K Number
K073405

Validate with FDA (Live)

Device Name
PLEXUR M
Manufacturer
Date Cleared
2008-03-03

(90 days)

Product Code
Regulation Number
888.3045
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PLEXUR M is intended for use in filling bony voids or gaps of the skeletal system (i.e., extremities, pelvis) that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PLEXUR M is resorbed/remodeled and is replaced by host bone during the healing process.

Device Description

PLEXUR M is a bone void filler product that contains, as a key constituent, processed human bone particles that are mixed with resorbable/biodegradeable non-tissue components. It is intended to be heated at the time of use, whereupon it becomes moldable, thus allowing the surgeon to pack it into the implant site or shape it to accommodate variations in the geometry and size of the particular implant site. As it cools down, the PLEXUR M returns to its normal hardened/rigid state and remains this way at body temperature. The surgeon may further shape, cut or grind PLEXUR M in the hardened state using conventional surgical instruments. PLEXUR M is packaged/provided for single use in a sterile form. PLEXUR M is resorbed/remodeled and is replaced by host bone during the healing process. A compact, single use heater is available as a means to heat the PLEXUR M at the time of use to make it pliable. Osteotech is seeking approval of this single use heater as part of this 510(k) submission. Alternatively, PLEXUR M may be heated in a water bath.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called PLEXUR M, a bone void filler. It details the device description, intended use, and its substantial equivalence to predicate devices, but it does not contain information about acceptance criteria or a study proving the device meets specific performance metrics in the way a diagnostic AI/ML device would.

The "Performance Data" section states: "The results of studies in animals showed that PLEXUR M supports bone ingrowth and new bone formation." This is a general statement about the biological performance of the device rather than a quantitative assessment against predefined acceptance criteria.

Therefore, the requested information cannot be fully provided based on the given text.

Here's an attempt to answer based on the available information, with many points indicating that the information is not present in the document:


Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Not specified in the document for PLEXUR M as a diagnostic or analytical device."results of studies in animals showed that PLEXUR M supports bone ingrowth and new bone formation."
No quantitative performance metrics (e.g., sensitivity, specificity, accuracy, precision, etc.) are provided for human use.No quantitative performance data for human use provided.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not applicable/Not specified. The document refers to "studies in animals" but does not give a sample size, and it's for biological performance, not a diagnostic test set.
  • Data Provenance: Studies were conducted "in animals." No country of origin is specified. The studies appear to be for biological evaluation, not clinical efficacy in humans as a "test set" in the context of AI/ML evaluation. Such studies are typically prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Not applicable. The "ground truth" for bone ingrowth and new bone formation in animal studies would typically be established through histological examination or other biological assessments by veterinary pathologists or other relevant scientists, but this information is not detailed.

4. Adjudication Method for the Test Set

  • Not applicable. This concept is relevant for human-read data or diagnostic interpretations, which is not the focus of the performance data presented here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • No. This document does not mention any MRMC study. The device is a bone void filler, not an imaging or diagnostic device that typically undergoes MRMC studies.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

  • No. The device is a physical bone void filler, not an algorithm or AI/ML system.

7. The Type of Ground Truth Used

  • The "ground truth" for the animal studies would be related to biological outcomes such as histological evidence of bone ingrowth and new bone formation.

8. The Sample Size for the Training Set

  • Not applicable. This device is not an AI/ML algorithm requiring a training set.

9. How the Ground Truth for the Training Set was Established

  • Not applicable. This device is not an AI/ML algorithm requiring a training set.

Summary of what the document does provide:

  • Device Name: PLEXUR M
  • Intended Use: For filling bony voids or gaps of the skeletal system (extremities, pelvis) that are not intrinsic to the stability of the bony structure.
  • Device Description: Contains processed human bone particles mixed with resorbable/biodegradable non-tissue components, becomes moldable when heated, then hardens.
  • Performance Data (limited): Animal studies showed it "supports bone ingrowth and new bone formation."
  • Viral Inactivation: Processing steps for allograft bone inactivate viruses (HIV, hepatitis B and C, CMV).
  • Substantial Equivalence: Based on materials (human allograft bone, resorbable polymer) and heat-moldable characteristics compared to predicate devices.

{0}------------------------------------------------

510(K) Pre Market Application

510(k) Summary 5.0

MAR - 3 2008

The following information is provided as required by 21 CFR § 807.87 for Osteotech's PLEXUR M 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.

The safety and effectiveness of PLEXUR M is based upon a determination of the substantial equivalence as well as the safety and effectiveness of its predicate device(s) .

Sponsor/ManufacturerOsteotech, Inc.51 James WayEatontown, NJ 07724
ContactChris TalbotOsteotech, Inc.51 James WayEatontown, NJ 07724Ph: (732) 542-2800Talbot@osteotech.com
Date PreparedNovember 30, 2007
Proposed ClassII
Proprietary NamePLEXUR M
Common NameBone void filler
Classification NameResorbable calcium salt bone void filler device
Regulation Number21 CFR 888.3045
Product CodeMQV
Predicate DevicePLEXUR M is substantially equivalent, for the purpose of this510(k), to other devices that have received 510(k) clearance for similar indications foruse.

PLEXUR M is intended for use in filling bony voids or gaps of the Intended Use skeletal system (i.e., extremities, pelvis) that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects

{1}------------------------------------------------

created from traumatic injury to the bone. PLEXUR M is resorbed/remodeled and is replaced by host bone during the healing process.

Device Description

PLEXUR M is a bone void filler product that contains, as a key constituent, processed human bone particles that are mixed with resorbable/biodegradeable non-tissue components. It is intended to be heated at the time of use, whereupon it becomes moldable, thus allowing the surgeon to pack it into the implant site or shape it to accommodate variations in the geometry and size of the particular implant site. As it cools down, the PLEXUR M returns to its normal hardened/rigid state and remains this way at body temperature. The surgeon may further shape, cut or grind PLEXUR M in the hardened state using conventional surgical instruments.

PLEXUR M is packaged/provided for single use in a sterile form. PLEXUR M is resorbed/remodeled and is replaced by host bone during the healing process. A compact, single use heater is available as a means to heat the PLEXUR M at the time of use to make it pliable. Osteotech is seeking approval of this single use heater as part of this 510(k) submission. Alternatively, PLEXUR M may be heated in a water bath.

Performance Data

The results of studies in animals showed that PLEXUR M supports bone ingrowth and new bone formation.

Viral Inactivation

In the production of PLEXUR M, the allograft bone is subjected to processing steps that have been shown to inactivate viruses, including HIV, hepatitis B and C and CMV.

Technical Comparison

PLEXUR M is substantially equivalent to one or more of the predicate devices with respect to materials. PLEXUR M contains human allograft bone tissue, as does one or more of the predicate devices. PLEXUR M also contains resorbable polymer of the same

{2}------------------------------------------------

type as those in one or more of the predicate devices. Also, like one or more of the predicate devices, PLEXUR M is provided sterile in various sizes that, upon heating (using a compact, single use sterile heater to be marketed by Osteotech), is made moldable and can be cut or shaped by the user into various shapes or sizes. It may be further shaped by the surgeon in its hardened state using conventional surgical instruments.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Osteotech, Inc. % Mr. Christopher Talbot Director of Regulatory Affairs 51 James Way Eatontown, New Jersey 07724

MAR - 3 2008

Re: K073405

Trade/Device Name: PLEXUR M Bone Void Filler Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler devices Regulatory Class: Class II Product Code: MOV Dated: November 30, 2007 Received: December 4, 2007

Dear Mr. Talbot:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 – Mr. Christopher Talbot

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement 4.0

K073405 510(k) Number:

Device Name: PLEXUR M

PLEXUR M is intended for use in filling bony voids or gaps of the Indications for Use: skeletal system (i.e., extremities, pelvis) that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PLEXUR M is resorbed/remodeled and is replaced by host bone during the healing process.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE

Mark A. Mckerson

(Division Sign-Of al. Pestorative, Division of Genera and Neurological Dences

510(k) Number

Page 1 of _1

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.