K Number
K071875

Validate with FDA (Live)

Device Name
V-CATH POLY PICC
Manufacturer
Date Cleared
2007-11-05

(122 days)

Product Code
Regulation Number
880.5970
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Power-V PICC is indicated for patients that require repeated venous access for infusion or injection therapy. The Power-V PICC is indicated for peripheral access to the central venous system for intravenous therapy. The Power-V PICC is indicated for dwell times less than or greater than 30 days. The maximum recommended infusion rate is 5 ml/sec. The maximum pressure of power injectors used with the Power-V PICC catheter may not exceed 300 psi.

Device Description

The Power-V PICC is a family of peripherally inserted central venous catheters designed to allow for contrast media studies. The catheters, made of radiopaque polyurethane tubing, are inserted pcriphcrally. Each Power-V PICC has a kink resistant, reverse tapered catheter design. The Power-V PICC is indicated for dwell times greater than 30 days. The Power-V PICC catheter assemblies have been tested to withstand power injection of worst-case viscosity injection media at 5 cc/sec with a maximum power injection pressure of 300 psi (scc Appendix B). The catheter is supplied sterile and non-pyrogenic in a variety of kit configurations (scc Appendix C).

The Power-V PICC is illustrated in the drawings at the end of this section. The Power-V PICC product line has catheters in 4 Fr and 5 Fr single lumen and 5 Fr and 6 Fr dual lumen. All catheters are 60 cm long. The catheters are attached to an injection-molded polyurethane hub that has extension legs with luer lock fittings for access attachment.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Power-V PICC device.

It's important to note that the provided text is a 510(k) summary for a medical device seeking substantial equivalence, not a detailed scientific study report. As such, information regarding rigorous clinical trial design, statistical analysis, and detailed performance metrics as you might find for a novel drug or a high-risk device is largely absent. The focus of a 510(k) is to demonstrate equivalence to a legally marketed predicate device.


Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state "acceptance criteria" in the form of numerical thresholds for a specific test. Instead, it refers to the device meeting "all functional requirements and specifications" and demonstrating substantial equivalence to predicate devices. The key performance aspect mentioned is related to power injection capabilities.

Acceptance Criteria (Implied)Reported Device Performance
Withstand power injection of worst-case viscosity injection media."The Power-V PICC catheter assemblies have been tested to withstand power injection of worst-case viscosity injection media at 5 cc/sec with a maximum power injection pressure of 300 psi (see Appendix B)." The "Intended Use" section reiterates: "The maximum recommended infusion rate is 5 ml/sec. The maximum pressure of power injectors used with the Power-V PICC catheter may not exceed 300 psi."
Biocompatibility (conformance with standards)."Biocompatibility testing on the device was in conformance with ISO 10993-1, and GLP TRIPARTIE."
Functional requirements and specifications."The Power-V met all functional requirements and specifications." (No specific details provided on these requirements.)
Substantial equivalence to predicate devices."The design, methods of manufacturing, and materials used in the Power-V device are substantially equivalent to the predicate devices." (Implies meeting similar performance as predicate devices, but no direct quantitative comparison is given).

Study Details (Based on the provided 510(k) Summary)

The text describes engineering and biocompatibility testing rather than a clinical study in the traditional sense, especially a study involving human subjects or extensive retrospective data analysis for performance metrics beyond physical device capabilities.

  1. Sample size used for the test set and the data provenance:

    • The document refers to "Power-V PICC catheter assemblies" being tested for power injection. It does not specify the number of catheter assemblies used in this test.
    • Data provenance: The testing appears to be prospective laboratory/bench testing conducted by the manufacturer (HDC Corporation) or a contracted lab. There is no mention of country of origin for the data or whether it's retrospective.
    • For biocompatibility, it states "Biocompatibility testing on the device was in conformance with ISO 10993-1, and GLP TRIPARTIE." This would involve specific biological tests on materials, not human testing or data from patients.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This concept is not applicable to the type of testing described. The "ground truth" for power injection performance or biocompatibility is established by the physical and chemical properties of the device and its behavior under controlled laboratory conditions, as measured by instruments and analyzed against engineering specifications and relevant standards. There is no involvement of clinical experts for "ground truth" similar to image interpretation.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies where human interpretation or classification is involved, often for clinical outcomes or image reading. The tests described are objective, physical/chemical tests.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices, particularly those involving AI or human interpretation, to assess reader performance. The Power-V PICC is an intravascular catheter with mechanical and biocompatibility performance as its primary focus.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, this is not applicable. The Power-V PICC is a physical medical device, not an algorithm or AI system. Its performance is measured directly, not through an algorithm's output.
  6. The type of ground truth used:

    • For power injection: Engineering specifications and measurement data. The ability to withstand 5 cc/sec at 300 psi is measured directly using appropriate laboratory equipment.
    • For biocompatibility: Conformance to international standards (ISO 10993-1, GLP TRIPARTIE), which specify acceptable biological responses to materials. This involves a series of in-vitro and in-vivo tests to assess toxicity, irritation, sensitization, etc.
  7. The sample size for the training set:

    • Not applicable. This device is not an AI or machine learning model, so there is no concept of a "training set."
  8. How the ground truth for the training set was established:

    • Not applicable for the reason above.

Summary of the "Study" Proving Acceptance:

The "study" in this context refers to bench testing and biocompatibility testing performed by or for HDC Corporation to demonstrate that the Power-V PICC meets its functional specifications and adheres to relevant safety standards. The core claim is not one of clinical superiority or a complex diagnostic performance, but rather that the device is substantially equivalent to existing, legally marketed predicate devices in terms of its design, materials, manufacturing, and basic functional capabilities, particularly its power injection functionality and biocompatibility. The regulatory pathway of a 510(k) relies heavily on this concept of substantial equivalence, meaning rigorous, comparative clinical trials are usually not required if equivalence can be demonstrated through non-clinical means.

{0}------------------------------------------------

510(k) Summary

General Information

Classification: Class II

Trade Name: V-Cath Poly PICC

NOV - 5 2007

Submitter: HDC, Corporation. 628 Gibraltar Court Milpitas, CA 95035, USA

Earl Smart QA Tel # (408) 942-7340

Device Identification

Contact:

  • A. Common Name: Power-V PICC
  • B. Classification Name: Percutaneous Implant, long-term intravascular catheter
  • C. Regulation Number: 880.5970
  • D. Panel: General Hospital
  • E. Product Code: LJS
  • F. Class: Class II

Predicate Devices

V-Cath Poly PICC from HDC Corporation, K033853 OMNIPICC P.I. from RITA Medical Systems, Inc., K051102 OMNIPICC P.I. from RITA Medical Systems, Inc., K062579 PowerPICC Catheter, Bard Access Systems, Inc., K033389

Device Description

The Power-V PICC is a family of peripherally inserted central venous catheters designed to allow for contrast media studies. The catheters, made of radiopaque polyurethane tubing, are inserted pcriphcrally. Each Power-V PICC has a kink resistant, reverse tapered catheter design. The Power-V PICC is indicated for dwell times greater than 30 days. The Power-V PICC catheter assemblies have been tested to withstand power injection of worst-case viscosity injection media at 5 cc/sec with a maximum power injection pressure of 300 psi (scc Appendix B). The catheter is supplied sterile and non-pyrogenic in a variety of kit configurations (scc Appendix C).

The Power-V PICC is illustrated in the drawings at the end of this section. The Power-V PICC product line has catheters in 4 Fr and 5 Fr single lumen and 5 Fr and 6 Fr dual lumen. All catheters are 60 cm long. The catheters are attached to an injection-molded polyurethane hub that has extension legs with luer lock fittings for access attachment.

{1}------------------------------------------------

Intended Use

The Power-V PICC is indicated for patients that require repeated venous access for infusion or injection therapy. The Power-V PICC is indicated for peripheral access to the central venous system for intravenous therapy. The Power-V PICC is indicated for dwell times less than or greater than 30 days. The maximum recommended infusion rate is 5 ml/sec. The maximum pressure of power injectors used with the Power-V PICC catheter may not exceed 300 psi.

Biocompatibility

Biocompatibility testing on the device was in conformance with ISO 10993-1, and GLP TRIPARTIE . The device is identical to the previously approved device OMNIPICC P.I. from RITA Medical Systems and made by Health Line International Corporation. The device is now made for HDC Corporation by Health Line International Corporation.

Device Performance/Product Testing

The Power-V met all functional requirements and specifications.

Technology Characteristics

The Power-V is equivalent technologically to the devices mentioned under predicate devices above.

Substantial Equivalence

The design, methods of manufacturing, and materials used in the Power-V device are substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The logo also includes the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" arranged in a circular fashion around the caduceus symbol. The text is in all caps and appears to be in a sans-serif font.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 5 2007

Mr. Earl Smart Quality Assurance HDC Corporation 628 Gibraltar Court Milpitas, California 95035

Re: K071875

Trade/Device Name: V-Cath (Polyurethane) Power PICC (Power -V) Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: October 3, 2007 Received: October 5, 2007

Dear Mr. Smart:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Smart

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D. fa

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K071875
1 of 1

Indications For Use

510 (K) NUMBER:

K071875

Device Name:

Indications For Use:

V-Cath (Polyurethane) Power PICC (Power-V)

The Power-V PICC is indicated for patients that require repeated venous access for infusion or injection therapy. The Power-V PICC is indicated for peripheral access to the central venous system for intravenous therapy. The Power-V PICC is indicated for dwell times less than or greater than 30 days. The maximum recommended infusion rate is 5 ml/sec. The maximum pressure of power injectors used with the Power-V PICC catheter may not exceed 300 psi.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OROver-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

vision Sign-Off

livision of Anestheslology, General fection Control, Dental Devices

510(k) Number: KD 71875

K071875 HDC Corporation Dated: July 12, 2007

Page 11-1

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”