K Number
K071759

Validate with FDA (Live)

Date Cleared
2008-07-30

(398 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LTC VAD.400 is an electrosurgical generator coupled with designated LTC electrosurgical instruments for bipolar tissue fusion. The LTC VAD.400 electrosurgical generator and designated LTC electrosurgical instruments (i.e., "the VAD System") are intended for use in open general surgical procedures where ligation of vessels and ducts is desired and as an alternative to mechanical clamping (clips or staples) or suturing. The LTC VAD System can be used on vessels up to 7mm in diameter, on ducts up to 2mm in diameter, and tissue bundles as large as will fit in the jaw electrodes of the instrument.

The LTC VAD system has not yet been tested for use in tubal sterilization or tubal coagulation for sterilization procedures; therefore, the company does not recommend the use of the LTC VAD system for such procedures. In addition, LTC recommends against the use of the VAD System for male circumcisions.

Device Description

The LTC VAD System consists of: a bipolar electrosurgical generator and two distinct and dedicated electrosurgical instruments.

AI/ML Overview

The provided text is a 510(k) summary for the Live Tissue Connect Inc.'s VAD System, which is an electrosurgical bipolar device. It focuses on demonstrating substantial equivalence to predicate devices rather than proving specific performance characteristics against a set of acceptance criteria in a detailed study. Therefore, the information requested in your prompt regarding acceptance criteria, study details, and specific performance metrics is largely not available in the provided document.

Here's a breakdown of what can and cannot be answered based on the input:

1. A table of acceptance criteria and the reported device performance

  • Not available. The document states "In all instances, the LTC VAD System functioned as intended and all results observed were as expected." This is a general statement of compliance, not a report against specific, quantified acceptance criteria. Acceptance criteria and detailed performance metrics are not provided.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not available. The document mentions "Bench, animal and clinical testing was conducted." However, it does not provide details on the sample sizes for these tests, the number of cases, or the provenance of the data (e.g., country of origin, retrospective/prospective nature).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable / Not available. The device is an electrosurgical tool, and its performance is typically evaluated by objective measures (e.g., burst pressure of sealed vessels, thermal spread) and clinical outcomes, rather than by expert review of diagnostic images to establish a "ground truth." The document does not describe any human expert review process for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable / Not available. As explained above, this type of adjudication is typically for diagnostic devices where human interpretation errors might be present. For an electrosurgical device, the adjudication method is not a relevant concept for evaluating its performance.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. The LTC VAD System is an electrosurgical device, not an AI-powered diagnostic tool that assists human readers. Therefore, an MRMC study or AI assistance is not relevant to its evaluation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is an electrosurgical device and not an algorithm. Its performance inherently involves human operation; there is no "standalone" algorithm performance to evaluate.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Implied ground truth through functional assessment. While not explicitly stated as "ground truth," the performance testing (bench, animal, clinical) would have based its assessment on objective functional criteria relevant to vessel sealing, such as:
    • Successful ligation of vessels/ducts up to 7mm/2mm respectively.
    • Lack of complications (e.g., excessive thermal spread, incomplete seal).
    • Achievement of intended surgical outcome.
      However, the document does not detail how "intended function" and "expected results" were objectively defined or measured to serve as a ground truth.

8. The sample size for the training set

  • Not applicable / Not available. This is not an AI/machine learning device, so the concept of a "training set" in that context does not apply. If "training set" refers to data used for initial development or optimization, it is not mentioned in the document.

9. How the ground truth for the training set was established

  • Not applicable / Not available. As explained above, there is no mention of a training set or its associated ground truth in the context of an electrosurgical device.

In summary, the provided 510(k) summary for the LTC VAD System is focused on demonstrating substantial equivalence to existing predicate devices based on intended use, technological characteristics, and general performance (bench, animal, and clinical testing showing it "functioned as intended"). It does not contain the detailed, quantitative efficacy study information typically found for devices requiring a more rigorously defined performance evaluation against specific acceptance criteria.

{0}------------------------------------------------

JUL 3 0 2008

K071759

510(k) SUMMARY

Pzei/f~

Live Tissue Connect Inc.'s VAD System

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Live Tissue Connect, Inc. 2813 Gaviota Street Los Olivos, CA 93441

Phone: 805-686-0937 Facsimile: 805-686-5399

Contact Person: Frank D. D'Amelio

Date Prepared: February 10, 2007

Name of Device and Name/Address of Sponsor

LTC VAD System Live Tissue Connect, Inc. 500 North Shoreline Corpus Christi, Texas 78741

Common or Usual Name

Electrosurgical Bipolar Device

Classification Name

Electrosurgical Cutting and Coagulation Device and Accessories

Predicate Devices

LigaSure Vessel Sealing System (K981916) LigaSure Advance (K063195) Gyrus General Purpose Electrosurgical System (K050550) Gyrus OPEN FORCEPS (K024286)

{1}------------------------------------------------

Page 2 of 2

Intended Use / Indications for Use

The LTC VAD.400 is an electrosurgical generator coupled with designated LTC electrosurgical instruments for bipolar tissue fusion. The LTC VAD.400 electrosurgical generator and designated LTC electrosurgical instruments (i.e., "the VAD System") are intended for use in open general surgical procedures where ligation of vessels and ducts is desired and as an alternative to mechanical clamping (clips or staples) or suturing. The LTC VAD System can be used on vessels up to 7mm in diameter, on ducts up to 2mm in diameter, and tissue bundles as large as will fit in the jaw electrodes of the instrument.

The LTC VAD system has not yet been tested for use in tubal sterilization or tubal coagulation for sterilization procedures; therefore, the company does not recommend the use of the LTC VAD system for such procedures. In addition, LTC recommends against the use of the VAD System for male circumcisions.

Technological Characteristics

The LTC VAD System consists of: a bipolar electrosurgical generator and two distinct and dedicated electrosurgical instruments.

Performance Data

Bench, animal and clinical testing was conducted using the LTC VAD System. In all instances, the LTC VAD System functioned as intended and all results observed were as expected.

Substantial Equivalence

The LTC VAD System is as safe and effective as the predicate devices. The LTC VAD System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the LTC VAD System and its predicate devices raise no new issues of safety or effectiveness. Thus, the LTC VAD System is substantially equivalent.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 30 2008

Live Tissue Connect, Inc. % Hogan & Hartson Mr. Howard M. Holstein 555 Thirteenth Street, NW Washington, District of Columbia 20004

Re: K071759

Trade/Device Name: LTC VAD System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: May 1, 2008 Received: May 1, 2008

Dear Mr. Holstein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. IIoward M. Holstein

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Mulhern

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K071759

Device Name: LTC VAD System

Indications for Use:

The LTC VAD.400 is an electrosurgical generator coupled with designated LTC electrosurgical instruments for bipolar tissue fusion. The LTC VAD.400 electrosurgical generator and designated LTC electrosurgical instruments (i.e., "the VAD System") are intended for use in open general surgical procedures where ligation of vessels and ducts is desired and as an alternative to mechanical clamping (clips or staples) or suturing. The LTC VAD System can be used on vessels up to 7mm in diameter, on ducts up to 2mm in diameter, and tissue bundles as large as will fit in the jaw electrodes of the instrument.

The LTC VAD system has not yet been tested for use in tubal sterilization or tubal coagulation for sterilization procedures; therefore, the company does not recommend the use of the LTC VAD system for such procedures. In addition, LTC recommends against the use of the VAD System for male circumcisions.

Prescription Use X (Part 21 C.F.R. 801 Subpart D) AND/OR

Over-The-Counter Use (21 C.F.R. 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark H. Millerson

(Division Sign.Off Division of General, Restorative, and Neurological Devices

510(k) Number. K071754

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.