K Number
K070046

Validate with FDA (Live)

Date Cleared
2007-02-07

(35 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powered Vinyl Patient Examination Gloves, 80 LYZ, and meets all requirements of ASTM standard D-5250-00€4.

AI/ML Overview

While the provided text describes a medical device (Powdered Vinyl Patient Examination Gloves) and its regulatory approval process, it is not an AI/ML device. Therefore, many of the requested fields related to AI/ML device testing (e.g., ground truth, MRMC study, training set) are not applicable.

Here's the information that can be extracted or deduced from the provided text, adapted for a non-AI device:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
ASTM Standard D-5250-00€4 requirements (Physical and Dimensions Testing)"All testing meets requirements for physical and dimensions testing conducted on gloves."
FDA 1000 ml. Water Fill Test (Pinhole Requirements)"The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level 1, meeting these requirements." and "meets pinhole FDA requirements"
Primary Skin Irritation"Primary Skin Irritation...testing was conducted with results showing no primary skin irritant reactions."
Skin Sensitization (Allergic Contact Dermatitis)"Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing...no sensitization reactions."
Residual Powder (for "powdered" claim)"A Residual Powder Test that based on ASTM D-6124-06 for Starch at finished inspection is conducted to insure that our gloves meet our 'powdered' claims (contain no more than 10mg/dm4)." (Implies they meet this criterion).
General Biocompatibility Requirements"meets...biocompatibility requirements" (No specific criteria detailed, but stated as met).
Labeling Claims (No special/hypoallergenic claims)"There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels." and "meets...labeling claims as shown by data in Section 7." (The criterion is that any claims made are supported, and in this case, the lack of certain claims is a key aspect).

2. Sample sized used for the test set and the data provenance

  • ASTM D-5250-00€4 (Physical and Dimensions Testing): "Inspection level S-2, AQL 4.0." (This specifies a sampling plan for quality control rather than a fixed "test set" size.)
  • FDA 1000 ml. Water Fill Test: "samplings of AQL 2.5, inspection level 1." (Again, a sampling plan.)
  • Primary Skin Irritation and Skin Sensitization: The number of subjects is not specified, but it's a "testing" conducted.
  • Residual Powder Test: Not specified beyond being "conducted."

Data Provenance: The tests were conducted by the manufacturer, Shijiazhuang Lianhe Shunda Plastic Products Co., Ltd. (China), based on specified ASTM standards and FDA requirements. The data is retrospective for the purpose of this 510(k) submission, as it describes tests already performed to demonstrate substantial equivalence.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as this is a physical medical device. The "ground truth" for its performance is determined by established physical and biological testing methods (e.g., water leak test, tensile strength tests, irritation tests) as per industry standards (ASTM) and regulatory requirements (FDA). There is no "ground truth" established by human experts in the way an AI diagnostic device would require.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Testing is based on objective measurements and established pass/fail criteria from standards. There is no human adjudication process involved in determining device performance against these physical and biological criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable, as this is a physical medical device, not an AI/ML diagnostic or assistive device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable, as this is a physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device is based on objective measurement standards defined by ASTM (e.g., D-5250-00€4 for physical properties like dimensions, tensile strength, elongation, and D-6124-06 for residual powder) and regulatory compliance tests (e.g., FDA 1000 ml Water Fill Test for pinholes, biocompatibility testing for irritation/sensitization). There isn't "ground truth" in the diagnostic sense; rather, it's about meeting pre-defined performance specifications.

8. The sample size for the training set

Not applicable, as this is not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable, as this is not an AI/ML device.

{0}------------------------------------------------

7 2007 FE3

EXHIBIT #1 Page 1 of 2

510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.

ko 70046 The assigned 510(K) number is:

1. Submitter's Identification:

Mr. Li Quanfu Shijiazhuang Lianhe Shunda Plastic Products Co., Ltd. Nanwuhui Village, Yuanshi County, Shijiazhuang, Hebei, China 051130

Date Summary Prepared: September 20, 2006

2. Name of the Device:

Shijiazhuang Lianhe Shunda Plastic Products Co., Ltd. Powdered Vinyl Patient Examination Gloves

3. Predicate Device Information:

Shijiazhuang Great Eagle Plastic Products Co., Ltd Powdered Vinyl Patient Examination Gloves (K983645)

4. Device Description:

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powered Vinyl Patient Examination Gloves, 80 LYZ, and meets all requirements of ASTM standard D-5250-00€4.

5. Intended Use:

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

{1}------------------------------------------------

EXHIBIT #1 Page 2 of 2

6. Comparison to Predicate Devices:

Shijiazhuang Lianhe Shunda Plastic Products Co., Ltd.'s Powdered Vinyl Patient Examination Gloves is substantially equivalent in safety and effectiveness to the Shijiazhuang Great Eagle Plastic Products Co., Ltd.'s Powdered Vinyl Patient Examination Gloves

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:

The standards used for Shijiazhuang Lianhe Shunda Plastic Products Co., Ltd.'s glove production are based on ASTM-D-5250-00 =4. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level 1, meeting these requirements.

Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

A Residual Powder Test that based on ASTM D-6124-06 for Starch at finished inspection is conducted to insure that our gloves meet our "powdered" claims (contain no more than 10mg/dm4).

8. Labeling:

There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.

9. Discussion of Clinical Tests Performed:

Not Applicable - There is no hypoallergenic Claim.

10. Conclusions:

Shijiazhuang Lianhe Shunda Plastic Products Co., Ltd.'s Powdered Vinyl Patient Examination Gloves conform fully to ASTM D-5250-00 €4 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

Public Health Service

7 2007

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Shijiazhuang Lianhe Shunda Plastic Products Company, Limited C/O Ms. Kathy Liu Official Correspondent Gloveco, Incorporated 3973 Schaefer Avenue Chino, California 91710 FEB

Re: K070046

Trade/Device Name: Powdered Vinyl Patient Examination Glove Regulation Number: 880.6250 Regulation Name: Patent Examination Glove Regulatory Class: I Product Code: LYZ Dated: December 28, 2006 Received: January 3, 2007

Dear Ms. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Attachment A

INDICATION FOR USE

510 (k) NUMBER (IF KNOW): 上口90046

APPLICANT: DEVICE NAME: Shijiazhuang Lianhe Shunda Plastic Products Co., Ltd. Powdered Vinyl Patient Examination Glove.

INDICATIONS FOR USE:

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter-Use X
(Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrent of CDRH, Office of Device Evaluation (ODE)

Shiela Murphy, L.C.

10000050000

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.