K Number
K073274

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-03-17

(117 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Spinal Sphere System is intended to be inserted between the vertebral bodies into the disc space from L3 to S1 to help provide stabilization and to help promote intervertebral body fusion. Properly used, this internal fixation device is intended and designed solely for holding bone parts in alignment while they heal. The Spinal Sphere System is intended to be used with bone graft. The indications and contraindications of spinal instrumentation systems should be well understood by the surgeon. It is not intended for use in motion-sparing, non-fusion procedures.

Device Description

The Spinal Sphere System consists of spheres manufactured from either CoCrMo or PEEK-OPTIMA. The device may be implanted from L3 to S1 to provide temporary stabilization in order to help promote fusion. The spheres are available in a variety of sizes.

AI/ML Overview

I apologize, but I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them based on the provided text.

The provided documents are a 510(k) summary and an FDA clearance letter for the "Spinal Sphere System" by Life Spine. These documents relate to the regulatory clearance of a medical device, specifically demonstrating substantial equivalence to a predicate device.

They do not contain information about:

  1. Acceptance criteria in terms of specific performance metrics for a software or AI device.
  2. A study proving device meets acceptance criteria with details such as sample sizes, ground truth establishment, expert qualifications, or comparative effectiveness studies.
  3. Performance metrics (e.g., sensitivity, specificity, accuracy) that would populate a table of acceptance criteria and reported performance for an AI/software device.

The context of the provided text is for a physical orthopedic implant (spinal spheres), not a diagnostic AI/software device. The FDA clearance is based on demonstrating the new device is "substantially equivalent" to an existing, legally marketed device, primarily in terms of intended use, design, function, and materials, rather than a performance study against specific AI-related acceptance criteria.

{0}------------------------------------------------

A final rule classifying spinal spheres for use in intervertebral fusion procedures, a preamendments, unclassified device type, into class III (premarket approval), was published on March 30, 2023. See here: https://www.federalregister.gov/documents/2023/03/30/2023-06566/medical-devices-orthopedic-devices-classification-of-spinalspheres-for-use-in-intervertebral-fusion. Concurrently, FDA published a final order to require the filing of a premarket approval application (PMA) for spinal spheres for use in intervertebral fusion procedures. See here: https://www.federalregister.gov/documents/2023/03/30/2023-06565/effective-date-of-requirement-for-premarket-approvalapplications-for-spinal-spheres-for-use-in.

Please refer to the aforementioned final rule and final order for current regulatory requirements for this device type.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Life Spine. The words "Life Spine" are written in a bold, sans-serif font. A curved line is located underneath the words. The logo is black and white.

073274

510(k) Summary

MAR 1 7 2008

510(k) Summary - Spinal Sphere System
Preparation Date:November 20, 2007
Submitted By:Life Spine2401 W. Hassell Road, Suite 1535Hoffman Estates, IL 60169Telephone: 847-884-6117Fax: 847-884-6118
Contact:Rebecca M. BrooksLife Spine2401 W. Hassell Road, Suite 1535Hoffman Estates, IL 60169Telephone: 847-884-6117Fax: 847-884-6118
Proprietary Trade Name:Spinal Sphere System
Classification Name:Intervertebral Fusion Device With Bone Graft, Solid-Sphere, Lumbar
Product Code:NVR
Predicate Device:Satellite™ Spinal System (Medtronic Sofamor Danek,K051320 and K060415)

Device Description:

The Spinal Sphere System consists of spheres manufactured from either CoCrMo or PEEK-OPTIMA. The device may be implanted from L3 to S1 to provide temporary stabilization in order to help promote fusion. The spheres are available in a variety of sizes.

Intended Use of the Device:

The Spinal Sphere System is intended to be inserted between the vertebral bodies into the disc space from L3 to S1 to help provide stabilization and to help promote intervertebral body fusion. Properly used, this internal fixation device is intended and designed solely for holding bone parts in alignment while they heal. The Spinal Sphere System is intended to be used with bone graft. The indications and contraindications of spinal instrumentation systems should be well understood by the surgeon. It is not intended for use in motion-sparing, non-fusion procedures.

Substantial Equivalence:

The Spinal Sphere System was shown to be substantially equivalent to previously cleared devices in indications for use, design, function, and materials used.

Image /page/1/Picture/11 description: The image contains a handwritten text. The text appears to be cursive and slightly faded. The text is "Paz 1.st."

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern around the symbol. The text is in all caps and is evenly spaced around the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Life Spine % Rebecca M. Brooks Project Coordinator 2401 W. Hassell Road, Suite 1535 Hoffman Estates, IL 60169

MAR 1 7 2008

K073274 Trade/Device Name: Spinal Sphere System Regulation Number: Pre-amendment Regulation Name: N/A Regulatory Class: Unclassified Product Code: NVR Dated: February 28, 2008 Received: February 29, 2008

Dear Ms. Brooks:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must be prominently displayed in all labeling, including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print:

The safety and effectiveness of this device for use in motion sparing, non-fusion procedures has not been established.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

{3}------------------------------------------------

Page 2 - Ms. Rebecca M. Brooks

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

VCGILL

Donna-Bea Tillman, Ph.D., M.P.A. Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Life Spine. The words "Life Spine" are written in a bold, sans-serif font. A curved line is underneath the words. The registered trademark symbol is to the right of the word Spine.

Indications for Use

510(k) Number (if known): __ 0732 74

Device Name: Spinal Sphere System

The Spinal Sphere System is intended to be inserted between the vertebral bodies into the disc space from L3 to S1 to help provide stabilization and to help promote intervertebral body fusion. Properly used, this internal fixation device is intended and designed solely for holding bone parts in alignment while they heal. The Spinal Sphere System is intended to be used with bone graft. The indications and contraindications of spinal instrumentation systems should be well understood by the surgeon. It is not intended for use in motion-sparing, non-fusion procedures.

Prescription Use x (Part 21 CFR 801 Subpart D)

And/Or

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R.P. Oyler fix mem

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K073274

N/A