K Number
K072250

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-11-09

(88 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SuperCable" Grip and Plate System is indicated for use where wire, cable, or band cerclage is used in combination with a trochanteric grip or bone plate. The SuperCable™ Grip and Plate System is intended to be used in conjunction with the SuperCable™ Iso-Elastic Cerclage System for reattachment of the greater trochanter following osteotomy or fracture, and for fixation of long bone fractures.

Device Description

The SuperCable" Grip and Plate System consists of trochanteric reattachment grips, cable-plates, and cortical bone screws that are intended to be used in conjunction with 1.5mm diameter SuperCable Iso-Elastic Cerclage polymer cables. The cables pass through the grips and plates and provide fixation by attaching these devices to fractured or osteotomized bone fragments. Cortical bone screws may be used in combination with the cable-plates for additional fixation as deemed necessary by the surgeon user. The system includes a range of cable grip, cable-plate, and bone screw sizes and material options, with associated manual surgical instrumentation.

Trochanteric reattachment grips are available in a minimum of four lengths and contain transverse holes for the passage of cables. A cable is passed through the transverse holes and then through its own cable locking clasp, which is part of the cable system. Locking clasps are generally positioned adjacent to the grip on the anterior or posterior surface of the proximal femur. The cable hole exit geometry is designed for optimal cable trajectory and cable contact stress when the grip is affixed to the greater trochanter. Each grip contains two proximal claws that hook into or over the proximal portion of the trochanter fragment and prevent the grip from migrating distally. Each grip also contains two smaller distal claws that penetrate the trochanter fragment distally for additional fixation and stability. The longer grips incorporate an extension to allow for transversely oriented cables around the diaphysis below the lesser trochanter to better resist trochanteric migration or rotation. These longer extensions contain slots for the insertion of compression or locked cortical bone screws into the bone. The grips are available in titanium alloy, cobalt-chromium allov, or stainless steel allov.

Cable-plates are available in a minimum of three lengths and contain transverse holes for the passage of cables. A cable is passed through the transverse plate holes and then through its own cable locking clasp, which is part of the cable system. Locking clasps are generally positioned adjacent to the plate. The plates also contain alternating slots for the insertion of compression or locked cortical bone screws into the bone. By combining locking screw holes with compression slots. the plates can be used as both locking devices and fracture compression devices. The plates resemble standard plates, but have figure-of-eight shaped slots that accommodate standard or locking screws. Thus the plate can be used, depending upon the fracture situation, as a compression plate, a locked internal fixator or as a system combining both techniques. The cable hole exit geometry is designed for optimal contact stress within the cable when the plate is affixed to a long bone. The cable-plates are available in titanium alloy, cobalt-chromium alloy, or stainless steel alloy.

Bone screws are standard self-tapping cortical bone screws and are available in multiple lengths. The screws heads are available in standard compression or locked designs. The bone screws are available in titanium alloy or stainless steel. The product labeling specifies the screw and plate/grip material combinations which may be used together.

Manual instrumentation includes a grip inserter/impactor, drill, depth gage, screwdriver, and sterilization case. All implants are supplied in a non-sterile condition.

AI/ML Overview

Here's an analysis of the provided information, focusing on acceptance criteria and the study proving adherence, as requested:

Acceptance Criteria and Device Performance for K072250

Based on the provided document, the device in question is the SuperCable™ Grip and Plate System. This is a premarket notification (510(k)) submission for a medical device, and as such, the "acceptance criteria" are primarily related to substantial equivalence to predicate devices rather than specific quantitative performance metrics like sensitivity or specificity that would be typical for an AI/diagnostic device.

The study employed to demonstrate this "acceptance" (i.e., substantial equivalence) is a performance analysis that relied on mechanical and finite element analyses.

Here's the breakdown of your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implicit from 510(k) Scope)Reported Device Performance
Material Equivalence: Materials used are identical or similar to predicate devices.Stated: "The SuperCable Grip and Plate System's material... are identical or similar to the predicate devices." (Predicate devices use titanium alloy, cobalt-chrome alloy, or stainless steel; SuperCable also offers these options.)
Design Equivalence: Design features are identical or similar to predicate devices.Stated: "The SuperCable Grip and Plate System's ...design... are identical or similar to the predicate devices." (All devices use cables, locking clasps, grips, and bone plates for trochanteric or long bone fracture fixation. All systems use a tensioning instrument.) Specific design elements like grip claws, cable hole exit geometry, and plate slots are described, implying they are consistent with or advancements of predicate designs without introducing new risks.
Sizing Equivalence: Device sizes are comparable to predicate devices.Stated: "The SuperCable Grip and Plate System's ...sizing... are identical or similar to the predicate devices." (Grips and cable-plates are available in a minimum of four and three lengths, respectively, similar to predicate systems.)
Indications for Use Equivalence: Intended use is identical or similar to predicate devices.Stated: "The SuperCable Grip and Plate System's ...indications are identical or similar to the predicate devices." (Indicated for use where wire, cable, or band cerclage is used in combination with a trochanteric grip or bone plate, for reattachment of greater trochanter and fixation of long bone fractures, in conjunction with SuperCable™ Iso-Elastic Cerclage System.) This directly matches the general applications of the predicate devices.
Safety: Device presents no new risks compared to predicate devices.Stated: "The SuperCable Grip and Plate System presents no new risks as compared to the predicate devices." This is a key declaration for 510(k) clearance. Implicitly, the performance analysis (mechanical and finite element analysis) supports this by showing adequate strength and function.
Effectiveness (Strength): Device demonstrates comparable or superior strength to predicate devices.Stated: "The subject devices were found to have greater strength compared to predicate devices based on mechanical and finite element analyses as described in Section 12.11." (This is the primary quantitative performance claim made, demonstrating superior strength, which supports safety and effectiveness for the intended use in a similar manner to predicate devices.) The exact numerical values or specific acceptance criteria for "greater strength" are not provided in this summary but would have been detailed in Section 12.11 of the full submission.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not applicable in the traditional sense of a clinical or imaging study. The "test set" here refers to the models and physical specimens used for mechanical and finite element analyses. The document does not specify the number of simulations or physical parts tested.
  • Data Provenance: Not applicable in the context of clinical data. The "data" comes from engineering analyses (mechanical testing and finite element modeling). This is not retrospective or prospective patient data from a specific country. This type of data provenance is typical for non-clinical performance studies of orthopedic implants.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not applicable. Ground truth for mechanical and finite element analyses is typically established by engineering principles, material science, and established biomechanical testing standards, not by clinical expert consensus in the way an AI diagnostic algorithm's ground truth would be.
  • Qualifications of Experts: While not explicitly stated as "experts for ground truth," the performance analysis would have been conducted or overseen by engineers and potentially biomechanical specialists familiar with mechanical testing and finite element analysis. Their qualifications would stem from their expertise in these engineering disciplines.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. There is no human interpretative "test set" requiring adjudication. The outcomes of mechanical tests and finite element analyses are based on physical measurements and computational models, not subjective human judgment.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • MRMC Study: No, an MRMC study was not performed, nor would it be appropriate for this type of device (an orthopedic implant). MRMC studies are typically used for evaluating diagnostic imaging or AI systems where human readers interpret medical images.
  • Effect Size of Human Readers: Not applicable.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

  • Standalone Performance: Not applicable. This device is an orthopedic implant, not an algorithm. Therefore, there is no "standalone" algorithm performance to measure. Its performance is inherent in its physical and mechanical properties.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The "ground truth" for the performance analysis was derived from:
    • Engineering principles and material properties: For Finite Element Analysis (FEA).
    • Physical measurements and established biomechanical test methods: For mechanical testing.
    • Comparison to predicate device performance benchmarks: To demonstrate "greater strength" or at least comparable performance.

8. The Sample Size for the Training Set

  • Training Set Sample Size: Not applicable. This device does not use machine learning or AI that would require a "training set." The design is based on engineering principles, materials science, and established medical needs, not data-driven algorithmic training.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable, as there is no "training set" for this type of mechanical device.

In summary, the provided document details a 510(k) submission for an orthopedic implant. The "acceptance criteria" revolve around demonstrating substantial equivalence to existing predicate devices in terms of materials, design, sizing, indications, and safety. The "study" proving this substantial equivalence was a performance analysis based on mechanical testing and finite element analysis, showing comparable or superior strength without introducing new risks. This is a standard non-clinical pathway for FDA clearance of Class II orthopedic devices.

{0}------------------------------------------------

K072250

6. 510(k) Summary

Applicant:Kinamed, Inc.
Address:820 Flynn Road, Camarillo, CA 93012 USA
Contact person:Vineet K. Sarin Ph.D. 805-384-2748 (phone), 805-384-2792 (fax)
Date prepared:August 10, 2007
Trade name:SuperCable™ Grip and Plate System

Substantial equivalence claimed to (see Section 9.7):

  • a) K961267 Pioneer, Greater Trochanter Reattachment Device
  • b) K972223 Pioneer, Bone Plate with Cables
  • c) K982545 Biomet, BMP Cable System
  • d) K000734 Pioneer, Extended Greater Trochanter Reattachment Device
  • e) K001709 Pioneer, Trochanteric Reattachment Device
  • f) K032559 Synthes, Titanium Locking Screws
  • g) K051986-Synthes, LCP Straight Reconstruction Plates

Device Description:

The SuperCable" Grip and Plate System consists of trochanteric reattachment grips, cable-plates, and cortical bone screws that are intended to be used in conjunction with 1.5mm diameter SuperCable Iso-Elastic Cerclage polymer cables. The cables pass through the grips and plates and provide fixation by attaching these devices to fractured or osteotomized bone fragments. Cortical bone screws may be used in combination with the cable-plates for additional fixation as deemed necessary by the surgeon user. The system includes a range of cable grip, cable-plate, and bone screw sizes and material options, with associated manual surgical instrumentation.

Trochanteric reattachment grips are available in a minimum of four lengths and contain transverse holes for the passage of cables. A cable is passed through the transverse holes and then through its own cable locking clasp, which is part of the cable system. Locking clasps are generally positioned adjacent to the grip on the anterior or posterior surface of the proximal femur. The cable hole exit geometry is designed for optimal cable trajectory and cable contact stress when the grip is affixed to the greater trochanter. Each grip contains two proximal claws that hook into or over the proximal portion of the trochanter fragment and prevent the grip from migrating distally. Each grip also contains two smaller distal claws that penetrate the trochanter fragment distally for additional fixation and stability. The longer grips incorporate an extension to allow for transversely oriented cables around the diaphysis below the lesser trochanter to better resist trochanteric migration or rotation. These longer extensions contain slots for the insertion of compression or locked cortical bone screws into the bone. The grips are available in titanium alloy, cobalt-chromium allov, or stainless steel allov.

Cable-plates are available in a minimum of three lengths and contain transverse holes for the passage of cables. A cable is passed through the transverse plate holes and then through its own cable locking clasp, which is part of the cable system. Locking clasps are generally positioned adjacent to the plate. The plates also contain alternating slots for the insertion of compression or

NOV

0 9 2007

{1}------------------------------------------------

locked cortical bone screws into the bone. By combining locking screw holes with compression slots. the plates can be used as both locking devices and fracture compression devices. The plates resemble standard plates, but have figure-of-eight shaped slots that accommodate standard or locking screws. Thus the plate can be used, depending upon the fracture situation, as a compression plate, a locked internal fixator or as a system combining both techniques. The cable hole exit geometry is designed for optimal contact stress within the cable when the plate is affixed to a long bone. The cable-plates are available in titanium alloy, cobalt-chromium alloy, or stainless steel alloy.

Bone screws are standard self-tapping cortical bone screws and are available in multiple lengths. The screws heads are available in standard compression or locked designs. The bone screws are available in titanium alloy or stainless steel. The product labeling specifies the screw and plate/grip material combinations which may be used together.

Manual instrumentation includes a grip inserter/impactor, drill, depth gage, screwdriver, and sterilization case. All implants are supplied in a non-sterile condition.

Intended use:

The SuperCable" Grip and Plate System is indicated for use where wire, cable, or band cerclage is used in combination with a trochanteric grip or bone plate. The SuperCable "Grip and Plate System is intended to be used in conjunction with the SuperCable™ Iso-Elastic Cerclage System for reattachment of the greater trochanter following osteotomy or fracture, and for fixation of long bone fractures.

Summary of technological characteristics compared to predicate devices:

The SuperCable Grip and Plate System's material, design, sizing, and indications are identical or similar to the predicate devices. All devices use cables, locking clasps, grips, and bone plates for trochanteric or long bone fracture fixation. All device systems use a tensioning instrument to apply compression and affix the cable to the grip, plate, and bone. Predicate devices are available in titanium alloy, cobalt-chrome alloy, or stainless steel. Predicate systems include grips, plates, and bone screws in a variety of lengths to accommodate common fracture patterns that occur clinically.

Performance Analysis:

The SuperCable Grip and Plate System presents no new risks as compared to the predicate devices. No clinical testing is required for determination of substantial equivalence. The subject devices were found to have greater strength compared to predicate devices based on mechanical and finite element analyses as described in Section 12.11.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 0 9 2007

Kinamed, Inc. % Vineet K. Sarin, Ph.D. Chief Operating Officer 820 Flynn Road Camarillo, CA 93012

K072250 Re: Trade/Device Name: SuperCable™ Grip and Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code(s): KTT, HWC, JDQ Dated: August 10, 2007 Received: August 13, 2007

Dear Dr. Sarin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Dr. Vineet K. Sarin

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely vours,

Barbara Buckup for
Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

8. Indications for Use

510(k) Number (if known): K07QQ50

Device Name:

SuperCable™ Grip and Plate System

Indications For Use:

The SuperCable" Grip and Plate System is indicated for use where wire, cable, or band cerclage is used in combination with a trochanteric grip or bone plate. The SuperCable™ Grip and Plate System is intended to be used in conjunction with the SuperCable™ Iso-Elastic Cerclage System for reattachment of the greater trochanter following osteotomy or fracture, and for fixation of long bone fractures.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices

Page 1 of
510(k) NumberK072250

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.