(63 days)
The Acme Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
The Acme Spinal System consists of monoaxial and polyaxial pedicle screws, locking plugs, spinal rods and rod to rod connectors. Acme Spine System can be used for single or multiple level fixations. All components are manufactured from titanium alloy (Ti-6Al-4V ELI) that conforms to ASTM F136. The modification included in this submission is the addition of the Acme Talon Pedicle Screw.
Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and study information:
Based on the provided document, the device in question is the "Acme Spinal System," and the submission focuses on the addition of the "Acme Talon Pedicle Screw." This is a medical device clearance (510(k)), which typically relies on demonstrating substantial equivalence to a predicate device rather than extensive clinical efficacy studies like those for drugs.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Mechanical Strength per ASTM F1717 | "Testing was conducted according to ASTM F1717 with adequate strength." |
Note: The document only explicitly mentions mechanical testing according to ASTM F1717. It does not provide specific numerical acceptance criteria or detailed performance results (e.g., specific load capacities, fatigue cycles survived) in this summary. It only states that the testing demonstrated "adequate strength."
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified for the non-clinical testing. The document states "Testing was conducted according to ASTM F1717," which is a standard for static and fatigue testing of pedicle screw spinal systems. This typically involves a defined number of test samples (e.g., screws, constructs) as per the standard's methodology.
- Data Provenance: Not explicitly stated, but assumed to be laboratory-based mechanical testing, likely conducted by or for the manufacturer (Acme Spine, LLC) in the USA. It is retrospective in the sense that the testing was performed, and the results are being reported for the 510(k) submission, not as a prospective clinical trial.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Not applicable. The ground truth for this device (Acme Spinal System with Talon Pedicle Screw) is established through mechanical testing against industry standards (ASTM F1717), not through expert clinical consensus for diagnostic or prognostic purposes.
4. Adjudication Method for the Test Set
- Not applicable. As noted above, the "test set" here refers to physical components undergoing mechanical evaluation, not clinical data requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. This is a 510(k) submission for a spinal implant, which relies on mechanical performance and substantial equivalence to a predicate device. MRMC studies are typically performed for medical imaging devices or diagnostic algorithms where human readers interpret data, often with and without AI assistance, to assess diagnostic performance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not applicable. This device is a physical implant, not an algorithm.
7. The Type of Ground Truth Used
- Mechanical Performance Standards: The ground truth for this device is based on its ability to meet the mechanical performance requirements specified in ASTM F1717 (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model), which ensures the device's structural integrity and durability.
8. The Sample Size for the Training Set
- Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set. The "training" for such devices involves design, prototyping, and manufacturing process refinement.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As mentioned, there is no "training set" in the context of an AI/ML algorithm. For physical medical devices like the Acme Spinal System, the "ground truth" for design and manufacturing is established through engineering principles, material science, and adherence to quality systems and relevant standards (e.g., ASTM F136 for titanium alloy, ISO standards for manufacturing processes) to ensure the device's safety and effectiveness.
{0}------------------------------------------------
510(k) Summary for the Acme Spinal System
SEP - 4 2007
This safety and effectiveness summary for the Acme Spinal System is provided as required per Section 513(i)(3) of the Food, Drug and Cosmetic Act.
Date Prepared: August 2, 2007
1. Submitter:
Acme Spine, LLC 9980 Indiana Avenue Unit 9 Riverside, CA 92503 909-352-9862
Contact Person:
J.D. Webb The OrthoMedix Group, Inc. 1001 Oakwood Blvd Round Rock, TX 78681 Telephone: 512-388-0199
| 2. Trade name: | Acme Spinal System |
|---|---|
| Common Name: | posterior pedicle screw system |
| Classification Name: | Pedicle screw spinal system |
| Class II | |
| 21 CFR 888.3070 | |
| MNI/MNH |
3. Predicate or legally marketed devices which are substantially equivalent:
- · Acme Spine System (K001044)
4. Description of the device:
The Acme Spinal System consists of monoaxial and polyaxial pedicle screws, locking plugs, spinal rods and rod to rod connectors. Acme Spine System can be used for single or multiple level fixations. All components are manufactured from titanium alloy (Ti-6Al-4V ELI) that conforms to ASTM F136.
The modification included in this submission is the addition of the Acme Talon Pedicle Screw.
5. Intended Use:
The Acme Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
6. Comparison of the technological characteristics of the device to predicate and legally marketed devices:
The Acme Talon pedicle screw is similar to the Acme pedicle screw in terms of material, design and indications.
{1}------------------------------------------------
7. Summary of Nonclincal Tests
Testing was conducted according to ASTM F1717 with adequate strength.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing right, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Acme Spine, LLC % The OrthoMedix Group, Inc. Mr. J. D. Webb 1001 Oakwood Blvd Round Rock, Texas 78681
SEP - 4 2007
Re: K071824 Trade/Device Name: Acme Spinal System Regulation Number 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH, KWP Dated: August 2, 2007 Received: August 6, 2007
Dear Mr. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Pagc 2 -- Mr. J. D. Webb
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Neil R. Oyler
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): Ko7182 4
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The Acme Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis).
| Prescription Use | X |
|---|---|
| (Part 2) CFR 801 Subpart D) |
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nil R.P. Ogden for mxn
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
K071824 510(k) Number