K Number
K070465

Validate with FDA (Live)

Date Cleared
2007-07-19

(153 days)

Product Code
Regulation Number
868.5120
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Epidural Catheter is intended for administration of local anesthetics into the epidural space. Busse recommends the removal and/or replacement of the catheter every 72 hours.

Device Description

The Busse Epidural Catheter is a single use device, which is sold as sterile individually packaged and sterile packaged inside a kit/procedure tray. The Catheter will also be sold as a non-sterile product to kit packers, who then may sterilize the catheter as part of a kit.

AI/ML Overview

The provided document is a 510(k) summary for the Busse Epidural Catheter, which primarily focuses on demonstrating substantial equivalence to a predicate device. It does not contain information about a study with acceptance criteria for device performance in the way a clinical trial or algorithm performance study would. Instead, it details biocompatibility testing for the materials used in the catheter.

Here's an breakdown based on the information available:

1. Table of Acceptance Criteria and Reported Device Performance

Test PerformedAcceptance CriteriaReported Device Performance
L929 Men Elution TestAcceptable for intended useMet the testing requirements and were found to be acceptable
Kligman Maximization TestAcceptable for intended useMet the testing requirements and were found to be acceptable
Intracutaneous Injection TestAcceptable for intended useMet the testing requirements and were found to be acceptable
Systemic Injection TestAcceptable for intended useMet the testing requirements and were found to be acceptable
Salmonella Typhimurium; and Escherichia Coli Reverse Mutation AssayAcceptable for intended useMet the testing requirements and were found to be acceptable
Hemolysis - Rabbit BloodAcceptable for intended useMet the testing requirements and were found to be acceptable
14 Day Repeat Dose Intravenous Toxicity Study (Sub - Chronic)Acceptable for intended useMet the testing requirements and were found to be acceptable
Intramuscular Implantation Test - ISOAcceptable for intended useMet the testing requirements and were found to be acceptable

Note: The document states "These materials have met the testing requirements and were found to be acceptable for the intended use" for all listed tests, implying that meeting the standard for each test constitutes the acceptance criteria.

2. Sample Size Used for the Test Set and Data Provenance

This document describes biocompatibility testing of materials, not a test set of data for an AI/algorithm. Therefore, concepts like sample size for a test set, and data provenance (country of origin, retrospective/prospective) are not applicable in this context. The tests listed are laboratory-based assessments of material interaction with biological systems.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not Applicable. Ground truth, in the context of expert review, is not relevant for the biocompatibility tests described. These tests have established scientific methods and pass/fail criteria.

4. Adjudication Method for the Test Set

Not Applicable. Adjudication methods are typically used when there's subjective interpretation, like in a clinical image review. Biocompatibility tests have objective results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC study was not done. This document is for a physical medical device (epidural catheter), not an AI/algorithm where human reader performance would be evaluated.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, this is not an AI/algorithm device, so standalone performance is not applicable.

7. The Type of Ground Truth Used

The "ground truth" for the biocompatibility tests is the established scientific methodology and criteria for each specific test (e.g., cell viability in L929, skin reaction scores in Kligman, hemolytic index in rabbit blood). Pathology would be relevant if the study involved tissue analysis from implantation, which is part of the Intramuscular Implantation Test. Outcomes data is not mentioned in these specific tests.

8. The Sample Size for the Training Set

Not Applicable. This is not an AI/algorithm, so there is no training set.

9. How the Ground Truth for the Training Set Was Established

Not Applicable. As there's no training set, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for "busse Hospital Disposables". The word "busse" is in a bold, sans-serif font, with the letters slightly overlapping each other. Below the word "busse", the words "Hospital Disposables" are written in a smaller, less bold, sans-serif font. The logo is in black and white.

K070465

SUMMARY

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS BUSSE EPIDURAL CATHETER

Regulatory Affairs Contact:Muhamad AnsariBusse Hospital DisposablesPO Box: 1106775 Arkay Dr.Hauppauge NY 11788JUL 19 2007
Telephone:631-435-4711 Ext: 254
Fax:631-435-2849
Date Summary Prepared:February 8th 2007
Product Trade Name:Busse Epidural Catheter
Common Name:Epidural Catheter
Classification Name:Anesthesia Conduction Catheter
Classification:Class II, 21 CFR 868.5120
Product Code:BSO
Predicate Device:Micor - Conduction Catheter -- (K001717)
Device Description:The Busse Epidural Catheter is a single use device,which is sold as sterile individually packaged andsterile packaged inside a kit/procedure tray.
Intended Use:The Epidural Catheter is intended for administration oflocal anesthetics into the epidural space. Busserecommends the removal and/or replacement of thecatheter every 72 hours.The epidural catheter will be sold sterile individuallypackaged, and as part of a sterile kit.The Catheter will also be sold as a non-sterile productto kit packers, who then may sterilize the catheter aspart of a kit.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for "busse Hospital Disposables". The word "busse" is in a bold, stencil-like font, with the letters close together. Below the word "busse" is the phrase "Hospital Disposables" in a smaller, regular font. The logo is black and white.

510(k) SUMMARY OF SUBSTANTIAL EQUIVALENCE

Summary of Testing:All materials used in the fabrication of the epidural catheterwere evaluated through biological qualification safety tests.The biocompatibility tests performed were L929 Men ElutionTest, Kligman Maximization Test,Intracutaneous Injection Test, Systemic Injection Test,Salmonella Typhimurium; and Escherichia Coli ReverseMutation Assay, Hemolysis - Rabbit Blood, 14 Day RepeatDose Intravenous Toxicity Study (Sub - Chronic),Intramuscular Implantation Test - ISOThese materials have met the testing requirements and werefound to be acceptable for the intended use.
Technological Characteristics:[21 CFR 807.92(a)(6)]The subject device has the same TechnologicalCharacteristics as a legally marketed predicate device.
Conclusion:[21 CFR 807.92(b)(3)]The above statements are accurate representations of the deviceBusse intents to market.Based on all the testing and comparison Busse believes thesubject device is substantially equivalent to the predicatedeviceAll data and information submitted in this premarketnotification is truthful and accurate and no material fact hasbeen omitted.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 9 2007

Mr. Muhamad Ansari Director of Regulatory Affairs Busse Hospital Disposables P.O. Box 11067 Hauppauge, New York 11788

Re: K070465

Trade/Device Name: Busse Epidural Catheter Regulation Number: 21 CFR 868.5120 Regulation Name: Anesthesia Conduction Catheter Regulatory Class: II Product Code: BSO Dated: July 6, 2007 Received: July11, 2007

Dear Mr. Ansari:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 -Mr. Ansari

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Cars

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): __k ك ك ك ك

Device Name: Busse Epidural Catheter.

Indication for Use: The Epidural Catheter is intended for administration of local anesthetics into the epidural space. Busse recommends the removal and/or replacement of the catheter every 72 hours.

Prescription Use X (Per 21 CFR 801Subpart D)

AND/OR

Over-The-Counter Use (Per 21 CFR 801Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Null J. Dadel

al Color Sign-Olt)
Division of Anesthesiology, General Hospital
infection Control, Dental Devices infection of Anesthesiology, Gene

510(k) Number: K070465

§ 868.5120 Anesthesia conduction catheter.

(a)
Identification. An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.(b)
Classification. Class II (performance standards).