(73 days)
The MicroMRI Wrist Coll is a transmitting and receiving device to be used in conjunction with GE Healthcare and Siemens 1.5T MRI systems to image the wrist. The images acquired using the MicroMRI Wrist Coil along multiple axes (axial, coronal, sagittal, and oblique) are a measure of the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the spatial image are proton density, spin-lattice (TI) and spin-spin (T2) relaxation times, flow velocity and chemical shift. These wrist MRU images when interpreted by a trained physician can be useful in determining a diagnosis. The MicroMRI Wrist Coil cannot be used with an MRI system that supports the use of only receive coils.
The MicroMRI Wrist Coil uses an elliptical birdcage design to provide the highest possible signal-to-noise ratio (SNR) for imaging a human wrist. The two piece structure facilitates the use of the coil for a wide range of wrist sizes. The elliptical design and ability to conform to the subject's wrist makes the coil unique and efficient in transmitting and receiving the RF signal to the anatomy (i.e., the wrist).
The accompanying immobilization system provides a means to stabilize the patient's arm during the scan. The immobilization system provides a means for positioning the patient's arm consistently in the same position and orientation.
Here's an analysis of the provided text regarding the MicroMRI Wrist Coil, focusing on acceptance criteria and study details:
Device: MicroMRI Wrist Coil
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Safety | Performance testing was conducted in accordance with IEC/UL/CSA/EN 60601-1 standards. |
| Performance | Performance testing was conducted in accordance with NEMA standards. The performance test results indicate that the MicroMRI Wrist Coil is comparable to the predicate devices (Mayo Clinic BC-10 1.5T wrist coil and Invivo HRW Array Coil). |
| Equivalence (Primary differences from predicates) | Vs. BC-10 1.5T Wrist Coil: MicroMRI Wrist Coil scans only the wrist, while BC-10 can scan hand and wrist. Vs. HRW Array Coil: MicroMRI Wrist Coil is a transmit and receive coil (quadrature birdcage design, requiring less power), while HRW Array Coil is a receive-only coil. Despite these differences, the device is considered substantially equivalent. |
| Intended Use | The device is a transmitting and receiving device for imaging the wrist with GE Healthcare and Siemens 1.5T MRI systems. Images are used by trained physicians for diagnosis. (Device performance meets this intended use implicitly through equivalence to predicate devices for similar wrist imaging). |
| No New Indications | The device does not include any new indications for use. |
| No New Safety/Efficacy Issues | The device does not present any new safety or efficacy issues. |
2. Sample Size Used for the Test Set and Data Provenance
The provided text does not specify a sample size for a test set. This submission is for a medical device (an MRI coil), and the performance evaluation focused on adherence to engineering and safety standards (NEMA, IEC/UL/CSA/EN 60601-1) and comparison to predicate devices, rather than a clinical study with a patient test set in the traditional sense for an AI/algorithm.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to this submission. The device is hardware (an MRI coil), not an AI algorithm that would require expert-established ground truth on a test set of medical images. The "ground truth" related to this device's performance would be derived from physical measurements according to engineering standards.
4. Adjudication Method for the Test Set
This information is not applicable for the reasons stated in point 3.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This device is an MRI coil, not an AI algorithm designed to assist human readers or for standalone diagnosis.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
No, a standalone performance study was not done. This device is a hardware component (an MRI coil), not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance relies on:
- Engineering Standards: Adherence to NEMA and IEC/UL/CSA/EN 60601-1 standards for safety and performance (e.g., signal-to-noise ratio, power requirements, electrical safety).
- Comparison to Predicate Devices: Demonstrating comparable performance characteristics to existing, legally marketed MRI wrist coils. This implicit ground truth is that if it performs like a legally marketed predicate device, it is acceptable.
8. The Sample Size for the Training Set
This information is not applicable. This device is a hardware component and does not utilize a "training set" in the context of machine learning or AI.
9. How the Ground Truth for the Training Set was Established
This information is not applicable for the reasons stated in point 8.
{0}------------------------------------------------
K073131
pg. 1 of 2
Section E-1
SECTION E - 510(K) SUMMARY
JAN 1 8 2008
(As Required By 21 CFR 807.92)
| Submitter Information | |
|---|---|
| Company: | MicroMRI, Inc.1429 Walnut Street, Suite 1102Philadelphia PA 19102Phone: (215) 564-1002 Fax: (877) 801-3876 |
| Contact Person: | Hrishikesh GadagkarPhone: (215) 564-1002 Ext 18 Fax: (877) 801-3876Email: hgadagkar@micromri.com |
Device Information
| Device Name: | MicroMRI Wrist Coil |
|---|---|
| Common Name: | Magnetic Resonance Coil |
| Classification: | Magnetic Resonance Imaging System |
| Product Code: | 90 MOS, Class II (21CFR892.1000) |
Predicate Devices
The MicroMRI Wrist Coil system for scanning human wrist is significantly equivalent to the currently marketed BC-10 1.5T (Mayo Clinic) wrist coil and HRW Array Coil (Invivo Corporation).
Device Description
The MicroMRI Wrist Coil uses an elliptical birdcage design to provide the highest possible signal-to-noise ratio (SNR) for imaging a human wrist. The two piece structure facilitates the use of the coil for a wide range of wrist sizes. The elliptical design and ability to conform to the subject's wrist makes the coil unique and efficient in transmitting and receiving the RF signal to the anatomy (i.e., the wrist).
The accompanying immobilization system provides a means to stabilize the patient's arm during the scan. The immobilization system provides a means for positioning the patient's arm consistently in the same position and orientation.
Intended Use
The MicroMRI Wrist Coil is a transmitting and receiving device to be used in conjunction with GE Healthcare and Siemens 1.5T MRI systems to image the wrist. The images acquired using the MicroMRI Wrist Coil along multiple axes (axial, coronal, sagittal, and oblique) are a measure of the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the spatial image are proton density, spin-lattice (T1) and spin-spin (T2) relaxation times, flow velocity and chemical shift. These wrist MRI images when interpreted by a trained
{1}------------------------------------------------
K673131
pg. 2 of 2
Section E-2
physician can be useful in determining a diagnosis. The MicroMRI Wrist Coil cannot be used with an MRI system that supports the use of only receive coils.
Predicate Comparison
The MicroMRI Wrist Coil is comparable to BC-10 1.5T wrist coil with the primary difference being that the BC-10 has the ability to scan the hand and wrist. The MicroMRI Wrist Coil is designed specifically to scan only the wrist.
The MicroMRI Wrist Coil is comparable to HRW Array Coil with the primary difference being that the HRW Array Coil is a receive only coil. The MicroMRI Wrist Coil utilizes a quadrature birdcage design and can transmit and receive RF signals, requiring less power.
Summary of Performance Testing
The performance testing of MicroMRI's Wrist Coil was conducted in accordance with the NEMA standards, and IEC/UL/CSA/EN 60601-1 standards. The performance test results indicate that the MicroMRI Wrist Coil is comparable to the predicate devices.
Conclusions
MicroMRI considers the MicroMRI Wrist Coil to be substantially equivalent to both Mayo Clinic BC-10 1.5T and Invivo HRW Array Coil. MicroMRI Wrist Coil does not include any new indications for use or present any new safety or efficacy issues.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing a three-lined globe.
JAN 1 8 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Hrishikesh Gadagkar Vice President of Engineering and Manufacturing MicroMRI. Inc. 1429 Walnut Street, Suite 1102 PHILADELPHIA PA 19102
Re: K073131
Trade/Device Name: MicroMRI Wrist Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: November 2, 2007 Received: November 7, 2007
Dear Mr. Gadagkar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
SECTION D - STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known): NA
Device Name: MicroMRI Wrist Coil
Indications for Use:
The MicroMRI Wrist Coll is a transmitting and receiving device to be used in conjunction with GE Healthcare and Siemens 1.5T MRI systems to image the wrist. The images acquired using the MicroMRI Wrist Coil along multiple axes (axial, coronal, sagittal, and oblique) are a measure of the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the spatial image are proton density, spin-lattice (TI) and spin-spin (T2) relaxation times, flow velocity and chemical shift. These wrist MRU images when interpreted by a trained physician can be useful in determining a diagnosis. The MicroMRI Wrist Coil cannot be used with an MRI system that supports the use of only receive coils.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lori M. White
(Division/Sign Off)
Division of Reproductive, Abdominal, an Radiological Devices 510(k) Number
N/A