K Number
K070317

Validate with FDA (Live)

Date Cleared
2007-07-05

(153 days)

Product Code
Regulation Number
866.3900
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Zeus Scientific Varicella Zoster Virus (VZV) IgM ELISA Test System is intended for the qualitative detection of IgM class antibodies to Varicella Zoster Virus in human serum as an aid in the diagnosis of primary infection or reactivation.

The assay performance in detecting antibodies to VZV in individuals vaccinated with the FDA licensed VZV vaccine is unknown.

The user of this assay is responsible for establishing the performance characteristics with VZV vaccinated individuals.

The assay performance in detecting antibodies to VZV in cord blood and neonates has not been established.

Device Description

The Zeus Scientific VZV IgM ELISA Test System is an enzyme linked immunosorbent assay intended for the qualitative detection of distict IgM antibody to the Varicella-zoster virus.

The test is designed to detect IgM antibody using inactivated VZV antigen: strain, Ellen.

AI/ML Overview

This submission describes the Zeus Scientific VZV IgM ELISA Test System, an in-vitro diagnostic device. As such, acceptance criteria and performance are typically measured in terms of diagnostic accuracy metrics (sensitivity, specificity, agreement) against a comparator method, rather than effect sizes of human reader improvement or standalone algorithm performance.

Here's a breakdown of the requested information based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined "acceptance criteria" as clear numerical thresholds for performance. However, it presents the "Agreement Summary" with a commercially distributed ELISA assay as the primary evidence of performance. The performance metrics are Positive % Agreement and Negative % Agreement.

MetricAcceptance Criteria (Not explicitly stated, inferred from results)Reported Device Performance (Prospective and Retrospective Samples: Combined Sites)
Positive % AgreementSufficiently high agreement with predicate (e.g., >90%)97.4% (95% CI: 86.5% to 99.9%)
Negative % AgreementSufficiently high agreement with predicate (e.g., >90%)95.6% (95% CI: 91.8% to 97.1%)

Note: The reported performance for "Prospective Samples: Combined Sites" also provides similar figures: Positive % Agreement = 100% (95% CI: 54.1% to 100.0%) and Negative % Agreement = 95.6% (95% CI: 92.6% to 97.6%). The table above uses the combined prospective and retrospective data as it represents a larger and more comprehensive dataset for agreement.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:
    • Prospective samples: 302
    • Retrospective samples: 36
    • Total combined samples: 338 (used for the primary agreement summary)
  • Data Provenance: Retrospective and Prospective. The document does not specify the country of origin of the data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The ground truth for the test set was established by comparison to a "commercially distributed VZV IgM ELISA test system" (predicate device). There is no mention of human experts establishing ground truth for the test set.

4. Adjudication Method for the Test Set

  • The document implies a direct comparison method, where the results of the Zeus Scientific VZV IgM ELISA Test System were compared against the results of the "commercially distributed VZV IgM ELISA test system." There is no mention of an adjudication process (like 2+1 or 3+1). The commercial ELISA served as the reference standard.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size of Human Readers Improve with AI vs. Without AI Assistance

  • No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic assay (ELISA kit), not an AI-assisted diagnostic tool involving human readers. Therefore, this section is not applicable.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, the performance presented is standalone for the assay. ELISA tests are inherently standalone, as they provide a direct result based on chemical reactions, without human interpretation other than reading the optical density and interpreting it against a cut-off (which is part of the assay's design).

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

  • Ground Truth Type: Comparison to a predicate commercial ELISA test system. This is a common method for establishing the performance of new in-vitro diagnostic assays, where a well-established and legally marketed assay serves as the reference standard.

8. The Sample Size for the Training Set

  • The document mentions smaller sets of samples used for various non-clinical performance aspects:
    • Cut-off Establishment: 25 known negative samples and a minimum of 5 known positive samples (total at least 30 samples). These were confirmed by a commercially distributed ELISA assay.
    • Interfering Substances: 3 samples (positive, borderline, negative) were tested with various interfering substances.
    • Cross-Reactivity: A minimum of 10 samples for each cross-reactive substance (EBV, CMV, Lyme, RF, Mumps, Toxo, Measles, Rubella) were tested (total at least 80 samples). These were confirmed negative for VZV IgM using the predicate device.
    • Precision: 6 samples were used, tested at three sites, over three days.
  • It's important to note that for IVDs like ELISA kits, the concept of a "training set" as understood in machine learning (where an algorithm learns from data) isn't directly applicable in the same way. These studies are for establishing performance characteristics rather than "training" an algorithm. The samples listed above are used to define the assay's characteristics and validate its function.

9. How the Ground Truth for the Training Set was Established

  • For studies like cut-off establishment and cross-reactivity, the ground truth was established by confirmation using a commercially distributed ELISA assay (the predicate device). For other studies like linearity, limits of detection, interfering substances, and precision, the ground truth is inherent to the experimental design (e.g., known dilutions, spiked samples, or reproducibility across replicates/sites).

{0}------------------------------------------------

510 (k) Summary

JUL - 5 2007 Zeus Scientific VZV IgM ELISA Test System

510 (k) 070317

As required by 21 CFR 807.92, the following 510(k) summary is provided:

Submitter Information ਜ

Contact:Ewa Nadolczak, Manager, Clinical Affairs Zeus Scientific, Inc. 200 Evans Way Branchburg, NJ 08876 Telephone: 908.526.3744 Direct: 347.731.0402 Email: enadolczak@zeusscientific.com Fax: 908.526.2058 Establishment Registration Number: 224236

Device Information 2

Proprietary Name: Zeus Scientific Varicella-Zoster (VZV) IgM ELISA Test Kit Classification Name: Varicella-zoster Virus Serological reagents Class: Class II CFR: 866.39 Product Code: LFY

Predicate Device Information 3

Manufacturer: Trinity Biotech Name: Captia VZV IgM ELISA Test Kit Methodology: ELISA 510{k} Number: Unknown

Device Description বা

The Zeus Scientific VZV IgM ELISA Test System is an enzyme linked immunosorbent assay intended for the qualitative detection of distict IgM antibody to the Varicella-zoster virus.

The test is designed to detect IgM antibody using inactivated VZV antigen: strain, Ellen.

5 Intended Use

The Zeus Scientific VZV IgM ELISA Test System is intended for the qualitative detection of IgM antibody to Varicella-Zoster virus in human serum as an aid in diagnosis of primary infection or reactivation.

The assay performance in detecting antibodies to VZV in individuals Vaccinated with the FDA licensed VZV vaccine is unknown. The user of this Assay is responsible for establishing the performance characteristics with VZV vaccinated individuals. The assay performance in detecting antibodies to VZV antibodies to VZV in cord blood and neonates has not been established.

{1}------------------------------------------------

6 Summary of Technological Characteristics

The Zeus Scientific VZV IgM ELISA Test System is designed to detect IgM class antibodies in human sera to Varicella-zoster virus. Wells of plastic microwell strips are sensitized by passive absorption with VZV antigen. The test procedure involves three incubation steps:

Test sera are diluted with the Sample Diluent provided. The Sample Diluent contains anti-human IgG that is intended to bind IgG and Rheumatoid factor to prevent non-specific binding to the immobilized antigen. During sample incubation, any antigen specific IgM antibody in the sample will bind to the immobilized antigen. The plate is washed to remove unbound antibody and other serum components.

Peroxidase conjugated goat anti-human (u chain specific) is added to the wells and the plate is incubated. The conjugate will react with IgM antibody immobilized in the solid phase I, step 1.

The microwells containing immobilized peroxidase conjugate are incubated with peroxidase substrate solution. Hydrolysis of the substrate by peroxidase produces a color change. After a period of time, the reaction is stopped and the color intensity of the solution is measured photometrically. The color intensity of the solution depends upon the antibody concentration in the original test sample.

7 Performance Data Non-Clinical Establishment and Verification of Cut-off

The cut-off corresponds roughly to the mean plus (X) times the Standard Deviation of a negative population, X being the multiplication factor necessary to optimize the assay results.

25 known negative samples, confirmed by a commercially distributed ELISA assay were tested to establish the cut-off. Additionally, a minimum of 5 known positive samples, also confirmed by a commercially distributed ELISA assay were tested. The results of the known positive samples were ascertained to exceed the theoretical cut-off as well as the negative samples were ascertained to fall below the theoretical cut-off.

Linearity

Four positive samples were tested neat and with two-fold serial dilutions using the Zeus Scientific VZV IgM ELISA Test System. Results verify the linearity of the assay cut-off.

Limits of Detection

Four strongly positive samples were serially diluted and tested using the Zeus Scientific VZV IgM Test System and a commercially available EUISA test system. Results demonstrate that the Zeus Scientific VZV IgM ELISA Test System has comparable limits of detection to the commercially available ELISA test system.

Interfering Substances

Interfering Substances had been done based on industry standard levels of

{2}------------------------------------------------

test concentrations recommended in CLSI EP7-A2. The data is presented on the following page:

Interfering Substance Study

SpikedSAMPLE 2SAMPLE 3
LevelVZV IgM% PositiveVZV IgM% PositiveVZV IgM% Control
PositiveSignalBorderlineSignalNegativeSignal
Control-PBSN/A3.670.880.07
ontrol-EthanolN/A3.590.820.07
BilirubinLow3.78103.16%0.93105.33%0.08123.53%
BilirubinHigh3.5997.93%0.90101.7%0.0692.65%
AlbuminLow3.6399.05%0.91103.63%0.0688.24%
AlbuminHigh3.82104.01%0.89100.45%0.07108.82%
lgGLow2.7169.6%0.7983.0%0.10245.0%
lgGHigh1.9848.6%0.5156.70%0.16400.0%
CholesterolLow3.5097.63%0.88107.6%0.07100.0%
CholesterolHigh3.60100.33%0.88107.6%0.07102.82%
TriglyceridesLow3.80105.94%0.87106.99%0.07100.0%
TriglyceridesHigh3.79105.61%0.88107.6%0.0792.96%
HemoglobinLow3.77102.81%0.94106.58%0.14201.47%
HemoglobinHigh4.06110.66%0.97109.64%0.11167.65%
IntralipidLow3.77102.73%0.8798.53%0.08120.59%
IntralipidHigh3.6298.66%0.8798.75%0.0798.53%
ControlN/A3.660.890.06

Zeus Scientific VZV IgM ELISA

As depicted in the table above, the positive samples showed a range of recovery from 110.66% with the high spike of hemoglobin to a low of 48.6% with the high spike of IgG. The negative sample showed a range of recovery from 400% with the high spike of IgG to a low of 88.24% with the low spike of albumin. The borderline sample showed a range of recovery of 109.64% with the high spike of hemoglobin to a low of 83% with the low spike of IgG.

Some elevation of signal in the presence of excess hemoglobin was noted. The anti-IgG absorbent (SaVE Diluent) has been found to functionally remove > 13.9 mg/mL IgG from human serum. Patients with an IgG level exceeding 14 mg/mL may require additional treatment to neutralize all IgG. Excessively high levels of IgG have been shown to reduce reactivity to VZV IgM antibody.

Cross-Reactivity

A minimum of 10 samples, negative for VZV IgM, were acquired and the reactivity confirmed using the predicate device. The 10 samples were subsequently tested for cross-reactivity. In all cases the specimens remained negative for VZV IgM. Please refer to the data below. All results are presented as Index Values except where noted.

{3}------------------------------------------------

SampleIDEBV VCA IgMELISA ResultZeus ScientificVZV IgMELISA ResultSample IDCMV IgMELISA ResultZeus ScientificVZV IgMELISA Result
EBV M 53.540.77CMV M 335.220.66
EBV M 134.950.50CMV M 344.420.53
EBV M 151.940.40CMV M 351.630.64
EBV M 163.420.33CMV M 361.450.21
EBV M 175.230.52CMV M 371.570.26
EBV M 192.240.18CMV M 412.650.42
EBV M 201.190.10RD39016.920.89
4310626.030.76001779.000.13
4304103.500.32429023.006.340.86
4304113.800.83429057.003.330.78
Zeus Scientific
SampleLyme IgMVZV IgM
IDELISA ResultELISA Result
Lyme M 24.410.43
Lyme M 73.760.40
Lyme M 135.270.56
Lyme M 161.160.21
Lyme M 173.430.37
Lyme M 181.590.21
Lyme M 203.270.32
Lyme M 234.350.64
430068.002.260.64
436804.003.510.61

and the comments of the comments of the comments of

IU/mLZeus Scientific
SampleRF IgMVZV IgM
IDELISA ResultELISA Result
RF M 224.20.08
ARF 116.70.06
ARF 293.90.72
ARF 365.50.53
ARF 445.40.40
ARF 519.50.06
ARF 671.40.17
43006685.80.66
43006793.90.39
43693225.60.21
SampleIDMumps IgMELISA ResultZeus ScientificVZV IgMELISA Result
Mumps15.400.17
Mumps25.300.14
Mumps34.800.16
Mumps44.400.13
Mumps52.180.09
Mumps61.480.08
Mumps74.880.11
Mumps84.170.13
Mumps93.650.11
Mumps102.810.10

:

:

{4}------------------------------------------------

SampleIDToxo IgMELISA ResultZeus ScientificVZV IgMELISA Result
Toxo M 381.420.36
Toxo M 452.950.15
Toxo M 463.340.71
Toxo M 473.180.41
SX360341.860.23
RD40241.640.79
4304721.950.18
4348301.990.25
4348312.060.21
4348322.110.26
SampleIDMeasles IgMELISA ResultZeus ScientificVZV IgMELISA Result
Measles12.490.31
Measles21.610.26
Measles31.530.23
Measles41.820.26
Measles51.320.24
Measles62.210.30
Measles71.640.24
Measles81.250.20
Measles92.140.30
Measles102.680.30
SampleIDRubella IgMELISA ResultZeus ScientificVZV IgMELISA Result
RM 181.150.20
RM 191.510.33
RM 201.960.28
RM 351.930.19
RM 361.470.04
RM 372.550.08
RD38472.290.63
RD48142.910.18
4377062.090.18
RD67662.240.10
Result Key:
Positive
Equivocal
Negative

IgM Destruction

A VZV IgM destruction experiment was performed to assure that the antibody which is detected by the Zeus Scientific VZV IgM ELISA Test System is indeed IgM antibody. 2% beta-mercaptoethanol was the IgM destroying agent used in this study .

The results of the VVV IgM destruction study clearly demonstrate that the IgM antibody was destroyed, capturing the information that the antibody detected by the Zeus Scientific VZV IgM ELISA Test System is indeed VZV IgM antibody.

IgG/RF Removal

The Zeus Scientific VZV IgM ELISA Test System provides sample diluent which binds IgG and RF antibodies which may potentially cross-react with IgM antibodies during assay procedure. An IgG/RF removal experiment was performed. Results presented below demonstrate that the Sample Diluent eliminates reactivity of IgG and RF.

{5}------------------------------------------------

IgG Sample DiluentIgM Sample Diluent
IgG ConjIgG Sample DiluentIgG ConjIgM Sample Diluent
Sample IDODISRODISRODISRODISR
VZG+72.9028.2210.1890.5360.0020.0060.110.311
VZG+ 19>3.08.4970.230.6510.00.00.0680.192
RF+ 50.130.3690.0160.0450.0060.0160.0170.047
VZM+ RD51612.9148.2540.7762.1970.00.0010.7282.063
VZM- 426642>3.08.4970.0760.2140.00.00.0230.006
VZM- 4185231.3353.780.1720.4870.0010.0030.0990.282
VZG7/RF52.3256.5860.2550.7210.00.00.080.227
VZG19/RF52.7687.840.2130.6040.00.00.0440.124

Precision

Six samples were prepared based on their activity with the Zeus Scientific VZV IgM ELISA Test System. Two samples selected were clearly negative, two were clearly positive and two were near the assay cut0off. This panel was split into six aliquots and tested at three sites. On each day of testing, each sample was diluted twice and each dilution run in quadruplicate, resulting in eight results. This was performed for three days at each facility. A summary of this testing and calculations for the mean, standard deviation and CV appear in the following tables:

Site 1Site 2Site 3
Day 1Day 2Day 3Day 1Day 2Day 3Day 1Day 2Day 3
Sample 1
mean3.383.523.503.604.213.693.433.543.38
sd0.100.060.080.030.120.060.080.100.07
%CV2.8%1.8%2.2%1.0%2.9%1.5%2.3%2.7%2.1%
Sample 2
mean2.892.902.862.963.182.922.992.972.95
sd0.030.020.090.050.040.060.030.080.07
%CV1.0%0.8%3.0%1.6%1.4%2.1%1.1%2.7%2.3%
Sample 3
mean0.260.230.240.340.400.330.190.200.18
sd0.010.010.010.010.010.010.010.010.01
%CV5.0%2.3%3.4%2.1%2.9%3.8%3.8%7.5%8.0%
Sample 4
mean0.130.110.100.150.140.130.120.130.10
sd0.000.000.000.010.010.010.010.010.01
%CV3.6%3.6%2.6%5.0%4.0%7.1%7.5%7.3%8.5%
Sample 5
mean0.910.890.940.820.890.910.950.910.93
sd0.010.020.020.020.010.020.020.030.02
%CV1.5%2.5%2.4%2.8%1.5%1.8%2.6%2.8%1.7%
Sample 6
mean0.900.890.880.730.870.911.000.930.91
sd0.020.020.010.020.020.010.030.020.03
%CV3.0%2.0%0.9%3.3%2.4%1.4%3.3%2.5%3.4%

{6}------------------------------------------------

Inter-Assay Precison SummaryBetween Site Summary
Site 1Site 2Site 3
Sample 1mean3.843.45Sample 1mean3.6
sd0.100.290.10sd0.3
%CV0.030.070.03%CV7.1%
Sample 2mean2.903.022.97Sample 2mean3.0
sd0.100.130.06sd0.1
%CV0.020.040.02%CV3.4%
Sample 3mean0.200.360.19Sample 3mean0.3
sd0.000.030.01sd0.1
%CV0.070.100.07%CV28.1%
Sample 4mean0.100.140.11Sample 4mean0.1
sd0.000.010.01sd0.0
%CV0.110.070.13%CV14.1%
Sample 5mean0.900.870.93Sample 5mean0.9
sd0.000.050.03sd0.0
%CV0.030.050.03%CV4.6%
Sample 6mean0.900.840.95Sample 6mean0.9
sd0.000.080.05sd0.1
%CV0.020.100.05%CV8.0%

Performance Data

Clinical

Expected Values

Data was obtained from 302 prospective samples. Two samples were submitted with no sex designated. The data is visualized in the table below:

{7}------------------------------------------------

April 2017 Spainen Grup Con VZV (gMPasilie 2017 VZV IgMegatie 2017 (gMEquical 2010) (gMExicolor 2010) (gMExicolor 2010) (gMB) 2010) (gMBC 2010) (gMB) 2010) (gMB) 2010) (gMB)
1-9Prospective Males14
Prospective Fernales3
ール、日本の中古、「アドリア」、「アイマン、アドレスト」(アイデン・ミード)」「アイバー
10-19Prospective Males110
Prospective Fernales191
:"中国"【行】【下】【下】【,】【有】【的】【不】【,】【不】【,】【
20-29Prospective Males13
Prospective Ferralesങ്കട2
Sex Unknown11 11 11 1939 14 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11
1984년 1월 1일 19일 10월 1일 11월 11일 11일 11일 10일 10월 10일 10월 10일 10월 10일 10월 10일 10월 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 10일 1
30-30Prospective Males14
Prospective Ferralesહિન્ત
Sex Unknown1
40-40Prospective Males7
Prospective Ferrales26
.
સ્ત્રન્સ્સProspective Males7
Propodite Fernales 19 Marceller Formalia 2017 Manufactures 2017 Marchiterial Comments 2017 Marine 2017 Press
റ്റാക്കുProspective Males4
Prospective Ferrales10
் பிரிக்கு இருக்கும் இருக்கும் இந்து முறைய வருகின்றன. இந்த திரைப்படுகின்றன. முறைய முறைய மாறு முறையில் பய
70+Prospective Malesਤੇ
Prospective Familes Prospective Formanານ ການປະກວດນ້າງງາມ ແລະ
Total:Prospective Males262
Prospective Ferrales2275
Sex Unknown이 대한민국 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학교 대학
Total:ి2915
tota300

Agreement Summary

The presence of VZV IgM antibody in 302 prospective samples and 36 retrospective samples was evaluated at three sites using the Zeus Scientific VZV IgM ELISA Test System and a commercially distributed VZV IgM ELISA test system.

.

{8}------------------------------------------------

Prospective Samples:Combined Sites

Commercial ELISA Results

+-+/-Totals
+64212
-281281
+/-729
Totals62924302

Positive % Agreement = 6/6 = 100% 95% Confidence Interval**= 54.1% to 100.0% Negative % Agreement = 281/294 = 95.6% 95% Confidence Interval**= 92.6% to 97.6%

us Scientific,

eus Scientific, I

Prospective and Retrospective Samples:Combined Sites

Commercial ELISA Results

+-+/-Totals
+384446
-282282
+/-17210
Totals392936338

Positive % Agreement = 38/39 = 97.4% 95% Confidence Interval**= 86.5% to 99.9% Negative % Agreement = 282/297 = 95.6% 95% Confidence Interval**= 91.8% to 97.1%

The test is for in vitro use only. The performance of this assay has not been established for neonatal, immunocomprimised populations, cord blood or pre-transplant patients. The use of whole blood or plasma is not established.

{9}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes extending from its back, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper half of the circle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUL - 5 2007

Ewa K. Nadolczak Manager Clinical Affairs ZEUS Scientific, Inc. P.O. Box 38 Raritan, New Jersey 08869

Re: K070317

Trade/Device Name: Zeus Scientific, VZV IgM ELISA Test System

Regulation Number: 21 CFR 866.3900 Regulation Name: Varicella-Zoster virus Serological Reagents Regulatory Class: Class II Product Code: LFY Dated: January 21, 2007 Received: February 6, 2007

Dear Ms. Nadolczak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other

{10}------------------------------------------------

Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sajastor

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Indications for Use

670317 510(k) Number (if known):

Device Name: Zeus Scientific Varicella Zoster Virus IgM ELISA Test System

Indications For Use:

The Zeus Scientific Varicella Zoster Virus (VZV) IgM ELISA Test System is intended for the qualitative detection of IgM class antibodies to Varicella Zoster Virus in human serum as an aid in the diagnosis of primary infection or reactivation.

The assay performance in detecting antibodies to VZV in individuals vaccinated with the FDA licensed VZV vaccine is unknown.

The user of this assay is responsible for establishing the performance characteristics with VZV vaccinated individuals.

The assay performance in detecting antibodies to VZV in cord blood and neonates has not been established.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Uwe Scheff

Page 1 of 1

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K070317

§ 866.3900 Varicella-zoster virus serological reagents.

(a)
Identification. Varicella-zoster virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to varicella-zoster in serum. The identification aids in the diagnosis of diseases caused by varicella-zoster viruses and provides epidemiological information on these diseases. Varicella (chicken pox) is a mild, highly infectious disease, chiefly of children. Zoster (shingles) is the recurrent form of the disease, occurring in adults who were previously infected with varicella-zoster viruses. Zoster is the response (characterized by a rash) of the partially immune host to a reactivation of varicella viruses present in latent form in the patient's body.(b)
Classification. Class II (performance standards).