K Number
K072680

Validate with FDA (Live)

Date Cleared
2007-10-25

(31 days)

Product Code
Regulation Number
892.2040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Drystar AXYS is a free-standing dry film printer used to print diagnostic images on transparent film for viewing on a standard view box. It may be used in any situation in which a hard copy of an image generated by a medical imaging device is required or desirable, including digital mammography.

The Drystar AXYS is a free standing device used to print diagnostic conventional and mammography images on transparent film for viewing on a standard view box. It may be used in any situation in which a hard copy of an image generated by a medical imaging device is required or desirable.

Device Description

The device is the new Drystar AXYS and it is a dry, B/W printer, using the direct thermal printing principle to produce continuous-tone images with medical diagnostic image quality onto plastic sheets which can be viewed on a light box. The printer is sold with two film input trays. Each tray can be adjusted to five different sizes (in inches) of film, including 8x10, 10x12, 11x14, 14x14 and 14x17. Three different types of film can be used in this new device, two for general purpose radiography and a new type of film for mammography, Drystar DT 2 M. The new mammography film comes in only two sizes 8x10 and 10x12. It is thicker than the general purpose radiography film in order to provide a wider range of optical densities. The printer also handles borders for mammography images in a different manner than for regular medical images.

AI/ML Overview

The provided document is a 510(k) summary for the Drystar AXYS, a medical image hardcopy device (printer). It primarily focuses on demonstrating substantial equivalence to a predicate device (Drystar 5500) rather than providing detailed acceptance criteria and a study proving device performance against those criteria in the context of clinical AI/software performance.

Therefore, many of the requested sections (e.g., sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable or not available in this specific document, as it describes a hardware device that prints images, not an AI or software algorithm that analyzes them.

Here's a breakdown based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

This document does not define specific "acceptance criteria" for clinical performance of an AI/software device, nor does it report performance metrics relevant to an AI/software's diagnostic accuracy (e.g., sensitivity, specificity, AUC).

Instead, the closest equivalent to "performance" for this device relates to its printing capabilities and compliance with manufacturing standards.

Acceptance Criteria (Implied)Reported Device Performance
Substantial Equivalence to Predicate Device (Drystar 5500)"The new Drystar AXYS is very similar to the cleared Drystar 5500 in both hardware and software." "The new Drystar AXYS has essentially the same technological characteristics as the cleared Drystar 5500."
Print Diagnostic Images on Transparent Film"The Drystar AXYS is a free-standing dry film printer used to print diagnostic images on transparent film for viewing on a standard view box."
Compliance with MQSA (for mammography)"The Drystar AXYS contains an automatic QC procedure that assures compliance with the Mammography Quality Standards Act (MQSA) of the FDA."
Safety and Electromagnetic Compatibility"It was also tested against and met a number of consensus standards for safety and electromagnetic compatibility."
Support for various film sizes"Each tray can be adjusted to five different sizes (in inches) of film, including 8x10, 10x12, 11x14, 14x14 and 14x17."
Support for mammography film (Drystar DT 2 M)"a new type of film for mammography, Drystar DT 2 M... comes in only two sizes 8x10 and 10x12."

2. Sample Size Used for the Test Set and Data Provenance

  • Not Applicable. This document describes a medical printer, not a diagnostic AI/software that would have a "test set" of patient data for performance evaluation. The "testing" mentioned refers to engineering and quality assurance activities for hardware and software functionality, as well as compliance with standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Not Applicable. As above, no clinical test set requiring expert-established ground truth is described.

4. Adjudication Method for the Test Set

  • Not Applicable. No clinical test set or adjudication is mentioned.

5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was Done

  • No. This document does not mention any MRMC studies. The device is a printer, not a diagnostic aid that would involve human readers interpreting images with and without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study was Done

  • Not Applicable. The device is a printer, not an algorithm. Therefore, there is no "standalone performance" study in the context of an AI/software algorithm.

7. The Type of Ground Truth Used

  • Not Applicable. For a printer, "ground truth" in the clinical AI sense (e.g., pathology, outcomes data) is not relevant. The "truth" would involve whether the printed image accurately reflects the digital image data, has appropriate density, resolution, and is free from artifacts, which would be assessed through technical specifications and quality control, not clinical ground truth.

8. The Sample Size for the Training Set

  • Not Applicable. This is a hardware printer, not an AI model that requires a training set of data.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As above, no training set or its ground truth is relevant for this device.

Summary of the Study (as described in the 510(k) summary):

The "study" described in the document is primarily an equivalence demonstration. It focuses on showing that the Drystar AXYS is substantially equivalent to a previously cleared predicate device, the Drystar 5500.

The key points of the equivalence demonstration and "testing" mentioned are:

  • Comparison to Predicate: Detailed comparison of hardware and software characteristics to the Drystar 5500 (though the detailed table 3.1 is not provided in this excerpt).
  • Intended Use Equivalence: Stating that the Drystar AXYS has the same indications for use as the Drystar 5500.
  • Technological Equivalence: Asserting that the technological characteristics are "identical" to the predicate device.
  • Quality Control Procedures: Stating that the device includes an automatic QC procedure to comply with MQSA. This is a crucial "test" for the mammography application.
  • Consensus Standards Compliance: Testing against and meeting "a number of consensus standards for safety and electromagnetic compatibility."

In essence, the "study" for this device is a regulatory submission proving that a new version of a printer performs its basic function (printing diagnostic images) and meets relevant safety and quality standards, effectively demonstrating it is as safe and effective as a device already on the market. It does not involve clinical performance studies in the way one would assess a diagnostic AI or software.

{0}------------------------------------------------

K072680

OCT 25 2007

510(k) Summary

Drystar AXYS

Common/Classification Name: Medical Image Hard Copy Device 21 CFR 892.2040

Sponsor:

Agfa Healthcare Corporation 10 South Academy Street Greenville, SC 29601

Contact:

Patrick Lynch Agfa Healthcare Corporation 10 South Academy Street Greenville, SC 29601

Prepared: September 19, 2007

A. Legally Marketed Predicate Devices.

The Drystar 5500 printer is legally marketed (cleared) under 510(k) number K023287. The new Drystar AXYS, the subject of the present 510(k), is very similar to the cleared Drystar 5500 in both hardware and software. From the point of view of the hardware and software, the Drystar AXYS is substantially equivalent to the Drystar 5500. From the point of view of the intended use of the device, dry printing general radiology and mammography images on film, the device is substantially equivalent to the Drystar 5500.

B. Device Description.

The device is the new Drystar AXYS and it is a dry, B/W printer, using the direct thermal printing principle to produce continuous-tone images with medical diagnostic image quality onto plastic sheets which can be viewed on a light box. The printer is sold with two film input trays. Each tray can be adjusted to five different sizes (in inches) of film, including 8x10, 10x12, 11x14, 14x14 and 14x17. Three different types of film can be used in this new device, two for general purpose radiography and a new type of film for mammography, Drystar DT 2 M.

The new mammography film comes in only two sizes 8x10 and 10x12. It is thicker than the general purpose radiography film in order to provide a wider

{1}------------------------------------------------

range of optical densities. The printer also handles borders for mammography images in a different manner than for regular medical images.

C. Intended Use.

The Drystar AXYS is a free-standing dry film printer used to print diagnostic images on transparent film for viewing on a standard view box. It may be used in any situation in which a hard copy of an image generated by a medical imaging device is required or desirable, including digital mammography.

D. Substantial Equivalence Summary.

The new Drystar AXYS has the same indications for use as the cleared Drystar 5500.

The new Drystar AXYS has essentially the same technological characteristics as the cleared Drystar 5500. The differences can be seen in Table 3.1 located in Section III.

This premarket notification has described the characteristics of the Drystar 5500 in sufficient detail to assure a substantial equivalence determination.

E. Technological Characteristics.

The technological characteristics of the new Drystar AXYS are identical to those of the predicate device, the Drystar 5500.

F. Testing

The Drystar AXYS contains an automatic QC procedure that assures compliance with the Mammography Quality Standards Act (MQSA) of the FDA. It was also tested against and met a number of consensus standards for safety and electromagnetic compatibility.

G. Conclusions.

This pre-market submission has demonstrated Substantial Equivalence as defined and understood in Sections 513(f)(1) and 513(l)(1) of the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the staff and a snake winding around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 5 2007

Mr. Patrick J. Lynch Manager of Regulatory Affairs AGFA Healthcare Corporation 10 S. Academy Street, P.O. Box 19048 GREENVILLE SC 29602-9048

Re: K072680

Trade/Device Name: Drystar AXYS Regulation Number: 21 CFR 892.2040 Regulation Name: Medical image hardcopy device Regulatory Class: II Product Code: LMC Dated: September 19, 2007 Received: September 21, 2007

Dear Mr. Lynch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C Brogdon

Nancy C. B rogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Drystar AXYS

Indications For Use:

The Drystar AXYS is a free standing device used to print diagnostic conventional and mammography images on transparent film for viewing on a standard view box. It may be used in any situation in which a hard copy of an image generated by a medical imaging device is required or desirable.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR Over-The-Counter Use

logu M. Whg

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

§ 892.2040 Medical image hardcopy device.

(a)
Identification. A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.