(15 days)
The GE Lunar DPX and Prodigy Series DEXA Bone Densitometers with Mobile Kit provides a measurement of bone mineral density and fat and lean tissue mass. The values can then be compared to a reference population at the sole discretion of the physician.
The GE Lunar DPX and Prodigy Series DEXA Bone Densitometers with Mobile Kit is a full-featured general purpose x-ray bone densitometer system. It consists of a scan table that provides x-ray source generation, x-ray detection and processing. The user interface includes a computer workstation. This modification will provide users the option of mounting the bone densitometer system into a vehicle through the use of a mechanical hardware mobile kit used during the installation of the scanner and by the Operator during transport.
The provided text describes a 510(k) premarket notification for a medical device and thus does not contain information on dedicated "acceptance criteria" and a "study" to prove the device met those criteria, as one might find in a clinical trial or performance study report. Instead, it focuses on demonstrating substantial equivalence to a predicate device.
Here's an analysis based on the information provided, framed to address your questions where possible:
1. Table of Acceptance Criteria and Reported Device Performance
No explicit acceptance criteria with numerical targets are provided in the document. The performance is assessed through "in vitro precision and accuracy values" which were within "design specifications" (generic statement).
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Precision | "Within design specifications" (in vitro on phantoms) |
| Accuracy | "Within design specifications" (in vitro on phantoms) |
| Safety | Conforms with applicable medical device safety standards |
| Effectiveness | No specific numerical performance metrics are reported; deemed effective through substantial equivalence to predicate device. |
| Technological Characteristics | Comparable to predicate device. |
| Key Safety & Effectiveness Features | Comparable to predicate device. |
| Design, Construction, Materials | Similar to predicate device. |
| Intended Uses & Basic Operating Modes | Same as predicate device. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not applicable. The document states that "No clinical tests were required to establish safety or effectiveness." The testing mentioned was "in vitro precision and accuracy values ... computed through a series of tests on phantoms." The sample size (number of phantoms or tests conducted) is not specified.
- Data Provenance: Not applicable, as no clinical human data was used for effectiveness testing. The "in vitro" tests were performed by the manufacturer, within the US (GE Medical Systems Lunar, Madison, WI).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. The in vitro tests on phantoms do not involve human expert adjudication for ground truth.
4. Adjudication Method for the Test Set
Not applicable. No expert adjudication method was used given the nature of the in vitro phantom testing.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance
Not applicable. This device is a bone densitometer, not an AI-assisted diagnostic tool, and no MRMC study was performed or required.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. The device itself is the "standalone" component in the context of phantom testing, but this is not an AI algorithm assessment. The document clarifies that the device includes a "computer workstation" for processing, but this is inherent to DEXA technology, not an AI application subject to standalone algorithm evaluation in the modern sense.
7. The Type of Ground Truth Used
For the in vitro precision and accuracy tests, the "ground truth" would have been established by the known physical properties of the phantoms used and the expected output values for those phantoms, based on standard metrology. This is an objective, physical ground truth.
8. The Sample Size for the Training Set
Not applicable. This device does not use machine learning or AI that would require a "training set" in the conventional sense. The "design specifications" would have been developed through engineering and physics principles, not data-driven training.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for an AI/ML algorithm.
{0}------------------------------------------------
SEP 1 4 2007
K 072439
Special 510(k) Premarket Notification
GE Medical Systems Lunar – DPX and Prodigy Series DEXA Bone Densitometers with Mobile Kit August 28th, 2007
510 (k) Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).
GE Healthcare
Image /page/0/Picture/7 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined within a circular border. The logo is in black and white, with the letters and border appearing in white against a black background.
Section a):
| 1. | Submitter: | GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC division ofGeneral Electric CompanyGE Medical Systems Lunar (business name)726 Heartland TrailMadison, WI 53717 |
|---|---|---|
| Contact Person: | James P. RaskobSafety and Regulatory Engineering ManagerTelephone: 608-826-7425; Fax: 608-299-2132 | |
| Date Prepared: | August 28th, 2007 | |
| 2. | Device Name: | GE Lunar DPX and Prodigy Series DEXA Bone Densitometers with Mobile Kit,21 CFR 892.1170, 90-KGI |
| 3. | Marketed Device: | DPX and Prodigy series bone densitometers: K915535 K982267 K983564K000826 K001756 K001812 K011917 K023554 currently in commercialdistribution. |
| 4. | Device Description: | The GE Lunar DPX and Prodigy Series DEXA Bone Densitometers with MobileKit is a full-featured general purpose x-ray bone densitometer system. Itconsists of a scan table that provides x-ray source generation, x-raydetection and processing. The user interface includes a computerworkstation. This modification will provide users the option of mounting thebone densitometer system into a vehicle through the use of a mechanicalhardware mobile kit used during the installation of the scanner and by theOperator during transport. |
| 5. | Indications for Use: | The GE Lunar DPX and Prodigy Series DEXA Bone Densitometers with Mobile Kitprovides an estimate of bone mineral density and fat and lean tissue mass. Thevalues can then be compared to a reference population at the sole discretion of |
the physician.
{1}------------------------------------------------
Special 510(k) Premarket Notification
GE Medical Systems Lunar ~ DPX and Prodiqy Series DEXA Bone Densitometers with Mobile Kit August 28th, 2007
6. Comparison with
| Predicate Device: The GE Lunar DPX and Prodigy Series DEXA Bone Densitometers with Mobile |
|---|
| Kit is of a comparable type and substantially equivalent to the current DPX and |
| Prodigy series bone densitometers. It has the same technological |
| characteristics, is comparable in key safety and effectiveness features, it |
| utilizes similar design, construction, and materials, and has the same intended |
| uses and basic operating modes as the predicate device. |
Section b):
-
- Non-clinical Tests: The device has been evaluated for safety, and has been found to conform with applicable medical device safety standards. In vitro precision and accuracy values were computed through a series of tests on phantoms and were within design specifications.
-
- Clinical Tests: No clinical tests were required to establish safety or effectiveness.
-
- Conclusion: Intended uses and other key features are consistent with previously cleared bone densitometers. The design and development process of the manufacturer conforms with 21 CFR 820 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards and compliance was verified through independent evaluation with ongoing factory surveillance. The GE Lunar DPX and Prodigy Series DEXA Bone Densitometer with Mobile Kit is substantially equivalent to currently marketed devices. No new safety and effectiveness questions are raised with the GE Lunar DPX and Prodigy Series DEXA Bone Densitometers with Mobile Kit option.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo on the left and the word "DEPARTMENT" on the right. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or feathers. The word "DEPARTMENT" is in a bold, sans-serif font, and it appears to be part of a larger heading or title.
SEP 1 4 2007
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Mr. James P. Raskob Safety and Regulatory Engineering Manager GE Medical Systems Lunar GE Healthcare 726 Heartland Trail MADISON WI 53717
Re: K072439
Trade/Device Name: GE Lunar DPX and Prodigy Series DEXA Bone Densitometer with Mobile Kit Regulation Number: 21 CFR 892.1190 Regulation Name: Bone densitometer Regulatory Class: II Product Code: KGI Dated: August 28, 2007 Received: August 30, 2007
Dear Mr. Raskob:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows a circular logo with the text "1906-2006" at the top and "Centennial" at the bottom. The letters "FDA" are prominently displayed in the center of the logo. Below the logo, there is some text in a decorative font that reads "Protecting and Promoting Public Health."
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely vours.
Nancy Chrogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Special 510{k} Premarket Notification GE Medical Systems Lunar – DPX and Prodigy Series DEXA Bone Densitometers with Mobile Kit August 28th, 2007
Attachment E
Indications for Use
The GE Lunar DPX and Prodigy Series DEXA Bone Densitometers with Mobile Kit provide a measurement of bone mineral density and fat and lean tissue mass. The values can then be compared to a reference population at the sole discretion of the physician.
Prescription Use XXX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(P) EASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANQTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hulu Lewen
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K072439
N/A