GENERAL SEMEN ANALYSIS KIT

K071737 · Dyn-Bioshaf (2006) , Ltd. · POV · Oct 25, 2007 · Hematology

Device Facts

Record IDK071737
Device NameGENERAL SEMEN ANALYSIS KIT
ApplicantDyn-Bioshaf (2006) , Ltd.
Product CodePOV · Hematology
Decision DateOct 25, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The GSA kit is for the evaluation of human semen and provides a set of reagents to evaluate semen quality by measuring five parameters, recommended by the World Health Organization (WHO), that are used to determine whether infertility is caused by abnormalities of one or more of them. The parameters include: Sperm Count, sperm motility, sperm viability, white blood cell (WBC) count, and Sperm head morphology (the subject of this 510(k) submission.

Device Story

General Semen Analysis (GSA) Kit provides reagents for laboratory evaluation of human semen quality. Device measures five WHO-recommended parameters: sperm count, motility, viability, white blood cell (WBC) count, and sperm head morphology. Used by clinical laboratory professionals to assess potential male infertility. Output assists clinicians in diagnosing causes of infertility based on semen abnormalities. Reagents facilitate standardized testing protocols.

Clinical Evidence

Prospective clinical trial across four USA clinics and one Israeli clinic (n=485 semen samples). Compared GSA Kit results for sperm head morphology against the WHO routine manual method. Results: 83.9% agreement for abnormal morphology (95% CI: 79.1-88.0) and 87.8% agreement for normal morphology (95% CI: 82.6-92.0). Overall agreement rate was 85.6% (95% CI: 82.1-88.6). Cross-platform validation performed on three flow cytometers (BD FACScan, Beckman Coulter Epics XL, Partec PAS III) showed high correlation (r=0.980-0.987) and minimal bias.

Technological Characteristics

Reagents include fluorescently labeled monoclonal antibodies (FITC) and enzymatic fluorophores. Sensing principle: fluorescence intensity detection via flow cytometry. Validated on Becton Dickinson FACScan, Beckman Coulter Epics XL, and Partec PAS III. Quantitative output based on fluorescence gating (e.g., ≥ 65% staining in gate R2 for normal DNA condensation).

Indications for Use

Indicated for the evaluation of human semen quality in patients undergoing infertility assessment, specifically measuring sperm count, motility, viability, white blood cell count, and sperm head morphology.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Related Devices

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K071737 B. Purpose for Submission: Traditional 510(k) for modification of a previously cleared device C. Measurand: Sperm head morphology parameter D. Type of Test: Quantitative immunohistochemical assay using flow cytometry E. Applicant: Dyn-BioShaf Ltd., Israel F. Proprietary and Established Names: General Semen Analysis (GSA) Kit G. Regulatory Information: 1. Regulation section: a. There is no regulation for semen analysis devices b. 21 CFR 864.5220, Automated differential cell counter 2. Classification: Class II 3. Product code: MNA, GKZ {1} 4. Panel: Hematology (81), OB-GYN (85) H. Intended Use: 1. Intended use(s): The GSA Kit is for the evaluation of human semen and provides a set of reagents to evaluate semen quality by measuring five parameters, recommended by the World Health Organization (WHO), that are used to determine whether infertility is caused by abnormalities of one or more of them. The parameters include: sperm count, sperm motility, sperm viability, white blood cell count, sperm head morphology. 2. Indication(s) for use: N/A 3. Special conditions for use statement(s): N/A 4. Special instrument requirements: Assay has been validated on three commercial flow cytometers: Becton Dickinson FACScan, Beckman Coulter Epics XL, and Partec PAS III (German) I. Device Description: The GSA Kit consists of a set of reagents which when used with an appropriate flow cytometer [(Three have been validated with this kit: Becton Dickinson FACScan, Beckman Coulter Epics XL, and Partec PAS III (German)], enables the laboratory technician to perform a set of tests to evaluate five semen parameters: sperm count, sperm motility, sperm vitality, white blood cell count, and sperm head morphology (DNA condensation). J. Substantial Equivalence Information: 1. Predicate device name(s): Dyn-BioShaf Ltd., General Semen Analysis Kit (GSA) 2. Predicate K number(s): K024337 {2} 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Principle | Specific reagents that label cell surface molecules with a fluorescent dye (monoclonal antibodies tagged with fluorescent dye) or a fluorophore reagent whose fluorescent intensity depends on the physiological activity of the cell The fluorescent signal is measured by a flow cytometer with software to calculate results. | Same | | Intended Use | The GSA Kit is for the evaluation of human semen and provides a set of reagents to evaluate semen quality by measuring parameters, recommended by the World Health Organization (WHO), that are used to determine whether infertility is caused by abnormalities of one or more of them. | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Measured Parameters | Sperm count, sperm motility, sperm vitality, white blood cell count. | Sperm count, sperm motility, sperm viability, white blood cell count, sperm head morphology. | K. Standard/Guidance Document Referenced (if applicable): N/A L. Test Principle: Fluorescence intensity of a specific reagent is used to measure physiological function of the {3} sperm cells. For example, the vitality reagent passively diffuses into the cells and if the cell is vitality has enzymatic activity that cleaves the reagent. As a result of this cleavage the reagent becomes fluorescent and this fluorescence is detected by the flow cytometer, thus allowing counting of vital cells. For other parameters not related to physiological function such as WBC count—specific monoclonal antibodies labeled with FITC are used. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: a. Precision/Reproducibility: N/A b. Linearity/assay reportable range: N/A c. Traceability, Stability, Expected values (controls, calibrators, or methods): N/A d. Detection limit: N/A e. Analytical specificity: N/A f. Assay cut-off: N/A ### 2. Comparison studies: a. Method comparison with predicate device: In a prospective study clinical trial carried out in four USA clinics and one Israeli clinic, the morphology of 485 semen samples were evaluated by the Routine Method (WHO manual method) and by the GSA Kit. Table 1. Agreement Between Sperm Morphology by the Routine Method and the GSA kit. {4} | Routine Analysis | GSA Kit | | | | --- | --- | --- | --- | | | Normal | Abnormal | Total | | Normal | 180 | 25 | 205 | | | 87.8% | 12.2% | | | Abnormal | 45 | 235 | 280 | | | 16.1% | 83.9% | | | Total | 225 | 260 | 485 | In the above table the agreement of the GSA Kit with the Routine Analysis for abnormal sperm morphology was $83.9\%$ (exact $95\%$ CI, 79.1---88.0). The agreement of the GSA kit with the Routine Analysis for normal sperm morphology was $87.8\%$ (exact CI, 82.6--92.0). The overall agreement rate was $85.6\%$ (exact CI, $82.1\%$ -- $88.6\%$ ) Table 2. Comparison of GSA Kit for Sperm Head Morphology With Three Flow Cytometers | | (Y) Partec-PAS vs. (X) BD-FACScan | (Y) BC-Epix vs. (X) BD-FACScan | (Y) BC-Epix vs. (X) Partec-PAS | | --- | --- | --- | --- | | Slope | 0.952 | 0.970 | 0.992 | | Intercept | 3.762 | 1.966 | 0.128 | | R² | 0.964 | 0.975 | 0.960 | | Correlation (r) | 0.982 | 0.987 | 0.980 | | Residual Std Dev | 3.352 | 2.847 | 3.565 | | | | | | | Bias @ 40(% bias) | 1.85 (4.6%) | 0.76 (1.9%) | -0.18 (-0.05%) | | Bias @ 65 (% bias) | 0.66 (1.0%) | 0.00 (<0.01%) | -0.38 (-0.01%) | | Bias @ 90 (% bias) | -0.53 (-0.6%) | -0.75 (-0.01%) | -0.57 (-0.01%) | | | | | | | Mean Difference | 0.126 | -0.338 | -0.465 | | SD of Difference | 3.409 | 2.855 | 3.513 | b. Matrix comparison: N/A 3. Clinical studies: a. Clinical Sensitivity: N/A {5} b. Clinical specificity: N/A c. Other clinical supportive data (when a. and b. are not applicable): N/A 4. Clinical cut-off: N/A 5. Expected values/Reference range: For sperm head morphology, ≥ 65% of fluorescence staining in gate R2 defines a sample with normal DNA condensation. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: 1. A substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...