THE PRECISION / OPTIUM POINT OF CARE BLOOD GLUCOSE TEST STRIPS
K070984 · Abbott Diabetes Care, Inc. · LFR · Nov 1, 2007 · Clinical Chemistry
Device Facts
| Record ID | K070984 |
| Device Name | THE PRECISION / OPTIUM POINT OF CARE BLOOD GLUCOSE TEST STRIPS |
| Applicant | Abbott Diabetes Care, Inc. |
| Product Code | LFR · Clinical Chemistry |
| Decision Date | Nov 1, 2007 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The Precision® and Optium® Point of Care Blood Test Strips quantitatively measure glucose (D-glucose) in fresh neonatal, venous, arterial and fingertip capillary whole blood. The test strips are for use outside the body (in vitro diagnostic use) by healthcare professionals for use in healthcare facilities with either the Precision Xtra® or Optium® Blood Glucose Monitoring Systems. Home users may use the system for capillary samples only. The test strips are not for use in diagnosis or screening of diabetes mellitus, but are to be used as an aid in monitoring the effectiveness of diabetes control programs.
Device Story
Test strips utilize amperometric biosensor technology; enzyme Glucose Dehydrogenase (GDH) catalyzes reaction with Nicotinamide Adenine Dinucleotide (NAD); reduced NAD (NADH) re-oxidized by electrochemical mediator 1,10 phenanthroline quinone (1,10PQ); electron transfer at electrode surface generates current proportional to glucose concentration. Meter translates current into glucose value after 20-second countdown; requires lot-specific information from ROM calibrator. Used in healthcare facilities by professionals or at home by patients for capillary blood monitoring. Output provides quantitative glucose concentration to aid in diabetes management.
Clinical Evidence
Performance verified through clinical and non-clinical laboratory testing. Evaluated parameters included altitude, dynamic range, precision, linearity, accuracy, interference, oxygen sensitivity, environmental factors, hematocrit, shipping, and shelf life. Results demonstrated substantial equivalence to the predicate device.
Technological Characteristics
Amperometric biosensor technology; enzyme-based (Glucose Dehydrogenase/NAD); electrochemical mediator (1,10 phenanthroline quinone). System includes test strips and meter with ROM calibrator for lot-specific data. Operates via electron transfer at electrode surface. No specific material standards or software architecture details provided.
Indications for Use
Indicated for quantitative glucose measurement in fresh neonatal, venous, arterial, and fingertip capillary whole blood. For use by healthcare professionals in clinical settings or by home users (capillary samples only) to monitor diabetes control programs. Not for diabetes diagnosis or screening.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- Precision®/ Optium® Point of Care Blood Glucose Test Strips (K021960)
Related Devices
- K171941 — FreeStyle Precision Neo Blood Glucose Test Strips · Abbott Diabetes Care, Inc. · Sep 27, 2017
- K962295 — PRECISION QID BLOOD GLUCOSE TEST STRIP · Medisense, Inc. · Jul 26, 1996
- K062829 — DIACHEX BLOOD GLUCOSE MONITORING SYSTEM · Tyson Bioresearch, Inc. · Oct 10, 2006
- K963676 — PRECISION G BLOOD GLUCOSE TESTING SYSTEM · Medisense, Inc. · Feb 20, 1997
- K140371 — FREESTYLE PRECISION NEO BLOOD GLUCOSE: MONITORING SYSTEM, TEST STRIPS (50CT), TEST STRIPS (100CT) · Abbott Diabetes Care, Inc. · Sep 30, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
K070984
Image /page/0/Picture/1 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a stylized letter "A" on the left, followed by the words "Abbott" and "Diabetes Care" stacked on top of each other on the right. The text is in a bold, sans-serif font.
Abbott Diabetes Care Inc. 1360 South I oop Road Alameda, CA 94502 t: (510) 749-5400
NOV - 1 2007
# 510(k) Summary
(as required by 21 CFR 807.92)
| Date Prepared: | October 09, 2007 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------|
| Company | Abbott Laboratories |
| Division | Abbott Diabetes Care Inc. |
| Street Address | 1360 South Loop Road |
| City, State Zip | Alameda, CA 94502 |
| Telephone No. | 510-749-5400 |
| Fax No. | 510-864-4791 |
| Contact Person: | Maria Trejo<br>Regulatory Affairs Associate<br>Tel No. 510-749-6384<br>Fax No. 510-864-4791<br>maria.trejo@abbott.com |
| Device Name: | Precision® Point of Care Blood Glucose Test Strip<br>Optium® Point of Care Blood Glucose Test Strip |
| Common Name: | Blood Glucose Test Strip |
| Classification Name: | Glucose Dehydrogenase, Glucose. Glucose Test System,<br>Class II (21 CFR §862.1345, Product code LFR) |
| Predicate Device: | Precision®/ Optium® Point of Care Blood Glucose Test<br>Strips (K021960) |
# Description of the Device:
The Precision® /Optium® Point of Care Blood Glucose Test Strips work on the principle of amperometric biosensor technology, by determination of glucose oxidised by the enzyme (Glucose Dehydrogenase, GDH) catalysed reaction with Nicotinamide Adenine Dinucleotide (NAD). The reduced form of NAD (NADH) is re-oxidized by reaction with the electrochemical mediator, 1,10 phenanthroline quinone (1,10PQ). The reduced mediator is re-oxidized via electron transfer at the electrode surface. This current is translated into a number by the meter, after applying lot specific information from the supplied ROM calibrator and after a 20 second countdown, a concentration value is presented to the user.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a stylized letter "a" on the left, followed by the words "Abbott" and "Diabetes Care" stacked on top of each other on the right. The text is in a bold, sans-serif font.
#### Intended Use:
The Precision®/Optium® Point of Care Blood Glucose Test Strips quantitatively measure I he I receision > Operain > > Sprain > > arterial and fingertip capillary whole blood. The test strips are for use outside the body (in vitro diagnostic use) by healthcare I no cost surpo are in healthcare facilities with either the Precision Xtra® or Optium® Blood Glucose Monitoring Systems. Home users may use the system for capillary samples only. The test strips are not for use in diagnosis or screening of diabetes mellitus, but are to be used as an aid in monitoring the effectiveness of diabetes control programs.
#### Summary of Technological Characteristics:
The Precision® Point of Care Blood Glucose Test Strips/ Optium® Point of Care Blood Glucose Test Strips have the same fundamental scientific technology and the same intended use as the current Precision® Point of Care Blood Glucose Test Strips and Optium® Point of Care Blood Glucose Test Strips which is based on amperometric biosensor technology. The Precision® Point of Care Blood Glucose Test Strips/ Optium® Point of Care Blood Glucose Test Strips are substantially equivalent to the current predicate device.
#### Assessment of Non-Clinical Performance Data:
The performance of the Precision®/Optium® Point of Care Blood Glucose Test Strips was verified through non clinical testing in the laboratory. These studies demonstrated that the Precision® Point of Care Blood Glucose Test Strips/ Optium® Point of Care Blood Glucose Test Strips are substantially equivalent to the current Precision® Point of Care Blood Glucose Test Strips/Optium® Point of Care Blood Glucose Test Strips for blood glucose measurements and are suitable for its intended use.
## Assessment of Clinical Performance Data:
The performance of the Precision®/Optium® Point of Care Blood Glucose Test Strips was also verified through clinical testing. Some of the test performed and passed were altitude, dynamic range, precision , linearity, accuracy, interference, oxygen sensitivity, environmental, hematocrit, shipping and shelf life. These studies demonstrated that the Precision® Point of Care Blood Glucose Test Strips/ Optium® Point of Care Blood Glucose Test Strips are substantially equivalent to the current Precision® Point of Care Blood Glucose Test Strips/Optium® Point of Care Blood Glucose Test Strips for blood glucose measurements and are suitable for its intended use.
### Conclusion:
Results of clinical and non-clinical testing demonstrate that the performance of the Precision® Point of Care Blood Glucose Test Strips and Optium® Point of Care Blood Glucose Test Strips, when used according to the intended use stated above, are acceptable and comparable to the performance of the previously mentioned predicate device for blood glucose testing.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Abbott Laboratories Abbott Diabetes Care Inc. c/o Ms. Maria Trejo Regulatory Affairs Associate 1360 South Loop Road Alameda, CA 94502
NOV - 1 2007
Re: k070984
> Trade/Device Name: Precision® Point of Care Blood Glucose Test Strips Optium® Point of Care Blood Glucose Test Strips Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System. Regulatory Class: Class II Product Code: LFR, NBW Dated: October 10, 2007 Received: October 15, 2007
Dear Ms. Trejo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known): K070984
Device Name: Precision® Point of Care Blood Glucose Test Strips Optium® Point of Care Blood Glucose Test Strips
Indications for Use:
The Precision® and Optium® Point of Care Blood Test Strips quantitatively measure glucose (D-glucose) in fresh neonatal, venous, arterial and fingertip capillary whole blood. The test strips are for use outside the body (in vitro diagnostic use) by healthcare professionals for use in healthcare facilities with either the Precision Xtra® or Optium® Blood Glucose Monitoring Systems. Home users may use the system for capillary samples only. The test strips are not for use in diagnosis or screening of diabetes mellitus, but are to be used as an aid in monitoring the effectiveness of diabetes control programs.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol Ben
Page 1 of 1
ion Sign-Off
ിfice of In Vitro Diagnostic Device I Valuation and Safety
K070984j