K Number
K063811

Validate with FDA (Live)

Date Cleared
2007-02-09

(49 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of the antimicrobial susceptibility of minute in the for Enterobacteriaceae and Non-Enterobacteriaceae Inculture analone bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.

This premarket notification is for cefazolin at concentrations of 0.5-32 µg/mL to Gram-negative ID/AST or AST only Phoenix panels with the removal of the limitations for Klebsiella oxytoca and Proteus mirabilis.

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software.
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
  • BD Phoenix AST Broth used for performing AST tests only.
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.

The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed in the instrument.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

AI/ML Overview

This document describes the validation of the BD Phoenix™ Automated Microbiology System for Cefazolin (0.5-32 µg/mL) against Gram-negative bacteria, focusing on the removal of limitations for Klebsiella oxytoca and Proteus mirabilis.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated with numerical targets for Essential Agreement (EA) and Category Agreement (CA) in the provided text. However, the study aims to demonstrate substantially equivalent performance when compared with the CLSI reference broth microdilution method, as defined in the FDA guidance "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems".

Based on common FDA guidance for AST systems, "substantial equivalence" often implies that the new system's performance (EA and CA) is non-inferior to and ideally very similar to the reference method. While specific thresholds are not provided in this excerpt, typical FDA expectations for AST devices often include:

  • Essential Agreement (EA): Generally ≥ 90%
  • Category Agreement (CA): Generally ≥ 90%
  • Minor Errors (mE): Typically ≤ 7-10%
  • Major Errors (ME): Typically ≤ 1.5-3%
  • Very Major Errors (VME): Typically ≤ 1.5-3%

The provided Table 1 is incomplete and does not contain the actual performance data for Cefazolin. It is presented as:

AntimicrobiaTHE LEART THE FREE LER Concentration197 15 Carl College011 - A0
Comments of Children and Children and Children and Sefazolıne on 118/mr145 " VallOK1 0 P . . Vall05

Therefore, without the actual numerical data from Table 1, the device's reported performance against the acceptance criteria cannot be fully extracted from the provided text. The text does state that "The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested." and "The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA guidance document". This implies the device met the unstated acceptance criteria for substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The text mentions "Clinical, stock and challenge isolates were tested." No specific total number of isolates for the overall clinical study is provided. However, Table 1 (incomplete) shows a column with "197" which might refer to the number of isolates or a subset of them.
  • Data Provenance: The data was collected from "multiple geographically diverse sites across the United States." The study included "Clinical, stock and challenge isolates." Clinical isolates are typically prospective or retrospectively collected real-world samples, while stock and challenge isolates are curated collections often used for specific performance assessments.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth for the test set was established by the CLSI (formerly NCCLS) reference broth microdilution method. This is an established standard laboratory procedure, not an opinion-based expert consensus. Therefore, the concept of "number of experts" and their "qualifications" in the traditional sense of subjective interpretation (e.g., radiologists reading images) does not directly apply here. The "experts" would be the trained laboratory personnel who meticulously performed the CLSI reference method.

4. Adjudication Method for the Test Set

Not applicable. The ground truth was established by the CLSI reference broth microdilution method, which is an objective measurement, not a subjective interpretation requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This study is an evaluation of an automated AST system against a reference method, not a multi-reader multi-case study involving human readers with/without AI assistance. The device itself is an automated system for antimicrobial susceptibility testing.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, this was a standalone performance study. The BD Phoenix™ Automated Microbiology System is an automated system, and its results were compared directly to the CLSI reference method. The device's performance is reported as its direct output (MIC values and categories S, I, R).

7. The Type of Ground Truth Used

The ground truth used was the CLSI reference broth microdilution method. This is a laboratory standard considered the gold standard for determining minimum inhibitory concentrations (MICs) and antimicrobial susceptibility categories. It represents a precise, objective, and well-established scientific method.

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning or algorithm development. The study described focuses on the validation of the device's performance against reference methods. Automated microbiology systems typically use predefined algorithms and pre-calibrated reagents based on extensive prior development and internal testing, rather than a "training set" in the typical AI sense during the final validation for substantial equivalence submission. If there was an internal development phase, the sample size for that "training" would not be disclosed in this type of regulatory summary.

9. How the Ground Truth for the Training Set was Established

As no specific "training set" is mentioned in the context of this regulatory submission, this question cannot be answered from the provided text. The performance evaluation is against the CLSI reference broth microdilution method as the ground truth.

{0}------------------------------------------------

510(K) SUMMARY

SUBMITTED BY:Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: 410-316-4938Fax: (410)-316-4499
CONTACT NAME:Janine MatlakRegulatory Affairs Specialist
DATE PREPARED:December 20, 2006
DEVICE TRADE NAME:BD Phoenix™ Automated Microbiology System -Cefazolin 0.5-32 µg/mL
DEVICE COMMON NAME:Antimicrobial susceptibility test system-short incubation
DEVICE CLASSIFICATION:Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645
PREDICATE DEVICES:VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin (K020321,May 23, 2002), Ofloxacin (K020323, April 14, 2002), andLevofloxacin (K020322, March 27, 2002).
INTENDED USE:The BD Phoenix™ Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

FEB $ 2007### DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • . BD Phoenix instrument and software.
    111

  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.

  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .

  • BD Phoenix AST Broth used for performing AST tests only. .

  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.

{1}------------------------------------------------

The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed in the instrument.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI (formerly NCCLS) reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to CLSI M7). The system has been evaluated as defined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", February 5, 2003.

Site Reproducibility

Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram-negative Phoenix panels containing this antimicrobial agent and associated reagents.

The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the gram-negative isolates tested.

BD Diagnostic Systems Becton, Dickinson and Company

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).

Table 1 summarizes the performance for the isolates tested in this study.

Table 1: Performance of BD Phoenix System for Gram-Negative Organisms by Drug

AntimicrobiaTHE LEART THE FREE LERConcentration19715Carl College011- A0
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Comments of Children and Children and Children andर्Sefazolıne on118/mr145 "VallOK1 0 P .. Vall05

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), and Cefazolin (K061867, August 10, 2006).

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the emblem.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Janine Matlak Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152

FEB $ 2007

Re: K063811 Trade/Device Name: BD Phoenix™ Automated Microbiology System Cefazolin (0.5-32 µg/mL) - Gram-Negative ID/AST or AST only with the removal of the limitations for Klebsiella oxytoca and Proteus mirabilis Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: December 20, 2006 Received: December 20, 2006

Dear Ms. Matlak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its tolli alle (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sally, anton

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number: KO6381

Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent cefazolin Device Natie. "DDT Hounds" Adoministration of AST only Phoenix panels with the removal of the limitations for Klebsiella oxytoca and Proteus mirabilis.

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of THE DD I nochia antimicrobial susceptionity of minute in the for Enterobacteriaceae and Non-Enterobacteriaceae Inculture analone bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.

This premarket notification is for cefazolin at concentrations of 0.5-32 µg/mL to Gram-negative ID/AST or I ms premarks nonicalloris to: exemoval of the limitations for Klebsiella oxytoca and Proteus AS 1 only 1 nocher panolo was active in vitro against most strains of microorganisms listed mindow, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Escherichia coli Proteus mirabilis Klebsiella species Enterobacter aerogenes

Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Freddie Poole

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

KDb3811

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”