The Eigen DSA 2000 product is used in vascular imaging applications. During X-ray exposures, the DSA 2000 is used to acquire video images from the video display chain provided by the X-ray manufacturer's system. The images are stored in the DSA 2000 solid state memory, and written to the hard disk medium. Images are processed in real-time to provide increased image usability. The processing is primarily subtraction, but also includes window and level adjustments, as well as optional noise reduction, landscaping, and pixel shifting. The Eigen DSA 2000 device is used in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists and cardiologists, with minimal intervention required by users to perform basic capture, playback, and archiving functions.
The Eigen Digital Subtraction Angiography® (DSA) is a real-time video acquisition device that can be added to an existing standard line-rate X-ray system and provides creation of photos and real-time digital subtraction taken from a mask image. The DSA acquires and transmits data to a DICOM workstation or PAC system. The data will then be available for display. The DSA output conforms to the DICOM 3.0XA Standard for lossless images. The Eigen DSA is assembled on a Hewlett Packard (HP)/Intel platform and uses the Microsoft Windows XP® operating system.
The Eigen Digital Subtraction Angiography (DSA) 2000 is an image processing system used in vascular imaging applications. The information provided outlines its intended use and a general conclusion of its performance rather than specific quantitative acceptance criteria or a detailed study plan.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Functional Equivalence to Predicate Device (Digital 8) with modifications. | "Actual device performance as tested internally conforms to the system requirements.""The test results support the conclusion that the DSA 2000 is substantially equivalent to its predicate device, D8." |
| Maintenance of Intended Use and Fundamental Scientific Technology. | "The modifications made to the DSA 2000 do not alter the intended use or the fundamental scientific technology of the device." |
| Real-time video acquisition and image processing. | The device "acquires video images from the video display chain," "Images are processed in real-time to provide increased image usability." |
| Specific Processing Functions (Subtraction, Window/Level, Noise Reduction, Landscaping, Pixel Shifting). | "The processing is primarily subtraction, but also includes window and level adjustments, as well as optional noise reduction, landscaping, and pixel Shifting." |
| Image Storage and Archiving. | "Images are stored in the DSA 2000 solid state memory, and written to the hard disk medium.""Data Archiving" is listed as a function. |
| DICOM 3.0XA Standard Conformance for Lossless Images. | "The DSA output conforms to the DICOM 3.0XA Standard for lossless images." |
| Enhance diagnostic capabilities for radiologists and cardiologists with minimal user intervention. | "The Eigen DSA 2000 device is used in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists and cardiologists, with minimal intervention required by users to perform basic capture, playback, and archiving functions." |
2. Sample size used for the test set and the data provenance
The document states: "Testing was performed at the module and system level according to written test protocols established before the testing was conducted." However, it does not provide any specific sample size for the test set (e.g., number of images, patients, or cases). It also does not specify the data provenance (e.g., country of origin, retrospective or prospective). The testing appears to be internal (hence "tested internally").
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document mentions that "Test results were reviewed by designated technical professionals before release of the software." However, it does not specify the number of experts used or their qualifications. It also does not explicitly state that these "technical professionals" established ground truth for a test set in a clinical context; their role appears to be reviewing the internal technical test results.
4. Adjudication method for the test set
The document does not describe any adjudication method (e.g., 2+1, 3+1, none) for a clinical test set. The review mentioned seems to be technical verification of internal test results.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was mentioned in the provided text. The device is presented as a tool to enhance diagnostic capabilities, but no study comparing human readers with and without the device's assistance is described, nor is an effect size provided.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The document describes the device as an "image processing system" that "enhances diagnostic capabilities of radiologists and cardiologists." This implies it is a human-in-the-loop system. No standalone algorithm performance study is indicated. The "Test Discussion" section refers to "module and system level" testing, which likely refers to the functional performance of the software and hardware components rather than a standalone clinical performance evaluation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The document does not describe the type of ground truth used for any clinical evaluation as it focuses on functional equivalence and internal system requirements rather than a clinical performance study with established ground truth.
8. The sample size for the training set
The device described is an "image processing system" and not explicitly an AI/ML device that requires a training set in the contemporary sense. It performs fixed algorithms (e.g., subtraction, window/level, noise reduction). Therefore, no training set is mentioned or applicable in the context of this device's description.
9. How the ground truth for the training set was established
As there is no training set mentioned or applicable for this type of image processing system, the method for establishing its ground truth is not relevant and therefore not provided.
{0}------------------------------------------------
eigen
D63846
Special 510(k) Summary
| Manufacturer: | Eigen LLC13366 Grass Valley AvenueGrass Valley, CA 95945USAPhone (530) 274-1240(888) 924-2020FAX (530) 274-3656 |
|---|---|
| --------------- | ----------------------------------------------------------------------------------------------------------------------------------------- |
JAN 26 2007
- Contact Person: Ms. Natalie J. Kennel Consultant NJK & Associates, Inc. 13721 Via Tres Vista San Diego, CA 92129 Phone: (858) 705-0350 Fax: (858) 764-9739 email: Natalie.Kennel@sbcglobal.net
Date Prepared: December 21, 2006
DEVICE INFORMATION
Trade/Proprietary Name: Eigen Digital Subtraction Angiography (DSA) 2000, also referred to as the Eigen DSA 2000
| Common/Classification Name: | Image Processing System21 CFR 892.2050Class IIDevice Product Code: LLZ |
|---|---|
| ----------------------------- | ------------------------------------------------------------------------------------ |
Predicate Device:
The Eigen DSA 2000 is substantially equivalent to another Eigen Imaging Device: the Digital 8. The DSA 2000 is a modification of Eigen's Digital 8, which was cleared under 510(k) K901956. The modifications made to the DSA 2000 do not alter the intended use or the fundamental scientific technology of the device.
Both the Eigen DSA 2000 and its predicate, the Digital 8, chhance images by performing the following functions: Sales Comment
Eigen DSA 2000 - Special 510(k) December 21, 2006
Company Confidential Page 19 of 76
{1}------------------------------------------------
- Noise Reduction
- I Window/Level Feature
- Edge Enhancement 트
- 트 Image Subtraction
- I Still Image Display
- 트 X-ray Lab Command Interface
- 미 X-ray Lab Video Interface
- Data Archiving 제
- 트 Data and Space Management
- I Study Information (e.g., patient, site, and date)
- 트 Configuration Menu
- Still Image Acquisition and Display
- 로 Sequence Image Acquisition and Display, and
- Automated Contrast l
The DSA 2000 device also offers the following additional features: Landscaping, Pixel Shifting, and Modality Work-List Support which are not in the predicate device. The DSA 2000 device does not offer the following functions which were in the original device: Magnification, Panning, Mix, Quantification and test patterns because it has been determined that there is not a customer demand for these features.
Product Description:
The Eigen Digital Subtraction Angiography® (DSA) is a real-time video acquisition device that can be added to an existing standard line-rate X-ray system and provides creation of photos and real-time digital subtraction taken from a mask image. The DSA acquires and transmits data to a DICOM workstation or PAC system. The data will then be available for display. The DSA output conforms to the DICOM 3.0XA Standard for lossless images.
The Eigen DSA is assembled on a Hewlett Packard (HP)/Intel platform and uses the Microsoft Windows XP® operating system.
Intended Use:
The Eigen DSA 2000 product is used in vascular imaging applications. During X-ray exposures, the DSA 2000 is used to acquire video images from the video display chain provided by the X-ray manufacturer's system. The images are stored in the DSA 2000 solid state memory, and written to the hard disk medium. Images are processed in real-time to provide increased image usability. The processing is primarily subtraction, but also includes window and level adjustments, as well as optional noise reduction, landscaping, and pixel Shifting. The Eigen DSA 2000 device is used in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists and cardiologists, with minimal intervention required by users to perform basic capture, playback, and archiving functions.
Eigen DSA 2000 Special 510(k) December 21, 2006
Company Confidential Page 20 of 76
{2}------------------------------------------------
Test Discussion
Testing was performed at the module and system level according to written test protocols established before the testing was conducted. Test results were reviewed by designated technical professionals before release of the software.
Conclusion:
The test results support the conclusion that the DSA 2000 is substantially equivalent to its predicate device, D8. Actual device performance as tested internally conforms to the system requirements.
Eigen DSA 2000 Special 510(k) December 21, 2006
Company Confidential Page 21 of 76
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol with three curved lines, resembling a person with outstretched arms, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The text is in all caps and is oriented to follow the curve of the circle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Eiden, LLC % Ms. Natalie J. Kennel Principal NJK & Associates, Inc. 13721 Via Tres Vista SAN DIEGO CA 92129
JAN 2 6 2007
Re: K063846
Trade/Device Name: Eigen Digital Subtraction Angiography (DSA) 2000 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: II Product Code: LLZ and IZI Dated: December 21, 2006 Received: December 27, 2006
Dear Ms. Kennel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the radications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, or and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act, Tro general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulterviton.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image shows a circular logo with the text "1906-2006" at the top and "FDA Centennial" at the bottom. The letters "FDA" are prominently displayed in the center of the logo. There are three stars below the word "Centennial". The logo appears to be a commemorative emblem for the centennial anniversary of the Food and Drug Administration.
Protecting and Promoting Public Health
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely vours.
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use Statement
14 01 3846 510(k) Number:
Device Name: Eigen Digital Subtraction Angiography (DSA) 2000
Indications For Use:
The Eigen DSA 2000 product is used in vascular imaging applications. During X-ray exposures, the DSA 2000 is used to acquire video images from the video display chain provided by the X-ray manufacturer's system. The images are stored in the DSA 2000 solid state memory, and written to the hard disk medium. Images are processed in real-time to provide increased image usability. The processing is primarily subtraction, but also includes window and level adjustments, as well as optional noise reduction, landscaping, and pixel shifting. The Eigen DSA 2000 device is used in X-ray cardiology and radiology labs to enhance diagnostic capabilities of radiologists and cardiologists, with minimal intervention required by users to perform basic capture, playback, and archiving functions.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over the Counter Use
(per 21 CFR 801.109)