K Number
K071013

Validate with FDA (Live)

Device Name
REDSENSE
Date Cleared
2007-10-18

(191 days)

Product Code
Regulation Number
876.5820
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Redsense device is intended to monitor for potential blood loss from the hemodialysis access site in hemodialysis patients undergoing continuous hemodialysis treatment up to 5 hours in the clinical setting. The device includes a blood sensor incorporated into an adhesive sensor patch. The sensor monitors potential blood leakage from the venous needle puncture site via an infrared light and will alarm if blood leakage is detected via absorption onto the device's sensor patch.

Device Description

Redsense is a system for monitoring the vein or arterial needle during hemodialysis. Redsense consists of an alarm unit and infrared sensor incorporated into an adhesive patch. The patch with the sensor is placed around the vein or arterial needle and detects any blood that drips onto the patch if the needle has been accidentally pulled out or if there is leakage during dialysis. If blood loss is detected, the device will alarm.

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the Redsense device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state quantitative acceptance criteria for the Redsense device's performance. Instead, it indicates that "Verification testing has been performed to verify that the Redsense device fulfills the Requirement Specifications" and "Results showed the device functioned successfully in the clinical environment."

Based on the device's functional description, the primary performance metric would be its ability to detect blood leakage and alarm. The closest stated performance characteristic is:

ParameterAcceptance Criteria (Implied)Reported Device Performance
Blood amount for detectionNot explicitly stated as acceptance criteria, but mentioned as device capabilityApproximately 2 ml
Functionality in clinical environmentDevice functions successfully in the clinical environment.Device functioned successfully in the clinical environment.

2. Sample Size for the Test Set and Data Provenance

  • Sample Size for the Test Set: Not explicitly stated. The document mentions "clinical testing was conducted" but does not specify the number of patients or cases.
  • Data Provenance: Prospective (clinical testing).
  • Country of Origin: Sweden (Departments of Nephrology/Medicine at Eksjö, Halmstad, Hässleholm, Skövde, and Varberg Hospitals).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. The study involved clinical testing in hospitals, implying medical professionals were involved in observing and confirming events, but the specific number or qualifications of experts establishing ground truth are not detailed.

4. Adjudication Method for the Test Set

This information is not provided in the document.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No. The document describes a study to verify the Redsense device's functionality, not a comparative effectiveness study involving human readers with and without AI assistance.
  • Effect Size of human readers improvement with AI vs without AI assistance: Not applicable, as no MRMC comparative effectiveness study was performed.

6. Standalone (Algorithm Only) Performance Study

Yes, a standalone study (clinical testing) was performed. The provided text describes the Redsense device, which includes a sensor and an alarm unit, as a standalone system designed to detect blood leakage. The clinical testing was conducted to verify that "the Redsense device fulfills the Requirement Specifications" and "functioned successfully in the clinical environment." This implies the device's performance was evaluated independently.

7. Type of Ground Truth Used

The ground truth for the clinical testing would have been clinical observation of actual blood leakage and confirmation by medical staff during hemodialysis treatments. This is inferred from the description of the device's purpose and the clinical study setting.

8. Sample Size for the Training Set

This information is not provided in the document. The Redsense device is described as a sensor-based system, and there is no mention of an "AI" or "algorithm" that would typically require a training set in the context of machine learning. The "Verification testing" and "Clinical testing" refer to the testing of the final, built device.

9. How the Ground Truth for the Training Set Was Established

This information is not provided and is likely not applicable, as the device is not described as utilizing a machine learning algorithm that requires a training set with established ground truth. It appears to be a direct physical sensor detecting infrared light absorption by blood.

{0}------------------------------------------------

PAGE 1 OF 3

Redsense Medical K071013

2. 510(k) Summary

Submitter: Redsense Medical AB Rörkullsvägen 4 Box 287 301 07 HALMSTAD SWEDEN

0CT 1 8 2007

Contact Information: Constance G. Bundy 6470 Riverview Terrace Fridley, MN 55432 Submission Date:

Device Name and Classification: Redsense, Class II, 876.5540, product code not known

Equivalent Device Identification: Neotrend (K972314) and Fresenius 2008 Touch panel control Dialysis system (K890824)

Device Description: Redsense is a system for monitoring the vein or arterial needle during hemodialysis. Redsense consists of an alarm unit and infrared sensor incorporated into an adhesive patch. The patch with the sensor is placed around the vein or arterial needle and detects any blood that drips onto the patch if the needle has been accidentally pulled out or if there is leakage during dialysis. If blood loss is detected, the device will alarm.

Intended Use: The Redsense device is intended to monitor for potential blood loss from the hemodialysis access site in hemodialysis patients undergoing continuous hemodialysis treatment up to 5 hours in the clinical setting. The device includes a blood sensor incorporated into an adhesive sensor patch. The sensor monitors potential blood leakage from the venous needle puncture site via an infrared light and will alarm if blood leakage is detected via absorption onto the device's sensor patch.

{1}------------------------------------------------

Comparison Table

ParameterProposed DevicePredicate device 1Predicate Device 2
DeviceRedsenseNeotrend MultiparameterSensorVein pressure supervision,integrated in Fresenius 2008Dialysis system
Intended UseMonitor bloodleakage fromvenous needlepuncture siteContinuous blood gasmonitoring system for pO2,temperature, pCO2 and pHBlood line separation (can becaused by vein needledislodgement)
ManufacturerRedsense MedicalDiametrics Medical ItdFresenius
TechnologyBlood detectionthrough IR lightabsorption in fiberoptics/ adhesiveabsorption padBlood gas sensor thoughtoptical fiber/ fluorescencequenching technology (fiberoptic photometricabsorption)Pressure detection
Blood amountfor detectionapproximately 2 mlNot specifiedNot specified, alarm pressure levelcan be adjusted
Area ofdetectionNA "Blood or noblood"pO2 20-500mmHg,temperature 10-42 °C, pCO210-160 mmHg and pH 6.80 -7.80-60 to 520 mmHG
ToleranceNApO2 +/- 3mmHg,temperature +/- 0,3 °C,pCO2 +/- 3 mmHg and pH+/- 0.03+/- 10 mmHg
Size of sensorD= 0.25 mmD= 0.5 mmNA
Optical FibertypePlasticPlasticNA
Light typeInfraredGreen and Red visibleNA
ComponentspartsFibre Optical blooddetector, adhesive,absorbentFiber Optical blood sensorwith chemical dyePressure sensor
Single useYesYesNo
SterileNoYesNA
Usage time5h72 h5h
PatientcontactSkinIntravascular accessNo

Summary of Testing:

Verification testing has been performed to verify that the Redsense device fulfills the Requirement Specifications.

Clincal testing was conducted at the Departments of Nephrology/Medicine at Eksjö, Halmstad, Hässleholm, Skövde, and Varberg Hospitals in Sweden. Results showed the device functioned successfully in the clinical environment.

{2}------------------------------------------------

Ko71013 PAGE 3 OF 3

Redsense Medical K071013

Conclusion: Redsense is similar in function and intended use to the vein pressure component of the Fresenius device and has sensing technology that is similar to the Neotrend device. Verification and clinical testing show the Redsense system to be safe and effective for intended use.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three curved lines representing its wings or feathers.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 1 8 2007

Redsense Medical AB c/o Ms. Constance G. Bundy Regulatory Consultant C.G. Bundy Associates, Inc. 6470 Riverview Terrace FRIDLEY MN 55432

Re: K071013

Trade/Device Name: Redsense Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: ODX Dated: October 16, 2007 Received: October 17, 2007

Dear Ms. Bundy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.

Sincerely vours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K071013

Device Name: Redsense

Indications For Use:

The Redsense device is intended to monitor for potential blood loss from the hemodialysis access site in hemodialysis patients undergoing continuous hemodialysis treatment up to 5 hours in the clinical setting. The device includes a blood sensor incorporated into an adhesive sensor patch. The sensor monitors potential blood leakage from the venous needle puncture site via an infrared light and will alarm if blood leakage is detected via absorption onto the device's sensor patch.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence, of CDRH, Office of Device Evaluation (ODE)

Hebert Werner

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 876.5820 Hemodialysis system and accessories.

(a)
Identification. A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient.
(2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860).
(3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860).
(4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
(b)
Classification. (1) Class II (performance standards) for hemodialysis systems and all accessories directly associated with the extracorporeal blood system and the dialysate delivery system.(2) Class I for other accessories of the hemodialysis system remote from the extracorporeal blood system and the dialysate delivery system, such as the unpowered dialysis chair, hemodialysis start/stop tray, dialyzer holder set, and dialysis tie gun and ties. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.