GENTIAN CYSTATIN C IMMUNOASSAY

K071388 · Gentian AS · NDY · Nov 6, 2007 · Clinical Chemistry

Device Facts

Record IDK071388
Device NameGENTIAN CYSTATIN C IMMUNOASSAY
ApplicantGentian AS
Product CodeNDY · Clinical Chemistry
Decision DateNov 6, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1225
Device ClassClass 2

Intended Use

Gentian Cystatin C Immunoassay is an in-vitro diagnostic test for quantitative determination of cystatin c in human serum and plasma. The measurement of cystatin c is used in the diagnosis and treatment of renal diseases.

Device Story

The Gentian Cystatin C immunoassay is a particle-enhanced turbidimetric immunoassay (PETIA) used for the quantitative determination of cystatin C in human serum and plasma. The device utilizes activated polystyrene microspheres covalently attached to avian anti-human cystatin C antibodies. When mixed with patient samples, the immunoparticles and cystatin C form aggregations, causing a change in the absorbance signal proportional to the cystatin C concentration. The assay is designed for use on commercially available automated clinical chemistry analyzers. Healthcare providers use the resulting cystatin C measurements to aid in the diagnosis and treatment of renal diseases. The use of avian antibodies prevents interference from rheumatoid factor, a common issue in other assays. The device provides a rapid, automated method for assessing kidney function, potentially improving diagnostic accuracy for patients with renal impairment.

Clinical Evidence

Bench testing only. Performance evaluated across 4 study sites comparing the Gentian assay to the predicate (K003503) on automated analyzers (Modular P, Architect ci8200). Regression analysis showed high correlation (R=0.986-1.00). Total imprecision CV was 4.2% over 20 days. Interference studies confirmed no significant impact from common substances (hemoglobin, intralipid, triglycerides, bilirubin) or rheumatoid factor. Recovery ranged from 99-110%. Linearity demonstrated across 0.3-8.0 mg/L range.

Technological Characteristics

Particle-enhanced turbidimetric immunoassay (PETIA). Reagents consist of activated polystyrene microspheres conjugated with avian anti-human cystatin C antibodies. Operates on automated clinical chemistry analyzers using light absorption detection. Measuring range 0.3-8.0 mg/L. Total analysis time 10 minutes. Reagent stability 18 months at 2-8°C.

Indications for Use

Indicated for the quantitative determination of cystatin C in human serum and plasma for the diagnosis and treatment of renal diseases in patients requiring clinical assessment of kidney function.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ NOV -- 6 2007 K071388 #### 510(k) Summary 5. # Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination #### Device description The Gentian cystatin C immunoassay is a particle enhanced turbidimetric immunoassay (PETIA). The immunoparticles are made from activated polystyrene microspheres to which avian anti-human cystatin C antibodies are covalently attached. The immunoparticles and cystatin C form aggregations that change the absorbance signal, depending on the amount of cystatin C present. Measurements obtained by this device are used for the determination of Cystatin C in human serum and plasma. The Gentian C assay is calibrated with human Cystatin C calibrators. Cystatin C controls are assayed for the verification of the accuracy and precision of the Gentian cystatin C immunoassay. #### Substantial Eguivalence The Gentian Cystatin C immunoassay is substantially equivalent to the Dade Behring, Inc., N Latex Cystatin C Test Kit (K003503) with respect to indications for use, device design and material. The basic difference between the new device and the Dade Behring predicate device is in assay technology and the instruments used for testing. The Gentian assay is a particle enhanced turbidimetric immunoassay (PETIA), while the Dade Behring assay is a particle enhanced immunonephlometric assay (PENIA). The Gentian device can be used on all commercially available automated clinical chemistry analyzers using a light absorption detection system, while the Dade Behring test is applicable only on the Dade Behring, Inc. Nephlometer Systems. In the Gentian device avian antibodies are used and it is known by one skilled in the art that there is no interaction between Rheumatoid Factor (RF) and avian antibodies. The Dade Behring predicate device use antibodies with the inherent possibility for a false reactions with Rheumatoid Factor (RF). #### Comparison to predicate device The substantial equivalence, safety and efficacy of the Gentian cystatin C immunoassay to Dade Behring N Latex Cystatin C assay (K003503) was evaluated in two studies and on two different automated clinical chemistry analyzers. In total 4 study sites were involved. The issues addressed in these studies were the comparison between Gentian cystatin C immunoassay and the predicate device Dade Behring N Latex Cystatin C assay (K003503) (Table1). In addition external validation performance of the new device was evaluated at two study sites (Table 2). | | Instrument<br>application | Slope<br>(95% CI) | Intercept<br>(95% CI) | R | Cystatin C<br>range (mg/L) | N | |---------|-----------------------------|-------------------|-----------------------|-------|----------------------------|-----| | Study 1 | Modular P vs.<br>BNII | 1.075 | 0.009 | 0.986 | 0.53 - 9.47 | 172 | | | Modular P vs. BN<br>ProSpec | 1.048 | 0.059 | 0.992 | 0.53 - 9.47 | 174 | | Study 2 | Modular P vs. BN<br>ProSpec | 1.021 | -0.025 | 1.00 | 0.61 - 6.44 | 76 | | | Architect vs. BN<br>ProSpec | 0.975 | -0.059 | 0.989 | 0.51 - 7.95 | 87 | #### Table 1. Summary of method comparison regression analysis #### Table 2. External validation performance data | Sample | Mean cystatin C (mg/L) | Within run CV (%) | Between day CV (%) | Between site CV (%) | Between lot CV (%) | Total precision CV (%) | Recovery (%) | |--------|------------------------|-------------------|--------------------|---------------------|--------------------|------------------------|--------------| |--------|------------------------|-------------------|--------------------|---------------------|--------------------|------------------------|--------------| {1}------------------------------------------------ | | 0.82 | - A B ------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>t T | AL BL COMMEN<br>3.54 | 0.86 | J.55 | 4.4<br>and and and the commend of the comments of the commend of the commend of the may of the may of the may be any and | 103.8 | |---|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--------------------------|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | V | of the contraction for frequent<br>1 | 49 | 2.36 | 0.84 | 0.54 | 2.78<br>( | 107.3<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | | 3.94 | a from the collection of the property to the contributed to the compare<br>.38 | of the property of the larges and<br>t<br>. | 0.0<br>LIBLE IL BOOK AND | and announce at The 's cloud of the comments of the<br>0.64 | ar<br>- 1 -<br>-<br>www.bangle.com/sport/supproam-shows/diving/div/a | 107.7 | #### Performance characteristics The measuring range is 0.3-8.0 mg/L. The reference range is 0.52-0.98 mg/L. Total analysis time is 10 minutes. Linearity is demonstrated over the whole assay range, the Gentian Cystatin C immunoassay is linear in the range 0.34 - 8.4 mg/L on Modular P and in the range 0.27 - 8.8 mg/L on Architect. Total imprecision CV, measured over 20 days with two lots, is 4,2%. Interference studies with drugs and anticoagulants show no significant interference, also there is no significant interference from hemoglobin (8 g/L), intralipid (11 q/L), triclycerides (14 g/L) and bilirubin (420 mg/L). Due to the use of avian antibodies, no interference with rheumatoid factor is detected. No carry-over is detected. The limit of detection (LoD) and limit of quantification (LoQ) (within CV 6%) are both within given acceptance criteria and below the lowest calibrator concentration. Sample stability is up to one month at 2-8°C. Stability of the reagents at 2-8°C is calculated to be at least 18. months. Stability of the reagents in use is minimum 4 weeks. Recovery is 99-110 %. Antigen hook effect was observed in samples with spiked cystatin C concentrations above 16 mg/L, this will have no significant impact on patient serum and plasma samples, since sample concentrations above 9 mq/l never have been reported. Serum and plasma evaluations give identical cystatin C results. Between instrument comparison regression analysis shows excellent agreement between Gentian Cystatin C when measured on Architect ci8200 and Modular P instruments. #### Conclusion When considering the comparison studies between Gentian C immunoassay and the Dade Behring, Inc., N Latex Cystatin C Test Kit (K003503) and the additional documentation supporting the Gentian cystatin C immunoassay, it can be concluded that the Gentian cystatin C immunoassay when measured on Architect ci8200 and Modular P analyzers is as safe and effective as, and substantially equivalent to the Dade Behring, Inc., N Latex Cystatin C assay. Submitted by. Emile Hammondi Oct. 5, 2007 Ronald G. Leonardi, Ph.D. President R & R REGISTRATIONS P.O. Box 262069, San Diego, CA 92196 Date {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three overlapping wing shapes, symbolizing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Gentian AS c/o Dr. Ronald Leonardi President of R & R Registrations 9915 Cam. Chirimolla San Diego, CA 92131 NOV - 6 2007 Re: k071388 Trade/Device Name: Gentian Cystatin C Immunoassay Regulation Number: 21 CFR 8862.1225 Regulation Name: Creatinine test system Regulatory Class: Class II Product Code: NDY Dated: October 05, 2007 Received: October 09, 2007 Dear Dr. Leonardi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean M. Cooper, M.S., D.V.M. Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indication for Use 510(k) Number (if known): K071388 Device Name: Gentian Cystatin C immunoassay Indication For Use: Gentian Cystatin C Immunoassay is an in-vitro diagnostic test for quantitative determination of cystatin c in human serum and plasma. The measurement of cystatin c is used in the diagnosis and treatment of renal diseases. Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or 1 Over the Counter Use (21 CFR Part 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Carol C. Benson Division Sign Off Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%