K Number
K072279

Validate with FDA (Live)

Date Cleared
2007-09-21

(36 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OASIS MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Device Description

The OASIS is a Magnetic Resonance Imaging System that utilizes a 1.2 Tesla superconducting open magnet in an open gantry design was based on the ECHELON MRI system. The OASIS has been designed to enhance clinical utility as compared to the ECHELON by taking advantage of open architecture.

AI/ML Overview

The provided text does not contain information related to specific acceptance criteria or a study proving the device meets those criteria, as typically understood in the context of device performance metrics, clinical trials, or comparative effectiveness studies.

Instead, this document is a 510(k) premarket notification for a Magnetic Resonance Imaging (MRI) system. The core of a 510(k) submission is to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving performance against specific quantitative acceptance criteria through a dedicated study.

Here's what can be extracted and inferred from the document:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated in terms of performance metrics (e.g., sensitivity, specificity, accuracy). The acceptance criteria for a 510(k) are implicitly met by demonstrating substantial equivalence to a predicate device, which means having the same intended use, similar technological characteristics, and safety and effectiveness that are at least as good as the predicate.
  • Reported Device Performance: Not reported in quantitative performance metrics. The document states:
    • "MRI is capable of producing high quality anatomical images without the associated risks of ionizing radiation." (General statement, not specific performance.)
    • "The technological characteristics of this device are similar to the primary predicate device. The primary difference is the 1.2 Tesla superconducting open magnet. The control and image processing hardware and the base elements of the system software are identical to the predicate device." (Focuses on similarity to predicate, not explicit performance.)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable / Not provided. No specific test set or associated data is mentioned for performance evaluation. This submission relies on demonstrating similarity to a predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable / Not provided. No test set requiring expert ground truth is described. The "ground truth" for this submission is established by the accepted performance and safety of the predicate device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable / Not provided. No test set or adjudication process is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. An MRMC study was not done. This device is an MRI system itself, not an AI-assisted diagnostic tool for readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No. This is an MRI system, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Implicitly the established safety and effectiveness of the predicate device. The "ground truth" for this application is that the predicate device (ECHELON Magnetic Resonance Imaging System, K052172) is considered safe and effective for its intended use, and the new device is substantially equivalent.

8. The sample size for the training set

  • Not applicable / Not provided. There is no mention of a training set for an algorithm or model.

9. How the ground truth for the training set was established

  • Not applicable / Not provided. No training set is discussed.

In summary: This 510(k) submission for the OASIS MRI system is a "substantial equivalence" claim. It does not present a study with explicit acceptance criteria, performance metrics, or data from a test set as would be found for a novel diagnostic algorithm or software. Its acceptance is based on demonstrating that it is as safe and effective as a previously cleared device with similar intended use and technological characteristics.

{0}------------------------------------------------

K072279
pg. 1 of 3

1 2007

Submitter Information

Submitter:Hitachi Medical Systems America, Inc.1959 Summit Commerce ParkTwinsburg, Ohio 44080-2371ph: (330) 425-1313fax: (330) 425-1410SEP 2
Contact:Douglas J. Thistlethwaite
Date:08/13/2007
Device Name
Classification Name:System, Nuclear Magnetic Resonance Imaging
Classification Number:90LNH
Trade/Proprietary Name:OASIS Magnetic Resonance Imaging System
Predicate Device(s):ECHELON Magnetic Resonance Imaging System(K052172)

Device Intended Use

The MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

{1}------------------------------------------------

Device Description

Function

The OASIS is a Magnetic Resonance Imaging System that utilizes a 1.2 Tesla superconducting open magnet in an open gantry design was based on the ECHELON MRI system. The OASIS has been designed to enhance clinical utility as compared to the ECHELON by taking advantage of open architecture.

Scientific Concepts

Magnetic Resonance Imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nuclei currently used in magnetic resonance imaging. When placed in a static magnetic field, these nuclei assume a net orientation or alignment with the magnetic field, referred to as a net magnetization vector. The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a re-orientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adjacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2.

A RF emission or echo that can be measured accompanies these relaxation events. The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varying the RF excitation, applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.

Physical and Performance Characteristics

MRI is capable of producing high quality anatomical images without the associated risks of ionizing radiation. The biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In MR imaging, difference in proton density, blood flow, and T1 and T2 relaxation times can all contribute to image contrast. By varying the pulse sequence characteristics, the resulting images can emphasize T1, T2, proton density, or the molecular diffusion of water or other proton containing molecules.

{2}------------------------------------------------

K172279
pg. 3 of 3

Device Technological Characteristics

The technological characteristics of this device are similar to the primary predicate device. The primary difference is the 1.2 Tesla superconducting open magnet. The control and image processing hardware and the base elements of the system software are identical to the predicate device.

Conclusions

It is the opinion of Hitachi Medical Systems America, Inc. that OASIS MRI system is substantially equivalent to the listed predicate device. The intended use is identical to the listed predicate device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

SEP 2 1 2007

Mr. Doug Thistlethwaite Manager, Regulatory Affairs Hitachi Medical Systems America, Inc. 1959 Summit Commerce Park TWINSBURG OH 44087

Re: K072279

Trade/Device Name: OASIS, w/V1.0A Operating System Software Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: August 13, 2007 Received: August 16, 2007

Dear Mr. Thistlethwaite:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affective your. device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/3/Picture/9 description: The image shows a circular logo with the text "1906-XXX" at the top. The letters "FDA" are prominently displayed in the center of the logo. Below the letters, the word "Centennial" is written in a smaller font. Three stars are arranged in a horizontal line beneath the word "Centennial."

Promoting and Promoting Public Health

{4}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology).240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.litml.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): Device Name: OASIS, w/V1.0A Operating System Software

Indications for Use:

The OASIS MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Anatomical Region:Head, Body, Spine, Extremities
Nucleus excited:Proton
Diagnostic uses:T1, T2, proton density weighted imagingDiffusion weighted imagingMR AngiographyImage processing

Prescription Use × AND/OR

Over-the-Counter Use

(21 CFR 801 Subpart C)

(Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Lornim Wh

Radiological De 510(k) Number

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.