(63 days)
The Exactech Equinoxe™ Reverse Shoulder System is indicated to relieve pain and restore function in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe™ Reverse Shoulder System is also indicated for failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
The Equinoxe™ fracture adapter tray connects to the Equinoxe™ fracture humeral stems via a morse taper and a non-breakaway locking screw using a 5/16-18 thread instead of the primary adapter tray connecting to the primary/revision humeral stems via a spherical taper and a breakaway locking screw using a M6 thread
This document is a 510(k) premarket notification for a medical device called the Exactech® Equinoxe Reverse Shoulder Fracture Humeral Adapter Tray and Fracture Humeral Adapter Tray Locking Screw. The purpose of this notification is to demonstrate that the new device is substantially equivalent to legally marketed predicate devices, meaning it is as safe and effective.
The provided text does not contain a study that proves the device meets specific acceptance criteria in the format requested. Instead, it describes an engineering evaluation conducted to verify the device's performance. The acceptance criteria and "reported device performance" would typically be found within the detailed engineering evaluation report, which is not fully provided here.
However, based on the provided text, I can infer some aspects and highlight what is missing from the request.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Inferred) | Reported Device Performance (From Text) |
|---|---|
| Mechanical Integrity / Durability: The fracture adapter tray and locking screw must maintain their connection and intended function under anticipated in vivo use conditions. (Likely criteria would involve load-bearing capacity, fatigue resistance, and secure locking mechanism.) | "Engineering evaluations were conducted to verify that the performance of the proposed components would be adequate for anticipated in vivo use." This is a high-level statement and does not provide specific numerical performance metrics against detailed acceptance criteria. The device description mentions a "morse taper and a non-breakaway locking screw using a 5/16-18 thread" for the new device, compared to a "spherical taper and a breakaway locking screw using a M6 thread" for the predicate. This suggests different mechanical designs were evaluated to meet functional requirements. |
| Biocompatibility: The materials used, and the device itself, must not elicit an adverse biological response when implanted in the human body. | Not explicitly stated in this summary. However, being a Class II device, biocompatibility testing would be a standard requirement for implants. It's likely that the materials are identical or equivalent to the predicate device, or specific biocompatibility test results were part of the full submission, but not summarized here. |
| Functional Equivalence: The device should perform its intended function of connecting to humeral stems and assisting in reverse shoulder arthroplasty outcomes comparably to predicate devices. | "Based on successful results discussed in this submission, we conclude that the proposed devices are substantially equivalent to the previously cleared predicates." This statement indicates that functional equivalence was demonstrated, implying the engineering evaluations confirmed the new design performs adequately for its intended use, similar to the predicate. |
Missing Information (Not Available in the Provided Text for the requested items):
The provided text is a summary of a 510(k) submission, which focuses on demonstrating substantial equivalence to a predicate device rather than detailing a clinical study with human subjects, or a standalone algorithm-only performance study. Therefore, most of the requested information for AI/imaging device studies is not applicable or not present in this type of document.
Below is a breakdown of why each point is missing or not applicable in this context:
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable / Not Provided: This device is a mechanical implant (shoulder prosthesis components), not an AI algorithm or a diagnostic imaging device. The "test set" for this type of device typically refers to samples or prototypes used in mechanical engineering evaluations (e.g., fatigue testing, tensile testing) rather than a dataset of patient images or clinical outcomes. The document refers to "engineering evaluations" but does not specify the number of samples tested or their provenance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable / Not Provided: Ground truth, in the context of expert consensus, is typically established for diagnostic devices or AI algorithms where interpretation of data (e.g., images) by human experts is key. For mechanical implants, "ground truth" relates to physical properties and performance measured against engineering standards, not expert interpretation. The "experts" involved would be engineering and materials specialists designing and testing the device, and potentially orthopedic surgeons providing clinical input on requirements, but not in the sense of adjudicating a "test set" as described in the request.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable: Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert interpretations (e.g., radiologists' readings). This is not relevant for the engineering evaluation of a mechanical implant.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable: MRMC studies are specific to diagnostic devices, particularly in imaging, to evaluate the performance of human readers, sometimes with AI assistance. This is a mechanical implant; therefore, an MRMC study is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable: This device is a physical implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Engineering Standards / Biomechanical Principles: For this type of device, the "ground truth" for performance would be based on established engineering standards (e.g., ISO or ASTM for orthopaedic implants), biomechanical principles derived from cadaveric studies or in-silico models, and clinical requirements for long-term implant function and durability. The summary states "Engineering evaluations were conducted to verify that the performance of the proposed components would be adequate for anticipated in vivo use." This implies performance was measured against such standards.
8. The sample size for the training set
- Not Applicable: As this is a mechanical implant and not an AI algorithm, there is no "training set" in the conventional sense of machine learning. The design and development of the device would involve iterative design, prototyping, and testing (the "engineering evaluations"), but not a data-driven training process.
9. How the ground truth for the training set was established
- Not Applicable: Since there is no training set, this question is not applicable. The design "ground truth" would be established through clinical need assessment, biomechanical analysis, material science principles, and existing predicate device performance.
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized human figure in profile, with three faces overlapping each other. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular pattern around the figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2016
Exactech, Incorporated Mr. Graham Cuthbert Regulatory Affairs Specialist 2320 NW 66th Court Gainesville, Florida 32653
Re: K073688
Trade/Device Name: Equinoxe Reverse Shoulder System Fracture Humeral Adapter Tray and Fracture Humeral Adapter Tray Locking Screw Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, KWT Dated: January 31, 2008 Received: February 1, 2008
Dear Mr. Cuthbert:
This letter corrects our substantially equivalent letter of February 29, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Summary of Exactech® Equinoxe Reverse Shoulder Fracture Humeral Adapter Tray and Fracture Humeral Adapter Locking Screw Special 510(k) - Indications for Use
510(k) Number: K073688
Device Name: Equinoxe Reverse Shoulder System Fracture Humeral Adapter Tray and Fracture Humeral Adapter Tray Locking Screw
INDICATIONS FOR USE:
The Exactech Equinoxe™ Reverse Shoulder System is indicated to relieve pain and restore ിറ്റ് Exation Equinox falled glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
Prescription Use × and/or (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sabare Buchind forxym
Division Sign-Off
Division of General, Restorative, and Neurological Devices
510(k) Number K073688
Section 5 Page 1 of 1
{3}------------------------------------------------
·
:
Summary of Exactech® Equinoxe Reverse Shoulder Fracture Humeral Adapter Tray and Fracture Humeral Adapter Locking Screw Special 510(k) - 510(k) Summary
| Sponsor: | Exactech® Inc. |
|---|---|
| 2320 N.W. 66th Court | |
| Gainesville, Florida 32653 |
| Phone: | (352) - 377 - 1140 |
|---|---|
| Fax: | (352) - 378 - 2617 |
FDA Establishment Number 1038671 ·
Graham Cuthbert Contact: Regulatory Representative
December 17, 2007 Date:
Section 9 Page 1 of 2
{4}------------------------------------------------
Summary of Exactech® Equinoxe Reverse Shoulder Fracture Humeral Adapter Tray and Fracture Humeral Adapter Locking Screw Special 510(k) - 510(k) Summary
Trade or proprietary or model name(s):
Equinoxe Reverse Shoulder Fracture Humeral Adapter Tray and Fracture Humeral Adapter Locking Screw
Common Name
Shoulder Joint Prosthesis
Classification name
Shoulder joint metal/polymer non-constrained cemented prosthesis (21 CFR 888.3650, Class II, Product Code KWT)
Prosthesis, Shoulder, Semi-constrained, metal/polymer cemented (21 CFR 888.3660, Class II, Product Code KWS)
Information on devices to which substantial equivalence is claimed:
| 510(k) Number | Trade or Proprietary or Model Name | Manufacturer |
|---|---|---|
| K063569 | Exactech EquinoxeTM Reverse Shoulder System | Exactech, Inc. |
Indications for Use:
The Exactech Equinoxe " Reverse Shoulder System is indicated to relieve pain and restore function in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe™ Reverse Shoulder System is also indicated for failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.
Device Description:
The Equinoxe™ fracture adapter tray connects to the Equinoxe™ fracture humeral stems via a morse taper and a non-breakaway locking screw using a 5/16-18 thread instead of the primary adapter tray connecting to the primary/revision humeral stems via a spherical taper and a breakaway locking screw using a M6 thread
Substantial Equivalency Conclusion:
Engineering evaluations were conducted to verify that the performance of the proposed components would be adequate for anticipated in vivo use. Based on successful results discussed in this submission, we conclude that the proposed devices are substantially equivalent to the previously cleared predicates.
Section 9 Page 2 of 2
N/A