Intended Use

Rugalift is indicated for cosmetic use.

Device Description

Rugalift works by delivering electrical microcurrents to stimulate the skin of the face in a non invasive way. The body of the device includes the electronics for the control of the device, and the power switch for the activation of the device by the user and a detachable applicator with 5 microelectrodes. Rugalift is switched on when the microelectrodes are on the skin and pressing the finger on the power switch. The output of the device increases linearly over the first 10 seconds of stimulation, then it remains constant for 13 seconds and stops. The output cannot be changed by the user. During the treatment, an acoustic signal is activated. The frequency of the acoustic signal increases as the output current increases. When the user removes the device from the face, it automatically resets and turns itself off. Rugalift is powered by a 12V power adapter which is supplied with the device.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a device called "Rugalift". It details its intended use, technological characteristics, and performance data, alongside its substantial equivalence claim to a predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based only on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with corresponding device performance metrics. Instead, it asserts conformity with several international standards related to safety and electrical equipment. The "Performance Data" section lists these standards as the basis for compliance.

Acceptance Criteria (Implied)Reported Device Performance
Safety Standards & Electrical Compatibility:
Conformity to EN 60601-1-2 ED. 2 (2001)Stated that Rugalift is in conformity.
Conformity to IEC 60601-1-2 (2001-09)Stated that Rugalift is in conformity.
Conformity to EN 60601-2-10 (2001-11)Stated that Rugalift is in conformity.
Conformity to EN 60601-2-10 (1990-08) and its amendmentsStated that Rugalift is in conformity.
Conformity to CEI 62-39 (1992-10)Stated that Rugalift is in conformity.
Usability & Safety after Self-Training:
Sufficiency of documentation for self-selection and safety of the device"The study provided evidence that the documentation included is sufficient for self-selection and safety of the device."
Ease of use, reliability, and safety"Rugalift is a device that can be used easily, is reliable and safe."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The text mentions only "An usability study has been performed". It does not specify the sample size for this usability study. The provenance of the data (e.g., country of origin, retrospective or prospective nature) is also not specified in the provided text.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The document does not mention the use of experts to establish ground truth for any test set. The usability study appears to be focused on user experience and safety after self-training, rather than requiring expert adjudication of clinical outcomes in the traditional sense.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Given that no experts are mentioned for establishing ground truth, there is no adjudication method described for a test set. The usability study is described in terms of its ability to "provide evidence that the documentation included is sufficient for self-selection and safety of the device," suggesting a direct assessment of the documentation's clarity and completeness by potentially end-users.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document does not describe a multi-reader multi-case (MRMC) comparative effectiveness study. The device, a Transcutaneous Electrical Nerve Stimulator, is not an AI-assisted diagnostic tool, so such a study would not be relevant in this context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. The Rugalift is a physical medical device, not an algorithm. Its performance is inherent in its electrical stimulation capabilities and its interaction with a user.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the usability study, the "ground truth" implicitly relates to the sufficiency of documentation for self-selection and safety, and the device being easy to use, reliable, and safe. This would likely be established through direct observation, user feedback, and potentially expert review of the documentation itself and the device's operational specifications against safety standards, rather than clinical outcomes data or pathology reports.

8. The sample size for the training set

The document does not mention a training set in the context of device development or evaluation. The usability study is the only study described, and it is not identified as a "training set."

9. How the ground truth for the training set was established

As no training set is mentioned, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for "Fatrotek". The logo consists of the word "Fatrotek" in a bold, sans-serif font, with a small registered trademark symbol to the right of the "k". Above the word "Fatrotek" is a series of six black ovals arranged in a curved line, resembling a footprint or a path.

KC7/573

510(k) SUMMARY

DEC 0 9 2008

Rugalift

Submitter

Fatrotek S.r.I. Via Case Rosse, 22 84131 Salerno Italy

Telephone: +39 (089) 3055911 Facsimile: +39 (089) 3055955 Contact Person: Francesco Peluso, Managing Director

Date Prepared: 09/26/2008

Device Trade Name

Rugalift

Classification Name

Transcutaneous Electrical Nerve Stimulator (21 C.F.R. 882.5890)

Predicate Devices

Salton, Inc.'s Rejuvenique® Facial Toning System, Model RJV-10 (K011935)

Intended Use

Rugalift is indicated for cosmetic use.

Technological Characteristics

Rugalift works by delivering electrical microcurrents to stimulate the skin of the face in a non invasive way. The body of the device includes the electronics for the control of the device, and the power switch for the activation of the device by the user and a detachable applicator with 5 microelectrodes. Rugalift is switched on when the microelectrodes are on the skin and pressing the finger on the power switch.

The output of the device increases linearly over the first 10 seconds of stimulation, then it remains constant for 13 seconds and stops. The output cannot be changed by the user. During the treatment, an acoustic signal is activated. The frequency of the acoustic signal increases as the output current increases. When the user removes the device from the face, it automatically resets and turns itself off. Rugalift is powered by a 12V power adapter which is supplied with the device.

Performance Data

Rugalift is in conformity with the requirements as set forth in the following standards:

  • EN 60601-1-2 ED. 2 (2001): Medical electrical equipment. Part 1-2: General Requirements for . safety.
  • IEC 60601-1-2 (2001-09): Medical electrical equipment. Part 1-2: General Requirements for safety. . Collateral standard: EMC Requirements and tests.
  • EN 60601-2-10 (2001-11): Medical electrical equipment. Part 2-10: Particular requirements for the . safety of nerve and muscle stimulators

Fatrotek SRL unipersonale Via Case Rosse nº 22 84131 Salerno - ITALIA --

fatrotek@fatrotek.com

Image /page/0/Picture/27 description: The image shows the text "Società Certificata ISO 9001:2000". Below the text are two logos. The text indicates that the company is certified under ISO 9001:2000 standards. The logos likely represent the certification bodies or quality assurance organizations.

P.IVA: IT 02984320651 CCIAA: SA 253149/94 Trib. n°27375/94

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "Fatrotek" in a bold, sans-serif font. Above the word are a series of black dots that appear to be arranged in a curved line. To the right of the word "Fatrotek" is the registered trademark symbol.

www.fatrotek.com

  • EN 60601-2-10 (1990-08): Medical electrical equipment. Part 1: General Requirements for safety ● and its amendments: EN60601-1/Ec (1994-07) + A1 (1993) + A1/Ec (1994-07) + A2 (1995-06) + A13 (1996-01).
  • CEI 62-39 (1992-10): Electrical equipment for aesthetic use. .

Rugalift operation is based on the principle of stimulating the skin with microcurrents delivered through microelectrodes.

The microelectrodes are positioned on the face skin, after the skin has been cleansed and dried. The effects on the stimulated region of the face can vary based on individual sensitivity and the duration of treatment. Many people notice a softening effect which can enhance the results of make-up and works against the effects of skin ageing.

The results are not permanent but can last several months following the appropriate maintenance program.

An usability study has been performed to evaluate Rugality and safety after self-training. The study provided evidence that the documentation included is sufficient for self-selection and safety of the device.

Rugalift is a device that can be used easily, is reliable and safe.

Substantial Equivalence

Rugalift is substantially equivalent to other legally marketed transcutaneous electrical nerve stimulator devices for cosmetic use. Specifically, Rugalift is substantially equivalent to Salton, Inc.'s Rejuvenique® Facial Toning System.

The Rejuvenique system delivers electric stimulation through 26 fixed position electrodes which are contained in a face mask shaped to fit over the user's face, and held into position by an adjustable headband. The electrodes are activated in sequence: each pair of electrodes delivers the stimulus for 20 seconds (similar to the Rugalift recommended stimulation time), then a microprocessor located in the device control unit automatically switches to the next pair of electrodes. A full cycle requires approximately 4 minutes. Unless stopped by the user, the product will repeat the cycle four times (approximately 20 minutes) and then automatically shuts off.

Although the mechanism of action of the two devices is the same (i.e. stimulation of the face using microcurrents), there are some differences, which are related to the specific principle of operation of the two devices.

  • The Rugalift is a handheld device which allows delivery of the stimulation precisely on the selected t area. This is not possible with the Reiuvenique device because the mask shape and electrodes position are fixed. whereas face anatomy is variable form subject. Therefore, in order to obtain the same effect, the current density delivered by Rejuvenique and/or stimulation time may need to be greater than that of the Rugalift.
  • With Rejuvenique the user may be forced to interrupt the treatment cycle if stimulation of a certain area . is not tolerated, and this may prevent completion of the treatment. With Rugalift, selective application on each face area and the possibility to interrupt and resument easily may allow completion of stimulation even in the most sensitive areas. The ability to interrupt the stimulation easily by lifting the device from the skin is an important advantage in terms of patient safety.
  • . Rugalitt is a simpler device as it does not require a control unit and does not have a display; therefore, it is more user-friendly.

The above mentioned differences in the principle of operation result in some differences in the technological characteristics of the two devices, but do not cause substantial differences in terms of performance and safety features. Rugalift has the same intended use as the Rejuvenique®. The device also has similar technological characteristics as its predicate devices provide local stimulation of the face using low intensity currents.

Minor differences in the technological characteristics of Rugalift and the Rejuvenique® do not raise any new issues of safety or effectiveness. Thus, Rugalift is substantially equivalent.

Fatrotek SRL unipersonale Via Case Rosse nº 22 84131 Salerno - ITALIA --

Tel. +39 089 3055.911

Fax. +39 089 3055.955

fatrotek @ fatrotek.com

P.IVA: IT 02984320651 CCIAA: SA 253149/94 Trib. nº27375/94

Società Certificata ISO 9001:2000

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA".

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Fatrotek S.r.1. % Dr. Francesco Peluso Managing Director Via Case Rosse nº 22 84131 Salerno Italy

DEC 0 9 2008

Re: K071573

Trade Name: RUGALIFT version 2.1 Regulation Number: 21 CFR 882.5890 Regulation Names: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: NFO Dated: September 5, 2008 Received: September 10, 2008

Dear Dr. Peluso:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Dr. Francesco Peluso

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indication for Use

510(k) Number (if known): K071573

Device Name: RUGALIFT version 2.1

Indication For Use: The RUGALIFT is indicated for cosmetic use.

Prescription Use (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use _X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nil RP Johnson

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

6071573 510(k) Number

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).