K Number
K070787

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-08-24

(155 days)

Product Code
Regulation Number
888.3160
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tornier Elbow Prosthesis is intended for total elbow arthroplasty. Prosthetic replacement with this device may be indicated to relieve severe pain or significant disability following the effects of primary or secondary osteoarthtitis and rheumatoid arthritis; correction of functional deformities; revision procedures where other treatments or devices have failed; treatment of fractures that are unmanageable using other techniques.

The Tornier Elbow Prosthesis is intended for cemented use only.

Device Description

Total Elbow replacement is used to treat a number of clinical conditions such as severe pain or significant disability in degenerative, rheumatic disease of the elbow joint. It is also used in revision procedures where other treatments or devices have failed and treatment of fractures that are unmanageable using other techniques. The usual goal of such surgery is to restore the cloow joint to its best working condition and to reduce or eliminate pain. The Latitude Tornier Elbow Prosthesis is intended to accomplish these goals. The Tornier Elbow prosthesis is intended for use as a cemented total elbow.

The Latitude Tornier Elbow Prosthesis is a 3-part system consisting of a humeral, an ulnar and a radial component. The humeral implant is modular and consists in the assembly of various sizes of humeral stem and humeral spool in order to better reproduce the functionality of the natural humerus.

The prosthesis is a non-constrained prosthesis and when it is used with the ulnar cap the prosthesis becomes a semi-constrained prosthesis.

The present device submission corresponds to a modification in the assembly of the humeral stem with the humeral spool. Included in the modification the material of the ring stop has changed from UHMWPE to PEEK-OPTIMA. The radial components and the ulnar components are not modified. The technological characteristics (design, materials, manufacturing, sterilization, sizing and indications) of the modified humeral spool and humeral screw components are similar or identical to the predicate devices.

AI/ML Overview

The provided text is a summary of safety and effectiveness information for a medical device (Latitude Tornier Elbow Prosthesis) and correspondence from the FDA regarding its 510(k) submission. It does not contain information about acceptance criteria, device performance metrics, or study details such as sample sizes, ground truth establishment, or expert qualifications for a clinical validation study.

The document primarily focuses on:

  • Device classification and regulatory information.
  • Device description, materials, and intended use.
  • The FDA's determination of substantial equivalence to predicate devices, allowing the device to be marketed.

Therefore, I cannot fulfill the request for information regarding acceptance criteria and the study that proves the device meets them, as this information is not present in the provided text. The document is a regulatory approval notice, not a clinical study report.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "TORNIER" in all capital letters, followed by a logo. The logo is a hexagon with the letter "T" inside of it. The letter "T" is also surrounded by a smaller hexagon.

K070787

AUG 2 4 2007

Summary of Safety and Effectiveness information Special 510(k) Premarket Notification - Latitude Tornier Elbow Prosthesis

Regulatory authority: Safe Medical Devices Act of 1990, 21 CRF 807.92

1) Device name
Trade name:Latitude Elbow Prosthesis
Common name:Elbow Prosthesis
Classification name:Elbow joint metal/polymer semi-constrained cemented prosthesis
Elbow joint metal/metal or metal/polymer constrained cementedprosthesis
Classification number:888.3160 and 888.3150
  1. Submitter Tornier Rue Doyen Gosse 38330 Saint Ismier - France

  2. Company contact Tornier Mrs Mircille Lémery Regulatory affairs Manager 161, rue Lavoisier - Montbonnot 38334 Saint Ismier Cedex - France Tel: 00 33 4 76 61 38 98 Fax: 00 33 4 76 61 35 33 e-mail: mireille.lemery@tornier.fr

  3. Classification Device class: Class II Classification panel: Orthopedic Product code: JDB and JDC

  4. Equivalent / Predicate device

Tornier Elbow Prosthesis, Tornier, K000003, K011567, K031218 and K050848

6) Device description

Total Elbow replacement is used to treat a number of clinical conditions such as severe pain or significant disability in degenerative, rheumatic disease of the elbow joint. It is also used in revision procedures where other treatments or devices have failed and treatment of fractures that are unmanageable using other techniques. The usual goal of such surgery is to restore the

Page 1/ page 2

TORNIER

161 rue Lavoisier. Montbonnot. 38334 SAINT-ISMIER Cedex. France Tél. : 33(0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33 www.tornier.com SAS au capital de 268 000 €. SIRET 070 501 275 000 13. R.C.S. Grenoble 070 501 275. Code APE 331 B ège social : chemin Doyen Gosse. 38330 Saint-Ismier. France

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "TORNIER" in all capital letters, with a logo to the right of the word. The logo is a hexagon with the letter "T" inside of it. The letter "T" is stylized with a smaller "T" inside of it.

cloow joint to its best working condition and to reduce or eliminate pain. The Latitude Tornier Elbow Prosthesis is intended to accomplish these goals. The Tornier Elbow prosthesis is intended for use as a cemented total elbow.

The Latitude Tornier Elbow Prosthesis is a 3-part system consisting of a humeral, an ulnar and a radial component. The humeral implant is modular and consists in the assembly of various sizes of humeral stem and humeral spool in order to better reproduce the functionality of the natural humerus.

The prosthesis is a non-constrained prosthesis and when it is used with the ulnar cap the prosthesis becomes a semi-constrained prosthesis.

The present device submission corresponds to a modification in the assembly of the humeral stem with the humeral spool. Included in the modification the material of the ring stop has changed from UHMWPE to PEEK-OPTIMA. The radial components and the ulnar components are not modified. The technological characteristics (design, materials, manufacturing, sterilization, sizing and indications) of the modified humeral spool and humeral screw components are similar or identical to the predicate devices.

7) Materials

The humeral stem is available in CoCr alloy. The humeral spool is available in CoCr alloy and PEEK-OPTIMA polymer. The humeral screw is available in stainless steel. The radial components are made of CoCr alloy and UHMWPE. The ulnar components are made of CoCr alloy and UHMWPE.

8) Indications

The Tornier Elbow Prosthesis is intended for total elbow arthroplasty. Prosthetic replacement with this device may be indicated to relieve severe pain or significant disability following the effects of primary or secondary osteoarthtitis and theumatoid arthritis; correction of functional deformities; revision procedures where other treatments or devices have failed; treatment of fractures that are unmanageable using other techniques.

The Tornier Elbow Prosthesis is intended for cemented use only.

Page 2/ page 2

TORNIER

161 rue Lavoisier. Montbonnot. 38334 SAINT-ISMIER Cedex. France Tél. : 33(0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33 www.tornier.com SAS au capital de 288 000 € SIRET 070 501 275 000 13. R.C.S. Grenoble 070 501 275. Code APE 331 B al : chemin Doyen Gosse. 38330 Saint-tsmier. France

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Tornicr % Mrs. Mireille Lémery Regulatory Affairs Manager 161, rue Lavoisier - Montbonnot 38334 Saint Ismier Cedex - France

Re: K070787

Trade/Device Name: Latitude Elbow Prosthesis Regulation Number: 21 CFR 888.3160 Regulation Name: Elbow joint metal/polymer semi-constrained prosthesis Regulatory Class: II Product Code: JDB, JDC Dated: July 20, 2007 Received: July 25, 2007

Dear Mrs. Lémery:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

AUG 2 4 2007

{3}------------------------------------------------

Page 2 - Mrs. Mireille Lémery

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to legally marketed predicate device results in a classification for your device and thus, perrorts your cevice to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation on ) } ped. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2011 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/dsma/dsmamain html

Sincerely yours.

Barbave BuellD

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

K070787 510(k) Number (if knøwn):

Device Name: Latitude Tornier Elbow Prosthesis

Indications For Use:

The Tornier Elbow Prosthesis is intended for total elbow arthroplasty. Prosthetic replacement with this device may be indicated to relieve severe pain or significant disability following the effects of primary or secondary osteoarthtitis and rheumatoid arthritis; correction of functional deformities; revision procedures where other treatments or devices have failed; treatment of fractures that are unmanageable using other techniques.

The Tornier Elbow Prosthesis is intended for cemented use only.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Deyice Evaluation (Division Sign-Off) Division of General. and Neurological Devices

Page 1 of 1

510(k) Number K070787

§ 888.3160 Elbow joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. An elbow joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an elbow joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a radial resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.