K Number
K073034

Validate with FDA (Live)

Date Cleared
2008-03-14

(140 days)

Product Code
Regulation Number
876.5980
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Initial placement of a perculaneous gastrostomy device may be indicated for patients with a functioning gut who requre long-term tube feeding. This includes patients in whom malnutrition already exists, or may result, seconda ry to concurrent conditions.

Device Description

Not Found

AI/ML Overview

I am sorry, but the provided text content does not contain the information required to answer the request. The document is an FDA 510(k) clearance letter for a medical device (AMT Initial Placement GI Kit) and primarily discusses regulatory approval, not a study evaluating acceptance criteria or device performance. There is no mention of acceptance criteria, device performance metrics, sample sizes, data provenance, expert qualifications, ground truth, or details about a statistical study.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract caduceus or a bird-like figure with three curved lines representing wings or ribbons.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Brian Baker Engineer Applied Medical Technology, Inc. 8000 Katherine Blvd. BRECKSVILLE OH 44141

MAR 1 4 2008

Re: K073034

Trade/Device Name: AMT Initial Placement GI Kit Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: February 22, 2008 Received: February 25, 2008

Dear Mr. Brian Baker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807);

{1}------------------------------------------------

Page 2 - Mr. Brian Baker

labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

In addition, we have determined that your device kit contains iodine prep swabsticks and watersoluble lubricant, which are subject to regulation as drugs.

Our substantially equivalent determination does not apply to the drug component[s] of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing these drugs, we suggest you contact:

Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification' (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (240) 276-3150, or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Heilut lemer for
Nancy C. Brogdon

Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

SECTION - D

c. - INDICATIONS FOR USE STATEMENT

073034

510(k) Number (if known): -

Trade Name: AMT Initial Placement GI Kit Common Name: Initial Placement GI Kit

Indications For Use (as shown in labeling):

Initial placement of a perculaneous gastrostomy device may be indicated for patients with a functioning gut who requre long-term tube feeding. This includes patients in whom malnutrition already exists, or may result, seconda ry to concurrent conditions.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use
(21 CFR 801 Subpart C)

(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K073034 Applied Medical Technology, Inc. -510(k) Submission

AMT Initial Placement GI Kit Section D - Page D.2

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.