K Number
K072363

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-01-31

(161 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BIOMET 3i dental implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading, or as a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.

BIOMET 31 NanoTite dental implants are intended for immediate function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function.

Device Description

BIOMET 3i NanoTite Dental Implants cleared by K051461, K063341 & K063286 have both the proprietary OSSEOTITE acid-etched surface and NanoTite CaP Discrete Crystalline Deposition (DCD) surface treatment.

The devices themselves have not been modified since their original clearance. Specific performance claims for NanoTite Dental Implants are being modified. Two independent prospective randomized controlled studies provide clinical data to support the requested claims.

AI/ML Overview

The provided text describes a 510(k) premarket notification for BIOMET 3i NanoTite™ Dental Implants. This is a medical device submission, and the content focuses on establishing substantial equivalence to a predicate device and supporting claims for use. It is not a study about an AI-powered device, and therefore, most of the requested information (acceptance criteria, device performance, sample sizes, ground truth, MRMC studies, standalone performance, training set details) is not applicable.

However, I can extract the relevant information regarding the device itself and the type of evidence provided:

Device: BIOMET 3i NanoTite™ Dental Implants
Classification Name: Endosseous dental implant
Product Code: DZE
Predicate Device: K051461 - 3i Osseotite (Nano CaP) Dental Implant

Study Information (as pertains to this 510(k) submission):

The submission refers to changes in "specific performance claims for NanoTite Dental Implants," which are supported by clinical data.

Description of Study:

"Two independent prospective randomized controlled studies provide clinical data to support the requested claims."

Type of Ground Truth Used: Clinical data (implied patient outcomes and immediate function assessment).

Conclusion of the Submission:

"The clinical data submitted for the NanoTite dental implant compared to the control implant supports the modifications to specific performance claims requested within this premarket notification."

Regarding your specific questions, this document does not contain the following information as it pertains to an AI/Software as a Medical Device (SaMD):

  1. A table of acceptance criteria and the reported device performance: This document reports modifications to performance claims, not specific quantitative acceptance criteria or performance metrics for an AI device.
  2. Sample sizes used for the test set and the data provenance: The document states "Two independent prospective randomized controlled studies" were conducted but does not provide sample sizes, country of origin, or specific details of the test sets.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable for a dental implant study. The ground truth refers to clinical outcomes.
  4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI device.
  7. The type of ground truth used: As mentioned, it's "clinical data" from patient studies, which would involve assessing implant success, immediate function, etc. This is not an expert consensus on interpreting images for an AI device.
  8. The sample size for the training set: Not applicable, as this is not an AI device.
  9. How the ground truth for the training set was established: Not applicable, as this is not an AI device.

In summary, this 510(k) document pertains to a physical dental implant device and its clinical performance, not an AI software device. Therefore, the requested information about AI model acceptance criteria, ground truth establishment by experts, and MRMC studies is not present.

{0}------------------------------------------------

K072363

MET31

Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted inaccordance with the requirements of 21 CFR § 807.92
SubmitterBiomet 3i, Inc.4555 Riverside DrivePalm Beach Gardens, FL 33410
ContactDiana TaylorManager, Regulatory AffairsBiomet 3i, Inc.4555 Riverside DrivePalm Beach Gardens, FL 33410Tel. 561-776-6857Fax. 561-776-6852JAN 31 2008
Date PreparedAugust 21, 2007
Device NameBIOMET 3i NanoTite™ Dental Implants
Classification NameEndosseous dental implant
ClassificationClass II, 21 CFR § 872.3640
Product CodeDZE
Predicate Device(s)K051461 - 3i Osseotite (Nano CaP) Dental Implant
Device DescriptionBIOMET 3i NanoTite Dental Implants cleared by K051461,K063341 & K063286 have both the proprietary OSSEOTITEacid-etched surface and NanoTite CaP Discrete CrystallineDeposition (DCD) surface treatment.
The devices themselves have not been modified since their originalclearance. Specific performance claims for NanoTite DentalImplants are being modified. Two independent prospectiverandomized controlled studies provide clinical data to support therequested claims.
Indications for UseBIOMET 3i dental implants are intended for surgical placement inthe upper or lower jaw to provide a means for prostheticattachment in single tooth restorations and in partially or fullyedentulous spans with multiple single teeth utilizing delayed orimmediate loading, or as a terminal or intermediary abutment forfixed or removable bridgework, and to retain overdentures.
BIOMET 3i NanoTite dental implants are intended for immediatefunction on single tooth and/or multiple tooth applications whengood primary stability is achieved, with appropriate occlusalloading, in order to restore chewing function.
ConclusionThe clinical data submitted for the NanoTite dental implantcompared to the control implant supports the modifications tospecific performance claims requested within this premarketnotification.

:

08/21/07

:

{1}------------------------------------------------

:

.

.

:

:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 31 2008

Ms. Diana Taylor Manager, Regulatory Affairs Biomet 31, Incorporated 4555 Riverside Drive Palm Beach Gardens, Florida 33410

Rc: K072363

Trade/Device Name: BIOMET 3i NanoTite Dental Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: January 11, 2008 Received: January 17, 2008

Dear Ms. Taylor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Taylor

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Syute y. Michaudmp

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): Unknown

Device Name: BIOMET 3i NanoTite Dental Implants

Indications for Use:

BIOMET 3i dental implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading, or as a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.

BIOMET 31 NanoTite dental implants are intended for immediate function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Perry

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page of

510(k) Number:

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.