21.3 inch (53 cm) Monochrome Digital Mammography LCD Monitor with additional sub-pixel driving techniques designed to improve image quality, MS51i2 (MDL2123A) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is to be used in digital mammography PACS and modalities including FFDM.
MS51i2 (MDL2123A) is a 21.3-inch (53cm) monochrome LCD monitor whose display resolution is 2048 x 7680 (landscape), 7680 x 2048 (portrait) supporting DVI (digital visual interface).
The provided text is a 510(k) summary for a medical display device, not an AI/ML medical device. Therefore, the typical acceptance criteria and study designs associated with AI/ML devices (e.g., sample sizes for test and training sets, ground truth establishment, MRMC studies) are not applicable or described in this document.
The document describes a medical display monitor (TOTOKU MS51i2 (MDL2123A)) and its substantial equivalence to a predicate device for displaying and viewing medical images, specifically for digital mammography. The acceptance criteria for such a device primarily revolve around its technical performance specifications to ensure it can accurately display medical images for diagnostic purposes.
Since the document does not contain the detailed study information typically requested for AI/ML device evaluations, the following breakdown is based on what can be inferred or directly stated from the provided text regarding a display device's performance and validation.
Acceptance Criteria and Device Performance for TOTOKU MS51i2 (MDL2123A) - Medical Display Monitor
This section describes the acceptance criteria and a "study" in the context of demonstrating substantial equivalence for a medical display monitor, not an AI/ML diagnostic algorithm. The "study" here refers to the comparisons made to a predicate device and the technical specifications provided, rather than a clinical trial evaluating diagnostic accuracy.
1. Table of Acceptance Criteria and Reported Device Performance
Given this is a display monitor, the "acceptance criteria" relate to its technical specifications and its ability to perform comparably to a predicate device. The document states "MS51i2 (MDL2123A) shares the same characteristics with our predicate device ME551i2 (K061447) except for a newly-developed Independent Sub-pixel Drive (ISD), AR (Anti-Reflection) special coating panel and firmware to display driver by FPGA." This implies that the device's performance meets or exceeds the predicate device's established performance without introducing new safety or effectiveness concerns.
| Acceptance Criterion (Implicit/Inferred) | Reported Device Performance |
|---|---|
| Display Resolution | 2048 x 7680 (landscape), 7680 x 2048 (portrait) |
| Screen Size | 21.3 inch (53cm) |
| Display Type | Monochrome LCD monitor |
| Input Interface Support | DVI (digital visual interface) |
| Intended Use | Displaying and viewing medical images for diagnosis by trained medical practitioners in digital mammography PACS and modalities including FFDM. |
| Image Quality Enhancements | Additional sub-pixel driving techniques (Independent Sub-pixel Drive - ISD) designed to improve image quality; AR (Anti-Reflection) special coating panel. |
| Substantial Equivalence | Demonstrated to be substantially equivalent to predicate device ME551i2 (K061447) in characteristics for its intended use, with specified improvements. |
2. Sample Size Used for the Test Set and Data Provenance
Not applicable to this type of device submission. There is no "test set" of medical images in the context of an AI/ML algorithm. The performance of the monitor is assessed based on its technical specifications, manufacturing controls, and comparison to a predicate device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. Ground truth establishment for a test set of medical images is not relevant for a display monitor's 510(k) submission.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods are used in studies involving human interpretation or AI/ML output, not for a display monitor's technical evaluation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. An MRMC study is not relevant for a medical display monitor's 510(k) submission. These studies evaluate the diagnostic performance of human readers, sometimes with and without AI assistance. This document describes a passive display device.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done
No. This is not an algorithm, but a display device. Standalone performance is not applicable.
7. The Type of Ground Truth Used
Not applicable for a display monitor. "Ground truth" for an image display would relate to the accuracy of its physical representation of image data, which is governed by technical standards, not expert labels on cases.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. This is not an AI/ML device that requires a training set.
{0}------------------------------------------------
TOTUKU
510(k) SUMMARY
77 -- 1
Submitter Information: TOTOKU ELECTRIC CO., LTD. 300 Oya, Ueda Nagano 386-0192 Japan
Contact Person: Mikio Hasegawa, General Manager Email: hasegawam@totoku.co.jp Tel: +81.268.34.5469 Fax: +81.268.34.5548
Date Prepared: June 29, 2007
21.3 inch (53cm) Monochrome Digital Mammography LCD Monitor Device Name: MS51i2 (MDL2123A)
Common Name: MS51i2 (MDL2123A)
Classification Name: Class II (Part 892 Radiology Devices Sec. 892.2050 Picture Archiving and Communication System)
5M Monochrome Digital Mammography LCD Monitor ME551i2 Predicate Device: (MDL2116A) (K061447)
Device Description: MS51i2 (MDL2123A) is a 21.3-inch (53cm) monochrome LCD monitor whose display resolution is 2048 x 7680 (landscape), 7680 x 2048 (portrait) supporting DVI (digital visual interface).
- 21.3 inch (53cm) Monochrome Digital Mammography LCD Monitor Indended Use: with additionalsub-pixel driving techniques designed to improve image quality, MS51i2 (MDL2123A) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is to be used in digital mammography PACS and modalities including FFDM.
- Substantial Equivalence: MS51i2 (MDL2123A) shares the same characteristics with our predicate device ME551i2 (K061447) except for a newly-developed Independent Sub-pixel Drive (ISD), AR (Anti-Reflection) special coating panel and firmware to display driver by FPGA.
MDL2123A Premarket Notification
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or bird with three stylized wing segments.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 11 2008
Mr. Mikio Hasegawa General Manager TOTOKU Electric Co., Ltd. 300 OYA, Ueda Nagano, 386-0192 JAPAN
Re: K071794
Trade/Device Name: 21.3 inch (53 cm) Monochrome Digital Mammography LCD Monitor MS51i2 (MDL2123A)
Regulation Number: 21 CFR 892.2050
Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 25, 2008 Received: January 28, 2008
Dear Mr. Hasegawa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrt/industry/support/index.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
INDICATIONS FOR USE
510(k) Number: Not Known
Device Name: 21.3 inch (53 cm) Monochrome Digital Mammography LCD Monitor MS51i2 (MDL2123A)
Indications for Use:
21.3 inch (53 cm) Monochrome Digital Mammography LCD Monitor with additional sub-pixel driving techniques designed to improve image quality, MS51i2 (MDL2123A) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is to be used in digital mammography PACS and modalities including FFDM.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jorge Th. W.
(Division Sign-Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
N/A