K Number
K072672

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-11-28

(68 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

When intended for pedicle screw fixation from T1 -S1, the Sequoia Thoracolumbar Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor and failed previous fusion.

As pedicle screw system places between 1.3 and S1, the indications include Grade 3 or Grade 4 spondylolisthesis, when utilizing autologous bone graft, when affixed to the posterior lumbosacral spine, and intended to be removed after the solid fusion is established.

When intended for non-pedicle, posterior screw fixation of the non-cervical spine (T1-S1), the indications are idiopathic scoliosis, neuromuscular scoliosis with associated paralysis or spasticity, scoliosis with deficient posterior elements such as that resulting from laminectomy or myelomeningocele, spinal fractures (acute reduction or late deformity), degenerative dise disease (back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies), tumor, spondylolisthesis, spinal stenosis and failed previous fusion.

When intended for anterolateral screw, rod and/or cable fixation of the T6-L5 spine, the indications are degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e. scoliosis, and/or lordosis), turnor and failed previous fusion.

After solid fusion occurs, these devices serve no functional purpose and should be removed. In most cases, removal is indicated because the implants are not intended to transfer or support forces developed during normal activities. Any decision to remove the device must be made by the physician and the patient, taking into consideration the patient's general medical condition and the potential risk to the patient of a second surgical procedure.

The use of posterior spinal instrumentation in children has been reported in the literature. The Sequoia Spinal System may be used for non-pedicle posterior use in this patient group.

Device Description

The Sequoia Spinal System consists of various open style screws, rods and connectors and is intended to provide temporary stabilization following surgery to fuse the spine. This system includes titanium spinal pods in varying lengths. Additionally included are screws with a polyaxial design, allowing the surgeon to use a top loading technique for dropping the spinal rod down to the fixation components into a u-shaped opening.

AI/ML Overview

The provided text describes the Abbott Spine Sequoia Spinal System, a spinal fixation device. It does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria through clinical or AI-based evaluations.

Instead, the document focuses on:

  • Substantial Equivalence: The primary method of demonstrating safety and effectiveness for this device is through substantial equivalence to a predicate device (InCompass® Spinal Fixation System). This means the new device is compared to an already legally marketed device with similar indications and technological characteristics.
  • Non-Clinical Performance Data: The submission states that "Laboratory and bench testing results demonstrate that the proposed device is substantially equivalent to the predicate device." This typically refers to mechanical and material testing to ensure the device meets engineering specifications and performs similarly to the predicate in a simulated environment.
  • Absence of Clinical Performance Data: Crucially, the document explicitly states: "Clinical data and conclusions were not needed for this device." This confirms that no clinical study was conducted to demonstrate the device's performance in humans.
  • Indications for Use: Detailed indications for when and how the device can be used in spinal fusion procedures are provided.

Therefore, since no clinical or AI performance study was conducted or reported in this document, I cannot fulfill your request for:

  • A table of acceptance criteria and reported device performance.
  • Sample sizes, data provenance, number of experts for ground truth, adjudication methods, or MRMC/standalone studies for either a test set or training set.
  • The type of ground truth used.

In summary, the provided document indicates that the Abbott Spine Sequoia Spinal System's acceptance was based on non-clinical (laboratory and bench) testing demonstrating substantial equivalence to a predicate device, and no clinical performance data was required or submitted.

{0}------------------------------------------------

K072672

ABBOTT SPINE 510(K) SUMMARY

SUBMITTER:Abbott Spine
ESTABLISHMENT REGISTRATIONNUMBER:1649384
CONTACT PERSON:David PadgettSpecialist, Regulatory AffairsTelephone: (512) 533-1998Fax: (512) 258-0995
NOV 2 8 2007
DATE:20 September 2007
TRADE NAME:Sequoia Spinal System
COMMON NAME:Spinal Fixation System
CLASSIFICATION NAME:SPINAL FIXATION SYSTEM
CLASSIFICATION REFERENCE:21 CFR § 888.3070 (MNH, MNI, NKB)
PREDICATE DEVICE:InCompass® Spinal Fixation System, K021564,Abbott Spine, cleared 10/25/02InCompass® Special 510(k), K023644,Abbott Spine, cleared 11/25/02

DEVICE DESCRIPTION:

The Sequoia Spinal System consists of various open style screws, rods and connectors and is intended to provide temporary stabilization following surgery to fuse the spine. This system includes titanium spinal pods in varying lengths. Additionally included are screws with a polyaxial design, allowing the surgeon to use a top loading technique for dropping the spinal rod down to the fixation components into a u-shaped opening.

{1}------------------------------------------------

INDICATIONS:

When intended for pedicle screw fixation from TI-S1, the indications include immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor, and failed previous fusion.

As a pedicle screw system placed between L3 and S1, the indications include Grade 3 or Grade 4 spondylolisthesis, when utilizing autologous bone graft, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is established.

When intended for non-pedicle, posterior screw fixation of the non-cervical spine (T)-SI), the indications are idiopathic scoliosis, neuromuscular scoliosis with associated paralysis or spasticity, scoliosis with deficient posterior elements such as that resulting from laminectomy or myelomeningocele, spinal fractures (acute reduction or late deformity), degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), tumor, spondylolisthesis, spinal stenosis and failed previous fusion.

When intended for anterolateral screw, rod and or cable fixation of the T6-L5 spine the indications are degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor and failed previous fusion

After solid fusion occurs, these devices serve no functional purpose and should be removed. In most cases, removal is indicated because the implants are not intended to transfer or support forces developed during normal activities. Any decision to remove the device must be made by the physician and the patient, taking into consideration the patient's general medical condition and the potential risk to the patient of a second surgical procedure.

The use of posterior spinal instrumentation in children has been reported in the literature. Sequoia may be used for non-pedicle posterior use in this patient group.

{2}------------------------------------------------

COMPARISON TO PREDICATE DEVICE:

The subject device is the result of modifications to the existing InCompass® Spinal Fixation System. The subject device has the same intended use and is substantially equivalent to the aforementioned predicate device.

PERFORMANCE DATA (NONCLINICAL AND/OR CLINICAL):

Non-Clinical Performance and Conclusions:

Laboratory and bench testing results demonstrate that the proposed device is substantially equivalent to the predicate device.

Clinical Performance and Conclusions:

Clinical data and conclusions were not needed for this device.

{3}------------------------------------------------

Image /page/3/Picture/13 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Abbott Spine % Mr. David Padgett Specialist, Regulatory Affairs 5301 Riata Park Court, Bldg. F Austin, Texas 78727

NOV 2 8 2007

Re: K072672

Trade/Device Name: Sequoia Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle Screw Spinal System Regulatory Class: III Product Code: NKB, MNI, MNH Dated: November 5, 2007 Received: November 6, 2007

Dear Mr. Padgett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 – Mr. David Padgett

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N Mulhearn

  • Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): TBD

Device Name: Abbott Spine Sequoia Spinal System

Indications for Use:

When intended for pedicle screw fixation from T1 -S1, the Sequoia Thoracolumbar Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor and failed previous fusion.

As pedicle screw system places between 1.3 and S1, the indications include Grade 3 or Grade 4 spondylolisthesis, when utilizing autologous bone graft, when affixed to the posterior lumbosacral spine, and intended to be removed after the solid fusion is established.

When intended for non-pedicle, posterior screw fixation of the non-cervical spine (T1-S1), the indications are idiopathic scoliosis, neuromuscular scoliosis with associated paralysis or spasticity, scoliosis with deficient posterior elements such as that resulting from laminectomy or myelomeningocele, spinal fractures (acute reduction or late deformity), degenerative dise disease (back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies), tumor, spondylolisthesis, spinal stenosis and failed previous fusion.

When intended for anterolateral screw, rod and/or cable fixation of the T6-L5 spine, the indications are degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e. scoliosis, and/or lordosis), turnor and failed previous fusion.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

510(k) NumberK07 2672
-------------------------

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): TBD

Device Name: Abbott Spine Sequoia Spinal System

After solid fusion occurs, these devices serve no functional purpose and should be removed. In most cases, removal is indicated because the implants are not intended to transfer or support forces developed during normal activities. Any decision to remove the device must be made by the physician and the patient, taking into consideration the patient's general medical condition and the potential risk to the patient of a second surgical procedure.

The use of posterior spinal instrumentation in children has been reported in the literature. The Sequoia Spinal System may be used for non-pedicle posterior use in this patient group.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE

Mark N. Milheran

(Division Sign-Division of Ceneral Restorative, and Neurological Devices

510(k) Number

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.