K Number
K071212

Validate with FDA (Live)

Date Cleared
2007-07-30

(90 days)

Product Code
Regulation Number
868.5895
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ESPRIT ventilatory is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult, pediatric and neonatal patients as prescribed by a physician. The ESPRIT Ventilator is intended for use in either invasive or noninvasive applications. The Speaking Mode software option allows tracheostomized adult and pediatric patients who meet the assessment criteria to vocalize without the need of a speaking valve.

Device Description

The ESPRIT ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult and pediatric patients as prescribed by a physician. The ESPRIT Ventilator is intended for use in either invasive or non-invasive applications. The Speaking Mode software option allows tracheostomized adult and pediatric patients who meet the assessment criteria to vocalize without the need of a speaking valve.

AI/ML Overview

The provided document, K071212, describes the Esprit Ventilator Speaking Mode Option. It outlines the device's indications for use and demonstrates substantial equivalence to a predicate device through performance testing.

Here's the requested information structured around acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The document provides a table titled "Substantial equivalence was established by performance testing. The Table describing the performance testing follows:". This table outlines specific parameters and their purpose, with all outcomes reported as "Pass." This indicates that the device met the pre-defined acceptance criteria for each tested parameter.

ParameterPurposeReported Device Performance (Pass/Fail)
Speaking Mode (SM) operabilitySM operates with VCV with A/C, SIMV, and CPAP; is not functional with NPPV; operates with Flow-Trak; operates with Pressure Support; operates on both adult and pediatric patient typesPass
Speaking Mode operabilitySM is not available for Neonatal patients; is not active after power cycling; is not available when specific alarms are activePass
Speaking mode and I-triggerI-trigger will automatically adjust to pressure triggering if flow triggering is the current trigger type; that only pressure triggering is allowed while in SM; test the accuracy of pressure triggeringPass
Speaking Mode graphical user interfaceTest the ventilator behavior when SM is not installed; confirm the SM related dialogs; test the ventilator behavior when SM is deactivated; confirm that PEEP is always set to zero when SM is activePass
Speaking Mode alarms and settingsConfirm alarm performance, including patient disconnectPass
Speaking Mode alarms and settingsConfirm alarm performance, including occlusion alarmPass
Apnea alarmConfirm that if apnea alarm occurs while in SM, exhalation valve opens during exhalationPass
Maneuvers not availableConfirm that Respiratory Mechanics and expiratory Hold Maneuvers are not available when SM is activePass
Displayed patient dataConfirm that data displays are appropriatePass
TrendingConfirm that trending values are trended/not trended when SM is activePass
Breath deliveryConfirm that exhalation valve is closed during inhalation and exhalation while in SMPass
CommunicationsConfirm that RS 232 and Vuelink communications are operablePass

2. Sample Size Used for the Test Set and the Data Provenance

The document does not explicitly state the sample size used for the performance testing. The provenance of the data is not specified (e.g., country of origin, retrospective or prospective). The testing appears to be a series of functional and alarm verification tests performed on the device itself, rather than a clinical study with patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not mention the use of experts to establish a "ground truth" for the test set, as the testing described focuses on functional aspects and alarm performance of the device's software option rather than subjective interpretation of medical data. The "Pass" results likely reflect objective verification against a documented set of design specifications or engineering requirements.

4. Adjudication Method for the Test Set

No adjudication method is described. Given the nature of the tests (functional verification of software and hardware interaction), results would likely be directly observable and measurable against specifications, without the need for expert adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not conducted or described in this document. The focus is on the substantial equivalence of a new software feature for an existing ventilator, not on the comparative effectiveness of human readers with or without AI assistance. Therefore, no effect size for human reader improvement with AI is applicable or stated.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Yes, the performance testing described appears to be a standalone assessment of the Esprit Ventilator's Speaking Mode Option. The tests verify the algorithm's functional behavior, alarm performance, and interaction with the ventilator's existing systems, without explicit human-in-the-loop performance evaluation beyond potentially operating the device during testing.

7. The Type of Ground Truth Used

The "ground truth" for the described performance testing appears to be the device's functional specifications, design requirements, and intended operational behavior. The tests confirm that the ventilator's Speaking Mode Option operates as designed and meets the established safety and performance parameters. This is not
expert consensus, pathology, or outcomes data.

8. The Sample Size for the Training Set

This document does not describe a training set. The "Speaking Mode Option" is a software feature for an existing ventilator, and the testing outlined is a verification of its functionality and integration, not an algorithm developed using machine learning or data-driven training.

9. How the Ground Truth for the Training Set Was Established

As no training set is described or implied by the document, the establishment of ground truth for a training set is not applicable.

{0}------------------------------------------------

Ko71212

Image /page/0/Picture/1 description: The image shows the word "RESPIRONICS" in all caps. Above the word, there are two curved lines. To the right of the word, there is a small circle with an R inside, indicating a registered trademark.

JUL 3 0 2007

510(k) Summary

Submitter: Respironics Inc. Address: 2271 Cosmos Court Carlsbad, CA 92011 Contact: Mara Caler Telephone: 760 918 7458 Fax: 760 918 0169 Email: mara.caler(@Respironics.com Date: 30 April 2007

Trade name: Esprit Ventilator Speaking mode Option Common Name (Device Type): Continuous Ventilator Classification Regulation (CFR): 21 CR 868.5895 ( Class: 2 Product Code: 73 - CBK Panel: Division of Anesthesiology, General Hospital, Infection Control and Dental Devices (73)

Indications for Use:

The ESPRIT ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult and pediatric patients as prescribed by a physician. The ESPRIT Ventilator is intended for use in either invasive or non-invasive applications. The Speaking Mode software option allows tracheostomized adult and pediatric patients who meet the assessment criteria to vocalize without the need of a speaking valve.

Substantial Equivalence to Predicate Devices:

Predicate device: Esprit Ventilator, K981072 The proposed Esprit Ventilator is identical to the existing Esprit Ventilator with the exception of the addition of the Speaking Mode option.

List of Similarities:

  • Similar intended use the intended use is unchanged .
    • The Esprit Ventilator is intended for use by qualified medical personnel in o providing continuous or intermittent ventilatory support for adult and pediatric patients as prescribed by a physician. The ESPRIT Ventilator is intended for use in either invasive or non-invasive applications.

{1}------------------------------------------------

  • Similar patient circuit the patient circuit is unchanged .
    • o The patient circuit is unchanged.
  • Same operating principle the operating principle is unchanged .
    • The ESPRIT ventilatory is a microprocessor controlled, electrically o powered, mechanical ventilator. There have been no changes to the operating principle of the equipment.
      • The breathing system is under microprocessor control 트
      • The user interface is under microprocessor control, featuring a touch screen and graphical user interface technology.
      • The integral air source is built into the ventilator, eliminating the 피 requirement for a central compressor and piped, medical grade wall air and/or an individual stand alone compressor for each ventilator.
      • The ability to provide variable oxygen concentrations (21% to 해 100% O2) from a 35 to 80 PSIG medical oxygen gas source, including medical grad gas cylinders with suitable regulators
      • The ability to operate on a re-chargeable primary battery for 30 트 minutes (nominal)
      • The ability to be powered by a secondary DC power source (24 L VDC) for up to 3 hours
  • Same technology the technology is unchanged ●
    • The ESPRIT ventilator is a microprocessor controlled, electrically o powered, mechanical ventilator. There have been no changes to the technology or equipment hardware.
  • Same manufacturing process the manufacturing process is unchanged .
    • There have been no changes to the manufacturing processes for the o equipment.

List of Differences:

  • Indications for Use ●
    • The Speaking mode software option allows tracheostomized adult and o pediatric patients who meet the assessment criteria to vocalize without the need of a speaking valve (e.g., the Passy-Muir valve).
  • Breathing modes: .
    • The breathing modes available during active Speaking Mode include ০ pressure control and volume control in A/C, SIMV and CPAP, Pressure Support and Flow Trak. Speaking Mode option is not available while in the NPPV mode. Neonatal option are not available when in Speaking Mode.

The currently marketed Esprit ventilator and the Esprit ventilator with Speaking Mode option both analyze pressure to trigger and cycle the ventilator. In currently marketed ventilators, an external (closed) Speaking Valve is installed to allow the exhaled gas to flow past the vocal cords. In the Speaking Mode option, the Esprit exhalation valve is closed, so that the exhaled gas is allowed to flow past the vocal cords. The exhaled gas pressure is measured to evaluate if (a) the tracheostomy tube has become disconnected or (b) if the tracheostomy tube has become occluded.

{2}------------------------------------------------

Substantial equivalence was established by performance testing. The Table describing the performance testing follows:

ParameterPurposePass/Fail
Speaking Mode (SM)operabilitySM operates with VCV with A/C,SIMV, and CPAP; is not functionalwith NPPV; operates with Flow-Trak;operates with Pressure Support;operates on both adult and pediatricpatient typesPass
Speaking ModeoperabilitySM is not available for Neonatalpatients; is not active after powercycling; is not available when specificalarms are activePass
Speaking mode and I-triggerI-trigger will automatically adjust topressure triggering if flow triggering isthe current trigger type; that onlypressure triggering is allowed while inSM; test the accuracy of pressuretriggeringPass
Speaking Modegraphical user interfaceTest the ventilator behavior when SMis not installed; confirm the SMrelated dialogs; test the ventilatorbehavior when SM is deactivated;confirm that PEEP is always set tozero when SM is activePass
Speaking Mode alarmsand settingsConfirm alarm performance, includingpatient disconnectPass
Speaking Mode alarmsand settingsConfirm alarm performance, includingocclusion alarmPass
Apnea alarmConfirm that if apnea alarm occurswhile in SM, exhalation valve opensduring exhalationPass
Maneuvers not availableConfirm that Respiratory Mechanicsand expiratory Hold Maneuvers arenot available when SM is activePass
ParameterPurposePass/Fail
Displayed patient dataConfirm that data displays areappropriatePass
TrendingConfirm that trending values aretrended/not trended when SM is activePass
Breath deliveryConfirm that exhalation valve isclosed during inhalation andexhalation while in SMPass
CommunicationsConfirm that RS 232 and Vuelinkcommunications are operablePass

Summary of Validation Testing

{3}------------------------------------------------

Conclusion:

:

This testing demonstrates that the device is as safe, as effective and performs as well as or better than the predicate device.

:

:

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Respironics California, Incorporated Ms. Mara Caler Regulatory Affairs Critical Care Division 2271 Cosmos Court Carlsbad, California 92011

JUL 3 0 2007

Re: K071212

Trade/Device Name: Esprit Ventilator Speaking Mode Option Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: July 12, 2007 Received: July 16, 2007

Dear Ms. Caler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Caler

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Suttie G. Michael DMD.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for use

510(k) Number (if known): K071212

Device Name: Esprit Ventilator (Speaking Mode Option)

Indications for Use:

The ESPRIT ventilatory is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult, pediatric and neonatal patients as prescribed by a physician. The ESPRIT Ventilator is intended for use in either invasive or noninvasive applications. The Speaking Mode software option allows tracheostomized adult and pediatric patients who meet the assessment criteria to vocalize without the need of a speaking valve.

X Prescription Use (Part 21 CFR 801, Subpart D) Over-the-Counter Use (Part 21 CFR 801, Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

ch (acting)
(Division Sign-Off)

Page 1 of 1

vision of Anesthesiology, General Hospital. tion Control. Dental Devices

510(k) Number: K071212

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).