Intended Use

The GSA kit is for the evaluation of human semen. The kit provides the user with a set of reagents used to evaluate semen quality by measuring four (4) parameters recommended by the World Health Organization (WHO) that are used to determine whether infertility is caused by abnormalities of one or more of them. The parameters include:

  • . Sperm count
  • Sperm motility .
  • Sperm vitality t
  • . White blood cell (WBC) count
Device Description

The GSA kit is an in vitro diagnostic kit intended for use in a clinical laboratory test setting such as a test facility capable of providing laboratory assessment of male infertility, by skilled laboratory technicians, familiar with the handling of semen samples. This product does not contain biological materials of human origin. The kit provides the user with a set of reagents used to evaluate semen quality by measuring four parameters, recommended by the World Health Organization (WHO), that are used to help determine whether infertility is caused by abnormalities of one or more of them. The parameters to be tested for include:

  • . Sperm count
  • Sperm motility
  • . Sperm vitality
  • White blood cell (WBC) count .
AI/ML Overview

The provided text describes the performance evaluation of the GSA kit for semen analysis, comparing it to the manual method recommended by the World Health Organization (WHO).

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state numerical acceptance criteria (e.g., "Sperm Count agreement must be >90%"). Instead, the acceptance criterion for the GSA kit is generally described as being "at least as accurate as the Routine Method" and providing "more consistent results that are less dependent on the skills of an individual technician."

ParameterAcceptance Criteria (Implied)Reported Device Performance (GSA Kit vs. Routine Method)
Sperm CountComparable accuracy and more precise results.Comparable accuracy. Higher precision (Flowcytometer counts ~100x more cells). Inter- and intra-technician agreement: 91.6% in normal samples. Significantly higher reproducibility.
Sperm MotilityComparable accuracy and more precise results.Comparable accuracy. Higher precision and statistically more reproducible due to large amount of sperm cells analyzed. Inter- and intra-technician agreement: 82.3% in normal samples. Significantly higher reproducibility.
Sperm VitalityComparable accuracy and significantly more precise results.Comparable accuracy. Significantly more precise results. Inter- and intra-technician agreement: Close agreement in normal samples. Significantly higher reproducibility.
WBC CountComparable accuracy and more precise results.Comparable accuracy and more precise results.
Intra-technician PrecisionSmaller typical standard deviation for each parameter than Routine Method.Typical intra-technician standard deviation for each parameter is smaller when using the GSA kit, indicating high intra-technician precision for Flowcytometry assay. (Based on 5 semen samples repeatedly tested by the same technician).
Inter-technician PrecisionSignificantly smaller typical standard deviation for each parameter than Routine Method.Typical inter-technician standard deviation for each parameter is significantly smaller when using the GSA kit, indicating high inter-technician precision for Flowcytometry assay. (Based on 6 semen samples tested by 3 different technicians).

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size:
    • For Inter- and Intra-technician experiments for Sperm Count, Motility, and Vitality: "semen samples with approximately normal sperm count." No specific number is given for these general comparisons.
    • For Intra-Technician Precision: 5 semen samples were repeatedly tested.
    • For Inter-Technician Precision: 6 semen samples were tested.
  • Data Provenance: Blind, multi-center, prospective clinical studies.
    • Country of Origin: Israel (IVF unit and outpatient clinic Lin Center, Carmel Hospital) and USA (Baylor College of Medicine, Houston Medical Center).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • The ground truth (referred to as the "Routine Method") was established by "skilled laboratory technicians."
  • For the inter-technician precision study, three technicians evaluated the samples using the conventional method.
  • The document states that the sites "many years of experience performing semen analysis using the Routine Method," implying their expertise. Specific years of experience per technician are not detailed.

4. Adjudication Method for the Test Set:

  • The document implies a comparison between two different observers/methods rather than an adjudication process for a single ground truth. "A skilled laboratory technician evaluated each semen sample by the Routine Method and by a different skilled laboratory technician using the Flowcytometry method." This suggests a direct comparison, not an adjudication of discrepancies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done in the context of human readers improving with AI assistance.
  • This study compares a new device (GSA kit using Flowcytometry) against a manual method, focusing on the performance of the device and its impact on technician variability, not on AI assistance for human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done:

  • This is not applicable in the context of AI. The GSA kit uses a "Flowcytometry method," which automates parts of the semen analysis, but it's a device/kit, not an AI algorithm in the modern sense. The "standalone" performance is the performance of the GSA kit without direct human manual counting for the parameters it evaluates.

7. The Type of Ground Truth Used:

  • Expert Consensus / Reference Method: The ground truth was established by the "Routine Method as recommended by the WHO," performed by skilled laboratory technicians. This serves as the reference standard against which the GSA kit's performance is compared.

8. The Sample Size for the Training Set:

  • The document does not mention a training set. The GSA kit is described as a "kit" providing "reagents" and utilizing "Flowcytometry." This suggests a laboratory assay with a fixed protocol, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

  • As no training set is mentioned or implied for this type of device, this question is not applicable.

{0}------------------------------------------------

JUL 03 2003 K024337

510(k) Summary of Safety and Effectiveness

Intended Use

The GSA kit is for the evaluation of human semen. The kit provides the user with a set of reagents used to evaluate semen quality by measuring four parameters, recommended by the World Health Organization (WHO), that are used to help determine whether infertility is caused by abnormalities of one or more of them. The parameters to be tested for include:

  • . Sperm count
  • Sperm motility
  • . Sperm vitality
  • White blood cell (WBC) count .

The GSA kit is an in vitro diagnostic kit intended for use in a clinical laboratory test setting such as a test facility capable of providing laboratory assessment of male infertility, by skilled laboratory technicians, familiar with the handling of semen samples. This product does not contain biological materials of human origin.

The performance of the GSA kit was established by comparison to the manual method as recommended by the WHO.

The GSA kit involves specimen acquisition by ejaculation of semen by the patient for testing purposes. This assay presents no more of a safety hazard than the routine tests where semen samples are obtained from subjects.

Test Performance

The clinical performance of the GSA kit was evaluated via blind, multi-center, prospective clinical studies. The primary objective was to determine whether evaluation of sperm cells using the GSA kit is at least as accurate as the Routine Method. Another objective was to show that using the GSA kit provides more consistent results that are less dependent on the skills of an individual technician.

The study took place at two sites:

  • IVF unit and outpatient clinic Lin Center, Carmel Hospital, Israel. .
  • Baylor College of Medicine, Houston Medical Center, USA. ●

Each of the above sites has many years of experience performing semen analysis using the Routine Method.

{1}------------------------------------------------

A skilled laboratory technician evaluated each semen sample by the Routine Method and by a different skilled laboratory technician using the Flowcytometry method. The evaluation was based on the parameters recommended by the WHO (see below).

Sperm Count

Comparison of the GSA kit and the Routine Method showed that using the GSA kit with the Flowcytometer provides comparable accuracy and more precise results. The Flowcytometer allows counting of approximately 100 times more cells as compared to the light microscope and counting chamber.

The inter- and intra-technician experiments showed that in semen samples with approximately normal sperm count there is 91.6% agreement between the two methods while the reproducibility achieved by the Flowcytometry method is significantly higher.

Sperm Motilitv

Comparison of the GSA kit with the Routine Method showed that using the GSA kit provides comparable accuracy and more precise results. Again, because of the large amount of sperm cells that can be analyzed and the higher precision in cell count, the determination of the percentage of motile cells is statistically more reproducible using the GSA kit.

The inter- and intra- technician experiments showed that in semen samples characterized by approximately normal sperm count, there is close agreement (82.3%) between the two methods while the reproducibility achieved by the GSA kit is significantly higher.

Sperm Vitality

Comparison of the GSA kit with the Routine Method showed that using the GSA kit provides comparable accuracy and significantly more precise results.

The inter- and intra-technician experiments showed that in semen samples characterized by approximately normal sperm count, there is close agreement between the two methods while the reproducibility achieved by the GSA kit is significantly higher.

White Blood Cells (WBC)

Comparison of the GSA kit with the Routine Method showed that using the GSA kit provides comparable accuracy and more precise results.

Intra-Technician Precision

In an attempt to determine and compare the intra-technician precision of the two methods, five semen samples were repeatedly tested by both conventional (light microscope and counting chamber) analysis and Flowcytometry analysis according to the WHO manual. For each analysis method the same technician repeatedly evaluated the

{2}------------------------------------------------

semen samples in order to evaluate the intra-technician variability obtained by each methodology. The results show that by using the GSA kit the typical intra-technician standard deviation for each parameter is smaller then by using the Routine Method indicating a high intra-technician precision for Flowcytometry assay.

Inter-Technician Precision

To compare the inter-technician precision of the two methods six semen samples were tested by both conventional analysis and Flowcytometry analysis according to the WHO manual method. For each analysis method three technicians evaluated all semen samples and the results obtained for each sample by each technician were compared in order to evaluate the inter-technician variability obtained by each methodology. The results show that by using the GSA kit the typical inter-technician standard deviation for each parameter is significantly smaller then by using the Routine Method indicating a high inter-technician precision for Flowcytometry assay.

Interfering Substances

The following table summarizes data obtained using human semen samples characterized by certain abnormalities.

Interfering SubstanceAmountResult
High level of WBC$\geq 4 \times 10^7$ /mlMay cause high staining background and bythis false WBC count result. A new samplemust be obtained after the infection is over.
Fungi infectionVisiblecloudinessMay cause auto-fluorescence which can resultin erroneous results. Thus, contaminatedsemen samples should not be tested.
Extremely viscoussemen sample.-In rare cases dilution of the sample ispractically impossible and dilution treatmentusing Bromolin may be required.

Product Stability

The stability of the kit was tested in real time as the components were stored refrigerated and the performance tested periodically. The shelf life of the kit has been determined to be 6 months at the specified storage conditions.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circle around the eagle. The eagle is depicted in a simple, line-art style, and the text is in a sans-serif font. The overall design is clean and professional.

10903 New Hampshire Avenue Silver Spring, MD 20993

DEC 21 2012

Devices & Diagnostics Consulting Group, Inc. c/o Mr. Thomas M. Tsakeris President 16809 Briardale Road Rockville, Maryland 20855

Re: K024337

Trade/Device Name: BioShaf Limited General Semen Analysis (GSA) Kit Regulation Number: 21 CFR § 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Date: April 28, 2003 Received: April 28, 2003

Dear Mr. Tsakeris:

This letter corrects our substantially equivalent letter of July 3, 2003.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

Page 2 - Mr. Thomas Tsakeris

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Sincerely yours,

marie m chan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number: K024337

Device Name: General Semen Analysis (GSA) Kit

Indications for Use:

The GSA kit is for the evaluation of human semen. The kit provides the user with a set of reagents used to evaluate semen quality by measuring four (4) parameters recommended by the World Health Organization (WHO) that are used to determine whether infertility is caused by abnormalities of one or more of them. The parameters include:

  • . Sperm count
  • Sperm motility .
  • Sperm vitality t
  • . White blood cell (WBC) count

In most fertility test laboratorics these parameters are manually evaluated on a routine basis using a light microscope and a counting chamber. The performance characteristics of the GSA kit is based on a comparison between the GSA kit and the routine manual method performed according to the latest edition of the Manual for the Examination of Human Semen and Sperm published by the WHO.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Evaluation and Safety (OIVDES)

Professional use:

Prescription Use:

(Per 21 CFR 801.109)

OR

Over-the-counter Use: (Optional Format 1-2-96)

:

:

Caroline Bartoli

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K024337

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”