K Number
K073196

Validate with FDA (Live)

Device Name
MASTER-VU
Manufacturer
Date Cleared
2007-12-19

(36 days)

Product Code
Regulation Number
892.1560
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Master-Vu High Resolution Ultrasound System is intended to be used for visualization by ultrasound of the eye and orbit by B-scan.

Device Description

The Master-Vu High Resolution Ultrasound System is intended to be used for visualization by ultrasound of the eye and orbit by B-scan. The system is PC-based, and is used with a 12 MHz transducer. The computer is a standard pc, and is user-supplied. Computer requirements are supplied in the instructions. The scanner is motor-driven. The system can also make measurements. The Master-Vu is a conventional ophthalmic B-scan system using a motor-driven transducer and angle sensor for scanning. It uses a motor-driven 12 MHz transducer. The system is PC-based, and the display is on the computer screen.

AI/ML Overview

The provided document is a 510(k) summary for the Sonomed Master-Vu Ophthalmic Ultrasound System. It describes the device, its intended use, and claims substantial equivalence to a predicate device. However, it explicitly states that no clinical tests were required for this submission. Therefore, it does not contain the detailed information necessary to answer most of your questions about acceptance criteria, study design, and performance data from a clinical trial.

Here's a breakdown of what can be answered and what cannot:

1. A table of acceptance criteria and the reported device performance

  • Cannot be provided. The document states, "(2) Clinical tests Not required". Therefore, there are no reported clinical performance data or acceptance criteria for such, as they were not evaluated for this 510(k) submission. The submission focuses on non-clinical tests (ultrasonic emissions and accuracy/validation tests) to demonstrate substantial equivalence to the predicate device.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Cannot be provided. No clinical test set was used or described.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Cannot be provided. No clinical test set requiring expert ground truth was performed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Cannot be provided. No clinical test set requiring adjudication was performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Cannot be provided. No MRMC study was done, nor is there any mention of AI assistance in this document. This is a conventional ultrasonic B-scan system.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Cannot be provided. This device is a conventional ultrasound system and does not appear to involve an "algorithm only" component in the way a diagnostic AI would.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Cannot be provided. No clinical ground truth was established as clinical tests were not required. The "ground truth" for non-clinical tests would involve physical measurements and technical specifications rather than clinical outcomes.

8. The sample size for the training set

  • Cannot be provided. The document does not describe any machine learning or AI components that would require a training set.

9. How the ground truth for the training set was established

  • Cannot be provided. No training set was involved.

Summary of available information related to acceptance criteria and study:

The Sonomed Master-Vu Ophthalmic Ultrasound System received 510(k) clearance based on demonstrating substantial equivalence to a predicate device (Sonomed EZ-Scan AB 5500+).

The "study" referenced in the document involved:

  • Non-clinical tests: Both ultrasonic emissions tests and accuracy and validation tests were done. The acceptance criteria for these tests would be compliance with recognized standards for medical ultrasound devices and performance meeting the specifications of the predicate device.
  • Clinical tests: Not required for this 510(k) application. This means the device met the regulatory requirements for safety and efficacy without needing to conduct a clinical trial demonstrating its performance on patient data against clinical acceptance criteria.

The conclusion was that the Master-Vu is "equivalent in safety and efficacy to the legally marketed predicate device" based on these non-clinical tests and technological characteristics.

{0}------------------------------------------------

K073196
page 1 of 2

Sonomed Inc. Special 510(k) Master-Vu Ophthalmic Ultrasound System

510(k) Summary October 24, 2007

(1) Submitter Information

Name: Sonomed Inc..

Address: 1979 Marcus Avenue Lake Success, NY 11042

Telephone Number: 516-354-0900

Contact Person: Dr. George Myers Medsys Inc. 377 Rt. 17 S Hasbrouck Heights, NJ 07604 201-727-1703

Date Prepared: October 23, 2007

(2) Name of Device:

Trade Name: Master-Vu Common Name: Opththalmic ultrasonic B scan system Classification Name: System, Imaging, Ultrasonic, Ophthalmic, 90IYO

  • (3) Equivalent legally-marketed devices:
    Sonomed EZ-Scan AB 5500+

(4) Description

The Master-Vu High Resolution Ultrasound System is intended to be used for visualization by ultrasound of the eye and orbit by B-scan.

The system is PC-based, and is used with a 12 MHz transducer . The computer is a standard pc, and is usersupplied. Computer requirements are supplied in the instructions. The scanner is motor-driven. The system can also make measurements.

(5) Intended Use

The Master-Vu High Resolution Ultrasound System is intended to be used for visualization by

DEC 1 9 2007

{1}------------------------------------------------

Sonomed Master-Vu

Image /page/1/Picture/2 description: The image shows the text "KO 73196 Page 2 of 2". The text appears to be handwritten. The text is black against a white background.

ultrasound of the eye and orbit by B-scan.

(6) Technological characteristics

The Master-Vu is a conventional ophthalmic B-scan system using a motor-driven transducer and angle sensor for scanning. It uses a motor-driven 12 MHz transducer. The system is PC-based, and the display is on the computer screen.

(b) Performance data

(1) Non-clinical tests

Both ultrasonic emissions tests and accuracy and validation tests have been done.

(2) Clinical tests

Not required

(3) Conclusions

The Sonomed Master-Vu is equivalent in safety and efficacy to the legally marketed predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the seal is a stylized image of three human profiles facing to the right, with flowing lines beneath them.

Public Health Service

MAR 1 1 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Sonomed, Incorporated % Dr. George Myers President Medsys, Incorporated 377 Route 17 S HASBROUCK HEIGHTS NJ 07604

Re: K073196

Trade/Device Name: Master-Vu Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO and ITX Dated: October 30, 2007 Received: November 13, 2007

Dear Dr. Myers:

This letter corrects our substantially equivalent letter of December 19, 2007.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Master-Vu Ultrasound System, as described in your premarket notification:

Transducer Model Number

12 MHz Transducer

{3}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

{4}------------------------------------------------

Page 3 – Dr. George Myers

If you have any questions regarding the content of this letter, please contact Ms. Lauren Hefner at (240) 276-3666.

Sincerely yours,

Hebert Semen for

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

Page

K073/96

510(k) Number (if known):

Device Name: Master-Vu

Intended Use:

The Master-Vu High Resolution Ultrasound System is intended to be used for visualization by ultrasound of the eye and orbit by B-scan.

CLINICALAPPLICATIONABMPWDCWDCOLORDOPPLERPOWER(AMPLITUDE)DOPPLERCOLORVELOCITYIMAGINGCOMBINED(SPECIFY)OTHER(SPECIFY)
OphthalmicPP (3D)
Fetal
Abdominal
Intra-operative(specify)
Intra-operative
Neurological
Pediatric
Small Organ(Specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Trans-esophageal
Trans-rectal
Trans-vaginal
Trans-urethral
Intra-luminal
Peripheral
Vascular
Laparoscopic
Musculo-Skeletal
Other (Specify)

N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription UseOROver-the-Counter Use
(Per 21 CFR 810.109)(Optional Format 1-2-96)

(Division Sign-Off)

Division of Reproductive, Abdominal,

and Radiological Devices

510(k) NumberK073196
------------------------

Mode of Operation

{6}------------------------------------------------

Sonomed Master-Vu

Diagnostic Ultrasound Indications for Use Form

Page1 of
510(k) Number (if known):K073196

Device Name: Master-Vu 12 MHz transducer

Intended Use:

The Master-Vu 12 MHz transducer is for use with the Master-Vu ultrasound system Mode of Operation

CLINICALAPPLICATIONABMPWDCWDCOLORDOPPLERPOWER(AMPLITUDE)DOPPLERCOLORVELOCITYIMAGINGCOMBINED(SPECIFY)OTHER(SPECIFY)
OphthalmicP
Fetal
Abdominal
Intra-operative(specify)
Intra-operativeNeurological
Pediatric
Small Organ(Specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Trans-esophageal
Trans-rectal
Trans-vaginal
Trans-urethral
Intra-luminal
Peripheral
Vascular
Laparoscopic
Musculo-Skeletal
Other (Specify)

N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments:

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 810.109)

OR

Over-the-Counter Use

(Optional Format 1-2-96)

Hubert Hunn

(Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Devices 510(k) Number.

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.