K Number
K071576

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-11-02

(147 days)

Product Code
Regulation Number
870.2360
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Textronics™ ECG Electrode is intended for use in general electrocardiograph monitoring and recording procedures.

The Textronics™ ECG Electrode is indicated for use with most ECG instruments on the market. It is indicated for applications including resting ECGs, exercise ECGs, and/or ambulatory monitoring. The device is indicated for both prescription and over-the-counter use.

Device Description

The Textronics ™ ECG Electrode consists of electrodes, an exterior housing for the electrodes that makes contact with the patient, and a means of connecting the device to an ECG instrument. The Textronics™ ECG Electrode has textilebased electrodes made of electrically conductive yarns knitted together with electrically non-conductive yarns. The conductive yarns are silver coated polyamide (nylon). The non-conductive yarns can include cotton, spandex, polyester, and/or nylon. The Textronics™ ECG Electrode's textile-based electrodes have an exterior housing made of fabric, which can be made out of common textile materials, including polypropylene, nylon, Lycra® Spandex, polyester, and/or cotton. This fabric exterior housing can take the shape of a variety of garments, such as chest straps, wrist bands, shirts, bras, or vests. The Textronics™ ECG Electrode connects to ECG instruments via a conductive snap.

AI/ML Overview

Here's an analysis of the provided information regarding the Textronics™ ECG Electrode's acceptance criteria and substantiating study:

This 510(k) summary does not provide explicit acceptance criteria in the typical sense of quantitative thresholds (e.g., minimum sensitivity, PPV). Instead, the "acceptance criteria" are implied to be "functional as intended" and "clarity and resolution of the tracings were observed to be substantially equivalent to the commercially available electrodes."

The study is a comparative performance study against commercially available ECG electrodes.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
Functional as intended (for general ECG monitoring and recording)."In all instances, the Textronics™ ECG Electrode functioned as intended."
Clarity and resolution of tracings substantially equivalent to commercial electrodes."...the clarity and resolution of the tracings were observed to be substantially equivalent to the commercially available electrodes." This also implies compatibility with "most ECG instruments on the market."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document mentions "ECGs were recorded," implying multiple recordings, but no specific number of patients or recordings is provided.
  • Data Provenance: Not explicitly stated, but the study was conducted "in the lab." This suggests a controlled environment, likely in the US where Textronics, Inc. is located. It is a prospective study as the device was tested against existing commercial electrodes.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Number of Experts: Not stated.
  • Qualifications of Experts: Not stated.

The evaluation of "clarity and resolution" would typically involve expert interpretation, but the document does not specify who performed this assessment or their qualifications.


4. Adjudication Method for the Test Set

  • Adjudication Method: Not stated. Given the qualitative nature of the performance assessment ("observed to be substantially equivalent"), it's possible a formal adjudication process was not used, or at least not described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No. This type of study is not described. The assessment focuses on the device's technical performance in generating ECG tracings, not on human reader performance with or without AI assistance.

6. Standalone Performance Study (Algorithm Only)

  • Standalone Performance Study: Yes, in the sense that the performance data presented is for the device itself.
    • The Textronics™ ECG Electrode was "tested in the lab using ECG recording devices with the commercially available ECG electrodes and the Textronics ™ ECG Electrode used simultaneously."
    • "ECGs were recorded, and the tracings were compared."
    • The device's output (ECG tracings) was directly assessed for functionality, clarity, and resolution.

7. Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for this device's performance is effectively concurrent comparison to commercially available, predicate ECG electrodes. The "ground truth" for what constitutes a "functional" and "clear, resolute" ECG tracing is established by the accepted performance of these predicate devices. It is based on expert observation/interpretation of these comparative tracings, rather than pathology or long-term clinical outcomes.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. The Textronics™ ECG Electrode is a hardware device (electrode), not an AI algorithm that requires a training set.

9. How Ground Truth for the Training Set Was Established

  • How Ground Truth for Training Set Was Established: Not applicable, as there is no training set for this hardware device.

{0}------------------------------------------------

K071576

510(k) SUMMARY

Textronics, Inc.'s ECG Electrode

NOV - 2 2007

Submitter's Name, Address, Telephone Number, Contact Person, and Date Prepared

Textronics, Inc. 3825 Lancaster Pike Wilmington DE 19805

Phone: (302) 351-2162 Facsimile: (302) 998-7870

Contact Person: Michael McBrearty, Ph.D.

Date Prepared: May 31, 2007

Name of Device

Textronics ™ ECG Electrode

Common or Usual Name

ECG Electrode

Classification Name

Electrode, Electrocardiograph

Page 1 of 3

{1}------------------------------------------------

Predicate Devices

Unomedical 4560M Unilect Short Term ECG Electrodes (K944497) Medi-Trace model 530 ECG Conductive Adhesive Electrodes (K945479) Accuheart™ Electrode belt (K043361)

Intended Use / Indications for Use

The Textronics™ ECG Electrode is intended for use in general electrocardiograph monitoring and recording procedures.

The Textronics™ ECG Electrode is indicated for use with most ECG instruments on the market. It is indicated for applications including resting ECGs, exercise ECGs, and/or ambulatory monitoring. The device is indicated for both prescription and over-the-counter use.

Technological Characteristics

The Textronics ™ ECG Electrode consists of electrodes, an exterior housing for the electrodes that makes contact with the patient, and a means of connecting the device to an ECG instrument. The Textronics™ ECG Electrode has textilebased electrodes made of electrically conductive yarns knitted together with electrically non-conductive yarns. The conductive yarns are silver coated polyamide (nylon). The non-conductive yarns can include cotton, spandex, polyester, and/or nylon. The Textronics™ ECG Electrode's textile-based electrodes have an exterior housing made of fabric, which can be made out of common textile materials, including polypropylene, nylon, Lycra® Spandex, polyester, and/or cotton. This fabric exterior housing can take the shape of a variety of garments, such as chest straps, wrist bands, shirts, bras, or vests. The Textronics™ ECG Electrode connects to ECG instruments via a conductive snap.

Page 2 of 3

{2}------------------------------------------------

Performance Data

The Textronics™ ECG Electrode has been tested in the lab using ECG recording devices with the commercially available ECG electrodes and the Textronics ™ ECG Electrode used simultaneously. ECGs were recorded, and the tracings were compared. In all instances, the Textronics™ ECG Electrode functioned as intended and the clarity and resolution of the tracings were observed to be substantially equivalent to the commercially available electrodes.

Substantial Equivalence

The Textronics™ ECG Electrode is substantially equivalent to several predicates, including the Unomedical 4560M Unilect Electrodes, the Medi-Trace model 530 ECG Conductive Electrodes and the Accuheart™ Electrode belt. The Textronics™ ECG Electrode has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the Textronics™ ECG Electrode and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Textronics™ ECG Electrode performs very similarly to commercially available ECG electrodes. Thus, the Textronics™ ECG Electrode is substantially equivalent.

{3}------------------------------------------------

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 2 2007

Textronics Inc. c/o Ms. Janice Hogan, Esq. Hogan & Hartson, LLP 1835 Market Street, 28th Floor Philadelphia, PA 19103

Re: K071576

Textronics™ ECG Electrode Regulation Number: 21 CFR 870.2360 Regulation Name: Electrocardiograph Electrode Regulatory Class: Class II (two) Product Code: DRX Dated: October 242007 Received: October 24, 2007

Dear Ms. Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Janice Hogan, Esq.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

B/yummumana for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

K071576 510(k) Number (if known):

Device Name: Textronics™ ECG Electrode

Indications for Use:

The Textronics ™ ECG Electrode is intended for use in general electrocardiograph monitoring and recording procedures.

The Textronics™ ECG Electrode is indicated for use with most ECG instruments on the market. It is indicated applications including resting ECGs, exercise ECGs, and/or ambulatory monitoring.

Prescription Use X (Part 21 C.F.R. 801 Subpart D) AND/OR

Over-The-Counter X (21 C.F.R. 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bhimmer

) vis Sign-Off ivision of Cardiovase >10(k) Nu

of Page

§ 870.2360 Electrocardiograph electrode.

(a)
Identification. An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.