K Number
K073288
Date Cleared
2008-06-12

(204 days)

Product Code
Regulation Number
876.4300
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Articulating Snare is intended for the electrosurgical removal and cauterization of gastrointestinal tract polyps through an endoscope.

Device Description

The Ethicon Endo Surgery Articulating Snare consists of a flexible wire cable and loop, which can be extended, rotated, articulated, and retracted from the flexible outer shaft using a three-finger actuator. It is passed through endoscopes having a 3.2 mm or larger working channels. When activated, the snare delivers a monopolar electrical current to cut and cauterize tissue within the loop. The device is supplied sterile for single-patient use.

AI/ML Overview

The provided 510(k) summary for the Ethicon Endo Surgery Articulating Snare describes bench testing as the performance data used to demonstrate that the device performs as intended. It does not include information about clinical studies with human subjects, AI algorithms, or ground truth establishment relevant to those types of studies.

Therefore, many of the requested categories are not applicable to the information contained in the document.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard Compliance)Reported Device Performance
AAMI HF 18, 2001: Electrosurgical DevicesDevice was tested to demonstrate compliance and performs as intended.
IEC/EN 60601-2-2, 2000: Particular Requirements for the Safety of Endoscopic EquipmentDevice was tested to demonstrate compliance and performs as intended.
IEC/EN 60601-2-18, 1996: Particular Requirements for the Safety of High Frequency of Surgical EquipmentDevice was tested to demonstrate compliance and performs as intended.
ISO 10993-1 (Biocompatibility)Device materials have been evaluated for biocompatibility and comply.

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not applicable. The performance data is based on bench testing of the device itself, not a test set of data from human subjects.
  • Data Provenance: Not applicable, as this refers to bench testing of the device, not patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. Ground truth for a test set is not relevant for bench testing of device compliance with standards.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not applicable. Adjudication methods are not relevant for bench testing of device compliance with standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This is a medical device (snare) and not an AI-powered diagnostic or assistive tool, so an MRMC study is not relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • No. This is a medical device (snare) and not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The "ground truth" for the bench testing was compliance with established industry standards (AAMI HF 18, IEC/EN 60601-2-2, IEC/EN 60601-2-18, and ISO 10993-1).

8. The sample size for the training set

  • Not applicable. This document describes a physical medical device, not a machine learning model requiring a training set.

9. How the ground truth for the training set was established

  • Not applicable. This document describes a physical medical device.

{0}------------------------------------------------

K073288
page 1 of 2

510(k) Summary

JUN 1 2 2008

Company Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, OH 45242

Contact Glenda C Marsh QS/RA Sr. Project Manager Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, OH 45242 Telephone: (513) 337-7659 (513) 337-2860 Fax: Email: gmarsh(@eesus.jnj.com

Date Prepared November 19, 2007

  • New Device Name Trade Name: Ethicon Endo Surgery Articulating Snare Common or Usual Name: Snare, Flexible Classification Name: Endoscopic electrosurgical unit and accessories (21 CFR 876.4300, Product Code FDI)
    Predicate Devices Rotatable Snare (K992477) Single-Use Polypectomy Snare (K941750)

Device Description The Ethicon Endo Surgery Articulating Snare consists of a flexible wire cable and loop, which can be extended, rotated, articulated, and retracted from the flexible outer shaft using a three-finger actuator. It is passed through endoscopes having a 3.2 mm or larger working channels. When activated, the snare delivers a monopolar electrical current to cut and cauterize tissue within the loop. The device is supplied sterile for single-patient use.

Indications for Use The Articulating Snare is intended for the electrosurgical removal and cauterization of gastrointestinal tract polyps through an endoscope.

Technological Characteristics The EES device has similar technologic characteristics to the predicate devices in that it consists of a flexible wire cable and loop that is used to remove polyps utilizing monopolar RF energy under endoscopic visualization. In all devices, the operator can deploy and retract the snare by using three fingers. As in the predicate devices, the EES device features rotation of the end-effector by manipulation of a rotation knob. In addition, the EES device features articulation of the end-effector to provide the clinician with improved tissue targeting capability.

{1}------------------------------------------------

K073288
Page 2 of 2

Performance Data. Bench testing was performed to demonstrate that the EES device performs as intended. The device materials have been evaluated for biocompatibility and comply with the requirements of ISO 10993-1. The device was tested to demonstrate compliance with the following standards:

  • AAMI HF 18, 2001 : Electrosurgical Devices

  • IEC/EN 60601-2-2, 2000 : Particular Requirements for the Safety of Endoscopic Equipment

  • IEC/EN 60601-2-18, 1996 : Particular Requirements for the Safety of High Frequency of Surgical Equipment

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

JUN 12 2008

Ms. Glenda Marsh Senior Project Manager, Quality Systems & Regulatory Affairs Ethicon Endo-Surgery, Incorporated 4545 Creek Road CINCINNATI OH 45242

Re: K073288

Trade/Device Name: Ethicon Endo Surgery® Articulating Snare Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: FDI Dated: May 29, 2008 Received: May 29, 2008

Dear Ms. Marsh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image shows a circular logo with the text "1906-2006" at the top. Below the text is a stylized "FDA" abbreviation. The word "Centennial" is written below the abbreviation. Three stars are at the bottom of the logo.

Protecting and Promoting Public Health

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You max obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Nancy C. Burgdon

Nancy C. Brogdon . Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): 13288

Device Name:

Ethicon Endo Surgery® Articulating Snare

Indications for Use:

The Articulating Snare is intended for the electrosurgical removal and cauterization of gastrointestinal tract polyps through an endoscope.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Posted November 13, 2003)

(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
510(k) NumberK073288

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).