K Number
K071657
Device Name
TRU FLU
Date Cleared
2007-11-15

(150 days)

Product Code
Regulation Number
866.3328
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TRU FLU is a rapid, qualitative, lateral-flow immunochromatographic assay for detecting both influenza A and influenza B viral nucleoprotein antigens in human nasal wash, nasopharyngeal aspirate and nasal and nasopharyngeal swab samples from symptomatic patients. The test is not intended for the detection of influenza C viruses A negative test is presumptive and it is recommended these results be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other clinical management decisions.

Device Description

TRU FLU is a single use immunoassay that consists of a Conjugate Tube, a Test Strip, and Sample Diluent. The Conjugate Tube contains a lyophilized bead of colloidal gold-linked monoclonal antibodies to influenza A and influenza B (detector antibodies). The Test Strip carries a nitrocellulose membrane with dried capture antibodies at separate lines for influenza B. The Test Strip holder caps the Conjugate Tube during testing and subsequent disposal to reduce exposure to potential pathogens.

The conjugate bead is first rehydrated in the Conjugate Tube with Sample Diluent, prior to the addition of patient specimen. The contents are mixed before the Test Strip is added. As the test is incubated at 20-25 C. influenza A or influenza B antigens, if present in the diluted sample, bind to the corresponding monoclonal antibody-colloidal gold conjugate as the sample moves up the Test Strip. The influenza A capture monoclonal antibody is bound to the Test Strip at the test-FLU A position of the device. When it binds the antigen-influenza A antibody-colloidal gold complex, it yields a visible pink-red line. Similarly, the influenza B capture monoclonal antibody bound to the assay membrane at the test-FLU B position will result in a pink-red line when it captures antibody-colloidal gold complexes. When no antigen is present, no complexes are formed and no pink-red line will appear at either the test FLU A or the test FLU B position of the Test Strip. An internal control line helps determine whether adequate flow has occurred through the Test Strip during a test run. A visible pink-red line at the Control position of the Test Strip should be present each time a specimen or control is tested. If no pink-red control line is seen, the test is considered invalid.

AI/ML Overview

Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:

Device: TRU FLU (Rapid, qualitative lateral-flow immunoassay for the detection of Influenza A and B antigens)

Indications for Use: Rapid, qualitative, lateral-flow immunochromatographic assay for detecting both influenza A and influenza B viral nucleoprotein antigens in human nasal wash, nasopharyngeal aspirate and nasal and nasopharyngeal swab samples from symptomatic patients. Not intended for detection of influenza C viruses. Negative results are presumptive and should be confirmed by cell culture; they should not be the sole basis for treatment or clinical management.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria (e.g., "Sensitivity must be >= 90%"). However, based on the performance evaluation and the 510(k) clearance, the implied acceptance criteria are that the device demonstrates substantial equivalence to the reference method (tissue culture) and the predicate device (Binax NOW Influenza A&B) in terms of sensitivity and specificity for both Influenza A and B across different sample types and patient demographics.

Here's the summarized reported device performance against the "Gold Standard" (Tissue Culture):

MetricSubgroupTRU FLU Influenza A Performance (vs. Tissue Culture)TRU FLU Influenza B Performance (vs. Tissue Culture)
Overall SensitivityFresh Wash/Aspirate87.2% (95% CI: 72.6 - 95.7%)63.6% (95% CI: 45.1 - 79.6%)
Fresh Swab86.3% (95% CI: 76.3 - 93.2%)58.5% (95% CI: 45.6 - 70.6%)
Frozen Wash/Aspirate85.0% (95% CI: 62.1 - 96.8%)50.0% (95% CI: 1.3 - 98.7%)
Overall SpecificityFresh Wash/Aspirate89.3% (95% CI: 85.9 - 92.6%)99.4% (95% CI: 97.8 - 99.9%)
Fresh Swab92.7% (95% CI: 88.7 - 95.6%)99.6% (95% CI: 97.8 - 100%)
Frozen Wash/Aspirate93.0% (95% CI: 80.9 - 98.5%)100.0% (95% CI: 94.1 - 100%)
ReproducibilityHigh Positive & Low Negative100% correct results100% correct results
High Negative (near LOD)Weakly positive results in 8/72 replicatesN/A (not specified as often for B)
Low Positive (just above LOD)1 negative reaction out of 72 replicatesN/A (not specified as often for B)

(Note: "N/A" in the table above for 95% CI when only 1 positive sample existed implies the CI was not calculable or meaningful for such a small n, which is a limitation of some individual site data, not the overall study.)


2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Total Samples: 697 prospectively collected fresh samples + 63 frozen samples = 760 samples (after excluding 1 invalid tissue culture result from the 697 prospective samples for sensitivity/specificity calculations).
    • Fresh Samples: 697 samples (366 wash/aspirate, 320 swab, 11 not defined).
    • Frozen Samples: 63 samples (all wash/aspirate).
  • Data Provenance:
    • Country of Origin: United States. The study states "Nine independent laboratories (in different geographic regions of the US) and the manufacturer participated in the study."
    • Retrospective or Prospective: Primarily prospective. "The test plan was designed to evaluate the performance of prospectively collected fresh and frozen samples from the 2006-7 season." The document also mentions "prospectively collected fresh and 63 frozen samples were tested." Tissue culture on frozen samples was performed at the time of collection.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Not specified. The ground truth (tissue culture) was performed at "each laboratory's established internal method." It does not explicitly state that experts established a consensus ground truth for the device evaluation; rather, a standard laboratory method was employed.
  • Qualifications of Experts: Not specified. It's assumed that standard certified laboratory personnel performed the tissue culture, but specific qualifications (e.g., "clinical microbiologist with X years of experience") are not detailed.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (like 2+1 or 3+1) for resolving discrepancies between the TRU FLU results and the tissue culture results. The tissue culture results appear to have been used directly as the reference standard without further expert adjudication mentioned.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study focuses on the standalone performance of the TRU FLU device against a gold standard (tissue culture) and compares it to a predicate device, not on assessing human reader improvement with or without AI assistance. The device in question is a rapid diagnostic test, not an AI-assisted diagnostic tool.


6. Standalone Performance Study

Yes, a standalone study was performed. The entire clinical performance evaluation section describes the algorithm's (TRU FLU device's) performance without human intervention, specifically comparing its results against the tissue culture gold standard. The device requires no instruments and is read directly by a user. The sensitivity and specificity values reported are representative of the device's standalone performance.


7. Type of Ground Truth Used

The type of ground truth used was tissue culture isolation. The document explicitly states: "Tissue culture isolation remains the gold standard for the detection of influenza, yet the procedure can take 7 days to complete". It further confirms: "Trial sites were instructed to test samples in parallel by tissue culture (standard reference), TRU FLU and the predicate device, NOW Influenza A&B. Tissue culture (reference method) was performed by the site using the site's method."


8. Sample Size for the Training Set

The document does not provide information regarding a separate "training set" or its sample size. This is common for traditional in-vitro diagnostic devices (like lateral-flow immunoassays) where the "training" analogous to machine learning models is typically the biochemical assay design and optimization, followed by verification and validation, rather than a distinct data-driven training phase on labeled data. The study described focuses on clinical validation of the final device.


9. How Ground Truth for the Training Set Was Established

As no separate "training set" is mentioned (see point 8), the document does not describe how ground truth for such a set was established. The clinical evaluation sets (fresh and frozen samples) used tissue culture as the ground truth, as detailed in point 7.

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS

K071657

IDENTIFICATION INFORMATION

SUBMITTER'S INFORMATION

This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.20.

SUBMITTER'S NAME AND ADDRESS: Meridian Bioscience, Inc. 3471 River Hills Drive Cincinnati, OH 45244

PHONE NUMBER: (513) 271-3700

FAX NUMBER: (513) 272-5213

CONTACT PERSON: Susan Rolih Official Correspondent

DATE SUMMARY PREPARED: November 13, 2007

TRADE NAME: TRU FLU

COMMON NAME: Rapid, qualitative lateral-flow immunoassay for the detection of Influenza A and B antigens

CLASSIFICATION NAME: Antigen, CF (including CF control), influenza virus A, B, C

REGULATION: 866.3330

INTENDED USES:

The TRU FLU is a rapid, qualitative, lateral-flow immunochromatographic assay for detecting both influenza A and influenza B viral nucleoprotein antigens in human nasal wash, nasopharyngeal aspirate and nasal and nasopharyngeal swab samples from symptomatic patients. The test is not intendedf or the detection of influenza C viruses A negative test is presumptive and it is recommended these results be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other clinical management decisions.

PREDICATE DEVICES:

TRU FLU is a modification of, and is intended to detect the same analytes as, ImmunoCard STAT! Flu A&B PLUS (K041626), Meridian Bioscience, Inc.

Binax NOW Influenza A&B, Inverness Medical (K062109)

BACKGROUND:

Influenza is a highly contagious, epidemic to pandemic acute viral respiratory disease caused by several genera of the Orthomyxoviridae family. Influenza virus A and Influenza virus B are the two genera most commonly associated with disease in humans. Influenza infection rates tend to be highest in pediatic populations, while serious complications from influenza disease are more common in the elderly. Clinical signs and symptoms begin after a 1-4 day incubation period and include cough, fever, myalgia and malaise. The clinical presentation of influenza can range from asymptomatic infection to fatal pneumonia. Influenza co-circulates with other respiratory pathogens; hence it is important to differentiate influenza from other respiratory diseases. Antiviral drugs in general have shown more significant clinical benefit when administered within 48 hours of the appearance of symptoms, which

NOV 1 5 2007

{1}------------------------------------------------

obviates the need for the rapid detection of influenza. Not all antiviral drugs are effective against both influenza A and influenza B; therefore it is important to distinguish between the two.

Influenza A and B can be detected in human respiratory samples by a variety of methods including tissue culture, immunofluorescent assay and enzyme immunoassay. Tissue culture isolation remains the gold standard for the detection of influenza, yet the procedure can take 7 days to complete immunofluorescent antibody-based tests are moderately sensitive, yet highly dependent on specimen quality and preparation. The rapid detection of influenza using enzyme and microparticle-based immunoassays has become an important aspect of patient management in patients of all ages with acute respiratory disease due to influenza. The results of such tests are used to support data available from the patient's clinical evaluation and assist the physician in determining the course of action.

Type of test

TRU FLU is a rapid, single-use, qualitative lateral-flow immunoassay screening test.

Specimen type

The following specimens have been found compatible with TRU FLU.

    1. Nasal wash
  • Nasopharyngeal aspirate 2.
  • Nasopharyngeal swab જં
    1. Nasal swab

Conditions for use

TRU FLU is designed for use by laboratory professionals under the normal environmental conditions. The assay, which is stored at 2-25 C when not in use, is brought to room temperature prior to use. Normal laboratory lighting, humidity and temperature do not affect the performance of the assay.

Contraindications

There are no contraindications associated with the use of this product.

Special instrument requirements

No instruments are used with this product.

Combination with other medical devices

No other medical devices are used in combination with this device.

{2}------------------------------------------------

JeviceI
and the first to the state a more a more a more a more and the many ofPredicate
I1
John Device Format ve "
Prior Ft
VC
Bearing States
-- TRII EI ''
charts --And Change of Children College of the
Comparison e1
Table
CharacteristicsTRU FLUImmunoCard STAT! Flu A&B PLUS(prior format)BinaxNOW Influenza (predicate)
Device Type
TechnologySingle use, rapid, lateral flowimmunoassaySingle use, rapid, lateral flowimmunoassaySingle use, rapid, lateral flowimmunoassay
In vitro diagnostic deviceYesYesYes
ControlPurchased separatelyPurchased separatelySupplied with kit
CalibratorNoNoNo
Intended Use
Detection of influenza AantigenYesYesYes
Detection of influenza BantigenYesYesYes
Screening testNoNoNo
Diagnostic testYesYesYes
Identification testNoNoNo
Monitoring therapyNoNoNo
Acceptable Samples
Swab -- NasalYesYesYes
Swab -- NasopharyngealYesYesYes
Wash -- NasalYesYesYes
Wash - NasopharyngealYesYesYes
Aspirate -- NasopharyngealYesYesNo
Reagents/ComponentsProvided
Nitrocellulose test stripYes (attached to plastic holder/tubeclosure)Yes (enclosed in plastic frame)Yes (enclosed in cardboard frame)
Conjugate reagentYes (supplied as dried bead inConjugate Tube)Yes (supplied in conjugate padattached to test strip)Yes (supplied in conjugate padattached to test strip)
Reading GuideYes (part of plastic holder/tubeclosure)Yes (part of plastic frame)Yes (part of cardboard frame)
CharacteristicsTRU FLUImmunoCard STAT! Flu A&B PLUS (prior format)BinaxNOW Influenza (predicate)
Sample Diluent/NegativeControl (external)YesYesNo
Internal procedural controlYesYesYes
External positive controlNo (Purchased separately --FLU/RSV PositiveControl,Catalogue 751110No (Purchased separately --FLU/RSV PositiveControl,Included in kit as dry swab.
Source of influenza AantibodiesMonoclonal M2110169, IVF8Monoclonal M2110169, IVF8Not known
Source of influenza BantibodiesMonoclonal 2/3, M2110171Monoclonal 2/3, M2110171Not known

Section 5, Page 3 of 19

{3}------------------------------------------------

Table 1 Continued

{4}------------------------------------------------

Meridian Bioscience, Inc.
Cincinnati, OH

Table 2 Comparison of TRU FLU method to predicate

Comparison of assay steps*TRU FLUImmunoCard STAT! Flu A&B PLUS(prior format)Binax NOW Influenza A&B(predicate)
TechnologyLateral-flow,immunoassaygold-based Lateral-flow,immunoassayLateral-flow immunochromatographic assay.
Test Reagents1. Test Strip (nitrocellulose membrane withimmobilized capture antibody). Top endis inserted into plastic frame or holder.2. Conjugate Tube containing antibody-colloidal gold conjugate (lyophilized bead)3. Sample Diluent/Negative Control1. Test Device (Test Card with nitrocellulosemembrane with immobilized antibody,conjugate pad with colloidal gold, plasticframe with reading/reaction window andsample port2. Sample Diluent/Negative Control1. Test Devices2. Transfer pipettes3. Positive Control Swab4. Negative control Swab5. Elution Solution Vials for Control Swabs
Specimen TypeExternal Positive Control (Flu/RSV PositiveControl) sold separately as adjunct reagent.1. Nasal wash2. Nasopharyngeal aspirate3. Nasal swabs4. Nasopharyngeal swabsExternal Positive Control (Flu/RSV PositiveControl) sold separately as adjunct reagent.1. Nasal wash2. Nasopharyngeal aspirate3. Nasal swabs4. Nasopharyngeal swabs1. Nasal Wash2. Nasal aspirate3. Nasopharyngeal swab
Equipment RequiredNoNoNo
Level of skill requiredComplexity: ModerateComplexity: ModerateComplexity: CLIA Waived
Assay steps1. Add 100 µL Sample Diluent to theConjugate Tube.2. Add 100 µL sample to the Conjugate Tubeand mix.3. Insert Test Strip to Conjugate Tube.4. Press down on cap of Test Strip to sealConjugate Tube.5. Incubate 15 min., 20-25 С.6. Read at end of incubation using guide onCap-Carrier.1. Add 4 drops Sample Diluent to a testtube2. Add 150 µl sample and mix3. Add 150 µL diluted specimen to TestDevice.4. Incubate 15 min., 20-25 C.5. Read at end of incubation using guide atreaction window.1. Aspirate sample with transfer pipette.2. Add sample in drop-wise manner to samplepad.3. Close cardboard cover and incubate 15 minat 20-25 C.4. Read immediately at end of 15 minutes.
End pointAppearance of pink-red color at Test and/orControl linesAppearance of pink-red color at Test and/orControl linesAppearance of pink-purple color at Test and/orControl lines
Interpretation of test resultFlu A Positive = appearance of pink-red lines atFlu A test and control positions (indicatespresence of Influenza A antigens)Flu B Positive = appearance of pink-red linesat Flu B test and control positions (indicatespresence of Influenza B antigens)Negative = no test line color with pink-redcontrol line (indicates absence of Influenza Aor B antigens)Flu A Positive = appearance of pink-red lines atFlu A test and control positions (indicatespresence of Influenza A antigens)Flu B Positive = appearance of pink-redlines at Flu B test and control positions(indicates presence of Influenza B antigens)Negative = no test line color with pink-redcontrol line (indicates absence of InfluenzaA or B antigens)Flu A Positive = appearance of pink-purple lines atFlu A test and control positions (indicates presenceof Influenza A antigens)Flu B Positive = appearance of pink-purplelines at Flu B test position and control positions(indicates presence of Influenza B antigens)Negative = no test line color with a pink-purplecontrol line (indicates absence of influenza A or Bantigens)
Note:Differences are underlined to facilitate their detection

{5}------------------------------------------------

DEVICE DESCRIPTION AND TECHNOLOGICAL PRINCIPLES

Reagents

TRU FLU is distributed as a test kit that includes the following reagents:

  • Test Strip: A test strip attached to a plastic frame or holder enclosed in a foil pouch with 1. desiccant. The test strip carries monoclonal anti-influenza B capture antibodies* for the test. The holder is used to stopper the Conjugate Tube. The paddle portion of the holder indicates where test and control lines should appear.
  • Coniugate Tube: A capped plastic tube containing a conjuqate bead. The tube is enclosed in a 2. foil pouch. The conjugate consists of gold-conjugated anti-influenza A and anti-influenza B which serve as the detector antibodies.
  • Sample Diluent/Negative Control: A buffered protein solution provided in a plastic vial. Sodium 3. azide (0.094%) added as a preservative. Use as supplied.
    1. Plastic transfer pipettes with 50, 100, 200 and 300 µL volume marks.

Equipment needed to use the device

There is no equipment needed to use this device.

Interfering substances

Whole blood, at concentrations greater than 0.5% may interfere with the interpretation of test results.

Calibrators

There are no calibrators used with this device.

Controls

The assay includes an internal procedural control line that is used to determine if the test has been performed correctly, proper flow has occurred and that reagents were reactive at the time of use. A clean background around the test and control lines also serves as a procedural control. Control or test lies that are obscured by a heavy background color may invalidate the test and may be an indication of reagent deterioration, use of inappropriate sample or improper test performance.

Positive Control Reagent is supplied separately. It is used in parallel with Sample Diluent/Negative Control as external controls. These reagents also serve as indicators that the test was performed correctly, that the capture and detector antibodies were active at the time of use, and that the membrane supports proper sample flow.

Failure of the internal and external control to produce the expected results suggests the test was not performed correctly (ie, incorrect volume of reagents added; incorrect incubation temperature or times used or that reagents were not brought to room temperature prior to testing).

Technological principles

TRU FLU is a single use immunoassay that consists of a Conjugate Tube, a Test Strip, and Sample Diluent. The Conjugate Tube contains a lyophilized bead of colloidal gold-linked monoclonal antibodies to influenza A and influenza B (detector antibodies). The Test Strip carries a nitrocellulose membrane with dried capture antibodies at separate lines for influenza B. The Test Strip holder caps the Conjugate Tube during testing and subsequent disposal to reduce exposure to potential pathogens.

{6}------------------------------------------------

The conjugate bead is first rehydrated in the Conjugate Tube with Sample Diluent, prior to the addition of patient specimen. The contents are mixed before the Test Strip is added. As the test is incubated at 20-25 C. influenza A or influenza B antigens, if present in the diluted sample, bind to the corresponding monoclonal antibody-colloidal gold conjugate as the sample moves up the Test Strip. The influenza A capture monoclonal antibody is bound to the Test Strip at the test-FLU A position of the device. When it binds the antigen-influenza A antibody-colloidal gold complex, it vields a visible pink-red line. Similarly, the influenza B capture monoclonal antibody bound to the assay membrane at the test-FLU B position will result in a pink-red line when it captures antibody-colloidal gold complexes. When no antigen is present, no complexes are formed and no pink-red line will appear at either the test FLU A or the test FLU B position of the Test Strip. An internal control line helps determine whether adequate flow has occurred through the Test Strip during a test run. A visible pink-red line at the Control position of the Test Strip should be present each time a specimen or control is tested. If no pink-red control line is seen, the test is considered invalid.

PERFORMANCE EVALUATION - CLINICAL/FIELD TRIALS

Study Objective

A clinical/field study was conducted to demonstrate that TRU FLU was substantially equivalent in performance to the standard reference method - tissue culture - and to the predicate device Binax NOW Influenza A&B (K062109) in a clinical laboratory setting using samples submitted for influenza testing.

Investigation Plan

The test plan was designed to evaluate the performance of prospectively collected fresh and frozen samples from the 2006-7 season. Nine independent laboratories (in different geographic regions of the US) and the manufacturer participated in the study. Patient specimens, de-linked from identification information, were used. Trial sites were instructed to test samples in parallel by tissue culture (standard reference), TRU FLU and the predicate device, NOW Influenza A&B. Tissue culture (reference method) was performed by the site using the site's method.

Sample population and selection

The sample population used in this study included respiratory samples from patients of any age provided the samples had been submitted for influenza testing. Such samples were assumed to be from symptomatic patients. The sample types included nasal wash, nasopharyngeal aspirate and nasal and nasopharyngeal swabs.

Influence of other disease states

There are no known disease states that would have an influenza test results other than infection with influenza A or B virus.

Patient exclusion criteria

Samples from asymptomatic patients were excluded from the trials. There were no exclusions based on patient age gender or other therapies.

Clinical trial test system

Clinical trial sites employed full production lots of TRU FLU test kits that were labeled for investigational use only. TRU FLU, the predicate device and record sheets were shipped to test sites throughout the study. Unused portions of the assays were returned to Meridian at the completion of the study. When not in use, test kits were stored according to the instructions in the provisional IFU.

{7}------------------------------------------------

Patient samples were prepared according the package inserts that accompanied the test and predicate device. Viral cultures were performed by each laboratory's established internal method. Three of the nine independent laboratories completed reproducibility studies prior to testing patient samples. Reproducibility studies are described later in this section. Clinical data was analyzed using Fisher's exact method and is presented in 2 x 2 tables.

Clinical study data

A total of 697 prospectively collected fresh and 63 frozen samples were tested. Tissue culture tests were performed on all frozen samples at the time of their collection. As shown in Table 3, 52% (366/697) of fresh samples were wash/aspirate samples and 46% (320/697) were swab samples. Sample type was not identified for the remaining 2% (11/697). All frozen samples were wash/asp. Table 4 identifies the age groups of the patients from whom fresh samples were collected during the study. 27% (185/697) of the fresh samples were from patients 22 years of age or older, 64% (447/697) from patients 12 years of age or less, and 8% (59/697) from children 13 years to 21 years of age. The age of patients submitting 6 samples was not recorded. Of the 63 frozen samples only, the majority (33/63 or 52%) were collected from patients aged 2 months to 2 years. The patient ages for the remaining frozen samples were 14% (9/63) for patients aged 1 month or less, 22% (14/63) patients aged 3-12 years, and 11% (7/63) patients aged 13-21 years. Tables 5a-b categorize patients by gender. Fifty three percent (367/697) of the fresh samples and 51% (32/63) frozen samples were from male patients, while 47% (325/697) fresh and 49% (31/63) frozen samples were from females. The gender of patients submitting 5 samples was not recorded. Neither patient age nor gender affected assay performance. The 2 x 2 tables that summarize test results for each site are given in Table 6. Table 7 stratifies the data by patient age.

Of the 697 prospective samples tested. 1 produced invalid results in tissue culture and was excluded from the calculations for either sensitivity or specificity (see Tables 6 and 7).

{8}------------------------------------------------

Table 3 Description of sample types evaluated in the clinical studies

Wash/NPASwabNot DefinedTotal
Clinical Site 1
Total tested500050
Total fresh500050
Total frozen0000
Clinical Site 2
Total tested30820112
Total fresh30820112
Total frozen0000
Clinical Site 3
Total tested637070
Total fresh0707
Total frozen630063
Clinical Site 4
Total tested71690140
Total fresh71690140
Total frozen0000
Clinical Site 8
Total tested180018
Total fresh180018
Total frozen0000
Clinical Site 10
Total tested52600112
Total fresh52600112
Total frozen0000
Clinical Site 11
Total tested90211103
Total fresh90211103
Total frozen0000
Clinical Site 12
Total tested097097
Total fresh097097
Total frozen0000
Clinical Site 13
Total tested493052
Total fresh493052
Total frozen0000
Clinical Site 14
Total tested6006
Total fresh6006
Total frozen0000
Clinical Site Totals
Total tested42932011760
Total fresh36632011697
Total frozen630063

Legend: NPA = nasopharyngeal aspirate

:

:

{9}------------------------------------------------

Patient Agebirth to 1month>1 monthto 2 years>2 yearsto 12years>12 yearsto 21years>21 yearsNotDefinedTotal
Clinical site 1
Total tested fresh21024131050
Clinical site 2
Total tested fresh611157685112
Clinical site 3
Total tested fresh0130307
Total tested frozen9331470063
Clinical site 4
Total tested fresh2063391530140
Clinical site 8
Total tested fresh011115018
Clinical site 10
Total tested fresh46138351112
Clinical site 11
Total tested fresh1150241350103
Clinical site 12
Total tested fresh125485097
Clinical site 13
Total tested fresh438730052
Clinical site 14
Total tested fresh1500006
Clinical site Totals
Total tested fresh49242156591856697
Total tested frozen9331470063
Total individual samples58275170661856760

Table 4 Categories of patients by age from which samples were collected for clinical studies

{10}------------------------------------------------

·

Table 5a Classification of fresh samples based on patient gender

MaleFemaleNot definedTotal
Clinical site 1
Total tested2129050
Total TRU FLU A positive1304
Total TRU FLU B positive2608
Total TRU FLU negative1819037
TRU FLU A & B Positive0101
Clinical site 2
Total tested58495112
Total TRU FLU A positive2522047
Total TRU FLU B positive0044
Total TRU FLU negative3327161
TRU FLU A & B Positive0000
Clinical site 3
Total tested4307
Total TRU FLU A positive0000
Total TRU FLU B positive0101
Total TRU FLU negative3205
TRU FLU A & B Positive1001
Clinical site 4
Total tested76640140
Total TRU FLU A positive1613029
Total TRU FLU B positive3508
Total TRU FLU negative56460102
TRU FLU A & B Positive1001
Clinical site 8
Total tested144018
Total TRU FLU A positive0000
Total TRU FLU B positive0000
Total TRU FLU negative144018
TRU FLU A & B Positive0000
Clinical site 10
Total tested60520112
Total TRU FLU A positive1213025
Total TRU FLU B positive139022
Total TRU FLU negative3125056
TRU FLU A & B Positive4509
Clinical site 11
Total tested55480103
Total TRU FLU A positive55010
Total TRU FLU B positive1102
Total TRU FLU negative4942091
TRU FLU A & B Positive0000
MaleFemaleNot definedTotal
Clinical site 12
Total tested4453097
Total TRU FLU A positive94013
Total TRU FLU B positive0303
Total TRU FLU negative3546081
TRU FLU A & B Positive0000
Clinical site 13
Total tested3319052
Total TRU FLU A positive5106
Total TRU FLU B positive0202
Total TRU FLU negative2816044
TRU FLU A & B Positive0000
Clinical site 14
Total tested2406
Total TRU FLU A positive2204
Total TRU FLU B positive0000
Total TRU FLU negative0202
TRU FLU A & B Positive0000
Clinical site Totals
Total tested3673255697
Total TRU FLU A positive75630138
Total TRU FLU B positive1927450
Total TRU FLU negative2672291497
TRU FLU A & B Positive06012

{11}------------------------------------------------

Table 5a Continued

Table 5b Classification of frozen samples based on patient gender

MaleFemaleNot definedTotal
Clinical site 1
Total tested3231063
Total TRU FLU A positive137020
Total TRU FLU B positive0101
Total TRU FLU negative1923042
Total invalid**0000

{12}------------------------------------------------

TRU FLU Influenza A- Data stratified by site and sample type
Site IDPositive SamplesNegative Samples
FreshWash/AspirateTRU A/Culture%Sensitivity95% CITRU A/Culture%Specificity95% CI
14/4100%39.8 - 100%45/4697.8%88.5 - 99.9%
213/1586.7%59.5 - 98.3%13/1586.7%59.5 - 98.3%
42/366.7%9.4 - 99.2%58/6885.3%74.6 - 92.7%
80/0N/AN/A18/18100%81.5 - 100%
108/988.9%51.8 - 99.7%34/4379.1%64.0 - 90.0%
112/366.7%9.4 - 99.2%78/8690.7%82.5 - 95.9%
131/1100%N/A43/4889.6%77.3 - 96.5%
144/4100%39.8 - 100%2/2100%15.8 - 100%
TOTAL34/3987.2%72.6 - 95.7%291/32689.3%85.9 - 92.6%
Site IDPositive SamplesNegative Samples
FreshSwabTRU A/Culture%Sensitivity95% CITRUA/Culture%Specificity95% CI
231/3491.2%76.3 - 98.1%47/4897.9%88.9 - 99.9%
30/0N/AN/A6/785.7%42.1 - 99.6%
413/1872.2%46.5 - 90.3%46/5190.2%78.6 - 96.7%
106/785.7%42.1 - 99.6%42/5379.2%65.9 - 89.2%
110/0N/AN/A2/2100%15.8 - 100%
1213/1492.9%66.1 - 99.8%83/83100%95.7 - 100%
130/0N/AN/A3/3100%29.2 - 100%
TOTAL63/7386.3%76.3 - 93.2%229/24792.7%88.7 - 95.6%
Site IDPositive SamplesNegative Samples
FrozenWash/AspirateTRU A/Culture%Sensitivity95% CITRU A/Culture%Specificity95% CI
317/2085.0%62.1 - 96.8%40/4393.0%80.9 - 98.5%

Table 6 Summary of TRU FLU results vs tissue culture – data stratified by site and sample type

{13}------------------------------------------------

Table 6 Cont'd.

:

TRU FLU Influenza B - Data stratified by site and sample type
Site IDFreshWash/AspiratePositive SamplesNegative Samples
TRUB/Culture%Sensitivity95% CITRUB/Culture%Specificity95% CI
18/1172.7%39.0 - 94.0%38/3997.4%86.5 - 99.9%
20/0N/AN/A30/30100%88.4 - 100%
42/450.0%6.8 - 93.2%67/67100%94.6 - 100%
80/10%N/A17/17100%80.5 - 100%
107/977.8%40.0 - 97.2%42/4397.7%87.7 - 99.9%
112/633.3%4.3 - 77.7%83/83100%95.7 - 100%
132/2100%15.8 - 100%47/47100%92.5 - 100%
140/0N/AN/A6/6100%54.1 - 100%
TOTAL21/3363.6%45.1 - 79.6%330/33299.4%97.8 - 99.9%

TRU EI II Influ B - Data stratified by site and sample ty

Table Continued – TRU FLU Influenza B - Data stratified by site and sample type

Site IDPositive SamplesNegative Samples
FreshSwabTRUB/Culture%Sensitivity95% CITRUB/Culture%Specificity95% CI
23/3100%29.2 - 100%78/7998.7%93.1 - 100%
32/450.0%6.8 - 93.2%3/3100%29.2 - 100%
47/1936.8%16.3 - 61.6%50/50100%92.9 - 100%
1023/3467.6%49.5 - 82.6%26/26100%86.8 - 100%
110/0N/AN/A2/2100%15.8 - 100%
123/560.0%14.7 - 94.7%92/92100%96.1 - 100%
130/0N/AN/A3/3100%29.2 - 100%
TOTAL38/6558.5%45.6 - 70.6%254/25599.6%97.8 - 100%
Site IDPositive SamplesNegative Samples
FrozenWash/AspirateTRUB/Culture%Sensitivity95% CITRUB/Culture%Specificity95% CI
31/250.0%1.3 - 98.7%61/61100.0%94.1 - 100%

{14}------------------------------------------------

Table 7 Data stratified by age of patient and sample type

Age GroupPositive SamplesNegative Samples
FreshTRU%95% CITRU%95% CI
Wash/AspirateA/CultureSensitivityA/CultureSpecificity
≤1 month4/4100.0%39.8 - 100%28/3384.8%68.1 - 94.9%
>1 month - 2yrs16/1984.2%60.4 - 96.6%146/16886.9%81.8 - 92.0%
3 - 12 years5/5100.0%47.8 - 100%62/6891.2%81.8 - 96.7%
13 - 21 years3/475.0%19.4 - 99.4%31/3296.9%83.8 - 99.9%
>21 years6/785.7%42.1 - 99.6%24/24100.0%85.8 - 100%
Not Determined0/0N/AN/A0/10.0%N/A
Age GroupPositive SamplesNegative Samples
Fresh SwabTRU%95% CITRU%95% CI
A/CultureSensitivityA/CultureSpecificity
≤1 month1/1100.0%N/A9/9100.0%66.4 - 100%
>1 month - 2yrs9/1181.8%48.2 - 97.7%36/4285.7%71.5 - 94.6%
3 - 12 years16/2080.0%56.3 - 94.3%52/5889.7%78.8 - 96.1%
13 - 21 years2/450.0%6.8 - 93.2%14/1877.8%52.4 - 93.6%
>21 years35/3794.6%81.8 - 99.3%113/11598.3%93.9 - 99.8%
Not Determined0/0N/AN/A5/5100.0%47.8 - 100%

TRU FLU Influenza A - Data stratified by age of patient and sample type

TRU FLU Influenza B - Data stratified by age of patient and sample type

Age GroupFreshWash/AspiratePositive SamplesNegative Samples
TRUB/Culture%Sensitivity95% CITRUB/Culture%Specificity95% CI
≤1 month0/0N/AN/A37/37100.0%90.5 - 100%
>1 month - 2yrs6/875.0%34.9 - 96.8%178/17999.4%96.9 - 100%
3 - 12 years8/1457.1%28.9 - 82.3%58/5998.3%90.9 - 100%
13 - 21 years6/875.0%34.9 - 96.8%28/28100.0%87.7 - 100%
>21 years1/333.3%0.8 - 90.6%28/28100.0%87.7 - 100%
Not Determined0/0N/AN/A1/1100.0%N/A
Age GroupFresh SwabPositive SamplesNegative Samples
TRUB/Culture%Sensitivity95% CITRUB/Culture%Specificity95% CI
≤1 month1/1100.0%N/A9/9100.0%66.4 - 100%
>1 month - 2yrs9/1181.8%48.2 - 97.7%42/42100.0%91.6 - 100%
3 - 12 years20/3951.3%34.8 - 67.6%39/39100.0%91.0 - 100%
13 - 21 years1/425.0%0.6 - 80.6%18/18100.0%81.5 - 100%
>21 years4/757.1%18.4 - 90.1%145/145100.0%97.5 - 100%

Prevalence of the infection

. .

:

· :::

..

The positivity rate for each laboratory will be dependent on several factors including the method of specimen collection, the handling and transportation of the time of year, and the prevalence of influenza A or B at the time of testing. US influenza A rates (as reported by the US Centers for Disease Control) during the 2006-7 period of TRU FLU clinical trials ranged from 4% in the months of October/November 2006 to a peak of 40% in February 2007. The influenza B positivity rate peaked at approximately 8% during March/April 2007. The influenza A positivity rates at the clinical trial sites (based on tissue culture) averaged 14% in the month of February and 10% in March. Influenza B rates varied from 11% to 18% during the same time. The prevalence of Influenza A and B antigens in individuals showing signs and symptoms, as determined by the TRU FLU Assay is summarized in Table 8.

{15}------------------------------------------------

TRU FLU Influenza A Data distributed by gender and age of patientTRU FLU Influenza B Data distributed by gender and age of patient
GendernTRU FLU A PositiveTRU FLU A NegativePrevalenceTRU FLU B PositiveTRU FLU B NegativePrevalence
Female3196925021.6%3328610.3%
Male3618128022.4%253366.9%
Unknown5050.0%4180.0%
Total68515053521.9%6262310.0%
Age (years)nTRU FLU A PositiveTRU FLU A NegativePrevalenceTRU FLU B PositiveTRU FLU B NegativePrevalence
<22876322422.0%172705.9%
3 - 212094316620.6%3617317.2%
> 211834314023.5%51782.7%
Unknown61516.7%4266.7%

Table 8: Prevalence of Influenza A and B antigens by patient age and sex as determined by TRU FLU

Reproducibility

Assay precision, intra-assay variability and inter-assay variability were assessed with a reference panel prepared from pools of negative samples spiked with specific virus. The reproducibility panel consisted of high positive (n = 2), low negative (n = 1), and low positive (n = 4) and high negative specimens (n =4). The latter were prepared near the assay limit of sensitivity. Each reference specimen was coded to prevent its identification during testing. Each was evaluated twice per day for three consecutive days by three different laboratories. The results of reproducibility evaluations are shown in Table 9 below. Calculations of inter and intra assay variability are given in Table 10.

High negative samples (viral load just below LOD) produced weakly positive results in 8 out of 72 high negative replicate tests performed with the samples prepared near the cutoff. It is expected that high negative samples tested at the cut-off will produce weakly positive results 50% of the time. (See EP12-A. User protocol for evaluation of qualitative performance; approved guideline; NCCLS/CLSI, Vol. 22, no.14, 2002.) Low positive samples (viral load just above LOD) produced 1 negative reaction out of 72 replicate results. The high positive and low negative samples produced the correct results 100% of the time.

{16}------------------------------------------------

:

CONCLUSIONS

:

. .

TRU FLU can be used reliably for the rapid detection of influenza A or B antigens in the sample types defined in product labeling

{17}------------------------------------------------

Image /page/17/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

NOV 1-5 2007

Susan Rolih Vice President, RA/QA Meridian Bioscience, Inc. 3471 River Hills Drive Cincinnati, OH 45244

K071657 Re: Trade/Device Name: TRU FLU Regulation Number: 21 CFR 866.3330 Regulation Name: Influenza Virus Serological Reagents Regulatory Class: Class I Product Code: GNX Dated: November 2, 2007 Received: November 2, 2007

Dear Ms. Rolih:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{18}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally attyms

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{19}------------------------------------------------

510(k) Premarket Notification TRU FLU

INDICATIONS FOR USE STATEMENT TRU FLU

510(K) Number: K071657

TRU FLU is a rapid, qualitative, lateral-flow immunochromatographic assay for detecting both influenza A and influenza B viral nucleoprotein antigens in human nasal wash, nasopharyngeal aspirate and nasal and nasopharyngeal swab samples in symptomatic patients. This test is not intended for the detection of influenza C viruses. A negative test is presumptive and it is recommended these results be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other clinical management decisions.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-OffSchel
---------------------------------------------------------------

Office of In Vitro Diagnostic Device

Evaluation and Safety

510(k) K071657

Section 4, Page 1 of 1

§ 866.3328 Influenza virus antigen detection test system.

(a)
Identification. An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.