K Number
K072845

Validate with FDA (Live)

Device Name
PRECISION FLOW
Manufacturer
Date Cleared
2008-07-17

(287 days)

Product Code
Regulation Number
868.5450
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Precision Flow™ is intended for use to add warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital, sub-acute institutions, and home settings. It adds heat and moisture to a blended medical air/oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

Device Description

The Precision Flow™ consists of two parts: The main unit which contains all the electrical and electronic components including the electronic blender and flow controllers. All the sensors are located in the main unit. The main unit has no water pathways and the gas pathway contains only dry gas at room temperature, and consequently does not need internal cleaning or disinfection. The disposable components comprising the disposable water module, vapor transfer cartridge and heated delivery tube. Conditions in the circulating water and gas streams are sensed remotely via the interface between the main unit and the disposable module.

AI/ML Overview

This 510(k) summary describes a medical device, the Vapotherm Precision Flow™, which is a respiratory gas humidifier. It is not an AI/ML powered device, therefore, many of the requested categories related to AI performance, ground truth, and expert evaluation are not applicable.

Here's an analysis of the provided text based on your request, highlighting the non-applicability of AI/ML specific criteria:

Acceptance Criteria and Device Performance (Non-AI/ML Device)

The Vapotherm Precision Flow™ is cleared based on demonstrating substantial equivalence to predicate devices, primarily through non-clinical performance data (testing) and a comparison of its technological characteristics to existing devices. The "acceptance criteria" here are implicitly that the device performs its intended functions (humidifying and blending medical gases) safely and effectively, and that these performance characteristics are comparable to legally marketed predicate devices without raising new safety or effectiveness concerns.

Since this is not an AI/ML device, performance metrics like sensitivity, specificity, accuracy, or AUC are not applicable or reported in this 510(k) summary. Instead, the performance is demonstrated through various engineering, safety, and biocompatibility tests.

Acceptance Criteria (Implied)Reported Device Performance
Biocompatibility: No adverse biological reactions.Biocompatibility Tests (LLNA, Intracutaneous Reactivity, MEM Elution): Performed, implying acceptable results (otherwise it wouldn't be marketed).
Emissions: No harmful volatile organic compounds or particulates.Volatile Organic Compounds (VOC), Particulate Matter tests: Performed to recognized standards, implying acceptable results.
Electrical Safety: Meets international safety standards.EMC Test, Emissions Test, Safety Inspection Test, IEC 60601-1, IEC 60601-1-4, IEC 60601-1-8: Performed, ensuring compliance with general requirements for safety, programmable electrical medical systems, and alarm systems.
Microbiological Safety: No detectable bacteria after extended use.Bio Burden, 30 Day Comparative Use Testing: "No detectible bacteria in any water condensation samples from the Precision Flow® devices either initially (3 days) or after 30 days of operation."
Software Functionality: Verified and validated.Ximedica TRP 1097, TRP 1071 (Black Box), Unit Test Case Listing - 3VAP1004/TRP 1092 (White Box): Software verified and validated in accordance with applicable FDA guidance.
Thermal Stability: Maintains temperature regulation.Thermal Stability: Performed, indicating proper temperature control.
Blender Accuracy: Maintains precise air/oxygen mixture.Blender Comparison Performance: Performed, indicating comparable performance to predicate blenders, assuring integrity of the precise air/oxygen mixture via an integral oxygen analyzer.
Substantial Equivalence: Comparable to predicate devices.The device has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. Performance data demonstrate it is as safe and effective as the predicate devices, with minor differences raising no new issues.

Non-Applicable (N/A) for AI/ML Specific Questions:

Given that K072845 describes a humidifier, not an AI/ML-powered device, the following categories are not applicable. The submission process for such a device focuses on engineering specifications, safety testing, and performance validation against established standards and predicate devices, rather than statistical performance against a ground truth dataset.

2. Sample size used for the test set and the data provenance: N/A (Not an AI/ML device, no "test set" in the context of an algorithm's classification/prediction performance). Testing involved physical hardware, software, and biological compatibility, not a data-driven test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (Not an AI/ML device, no ground truth established by experts for algorithmic performance).

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: N/A (Not an AI/ML device, no adjudication of algorithm outputs).

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A (Not an AI/ML device. No human-in-the-loop AI assistance involved).

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: N/A (Not an AI/ML device, no standalone algorithm performance to evaluate).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): N/A (Not an AI/ML device). The "ground truth" for this device relates to physical measurements meeting specifications (e.g., oxygen percentage, temperature, bacterial counts being zero).

8. The sample size for the training set: N/A (Not an AI/ML device, no "training set"). Software verification and validation are described, but this refers to traditional software engineering processes, not machine learning model training.

9. How the ground truth for the training set was established: N/A (Not an AI/ML device, no "training set" or ground truth in the ML context).

{0}------------------------------------------------

K072845

510(k) SUMMARY

Vapotherm, Inc.'s Precision Flow™

JUL 1 7 2008

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:

Vapotherm, Inc. 198 Log Canoe Circle Stevensville, Maryland 21666

Phone: 410-604-3977 Facsimile: 410-604-3978

Contact Person: Gregory A. Whitney

April 15, 2008 Date Prepared:

Name of Device and Name/Address of Sponsor

Precision Flow™ Vapotherm, Inc. 198 Log Canoe Circle Stevensville, Maryland 21666

A.. Common or Usual Name

Humidifier, Respiratory Gas (Direct Patient Interface)

B. Classification Name

Humidifier, Respiratory Gas (Direct Patient Interface) Anesthesiology Panel (868.5450) Class II

C. Product Code

BTT

D. Predicate Devices

Vapotherm. Inc. Medtec Bird Products

2000i& 2000h Oxygen Sensor Air-Oxygen Blender

K000401, K0142245, K042245 K063488 K911962

{1}------------------------------------------------

E. Intended Use / Indications for Use

The Precision Flow™ is intended to be used for adding warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital, sub-acute institutions, and home settings. It adds heat and moisture to a blended medical air/oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

F. Technological Characteristics

The Precision Flow™ consists of two parts:

The main unit which contains all the electrical and electronic components including the electronic blender and flow controllers. All the sensors are located in the main unit. The main unit has no water pathways and the gas pathway contains only dry gas at room temperature, and consequently does not need internal cleaning or disinfection.

The disposable components comprising the disposable water module, vapor transfer cartridge and heated delivery tube. Conditions in the circulating water and gas streams are sensed remotely via the interface between the main unit and the disposable module.

1. Main unit:

  • · The flow of oxygen and air are measured by mass flow sensors. The operating software calculates the required flow of each needed to reach the target flow and oxygen percentage set by the operator. The system controls gas flows accordingly by adjusting proportional solenoid valves on the gas lines. An oxygen sensor monitors the gas mixture and signals any discrepancy between target and measured percentage. The oxygen sensor is automatically calibrated with oxygen at power-up and every 24 hours.
  • · Firmware running in the main unit uses sensors to monitor gas pressure, water level and water temperature, and to detect air leaks into the water circuit (bubble detector) and water leaks into the gas circuit (droplet detector). Alarms are displayed if any parameters are outside the normal range. Other indicators show low charge in the backup battery, and the type of cartridge installed. An internal battery backup will maintain the set flow and oxygen blend for at least 15 minutes without AC power.

{2}------------------------------------------------

Disposable components: 2.

  • · Vapor Transfer Cartridge. In the cartridge blended gas passes through the lumens of hundreds of parallel hollow fibers made of a specially developed polymer. Warm water circulates around the fibers and diffuses as vapor through the fiber material into the gas stream flowing through each fiber. Unlike most humidifiers, there is no direct contact between the water and gas streams. The gas stream leaves the cartridge essentially saturated with vapor at the set temperature.
  • · Triple-lumen heated delivery tube. The warmed humidified gas passes through the center lumen. The center lumen is surrounded by two outer lumens carrying warmed water to maintain the temperature of the inner lumen and to minimize rain-out. A proprietary short nasal cannula is connected to the end of the delivery tube and passes the humidified breathing gas to the patient's nose.
  • · Disposable module. The module houses a water reservoir, pump, connections for the cartridge and delivery tube, and sensor interfaces to the main unit. Water is pumped past a heater plate through the outer lumens of the delivery tube. Returning water passes through the outer jacket of the Vapor Transfer Cartridge where some water is lost as vapor to the gas stream. There is no direct contact between water and gas flows. The water then returns to the pump reservoir. Heater power is adjusted continuously to maintain the set temperature. Water flows into the circuit from the water bag to replace evaporative losses in the Vapor Transfer Cartridge. Air is purqed to atmosphere from the circulation via a hydrophobic filter membrane.

3. Summary of Non-Clinical Performance Data

The following table lists the non-clinical Performance Testing Data that was performed on the Precision Flow™:

NumberTest Description
1Biocompatibility – Murine Local Lymph Node Assay (LLNA) (Sensitization)
2Biocompatibility - ISO Intracutaneous Reactivity Test (Irritation)
3Biocompatibility – MEM Elution (Cytotoxicity)
4Volatile Organic Compounds (VOC), EPA Compendium Method TO-15
5Particulate Matter, NIOSH Method 0500
6EMC Test
7Emissions Test
8Safety Inspection Test
9IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety
10IEC 60601-1-4, Medical Electrical Equipment Part 1-4 - Collateral Standard:Programmable Electrical Medical Systems

{3}------------------------------------------------

(------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
NumberTest Description
11IEC 60601-1-8, Medical electrical equipment, General requirements for basic safety andessential performance - Collateral standard: General requirements, tests and guidancefor alarm systems in medical electrical equipment and medical electrical systems
12Bio Burden, 30 Day Comparative Use Testing
13Ximedica TRP 1097, “Vapotherm Precision Flow Hyperthermic Humidification System,System Software Verification & Validation Test Plan"
14TRP 1071 "System Software Test" (Black Box)
15Unit Test Case Listing - 3VAP1004/TRP 1092 (White Box)
16Thermal Stability
17Blender Comparison Performance
  • Bio Burden, 30 Day Comparative Use Testing: After a thirty day . comparative use testing, there were no detectible bacteria in any water condensation samples from the Precision Flow® devices either initially (3 days) or after 30 days of operation.
  • The Precision Flow™ software was verified and validated in . accordance with applicable FDA guidance documents.
  • . The Volatile Organic Components and Particulate Mater tests were performed to recognized standards.

Substantial Equivalence

The Precision Flow™ is as safe and effective as the Vapotherm 2000i and 2000h, the Bird Mircoblender, and the Medtec Oxygen Sensor. The Precision Flow™ has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the Precision Flow™ and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Precision Flow™ is as safe and effective as the predicate devices. Thus, the Precision Flow™ is substantially equivalent.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" arranged around the perimeter. In the center of the circle is a stylized eagle emblem, with its wings spread and head turned to the side. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Vapotherm, Incorporated C/O Mr. Jonathan S. Kahan Hogan & Hartson L.L.P. 555 Thirteenth Street, NW Washington DC 20004

JUL 1 7 2008

Re: K072845

Trade/Device Name: Precision FlowTM Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: June 27, 2008 Received: June 30, 2008

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Kahan

Please be advised that FDA's issuance of a substantial equivalence determination does not mcan that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to; registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your tevice to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Susan Quars

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for use

510(k) Number (if known):

Device Name: Precision Flow™

Indications for Use:

Precision Flow™ is intended for use to add warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital, sub-acute institutions, and home settings. It adds heat and moisture to a blended medical air/oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

Precision Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Will Maly for M. Husband Page _ of _

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: 1< 072 845

§ 868.5450 Respiratory gas humidifier.

(a)
Identification. A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.(b)
Classification. Class II (performance standards).