(212 days)
The HF54 interferential therapy and premodulated therapy is indicated for: Pain relief for: Symptomatic relief of chronic intractable pain and/or management of traumatic or post surgical pain.
Ultrasound therapy is available from the HF54 and indicated for: Applying therapeutic deep heat within body tissues for the treatment of selected chronic and sub-chronic medical conditions such as: Relief of pain joint contractures This can be done using One or two ultrasound applicators. The second ultrasound applicator is optional. Ultrasound therapy, either alone or in combination with interferential therapy or premodulated therapy can be used in a scanned manner (where the therapist moves the applicator) or for specific treatment areas identified in the operator's manual, the applicator can be used in a hands-free manner (where the applicator remains stationary).
An Infrared Therapy Probe is available as an optional accessory (HF023) for use with the Hill Laboratories HF54 Combination Ultrasound Interferential and premodulated stimulation system or for stand-alone use with an optional external medical grade, isolated power supply. It is used to provide topical heating for: Temporary increases in local blood flow and circulation. Temporary relief of minor muscle and joint aches Temporary relief of pain and stiffness Relaxation of muscles. Temporary relief (or relaxation) of muscle spasms. Temporary relief of minor pain and stiffness associated with arthritis
The HF54 Combination Ultrasound Interferential and Premodulated Stimulation System with optional hands-free operation is a device that provides ultrasonic diathermy, interferential current therapy, and infrared lamp therapy. It can be used with one or two ultrasound applicators, and the ultrasound therapy can be applied in a scanned or hands-free manner. An optional infrared light probe is also available.
The provided 510(k) summary for the Hill Laboratories HF54 Combination Ultrasound Interferential and Premodulated Stimulation System with optional hands-free operation (K072256) does not describe a study involving a test set, expert adjudication, MRMC comparative effectiveness, or stand-alone algorithm performance in the typical sense of a clinical trial or algorithm validation study for diagnostic devices.
Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (K062256) by affirming compliance with mandatory and consensus standards, and by performing risk analysis for the expanded indication of "hands-free" operation for ultrasound. The "acceptance criteria" here refer to safety thresholds and compliance with established performance standards for similar medical devices.
Here's an breakdown based on the provided text, addressing your points where applicable:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Safety/Performance Standard) | Reported Device Performance (as demonstrated by study/analysis) |
|---|---|
| Temperatures in tissues above 45 °C | With an intensity of ≤ 10W, the temperature of Tissue or Bone would not rise above 45 °C when used as directed. (See Attachment C-1, not provided in this excerpt) |
| Temperature rise in bone | With an intensity of ≤ 10W, the temperature of Tissue or Bone would not rise above 45 °C when used as directed. (See Attachment C-1, not provided in this excerpt) |
| Presence of Standing Waves | The ultrasound output is pulsed and will not support standing waves. |
| Self-heating of the ultrasound applicator | Measurements during the original submission (K062256) demonstrated that the applicator surface temperature did not exceed 45 °C when stationary, over the whole treatment time. Measurements were made at equal distances across the surface of the treatment surface. |
| Compliance with 21 CFR 1050.10 (Therapeutic Ultrasound) | The HF54 Combination Ultrasound Interferential and Premodulated Stimulation System with optional hands-free operation complies with this mandatory standard. |
| Compliance with IEC 60601-1 +A1, +A2 (General Safety) | The device is designed to comply with this consensus standard. |
| Compliance with UL 60601-1 (General Safety) | The device is designed to comply with this consensus standard. |
| Compliance with IEC 60601-1-2 (EMC) | The device is designed to comply with this consensus standard. |
| Compliance with IEC 60601-2-5 (Safety of Ultrasonic Physiotherapy Equipment) | The device is designed to comply with this consensus standard. |
| Compliance with IEC 60601-2-10 (Safety of Nerve and Muscle Stimulators) | The device is designed to comply with this consensus standard. |
| Operator monitoring during hands-free use (patient falling asleep) | The operator's manual requires that the operator remain in the same area as the patient and must monitor the patient to ensure that the patient has not fallen asleep. (This is a procedural control, not a device performance metric). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not describe a "test set" in the context of clinical data or images. The "tests" mentioned are primarily engineering and electrical safety measurements (e.g., thermal measurements, pulsed output analysis) for the device itself.
- Sample Size: Not applicable in the context of a clinical test set. The measurements were likely performed on a single or a small number of device units.
- Data Provenance: The measurements related to thermal output and standing waves were conducted internally by Hill Laboratories, likely in a laboratory setting. This is a technical validation, not clinical data from patients.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. Ground truth, in the sense of expert consensus on clinical diagnoses or outcomes, is not described as part of this submission. The "ground truth" for the engineering performance aspects would be the physical laws and measurement standards.
4. Adjudication Method for the Test Set
Not applicable. There is no clinical test set requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. This submission is for an expanded indication for a physical therapy device, focusing on safety and technical equivalence rather than a diagnostic or screening efficacy study involving human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical therapy equipment, not an AI algorithm for diagnosis or image analysis. Its operation inherently involves a human operator (therapist). The "hands-free" aspect refers to the applicator staying stationary on the patient, but the therapist still monitors the patient.
7. The Type of Ground Truth Used
The "ground truth" here is based on:
- Engineering and physical measurements: For thermal output, acoustic properties (standing waves), and electrical safety.
- Compliance with established standards: Mandatory (21 CFR 1050.10) and consensus standards (IEC 60601 series, UL 60601-1).
- Risk analysis: Identifying potential hazards and mitigation strategies (e.g., operator manual instructions).
8. The Sample Size for the Training Set
Not applicable. There is no training set in the context of machine learning or AI algorithms described in this submission.
9. How the Ground Truth for the Training Set was Established
Not applicable. As there is no training set, there's no ground truth established for one.
{0}------------------------------------------------
510(k) SUMMARY
For
Hill Laboratories
HF54 Combination Ultrasound Interferential and Premodulated Stimulation System with optional hands-free operation
l. Submitter's Name and Address
| Submitter's Name: | Hill Laboratories | MAR 12 2008 |
|---|---|---|
| Address: | 3 Bacton Hill Rd | |
| City, State, and Zip: | Frazer, PA 19355 |
2. Contact Person
| Name: | Brady Aller | |
|---|---|---|
| Title: | Sales/Service Manager | |
| Telephone: | (610) | 644-2867 |
| Facsimile: | (610) | 647-6297 |
| E-mail: | bradyaller@hilllabs.com |
3, Manufacturing Facility Address
| Manufacturer: | Hill Laboratories |
|---|---|
| Address: | 3 Bacton Hill Rd |
| City, State, and ZIP: | Frazer, PA 19355 |
4. Establishment Registration Number
Establishment Registration 2510425 Number:
5. Reason for Submission
Expanded Indications for Use
6. Device Details
| Proprietary or TradeName: | HF54 Combination Ultrasound Interferential andPremodulated Stimulation System with optionalhands-free operation |
|---|---|
| Common Name: | Ultrasonic Diathermy |
{1}------------------------------------------------
7. Device Common Name, Classification, Product Code & CFR No.
| Common Name | Class | ProCode | CFR |
|---|---|---|---|
| Ultrasonic Diathermy | 2 | IMI | 890.5300 |
| Interferential Current Therapy | 2 | LIH | |
| Infrared lamp | 2 | ILY | 890.5500 |
8. Classification Name
- (i) diathermy, ultrasonic, for use in applying therapeutic deep heat
- (ii) interferential current therapy
- lamp, infrared, therapeutic heating (iii)
9. Device Classification Panel
Physical Medicine & Neurology
10. Indications for Use
Interferential and Premodulated Modes 10.1
- Pain relief for:
- Symptomatic relief of chronic intractable pain and/or management of traumatic . or post surgical pain.
10.2 Ultrasound Therapy
Ultrasound therapy is available from the HF54 and indicated for:
Applying therapeutic deep heat within body tissues for the treatment of selected chronic and sub-chronic medical conditions such as:
- . Relief of pain
- . Joint contractures
This can be done using One or two ultrasound applicators. The second ultrasound applicator is optional.
Ultrasound therapy, either alone or in combination with interferential therapy or premodulated therapy can be used in a scanned manner (where the therapist moves the applicator) or for specific treatment areas identified in the operator's manual, the applicator can be used in a hands-free manner (where the applicator remains stationary).
Infrared Therapy 10.3
An Infrared light probe is available as an optional accessory (HF023) for use with the Hill Laboratories HF54 Combination Ultrasound Interferential and premodulated stimulation system or with an optional external medical grade, isolated power supply. It is used to provide topical heating for:
- Temporary increases in local blood flow and circulation .
- . Temporary relief of minor muscle and joint aches
- . Temporary relief of pain and stiffness
- . Relaxation of muscles
- . Temporary relief (or relaxation) of muscle spasms
- . Temporary relief of minor pain and stiffness associated with arthritis
December 10, 2007
{2}------------------------------------------------
11. Standards
11.1 Mandatory Standards
21 CFR 1050.10 is applicable to therapeutic ultrasound. The HF54 Combination Ultrasound Interferential and Premodulated Stimulation System with ontional hands-free operation complies with this mandatory standard.
11.2 Consensus Standards
The HF54 Combination Ultrasound Interferential and Premodulated Stimulation System with optional hands-free operation is designed to comply with the following Consensus Standards:
| STANDARD NO. | TITLE |
|---|---|
| IEC 60601-1 +A1, +A2 | Medical Electrical Equipment- Part 1: General Requirements forSafety |
| UI. 60601-1 | Medical Electrical Equipment- Part 1: General Requirements forSafety |
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General requirements forsafety - Collateral standard: Electromagnetic compatibility -Requirements and tests |
| IEC 60601-2-5 | Medical electrical equipment - Part 2-5: Particular requirements forthe safety of ultrasonic physiotherapy equipment |
| IEC 60601-2-10 | Medical electrical equipment. Part 2: Particular requirements for thesafety of nerve and muscle stimulators |
12. Predicate Devices
| 510(k) Number | Trade or Proprietary or ModelName | Manufacturer | Class |
|---|---|---|---|
| K062256 | HF 54 | Hill Laboratories | 2 |
{3}------------------------------------------------
Substantial Equivalence (SE) Rationale 12.1
12.1.1 Technology
The HF54 with optional hands-free operation offers the same electrical stimulation, ultrasonic therapy and/or combination of the two and shares the same characteristics including waveforms, operating frequencies and the same use in physical medicine and neurology as the predicate device. The device with its proposed extended indications for use is intended to be used only by a qualified therapist.
12.1.2 Standards
The ultrasound output and electrical stimulation currents are consistent with FDA guidance and international standards. The new device is in compliance to the same standards.
12.1.3 Materials
The materials used in construction of the device and the method of information display are identical. The measured parameters for the proposed HF54 are the identical to those displayed on the device cleared under K062256. The software has not been changed from the cleared device.
12.1.4 Risk Analysis
The expanded indications for use (stationary ultrasound) needed a review of the risks posed by a stationary technique. The primary risks were identified as:
- a) Temperatures in tissues above 45 °C
- b) Temperature rise in bone
- Standing waves c)
- d) Self-heating of the ultrasound applicator
- The patient falling asleep during hands-free operation possibly leading to e) overheating of tissues.
Temperatures in Tissues and In Bone 12.1.4.1
From Measurement and Thermal Index Calculations, it was shown that with an intensity of ≤ 10W, the temperature of Tissue or Bone would not rise above 45 °C when used as directed. See Attachment C-1. There is no new safety issue.
12.1.4.2 Standing Waves
The ultrasound output is pulsed and will not support standing waves. There is no new safety issue.
12.1.4.3 Self Heating of the Ultrasound Applicator
Measurements made during the original submission for the HF54 included thermal measurements of the ultrasound applicator surface. The tests demonstrated that the applicator surface temperature did not exceed 45 °C when stationary, over the whole treatment time. Measurements were made at equal distances across the surface of the treatment surface. There is no new safety issue.
HF54 With Optional Hands-free operation
{4}------------------------------------------------
Patient falling asleep during treatment 12.1.4.4
The operator's manual requires that the operator remain in the same area as the patient and must monitor the patient to ensure that the patient has not fallen asleep.
There are no new treatment modes, identical circuitry and software are used. The changes therefore do not change the effectiveness of the device.
12.2 Conclusion
The proposed HF54 with Optional Hands-free operation when used as directed in the operator's manual presents no new safety or effectiveness concerns and is Substantially Equivalent to the current version of the HF54 cleared under K062256.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 5, 2014
Hill Laboratories c/o Mr. Brady Aller 3 Bacton Hill Rd. Frazer, PA 19355
Re: K072256
Trade/Device Name: HF54 Combination Ultrasound Interferential and Premodulated Stimulation System with optional hands-free operation Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic Diathermy Regulatory Class: II Product Code: PFW; ILY; LIH Dated: December 10, 2007 Received: December 13, 2007
Dear Mr. Aller:
This letter corrects our substantially equivalent letter of March 12, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 - Mr. Brady Aller
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Joyce M. Whang -S
for Carlos Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) No. If known
Indications For Use statement - Interferential and Premodulated Modes
Device Name:
HF54 Combination Ultrasound Interferential and Premodulated Stimulation System with optional hands-free operation
Indications For Use:
The HF54 interferential therapy and premodulated therapy is indicated for:
Pain relief for:
Symptomatic relief of chronic intractable pain and/or management of traumatic or post surgical pain.
Continued on next page
Prescription Use ×
AND/OR
Over-The-Counter Use
(Per 21 CFR 801 Subpart D
(Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nelke Dsl for nxm
Division Size: 18
Division of General Restorative
and Neurological Devices
510(1) * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * *
HF54 With Optional Hands-free operation
December 10, 2007
Page 14 of 70
{8}------------------------------------------------
510(k) No. If known
Indications For Use statement - Ultrasound Therapy
| Device Name: | HF54 Combination Ultrasound Interferential and Premodulated Stimulation System with optional hands-free operation |
|---|---|
| Indications For Use: | Ultrasound therapy is available from the HF54 and indicated for:Applying therapeutic deep heat within body tissues for the treatment of selected chronic and sub-chronic medical conditions such as:Relief of painjoint contracturesThis can be done using One or two ultrasound applicators. The second ultrasound applicator is optional.Ultrasound therapy, either alone or in combination with interferential therapy or premodulated therapy can be used in a scanned manner (where the therapist moves the applicator) or for specific treatment areas identified in the operator's manual, the applicator can be used in a hands-free manner (where the applicator remains stationary). |
Continued on next page
| Prescription Use | X | AND/OR | Over-The-Counter Use |
|---|---|---|---|
| ------------------ | --- | -------- | ---------------------- |
Prescription Use X Over-The-Counter Use
(Per 21 CFR 801 Subpart D
(Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Nilpeaglh formxm
Division of General, Restorative, and Neurological Devices
510(k) Number: K072256
HF54 With Optional Hands-free operation
December 10, 2007
{9}------------------------------------------------
510(k) No. If known
Indications For Use statement -Infrared Applicator
Device Name:
Infrared Therapy (using optional HF023)
Indications For Use:
An Infrared Therapy Probe is available as an optional accessory (HF023) for use with the Hill Laboratories HF54 Combination Ultrasound Interferential and premodulated stimulation system or for stand-alone use with an optional external medical grade, isolated power supply. It is used to provide topical heating for:
- Temporary increases in local blood flow and circulation .
- Temporary relief of minor muscle and joint aches
- Temporary relief of pain and stiffness
- Relaxation of muscles ●
- . Temporary relief (or relaxation) of muscle spasms
- . Temporary relief of minor pain and stiffness associated with arthritis
| Prescription Use | X | AND/OR |
|---|---|---|
| (Per 21 CFR 801 Subpart D |
Over-The-Counter Use
(Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil Ogden for mxm
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number K072256
HF54 With Optional Hands-free operation
December 10, 2007
Page 16 of 70
N/A