K Number
K073573

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-01-30

(41 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AgaMatrix WaveSense™ KeyNote Codeless Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

Device Description

The AgaMatrix WaveSense™ KeyNote Codeless Blood Glucose Monitoring System includes a meter with batteries, compact carrying case, lancing device, lancets, control solution and owner's booklet. Test Strips are sold separately. The meter is a portable, battery-operated instrument.

AI/ML Overview

The provided text describes the AgaMatrix WaveSense™ KeyNote Codeless Blood Glucose Monitoring System, which states that it complies with ISO 15197:2003. This standard specifies the requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. While the summary states compliance, it does not explicitly provide a table of acceptance criteria or reported device performance against specific targets from ISO 15197. Therefore, the following information is based on the general understanding of ISO 15197 for blood glucose meters, as the specific performance data is not detailed in the provided document.

1. Table of Acceptance Criteria and Reported Device Performance

As specific performance data against ISO 15197:2003 criteria is not provided in the document, a general representation based on the standard's requirements for accuracy is used.

Performance CharacteristicAcceptance Criteria (from ISO 15197:2003 for glucose values ≥ 4.2 mmol/L (75 mg/dL) and < 4.2 mmol/L (75 mg/dL))Reported Device Performance
Accuracy (System)For glucose values ≥ 4.2 mmol/L (75 mg/dL):- 95% of all measured glucose values shall fall within ±20% of the YSI reference measurement.For glucose values < 4.2 mmol/L (75 mg/dL):- 95% of all measured glucose values shall fall within ± 0.83 mmol/L (± 15 mg/dL) of the YSI reference measurement.Stated to comply with ISO 15197:2003. (Specific data not provided in the document)
Precision (Repeatability)Coefficient of variation (CV) ≤ 5% for glucose concentrations ≥ 4.2 mmol/L (75 mg/dL), and standard deviation (SD) ≤ 0.28 mmol/L (5 mg/dL) for glucose concentrations < 4.2 mmol/L (75 mg/dL).Stated to comply with ISO 15197:2003. (Specific data not provided in the document)

2. Sample Size Used for the Test Set and Data Provenance

The document states compliance with ISO 15197:2003, which typically requires a minimum number of subjects and blood samples for accuracy evaluation. However, the exact sample size used for the test set in the study confirming compliance and the data provenance (e.g., country of origin, retrospective or prospective) are not specified in the provided text.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not provide information on the number of experts used or their qualifications to establish ground truth for the test set. For blood glucose monitoring systems, ground truth is typically established by a laboratory reference method (e.g., YSI analyzer) operated by trained laboratory personnel, rather than expert clinicians.

4. Adjudication Method for the Test Set

Not applicable. For blood glucose meter accuracy studies, ground truth is typically a reference laboratory measurement, not subject to clinical adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This is a blood glucose monitoring system, which does not involve human readers interpreting medical images or data that would typically be evaluated in an MRMC study with AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device is a standalone blood glucose monitoring system. Its performance is evaluated intrinsically, meaning the system's output (blood glucose reading) is directly compared to a reference method, without human intervention as an interpretive step.

7. The Type of Ground Truth Used

Based on the nature of blood glucose monitoring systems and the reference to ISO 15197, the ground truth used for evaluating device performance would be laboratory reference measurements, typically from a highly accurate and precise instrument like a YSI glucose analyzer, applied to the same blood samples as tested by the device.

8. The Sample Size for the Training Set

The document does not provide information about a "training set" or its sample size. Blood glucose meters are typically calibrated during manufacturing rather than "trained" with data in the way an AI algorithm would be. The design and manufacturing process would be validated against established standards.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as a "training set" in the context of machine learning for an algorithm is not relevant to this type of device. Calibration and validation for blood glucose meters rely on precisely prepared glucose solutions and comparison to reference methods.

{0}------------------------------------------------

K073573

10 Manor Parkway Salem, NH 03079 Office: +1 (603) 328-6000 Fax: +1 (603) 893-4191

JAN 30 2008

510 (k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: unknown

    1. Submitter's Name, Address, Telephone Number, Contact Person, and date the summary was prepared.
Submitter's Name:AgaMatrix, Inc.10 Manor ParkwaySalem, NH 03079
Contact Person:Connie HertelDirector Quality & Regulatory Affairs
Telephone:(603) 328-6051
Fax:(603) 893-4191

Date the summary prepared: December 6, 2007

2. Device Name

Trade/Propritary Name:WaveSense™ KeyNote Codeless Blood Glucose
Monitoring System
Common/Usual Name:Blood Glucose Monitoring System
Classification Name:Glucose Test System (per 21 CFR 862.1345)
Class:II
Panel:Chemistry

3. Modification Device

The modification of the WaveSense Jazz TM Blood Glucose Monitoring System (Codeless) excludes two User Interface features: meal time tagging and positive feedback (smiley face). In addition, the meter's shell has been modified.

4. Description of the Device:

The AgaMatrix WaveSense™ KeyNote Codeless Blood Glucose Monitoring System includes a meter with batteries, compact carrying case, lancing device, lancets, control solution and owner's booklet. Test Strips are sold separately.

AgaMatrix

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for AgaMatrix. The logo consists of the word "AgaMatrix" in a simple, sans-serif font. A curved line underlines the text. To the right of the word, there are four small squares arranged in a curved line, suggesting a matrix or a connection of elements.

10 Manor Parkway Salem, NH 03079 Office: +1 (603) 328-6000 Fax: +1 (603) 893-4191

It is intended for over-the counter home use by diabetics to monitor their blood glucose levels, or for use in a clinical setting by healthcare professionals. The system tests fresh capillary whole blood. The meter is a portable, battery-operated instrument.

5. Intended use of the device:

The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System is intended to quantitatively measure blood glucose levels, also known as blood sugar, from fresh capillary whole blood samples taken from the fingertips, palm, or forearm. The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System test strips are for in vitro diagnostic (outside of the body) use only. The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System is not intended for use with neonates.

Testing:

The manufacturer of the WaveSense™ KeyNote Codeless Blood Glucose Monitoring System certifies that the device complies with the following:

ISO 15197:2003 In vitro diagnostic test systems - Requirements for bloodglucose monitoring systems for self-testing in managing diabetes mellitus

ISO 14971:2000 Medical devices – Application of risk management to Medical devices

IEC 61010-1 Medical electrical equipment – General·requirements for safety

IEC 61010-2-101 Safety requirements for electrical equipment for measurement, control and laboratory use - particular requirements for in vitro diagnostic (IVD) medical equipment

IEC 61000-4-3 Electromagnetic compatibility (EMC)

8. Conclusions

Based upon the testing and comparison to the predicate device, the WaveSense™ KeyNote Codeless Blood Glucose Monitoring System Blood Glucose Monitoring System has the same intended uses, with similar technological characteristics as the predicate device. The system performs as intended and raises no new safety or Effectiveness issues.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and three stripes extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

JAN 30 2008

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AgaMatrix, Inc. c/o Ms. Connie Hertel Director Quality & Regulatory Affairs 10 Manor Parkway Salem, NH 03079

Re: K073573

Trade Name: WaveSense™ Keynote Codeless Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Monitoring System Regulatory Class: Class II Product Codes: NBW, CGA Dated: January 16, 2008 Received: January 18, 2008

Dear Ms. Hertel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean M. Cooper, M.S., D.V.M.

Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indication for Use

510(k) Number (if known): K073573

Device Name: AgaMatrix WaveSense™ KeyNote Codeless Blood Glucose Monitoring System

Indication For Use:

AgaMatrix WaveSense™ KeyNote Codeless Blood Glucose Monitoring System: AgaMatrix WaveSense KeyNote Codeless Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use X. (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Carl C. Benson

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K073573

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.