/8365.1
41-50Female8/869.378/8690.760/8669.8
Male16/9616.778/9681.364/9666.7
51-60Female5/578.854/5794.737/5764.9
Male26/9926.391/9991.962/9962.6
61-70Female12/4427.338/4486.436/4481.8
Male15/5030.049/5098.035/5070.0
71+Female8/1266.712/12100.012/12100.0
Male5/1435.712/1485.710/1471.4
Unknown Age and/orGender1/1010.08/1080.06/1060.0
Total187/120015.61066/120088.8710/120059.2

Note: There were 5 equivocal results for T. gondii, 31 equivocal results for Rubella, and 1 equivocal result for CMV. Comments of the country of the county of the station of

. . . . . .

.

134 4 1 1 - 4 -

:

. 1

の 2007年 - 1000 - 100 - 1000 - 100000 - 1000 - 1000

:

Bio-Rad Laboratories, Inc. 510(k) Additional Information – K080008

.

.

.

a de la product of the production of

{13}------------------------------------------------

Prevalence of Dual Assay Positive in Samples Submitted for ToRC IgG Testing

AgeGenderT.gondii IgG/ Rubella IgGT.gondii IgG/ CMV IgGRubella IgG /CMV IgG
Pos/Total% PrevalencePos/Total% PrevalencePos/Total% Prevalence
1-10Female1/911.10/90.02/922.2
1-10Male1/812.50/80.00/80.0
11-20Female2/633.22/633.228/6344.4
11-20Male1/313.21/313.214/3145.2
21-30Female29/23212.524/23210.3120/23251.7
21-30Male5/568.93/565.431/5655.4
31-40Female30/25012.017/2506.8115/25046.0
31-40Male15/8318.113/8315.743/8351.8
41-50Female8/869.36/867.054/8662.8
41-50Male14/9614.612/9612.552/9654.2
51-60Female5/578.84/577.035/5761.4
51-60Male24/9924.215/9915.257/9957.6
61-70Female12/4427.312/4427.330/4468.2
61-70Male15/5030.011/5022.034/5068.0
71+Female8/1266.78/1266.712/12100.0
71+Male5/1435.75/1435.79/1464.3
Unknown Age and/orGender1/1010.01/1010.04/1040.0
Total176/120014.7134/120011.2640/120053.3

Prevalence of CMV in Immunocompromised/AIDS Patient Samples Submitted for CMV IgG Testing

,

.:

රි.

AgeGenderCMV IgG
Pos/Total% Prevalence
1-10Female0/00.0
Male1/1100
11-20Female4/850.0
Male2/633.3
21-30Female2/2100
Male5/5100
31-40Female10/10100
Male12/1485.7
41-50Female8/8100
Male21/2391.3
51-60Female3/475.0
Male10/1190.9
61-70Female4/4100
Male4/4100
71+Female0/00.0
Male0/00.0
Total86/10086.0

Bio-Rad Laboratories, Inc. 510(k) Additional Information - K080008

.

February 18, 2009 - Page - Call - Ca 11:40 PM IST 11:00

ﺮ ﺍ

:

{14}------------------------------------------------

G. Performance Characteristics

Prospective Study: Performance of the ToRC IgG kit was evaluated against corresponding commercially available T. gondii, Rubella, and CMV immunoassays. U.S. clinical sites tested a combined: 300 prospective samples from pregnant women (150 U.S. and 150 Europe), 1200 prospective samples submitted for T. gondii, Rubella, and/or CMV IgG testing, and 100 prospective samples from immunocompromised/AIDS patients submitted for CMV testing. Results from all sites are shown and summarized in the tables below.

.. ... . . . . . . . . . . . . . . . . . .

Predicate Rubella IgG AssayBioPlex 2200 Agreement Excluding Equivocal ResultsBioPlex 2200 Agreement Including Equivocal Results
Antibody/PopulationPos (+)Neg (-)EquivocalPos (+)% Agreement95% ConfidenceIntervalNeg (-)% Agreement95% ConfidenceIntervalPos (+)% Agreement95% ConfidenceIntervalNeg (-)% Agreement95% ConfidenceInterval
BioPlex 2200 ToRC IgGRubella IgGPregnantWomen(N = 300)Pos (+)2760097.2%(276/284)94.5-98.6%100%(8/8)67.6-100%94.5%(276/292)91.3-96.6%100%(8/8)67.6-100%
Neg (-)882
Equivocal600
Total29082
ClinicalSamplesSubmittedfor Testing(N = 400)Pos (+)3580196.5%(358/371)94.1-97.9%100%(12/12)75.8-100%92.7%(358/386)89.7-94.9%85.7%(12/14)61.1-96.0%
Neg (-)13123
Equivocal1210
Total383134

BioPlex Rubella IgG vs. EIA: Prospective

*Due to the low prevalence of Rubella IgG negative samples, a retrospective study also was conducted.

{15}------------------------------------------------

PredicateT. gondii IgG AssayBioPlex 2200 Agreement
Antibody/PopulationPos (+)Neg (-)EquivocalPos (+)% Agreement95%ConfidenceIntervalNeg (-)% Agreement95%ConfidenceInterval
BioPlex 2200 ToRCT. gondii IgGTotal(N = 700)Pos (+)1180697.5%(118/121)93.0-99.2%98.8%(569/576)97.5-99.4%
Neg (-)15691
Equivocal113
Total12057010

BioPlex T. gondii IgG vs. EIA: Prospective

:

i

BioPlex CMV IgG vs. ElA: Prospective

PredicateCMV IgG AssayBioPlex 2200 Agreement
Antibody/PopulationPos (+)Neg (-)EquivocalPos (+)% Agreement95% ConfidenceIntervalNeg (-)% Agreement95% ConfidenceInterval
BioPlex 2200 ToRC IgGCMV IgGTotal(N = 700)Pos (+)3942299.0%(394/398)97.4-99.6%98.7%(298/302)96.6-99.5%
Neg (-)42980
Equivocal000
Total3983002
HIV+(N = 100)Pos (+)8600100%(86/86)95.8-100%100%(14/14)76.8-100%
Neg (-)0140
Equivocal000
Total86140

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

·

です。 2017年 10月 10:00 10:00 10:00 10:00 10:00 1 . . . - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -Bio-Rad Laboratories, Inc. 510(k) February 18, 2009 Additional Information - K080008 .

and the comments of the country

: 上一篇:

{16}------------------------------------------------

Retrospective Study: Performance of the ToRC IgG kit was evaluated using samples with confirmed known Rubella IgG status against a commercially available Rubella IgG immunoassay. Rubella IgG status was determined in clinical laboratories using an FDA-cleared assay for Rubella IgG. Two (2) U.S. clinical sites tested a combined: 50 Rubella IgG low positive samples (10-20 IU/mL), 50 Rubella IgG high positive samples (> 20 IU/mL) and 130 Rubella IgG negative samples (< 10 IU/mL). All samples were preselected based on the assay result falling in the ranges described. Results are shown in the table below.

Preselected SamplesPredicateRubella IgG AssayBioPlex 2200 Agreement ExcludingEquivocal ResultsBioPlex 2200 Agreement IncludingEquivocal Results
Pos (+)Neg (-)EquivocalPos (+)% Agreement95% ConfidenceIntervalNeg (-)% Agreement95% ConfidenceInterval
BioPlex 2200 ToRC IgGRubella IgGPos (+)900192.8%(90/97)85.8-96.5%100%(124/124)97.0-100%
Neg (-)71242
Equivocal501
Total1021244

BioPlex Rubella IgG vs. EIA: Retrospective

*Most of the discordant samples were low positive or close to the cut-off for the predicate assay.

Bio-Rad Laboratories, Inc. 510(k) Additional Information - K080008 510(k) Summary 17 of 17

{17}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Cynthia Sinclair, RAC Principal Consultant, Regulatory Services Medical Device Consultants, Inc. 49 Plain Street North Attleboro, MA 02760

FEB 2 3 2009

Re: K080008

Trade/Device Name: BioPlex™ 2200 ToRC IgG Regulation Number: 21 CFR 866.3510 Regulation Name: Rubella virus serological reagents Regulatory Class: Class II Product Code: OMI, JIX, JJY Dated: February 18, 2009 Received: February 19, 2009

Dear Ms. Sinclair:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your

{18}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally attayma

Sally A. Hoivat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{19}------------------------------------------------

Indications for Use

510(k) Number (if known): K080008

Device Name: BioPlex™ 2200 System ToRC IgG Kit on the BioPlex™ 2200 Multi-Analyte Detection System BioPlex™ 2200 ToRC IgG Calibrator Set BioPlex™ 2200 ToRC IgG Control Set

Indications for Use:

BioPlex™ 2200 ToRC IgG Kit

The BioPlex 2200 ToRC IgG kit is a multiplex flow immunoassay intended for the quantitative detection of IgG antibodies to Toxoplasma gondii (T. gondii) and Rubella, and the qualitative detection of IgG antibodies to Cytomegalovirus (CMV) in human serum and EDTA or heparinized plasma.

The ToRC IgG kit is intended for use with the Bio-Rad BioPlex 2200 System.

This kit is intended as an aid in the determination of serological status to T. gondii, Rubella and CMV. This kit is not intended for use in screening blood or plasma donors.

Performance characteristics for T. gondii and Rubella have not been evaluated in immunocompromised or immunosuppressed individuals. Performance characteristics for CMV have not been evaluated in immunosuppressed or organ transplant individuals. Performance characteristics of this kit have not been established for use in neonatal screening or for use at a point of care.

BioPlex™ 2200 ToRC IqG Calibrator Set

The BioPlex 2200 ToRC IgG Calibrator Set is intended for the calibration of the BioPlex 2200 ToRC IgG Reagent Pack.

BioPlex™ 2200 ToRC IqG Control Set

The BioPlex 2200 ToRC IgG Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex 2200 Instrument and BioPlex ToRG IgG Reagent Pack in the clinical laboratory. The performance of the BioPlex 2200 ToRC IgG Control Set has not been established with any other Toxoplasma gondii, Rubella or Cytomegalovirus (CMV) IgG antibody assays.

Prescription Use × (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

{20}------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Um Schaf

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

2080008 510(k)

§ 866.3510 Rubella virus serological reagents.

(a)
Identification. Rubella virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rubella virus in serum. The identification aids in the diagnosis of rubella (German measles) or confirmation of a person's immune status from past infections or immunizations and provides epidemiological information on German measles. Newborns infected in the uterus with rubella virus may be born with multiple congenital defects (rubella syndrome).(b)
Classification. Class II. The special controls for this device are:(1) National Committee for Clinical Laboratory Standards':
(i) 1/LA6 “Detection and Quantitation of Rubella IgG Antibody: Evaluation and Performance Criteria for Multiple Component Test Products, Speciment Handling, and Use of the Test Products in the Clinical Laboratory, October 1997,”
(ii) 1/LA18 “Specifications for Immunological Testing for Infectious Diseases, December 1994,”
(iii) D13 “Agglutination Characteristics, Methodology, Limitations, and Clinical Validation, October 1993,”
(iv) EP5 “Evaluation of Precision Performance of Clinical Chemistry Devices, February 1999,” and
(v) EP10 “Preliminary Evaluation of the Linearity of Quantitive Clinical Laboratory Methods, May 1998,”
(2) Centers for Disease Control's:
(i) Low Titer Rubella Standard,
(ii) Reference Panel of Well Characterized Rubella Sera, and
(3) World Health Organization's International Rubella Standard.