K Number
K073111

Validate with FDA (Live)

Date Cleared
2008-03-27

(143 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Portrait® PSR3 System is intended for treatment of the following dermatological conditions:

  • Treatment of wrinkles and rhytides
  • Superficial skin lesions
  • Actinic Keratosis
  • Viral Papillomata
  • Seborrhoeic Keratosis
  • Acne Scars
Device Description

The Portrait® PSR is an electro-surgical device for use in dermatological applications. The effect of the device is achieved by heating the outer layer of the skin so that part or all of the epidermis becomes non-viable and there is controlled damage to the underlying dermis. Similar biological changes are produced as with established laser-based dermatological surgical and skin resurfacing technologies.

The system includes:

  • Portrait® PSR3 Generator
  • Footswitch
  • Handpiece and Cable Assembly
  • Treatment Pack comprising a disposable Nozzle and an electronic "treatment pack key"
  • Nitrogen Gas Cylinder
AI/ML Overview

Here's an analysis of the provided text regarding the Rhytec, Inc. Portrait® PSR3 device, focusing on acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document (a 510(k) Summary) does not explicitly define specific "acceptance criteria" in terms of quantitative performance metrics (e.g., a specific percentage of improvement, a reduction in lesion size by X%). Instead, the acceptance is based on the demonstration of substantial equivalence to predicate devices and the safety and effectiveness for its stated indications for use.

Therefore, the "acceptance criteria" are implied by the regulatory standard for a 510(k) submission:

  • Substantial Equivalence: The device must be as safe and effective as a legally marketed predicate device.
  • Indications for Use: The device must be shown to be safe and effective for its stated indications.

The reported device performance is described qualitatively:

Acceptance Criteria (Implied)Reported Device Performance
Substantial Equivalence to predicate devices (K071786, K963339, K030453, K053047)"Portrait® PSR3 utilizes the same technology as previously cleared for Portrait® PSR3. There have been no changes to the device design. The substantial equivalence of this device is supported by the prior approval of this device and the clinical data submitted."
Safety and Effectiveness for labeled Indications for Use:"Clinical studies to support clinical performance of the Portrait® PSR3 to improve acne scarring were conducted... These studies conclude that the Portrait® PSR³ is safe and effective for treating acne scars." "Clinical studies demonstrated that the Portrait® PSR3 is capable of producing tissue effects similar to that seen with carbon dioxide laser treatment."
Treatment of wrinkles and rhytidesImplied by general safety and effectiveness claims. (No specific study details for this indication provided in the summary)
Superficial skin lesionsImplied by general safety and effectiveness claims. (No specific study details for this indication provided in the summary)
Actinic KeratosisImplied by general safety and effectiveness claims. (No specific study details for this indication provided in the summary)
Viral PapillomataImplied by general safety and effectiveness claims. (No specific study details for this indication provided in the summary)
Seborrhoeic KeratosisImplied by general safety and effectiveness claims. (No specific study details for this indication provided in the summary)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The 510(k) Summary does not specify the sample size for the clinical studies. It mentions "Clinical studies... to improve acne scarring," but the number of subjects or cases is not provided.
  • Data Provenance: Not explicitly stated. The studies were conducted "in compliance with the Code of Federal Regulations, as specified in 21 CFR 812," which governs investigational device exemptions in the US. This implies the studies were likely conducted in the United States and were prospective clinical studies.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • The 510(k) Summary does not provide any information on the number of experts or their qualifications used to establish ground truth for the test set.

4. Adjudication Method for the Test Set

  • The 510(k) Summary does not provide any information on the adjudication method used for the test set.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

  • The 510(k) Summary does not mention a Multi Reader Multi Case (MRMC) comparative effectiveness study. The studies mentioned focus on evaluating the device itself, not comparing human readers with or without AI assistance. The device is a direct treatment device, not an AI-assisted diagnostic tool.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • This question is not applicable as the Portrait® PSR3 System is an electrosurgical device for dermatological treatment, not an algorithm or AI-based diagnostic tool. It requires human operation.

7. The Type of Ground Truth Used

  • The ground truth for the clinical studies evaluating "acne scarring" would typically be based on clinical assessment/outcomes data by clinicians (e.g., visual grading scales, objective measurements of scar reduction, patient satisfaction). The summary states "Clinical studies... conclude that the Portrait® PSR³ is safe and effective for treating acne scars," implying clinical endpoints were used.
  • For the claim of "producing tissue effects similar to that seen with carbon dioxide laser treatment," the ground truth would likely involve histopathological analysis (biopsies) of treated tissue, potentially compared to tissue treated with CO2 lasers.

8. The Sample Size for the Training Set

  • The 510(k) Summary does not mention a "training set" in the context of machine learning or AI algorithms. The device described is a physical medical device, not an AI model requiring a training set.

9. How the Ground Truth for the Training Set Was Established

  • This question is not applicable as the device is not an AI model that would require a training set and associated ground truth establishment.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the words "MAR 27 2008 RHYTEC TRUE REGENERATIVE SCIENCE". The date is March 27, 2008. The word "RHYTEC" is in bold, and the words "TRUE REGENERATIVE SCIENCE" are stacked on top of each other to the right of the word "RHYTEC".

KO 73111

510(k) Summary

Page 1 of 3

For the Rhytec, Inc. Portrait® PSR3

General Information

Submitter:

Rhytec, Inc. 130 Turner Street Building Two Waltham, MA 02453

Contact Person

Robert Zoletti Rhytec, Inc. 130 Turner Street Building Two Waltham, MA 02543 Telephone: 781-419-9482 Email: rzoletti@rhytec.com

Summary Preparation Date: March 24, 2008

Names

Trade Name: Common Name: Classification Name: Product Code:

Portrait® PSR3 System (None) Electrosurgical cutting and coagulation device GEI

Legally Market Predicate Devices

K071786 Portrait® PSR³ K963339 Coherent UltraPulse K030453 CoolTouch ND:YAG K053047 Fraxel ISR Laser System

Device Description

The Portrait® PSR is an electro-surgical device for use in dermatological applications. The effect of the device is achieved by heating the outer layer of the skin so that part or all of the epidermis becomes non-viable and there is controlled damage to the underlying dermis. Similar biological changes are produced as with established laser-based dermatological surgical and skin resurfacing technologies.

{1}------------------------------------------------

K073111

SCIENCE

Page 2 of 3

The system includes:

  • Portrait® PSR3 Generator-a mobile unit powered from 110/230 VAC standard wall socket . comprising a trolley and lift-out section referred to as a "generator section". Treatment parameters are displayed on the control panel.
  • Footswitch-a single pedal unit for activation ♥
  • . Handpiece and Cable Assembly-this carries RF power and Nitrogen gas to the Nozzle.
  • Treatment Pack comprising a disposable Nozzle that is connected to the Handpiece and . an electronic "treatment pack key" that is used by the generator to ensure the Nozzle is not used beyond its validated operational life. The instrument does not touch the skin during treatment.
  • . Nitrogen Gas Cylinder - the gas is provided in a cylinder housed in the generator enclosure. Refill cylinders containing the required purity of Nitrogen are obtained by the user from a commercial gas product supplier.

Indications for Use Statement

The Portrail® PSR3 System is intended for treatment of the following dermatological conditions:

  • Treatment of wrinkles and rhytides .
  • Superficial skin lesions .
  • Actinic Keratosis ●
  • Viral Papillomata .
  • Seborrhoeic Keratosis 4
  • . Acne Scars

Technological Characteristics

The technology involved in the Portrait® PSR3 System is not changed from previous Portraille PSR3 submissions.

UHF energy from the generator converts Nitrogen gas into plasma within the Handpiece. The plasma emerges from the Nozzle at the distal end of the Handpiece and is directed onto the skin to be treated. Rapid heating of the skin occurs as the excited gas gives up energy to the skin. Through the combination within the Handpiece of precisely controlled energy and Nitrogen gas, individual plasma pulses are produced that will give predictable tissue effects.

The Portrait® PSR3 uses Nitrogen plasma as compared to laser light used by UltraPulse, CoolTouch ND:YAG and Fraxel ISR Laser System.

{2}------------------------------------------------

K073111

RHYTEC TRUE REGENERATIVE SCIENCE

Page 3 of 3

Substantial Equivalence Comparison

Portrait® PSR3 utilizes the same technology as previously cleared for Portrait® PSR3. There have been no changes to the device design. The substantial equivalence of this device is supported by the prior approval of this device and the clinical data submitted. Other devices used to treat acne scars include the 1320-nmNd:YAG (K030453), Fraxel SR 1500 (K070284) and CO2 Laser (K963339).

Clinical and Non-Clinical Data

Clinical studies demonstrated that the Portrait® PSR3 is capable of producing tissue effects similar to that seen with carbon dioxide laser treatment. Clinical studies to support clinical performance of the Portrait® PSR3 to improve acne scarring were conducted as Non-Significant Risk studies and undertaken in compliance with the Code of Federal Regulations, as specified in 21 CFR 812. These studies conclude that the Portrait® PSR³ is safe and effective for treating acne scars.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle faces right and is positioned within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper half of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 7 2008

Rhytec Incorporated % Mr. Robert Zoletti Director RA 130 Turner Street Building Two Waltham, Massachusetts 02543

K073111 Trade/Device Name: Portrait® PSR3 System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: February 29, 2008 Received: March 04, 2008

Dear Mr. Zoletti:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Robert Zoletti

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication for Use

510(k) Number: K073111

Device Name: Portrait® PSR³

Indications for use:

The Portrait® PSR3 is intended for treatment of the following dermatological conductions:

  • Treatment of wrinkles and rhytides 1.
  • Superficial skin lesions 2.
  • Actinic keratosis 3.
  • Viral papillomata 4.
  • Seborrhoeic keratosis 5.
  • Acne Scars 6.

Prescription Use X

And/Or

Over the Counter Use

(21 CFR Part 801 Subpart D) Subpart C)

(21 CFR Part 801

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Outta
S. Sign Off

Division of General, Restorative, and Neurological Devices

510(k) Number K073111

Rhytec, Inc. November 2007

Page 34

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.