(36 days)
The device is intended for medical purposes to provide mobility to persons restricted to a seated position.
The COMFORT POWERED WHEELCHAIR, LY-EB206 is an indoor / outdoor Electric Wheelchair that is battery operated. It has a base with four-wheeled with a seat. The movement of the Wheelchair is controlled by the rider who uses hand controls located at the top of the steering column. The device can be disassembled for transport and is provided with an onboard battery charger.
This 510(k) submission (K072711) describes a Powered Wheelchair (COMFORT POWERED WHEELCHAIR, LY-EB206). The submission focuses on demonstrating substantial equivalence to a predicate device (COMFORT Powered Wheelchair, LY-EB103, K030356), rather than establishing new performance criteria through a standalone study. Therefore, the information provided mainly pertains to the comparison with the predicate device and the standards met, rather than a traditional "acceptance criteria" and "study proving it meets acceptance criteria" as might be found for novel diagnostic AI devices.
However, I can extract the relevant information and present it in a structured way based on the provided text.
Here's an analysis of the provided information:
Acceptance Criteria and Device Performance (Based on Substantial Equivalence and Standards)
Given that this is a 510(k) for a powered wheelchair, the "acceptance criteria" are implicitly defined by the safety and performance standards met by the device, as well as its substantial equivalence to a legally marketed predicate device. The "device performance" is primarily discussed in terms of meeting these standards and being comparable to the predicate.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implied by Standards & Predicate Comparison) | Reported Device Performance (COMFORT POWERED WHEELCHAIR, LY-EB206) |
|---|---|
| Intended Use (Mobility for persons restricted to a seated position) | Intended for medical purposes to provide mobility to persons restricted to a seated position. (Same as predicate) |
| Mainframe Strength & Fatigue | Mainframes meet strength and fatigue tests. (Similar to predicate) |
| Material Aspects | Similar mainframe materials to predicate. |
| Back Upholstery Ignition Resistance | Back upholstery passed resistance ignition test by SGS. (Same material as predicate) |
| Electronic Systems Safety (controller, batteries, motor, charger) | Electronic systems are the same as the predicate and all passed UL certification. (Same safety level assured) |
| EMC Compliance | Complies with EMC Report ANSI / RESNA WC/Vol.2-1998, CISPR 11: 1990, EN61000-3-2: 1995, IEC61000-3-3: 1995. |
| Maximum Speed (Under 6 mph limit) | Maximum speed is 4.75 mph. (Within the 6 mph limit, compared to predicate's 5.31 mph) |
| Maximum Range per Charge | Maximum range per charge is 23.8 miles. (Different from predicate's 20 miles, but considered substantially equivalent in real-life use) |
| Other Features (Suspension, footplates, incline degree 12°, armrest type, weight limit, warranty) | All are the same as the predicate device. |
| Overall Dimensions | Minor differences from the predicate, but not considered a safety aspect. |
| Tire Sizes | Differences in tire sizes from the predicate, but not considered a safety aspect. |
2. Sample Size for the Test Set and Data Provenance
- Sample Size for Test Set: Not applicable in the context of a clinical test set for AI/diagnostic performance. The "test set" here refers to the device itself undergoing engineering and performance testing. The document describes the testing of the wheelchair as a single unit or representative units, not a patient-centric data set.
- Data Provenance: The referenced standards (ANSI / RESNA WC/Vol.2-1998, CISPR 11: 1990, EN61000-3-2: 1995, IEC61000-3-3: 1995) are international and national technical specifications for power wheelchairs and electrical compatibility. The ignition resistance test was conducted by SGS, a global inspection, verification, testing, and certification company. The electronic components were UL certified. The manufacturing country is Taiwan, R.O.C. (Comfort Orthopedic Co., LTD.). This is a submission for a physical device, so "retrospective or prospective" data provenance regarding patient cohorts is not relevant.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: Not applicable. For a physical device like a wheelchair, "ground truth" is established through engineering specifications, validated test methods (e.g., those specified in ANSI/RESNA, CISPR, EN, IEC standards), and certifications from recognized bodies (e.g., UL, SGS). There's no mention of human experts establishing a "ground truth" in the diagnostic sense.
- Qualifications of Experts: N/A. The "experts" are the engineers and technical specialists involved in the design, testing, and certification processes according to established industry standards.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. This concept is typically relevant for medical imaging or diagnostic AI where discordant expert opinions need resolution for ground truthing. For a physical device, compliance is determined by meeting predefined engineering specifications and passing standardized tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- MRMC Study: No. This submission does not include an MRMC study. MRMC studies are used to evaluate the diagnostic performance of a system (often AI-assisted) compared to human readers over a diverse set of cases. This is a submission for a physical medical device (powered wheelchair) where such studies are not typically performed.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done
- Standalone Study: Not applicable. This device is a physical powered wheelchair, not an algorithm. Its performance is measured directly through mechanical, electrical, and safety tests, not through an AI algorithm operating in a standalone capacity. The "standalone performance" here would be the physical performance of the wheelchair itself as tested against the standards mentioned.
7. The Type of Ground Truth Used
- Type of Ground Truth: The "ground truth" for this device is based on engineering specifications, compliance with recognized industry standards (e.g., ANSI/RESNA, CISPR, EN, IEC) for safety and performance, and certifications from accredited testing laboratories (e.g., UL, SGS). The primary "ground truth" for substantial equivalence is the performance and safety profile of the predicate device.
8. The Sample Size for the Training Set
- Sample Size for Training Set: N/A. This is not an AI/ML device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
- How Ground Truth for Training Set Was Established: N/A. There is no training set for this type of device.
{0}------------------------------------------------
COMFORT ORTHOPEDIC CO., LTD. www.comfort.com.tw No.120,Nan Shiang Tsuen,Shoei Shang Shiang,Chia-yi,Taiwan,R.O.C.608 TEL : 886-5-2892093 FAX : 886-5-2890070
.
510(k) SUMMARY " 66
Submitter's Name: COMFORT ORTHOPEDIC CO., LTD.
N0. 120, NAN SHIANG TSUEN, SHOEI SHANG SHIANG, CHIA-YI, 608, TAIWAN, ROC
Date summary prepared:
September 21, 2007
Device Name:
Proprietary Name: Common or Usual Name: Classification Name:
COMFORT POWERED WHEELCHAIR, LY-EB206 Powered Wheelchair Powered Wheelchair, Class II, 21 CFR 890.3860
Indications for Use:
The device is intended for medical purposes to provide mobility to persons restricted to a seated position.
Description of the device:
The COMFORT POWERED WHEELCHAIR, LY-EB206 is an indoor / outdoor Electric Wheelchair that is battery operated. It has a base with four-wheeled with a The movement of the Wheelchair is controlled by the rider who uses hand seat. controls located at the top of the steering column. The device can be disassembled for transport and is provided with an onboard battery charger.
Performance Testing:
EMC Report ANSI / RESNA WC/Vol.2-1998, CISPR 11: 1990, EN61000-3-2: 1995, IEC61000-3-3: 1995 (Electrically Powered Wheelchairs, controller, and their chargers - requirements and test methods)
Legally marketed device for substantial equivalence comparison: COMFORT Powered Wheelchair, LY-EB103 (K030356)
{1}------------------------------------------------
Summary for substantial equivalence comparison:
We can know from the above table that the intended use between the two devices is Mainframes of two devices are foldable. Mainframes materials of the two the same. devices all meet the strength and fatigue tests and they are similar for the material aspects. The overall dimensions are similar. The back upholstery is the same material, and also passed the resistance ignition test by SGS. Moreover, the suspension of cross brace, footplates. incline degree 12°, armrest type, weight limit, and warranty are all the same. They are substantially equivalent.
Especially the electronic systems between two devices are the same and all passed by the UL certificated, for instance the electronic controller, batteries, motor, and recharge. Thus the same safety level for the two devices is assured.
The major differences existing are the overall dimension and the sizes of tires are differences between the two devices. The overall appearance differences are not safety aspect. So the new device is substantially equivalent to the predicate devices in this aspect.
The maximum speed is 4.75 mph for the new device and 5.31 mph for the predicate device that are also under the 6 miles maximum speed limited. Those speeds can be continuous adjusted by the throttle tiller. The operators can set the adequate speed according to their feeling and need, and the maximum speed differences do not mean any performance differences. They are substantially equivalent.
Maximum range per charge is different, and new device is 23.8 miles and 20 miles for Certainly the real range depends on the practical environments, i.e., the predicate device. weight, surface, incline, and temperature. For the real life use, the two devices are substantially equivalent.
Based on the above the information and the analysis, we know that the subject device and the predicate device have the same intended use the same technological aspects and only minor dimensions and material differences exist. We believe that FDA can decide the subject device and the predicate device are substantially equivalent.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three abstract human figures or faces forming the staff and a serpent-like shape winding around them.
.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Comfort Orthopedic Co., LTD. % ROC Chinese European Industrial Research Society Dr. Jen, Ke-Min No. 58 Fu-Chiun Street Hsin-Chu City, Taiwan, ROC
Re: K072711
Trade/Device Name: Comfort Powered Wheelchair, LY-EB206 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: II Product Code: ITI Dated: September 21, 2007 Received: September 25, 2007
Dear Dr. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Dr. Jen, Ke-Min
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sincerely yours,
Mark A. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
UIIIruKI
COMFORT ORTHOPEDIC CO., LTD.
www.comfort.com.tw No.120,Nan Shiang Tsuen,Shoei Shang Shiang,Chia-yi,Taiwan,R.O.C.608 TEL : 886-5-2892093 FAX:886-5-2890070
Indications for Use
510 (K) Number ( If Known ): _ _ K
Device Name: COMFORT POWERED WHEELCHAIR, LY-EB206
Indications for Use:
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Mellman
(Division Sign-Off) Division of General. Restorative, and Neurological Devices Page l of 510(k) Number F1
N/A