K Number
K070629

Validate with FDA (Live)

Date Cleared
2007-03-26

(19 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the Flex and Loop Coil Set 1.5 T is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the human body.

Used in the MAGNETOM Avanto, Espree and Symphony a Tim System, the Flex and Loop Coils are indicated for use as a diagnostic imaging device to produce transversal, sagittal, coronal and oblique images of the internal structures of the body. The images produced by the MAGNETOM Avanto, Espree and Symphony a Tim System with the Flex and Loop Coils reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance.

When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

The intended use of the MAGNETOM Avanto, Espree and Symphony a Tim System is not affected in any way by the use of the new Flex and Loop Coil Set 1.5 T.

Device Description

The Flex and Loop Coil Set 1.5 T consists of five receive-only coils and one interface, which contains the preamplifiers, cable traps and a combiner network. The coils are two 4-channel flexible coils (Flex Coil Large and Flex Coil Small) and three linear polarized (LP) flexible coils (Loop Coil 4 cm, 7 cm and 11 cm).

AI/ML Overview

Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and study for the Siemens Flex and Loop Coil Set 1.5 T:

This submission (K070629) is for a device accessory – specifically, the Flex and Loop Coil Set 1.5 T, which is a receive-only coil set used with MRI scanners. The primary focus of the evaluation is on safety and performance parameters relevant to the coil's function within an MRI system, not on a diagnostic algorithm's accuracy.

Therefore, many of the typical acceptance criteria and study components you'd see for an AI-powered diagnostic device (like sample sizes for test/training sets, ground truth establishment by experts, MRMC studies, or standalone algorithm performance) are not applicable here. This is a hardware component.

The core of the submission relies on demonstrating substantial equivalence to an existing predicate device (the Flex Loop Coil Set for 3T systems, K063373) by showing that the new device does not raise new questions of safety or effectiveness.


Acceptance Criteria and Reported Device Performance

The acceptance criteria here are defined by general safety and performance parameters for MR diagnostic devices, as well as specific parameters for the coil set, with the implicit acceptance criterion being that the new device is "equivalent" or "unaffected" compared to the predicate or existing standards.

Acceptance Criteria CategorySpecific ParameterReported Device Performance
SafetyMaximum Static FieldUnaffected by modifications
Rate of Change of Magnetic FieldUnaffected by modifications
RF Power DepositionUnaffected by modifications
Acoustic Noise LevelUnaffected by modifications
BiocompatibilityNo new materials, therefore no new tests performed; implicitly safe.
PerformanceGeometric DistortionUnaffected by modifications
Slice Profile, Thickness and GapUnaffected by modifications
High Contrast Spatial ResolutionUnaffected by modifications
Signal to Noise Ratio (SNR)Tested, results show equivalence with predicate devices (acc. to NEMA MS-6)
Image UniformityTested, results show equivalence with predicate devices (acc. to NEMA MS-6)

Study Details:

Given that this is a hardware accessory being evaluated for substantial equivalence rather than a diagnostic algorithm, much of the requested information (like sample sizes for test/training sets, expert ground truth, adjudication, MRMC studies, standalone algorithm performance) is not relevant or applicable to this 510(k) submission.

  1. Sample size used for the test set and the data provenance: Not applicable. The "test set" in this context refers to laboratory testing and measurements performed on the physical coil set, not a dataset of patient images.

    • Data Provenance: The "study" appears to be internal laboratory testing by Siemens AG (Germany) and Siemens Medical Solutions USA, Inc.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically defined for diagnostic algorithms (e.g., presence/absence of disease) is not part of this submission. The "ground truth" for hardware performance is established by metrology, engineering specifications, and adherence to standards like NEMA MS-6.

  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, an MRMC study was not done, as this is not an AI/diagnostic algorithm device.

  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: No, a standalone algorithm performance study was not done, as this is a hardware accessory.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable in the traditional sense. The "ground truth" for the device's technical performance parameters (SNR, image uniformity) would be derived from engineering standards and physical measurements, often using phantoms.

  7. The sample size for the training set: Not applicable. There is no "training set" for this type of device.

  8. How the ground truth for the training set was established: Not applicable.


Summary of the "Study" for K070629:

The "study" described in this 510(k) is a series of laboratory tests performed on the Flex and Loop Coil Set 1.5 T to measure its technical performance (specifically Signal to Noise Ratio and Image Uniformity) according to the NEMA MS-6 standard. These measured performance characteristics were then compared to those of the predicate device.

The conclusion of the study was that the results obtained for the new device "show that they are equivalent with the predicate devices" for SNR and image uniformity. For other safety and performance parameters (Static Field, Rate of Change of Magnetic Field, RF Power Deposition, Acoustic Noise Level, Geometric Distortion, Slice Profile, Thickness and Gap, High Contrast Spatial Resolution), the modifications were deemed to have no effect. Biocompatibility was assessed as acceptable because no new materials were used.

This approach is standard for demonstrating substantial equivalence for hardware accessories where the primary concern is that the new device performs within established limits and does not introduce new safety or effectiveness concerns compared to a legally marketed predicate device.

{0}------------------------------------------------

K070629

Section 8 510(k) Summary

8 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.

I. General Information MAR 2 6 2007 Date of summary preparation: February 13, 2007 Manufacturer Siemens AG, Medical Solutions Henkestrasse 127 D-91052 Erlangen, Germany Headquarters: Siemens AG Wittelsbacherplatz 2 D-80333 Munich, Germany Registration Number 8010024 Importer/Distributor Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Registration Number 2240869 Contact Person Mr. Helmut Neumann Regulatory Affairs Manager Henkestrasse 127 D-91052 Erlangen, Germany

Siemens 510(k) Premarket Notification February 13, 2007

e-mail: helmut.neumann@siemens.com

Phone: +49 (9131) 84-7445 Fax: +49 (9131) 84-2200

{1}------------------------------------------------

Classification and Device Name

Classification Panel:Radiology
Classification Name:Magnetic Resonance Diagnostic Device
Accessory
Device Class:Class II [21 CFR § 892.1000]
Product Code:MOS
Common Name:Special Purpose Coil
Trade Name:Flex and Loop Coil Set 1.5 T

Safety and Effectiveness Information Supporting Substantial II. Equivalence

Intended Use

The intended use of the Flex and Loop Coil Set 1.5 T is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the human body. Used in the MAGNETOM Avanto, Espree and Symphony a Tim System, the Flex and Loop Coils are intended to produce transversal, sagittal, coronal and oblique images of the internal structures of the body.

Device Description

The Flex and Loop Coil Set 1.5 T consists of five receive-only coils and one interface, which contains the preamplifiers, cable traps and a combiner network. The coils are two 4-channel flexible coils (Flex Coil Large and Flex Coil Small) and three linear polarized (LP) flexible coils (Loop Coil 4 cm, 7 cm and 11 cm).

Image /page/1/Figure/8 description: The image shows two lines labeled (1) and (2). According to the text, (1) is the coil socket and (2) is the plug for the coil socket on the patient table. The lines point to the respective parts in the image.

Figure 8.II.1: Flex loop interface

{2}------------------------------------------------

Image /page/2/Figure/1 description: The image shows three diagrams of a coil loop and coil plug. Each diagram has two labeled parts: (1) Coil loop and (2) Coil plug. The diagrams are simple line drawings, and the labels are placed near the corresponding parts of the coil.

Figure 8.II.2: Loop coils 4 cm, 7 cm, and 11 cm

Image /page/2/Figure/3 description: The image shows two diagrams of an electronic component, possibly a circuit board or a similar device. Each diagram has labels indicating different parts: (1) Center markings, (2) Electronic boxes, and (3) Coil plug. The diagrams appear to be technical illustrations, providing a visual representation of the component's structure and key features.

Figure 8.II.3: Flex Coil large and small

77

{3}------------------------------------------------

Equivalency Information

Siemens believes that the Flex and Loop Coil Set 1.5 T for MAGNETOM Avanto, Espree and Symphony a Tim System is substantially equivalent to the Flex Loop Coil Set for MAGNETOM 3 T systems described in the following submission:

Device NameFDA Clearance NumberFDA Clearance Date
Flex Loop Coil Set3 TK063373Nov 17, 2006

{4}------------------------------------------------

Summary of Technological Characteristics of the Principal Device as Compared with the Predicate Device

Although the new Flex and Loop Coils are designed for a field strength of 1.5 T, Siemens believes that they are substantially equivalent to the predicate Flex Loop Coils for the 3 T MAGNETOM Trio a Tim System.

General Safety and Effectiveness Concerns

The following safety and performance parameters:

[Safety]

  • Maximum Static Field
  • Rate of Change of Magnetic Field
  • RF Power Deposition
  • Acoustic Noise Level

[Performance]

  • Geometric Distortion
  • Slice Profile, Thickness and Gap
  • High Contrast Spatial Resolution

specified by the FDA Guidance document for MR Diagnostic Devices are unaffected by the modifications described within this notification.

The following parameters were considered for the new Flex and Loop Coil Set 1.5 T

[Safety] - Biocompatibility

[Performance] - Signal to Noise Ratio - Image Uniformity

No new materials were used for the new Flex and Loop Coil Set 1.5 T compared to their predicate device. Therefore no biocompatibility tests were performed. Signal to Noise Ratio (SNR) and image uniformity tests acc. to NEMA MS-6 standard were performed for the new Flex and Loop Coil Set 1.5 T and the results presented in this submission show that they are equivalent with the predicate devices.

Conclusion as to Substantial Equivalence

Laboratory testing was performed to support this claim of substantial equivalence and to show that the technological differences do not raise any new questions pertaining to effectiveness.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem resembling an eagle or bird in flight, composed of three curved lines that suggest feathers or wings. The logo is presented in black and white, with the text and emblem appearing in black against a white background.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Siemens Medical Solutions % Mr. Stefan Preiss Responsible Third Party Official TÜV Product Service 1775 Old Hwy 8 NW, Ste 104 NEW BRIGHTON MN 55112-1891

MAR 2 6 2007

Re: K070629

Trade/Device Name: Flex and Loop Coil Set 1.5T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: March 7, 2007 Received: March 7, 2007

Dear Mr. Preiss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/5/Picture/10 description: The image is a black and white circular logo. The logo contains the text "1886-1986" at the top, followed by the letters "PA" in a large, bold font. Below the letters, the word "Centennial" is written in a cursive font. There are four stars at the bottom of the logo. The logo appears to be a commemorative emblem for the centennial anniversary of an organization or event.

Protesting and Promoting Public Health

{6}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely vours.

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

2 Indications for Use Statement

510(k) Number (if known) _ K 070629

Device Name: Flex and Loop Coil Set 1.5 T

Indications for Use:

The intended use of the Flex and Loop Coil Set 1.5 T is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the human body.

Used in the MAGNETOM Avanto, Espree and Symphony a Tim System, the Flex and Loop Coils are indicated for use as a diagnostic imaging device to produce transversal, sagittal, coronal and oblique images of the internal structures of the body. The images produced by the MAGNETOM Avanto, Espree and Symphony a Tim System with the Flex and Loop Coils reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance.

When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

The intended use of the MAGNETOM Avanto, Espree and Symphony a Tim System is not affected in any way by the use of the new Flex and Loop Coil Set 1.5 T.

Prescription Use __ X (Part 21 CFR 801 Subpart D) Over-The-Counter Use

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel h. Sepermm

sion of Reproductive, adiological Devices

Siemens 510(k) Premarket Notification

February 13, 2007

Page 2-1

N/A