K Number
K063655

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-02-15

(69 days)

Product Code
Regulation Number
866.5210
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Dimension Vista™ CER Flex® reagent cartridge: The CER method is an in vitro diagnostic test for the quantitative determination of Ceruloplasmin in human serum, heparinized plasma or EDTA plasma on the Dimension Vista System. Measurements of ceruloplasmin aid in the diagnosis of copper metabolism disorders.

Dimension Vista™ Protein 1 Calibrator: PROT1 CAL is an in vitro diagnostic product for the calibration of the C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Immunoglobulin A (IGA), Immunoglobulin G (IGG), Immunoglobulin M (IGM) and Prealbumin (PREALB) methods on the Dimension Vista® System.

Dimension Vista" Protein 1 Control L, M and H: PROT1 CON L, M and H are assayed intralaboratory quality controls for assessment of precision and analytical bias in the determination of C3 Complement (C3), C4 Complement (C4), ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB) on the Dimension Vista® System.

Device Description

Dimension Vista™ CER Flex® reagent cartridge: Proteins contained in human body fluids form immune complexes in an immunochemical reaction with specific antibodies. These complexes scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.

Dimension Vista™ Protein 1 Calibrator: Protein 1 Calibrator is a multi-analyte, liquid, human serum based product containing C3 complement, C4 complement, ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB).

Dimension Vista™ Protein 1 Control L, M and H: Protein 1 Control L, M and H are multi-analyte, liquid, human serum based products containing C3 complement, C4 complement, ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB).

AI/ML Overview

1. Acceptance Criteria and Reported Device Performance:

The primary acceptance criteria for the Dimension Vista™ CER assay are based on demonstratincorr correlation and equivalent performance with the predicate device (Dade Behring N Antisera to Human Ceruloplasmin assay on the BN ProSpec System). While explicit numerical thresholds for acceptance (e.g., minimum correlation coefficient, maximum acceptable slope/intercept deviation) are not provided, the reported performance is presented to demonstrate that these criteria have been met.

MetricAcceptance Criteria (Implied)Reported Device Performance
Correlation CoefficientHigh correlation with predicate system (N Antisera on BN ProSpec®)0.994
SlopeClose to 1.0 (relative to predicate system)1.009
Intercept (g/L)Close to 0.0 (relative to predicate system)0.008

2. Sample Size and Data Provenance:

  • Test Set Sample Size: 130 samples.
  • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. It states "evaluating serum and plasma samples," implying collection for the purpose of the study, which could suggest prospective, but this is not explicitly confirmed.

3. Number of Experts and Qualifications for Ground Truth:

Not applicable. This study involves a direct comparison of a new assay against a legally marketed predicate assay, not against a ground truth established by experts interpreting various forms of data (e.g., images, clinical symptoms). The "ground truth" or reference for comparison is the performance of the predicate device.

4. Adjudication Method:

Not applicable. There was no need for adjudication as the study directly compares quantitative assay results from two different systems.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done. This study is a method comparison for an in vitro diagnostic (IVD) assay, not a study evaluating human reader performance with or without AI assistance.

6. Standalone Performance Study:

Yes, a standalone performance study was conducted in the sense that the Dimension Vista™ CER assay's performance was directly measured and compared against the predicate device. This is typical for IVD device submissions, where the "standalone" performance refers to the device's output without human interpretation in the results generation process. The comparison is between the Dimension Vista™ CER assay results and the predicate assay results for the same samples.

7. Type of Ground Truth Used:

The "ground truth" for this method comparison study is the results obtained from the legally marketed predicate device, the Dade Behring N Antisera to Human Ceruloplasmin assay on the BN ProSpec System. This is a common approach for demonstrating substantial equivalence for new IVD assays.

8. Sample Size for the Training Set:

Not applicable. This is a method comparison study for an IVD device, not a machine learning model that requires a "training set." The device is a reagent cartridge and calibrator system for a laboratory instrument, not an AI algorithm learned from data.

9. How Ground Truth for Training Set was Established:

Not applicable, as there is no training set for this type of device and study.

{0}------------------------------------------------

Dade Behring Inc. Dimension Vista™ CER 510(k) Notification

FEB 1 5 2007

510(k) Summary for Dimension Vista" CER Flex reagent cartridge Dimension Vista™ Protein 1 Calibrator Dimension Vista" Protein 1 Control L, M and H

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807,92.

The assigned 510(k) number is: 1003655

  • Manufacturer's Name, Address, Telephone, and Contact Person, Date of 1. Preparation:
Manufacturer:Dade Behring Marburg GmbH
Emil-von-Behring Str. 76
35041 Marburg, Germany

Contact Information: Dade Behring Inc. P.O. Box 6101 Newark, Delaware 19714-6101 Attn: Kathleen Dray-Lyons Tel: 781-826-4551 Fax: 781-826-2497

Preparation date: December 7, 2006

  • Dimension Vista" CER Flex® reagent cartridge
    Dimension Vista™ Protein 1 Calibrator 2. Device Name: Dimension Vista" Protein 1 Control L Dimension Vista" Protein 1 Control M
    Dimension Vista" Protein 1 Control M
    Classification: Class II; Class II; Class I Product Code: DDB; JIX; JJY Panel: Immunology (82) and Clinical Chemistry (75)

3. Identification of the Legally Marketed Devices:

Dade Behring N Antisera to Human Ceruloplasmin~ K053074 Dade Behring N Protein Standard SL - K012470 Dade Behring N/T Protein Control SL - K012468

{1}------------------------------------------------

Dade Behring Inc. Dimension Vista™ CER 510(k) Notification

4. Device Descriptions:

Dimension Vista™ CER Flex® reagent cartridge

Proteins contained in human body fluids form immune complexes in an immunochemical reaction with specific antibodies. These complexes scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.

Dimension Vista™ Protein 1 Calibrator

Protein 1 Calibrator is a multi-analyte, liquid, human serum based product containing C3 complement, C4 complement, ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB).

Dimension Vista™ Protein 1 Control L, M and H

Protein 1 Control L, M and H are multi-analyte, liquid, human serum based products containing C3 complement, C4 complement, ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB).

5. Device Intended Uses:

Dimension Vista™ CER Flex® reagent cartridge:

The CER method is an in vitro diagnostic test for the quantitative determination of ceruloplasmin in human serum, heparinized plasma on the Dimension Vista System. Measurements of ceruloplasmin aid in the diagnosis of copper metabolism disorders.

Dimension Vista™ Protein 1 Calibrator:

PROT1 CAL is an in vitro diagnostic product for the calibration of the C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Immunoglobulin A (IGA), Immunoglobulin G (IGG), Immunoglobulin M (IGM) and Prealbumin (PREALB) methods
on the Dimension Vista "System.

Dimension Vista " Protein 1 Control L, M and H:

PROT1 CON L, M and H are assayed intralaboratory quality controls for assessment of precision and analytical bias in the determination of C3 complement (C3). C4 complement (C4), ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB) on the Dimension Vista "System.

6. Medical device to which equivalence is claimed and comparison information:

The Dimension Vista" CER assay, like Dade Behring N Antisera to Human Ceruloplasmin is an in vitro diagnostic test for the quantitative measurement of Ceruloplasmin in human serum and plasma.

7. Device Performance Characteristics:

The Dimension Vista" CER assay was compared to the Dade Behring N Antisera to Human Ceruloplasmin assay on the BN ProSpec System by evaluating serum and plasma samples with concentrations ranging from 0.09 g/L to 2.20 g/L. Regression analysis of these results yielded the following equation.

{2}------------------------------------------------

Comparative MethodnSlopeInterceptg/LCorrelationCoefficient
N Antisera to HumanCeruloplasmin on BN ProSpec®1301.0090.0080.994

Method Comparison Study

8. Conclusion:

These studies demonstrate correlation and equivalent performance between the Dade
Behring N Antisera to Human Ceruloplasmin assay and the Dimension Vista" CER assay

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged in a circular fashion around the eagle. The text is in all capital letters and is relatively small compared to the eagle symbol. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Dade Behring Inc. c/o Ms. Kathleen Dray-Lyons Regulatory Affairs and Compliance Manager Glasgow Site P.O. Box 6101 Newark, DE 19714-6101

FEB 1 5 2007

Re: K063655

Trade/Device Name: Dimension Vista™ Ceruloplasmin Flex reagent cartridge Dimension Vista™ Protein 1 Calibrator Dimension Vista™ Protein 1 Control L Dimension Vista™ Protein 1 Control M Dimension Vista™ Protein 1 Control H Regulation Number: 21 CFR 866.5210 Regulation Name: Ceruloplasmin Immunological Test System Regulatory Class: Class II Product Code: CHN, JIX, JJY Dated: December 7, 2006 Received: December 8, 2006

Dear Ms. Dray-Lyons:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 –

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements sset forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Robert Z. Beebech

Robert L. Becker, Jr., M.D., Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Dimension Vista™ CEF

Indications Statement

K063655

Device Name:

Dimension Vista™ CER Flex® reagent cartridge " Protein 1 Calibrator Dimension Vista" Protein 1 Calibrator Dimension Vista™ Protein 1 Control M Dimension Vista™ Protein 1 Control H

Indications for Use:

Dimension Vista™ CER Flex® reagent cartridge:

The CER method is an in vitro diagnostic test for the quantitative determination of Ceruloplasmin in human serum, heparinized plasma or EDTA plasma on the Dimension Vista System. Measurements of ceruloplasmin aid in the diagnosis of copper metabolism disorders.

Dimension Vista™ Protein 1 Calibrator

PROT1 CAL is an in vitro diagnostic product for the calibration of the C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Immunoglobulin A (IGA), Immunoglobulin G (IGG), Immunoglobulin M (IGM) and Prealbumin (PREALB) methods on the Dimension Vista® System.

Dimension Vista" Protein 1 Control L, M and H

PROT1 CON L, M and H are assayed intralaboratory quality controls for assessment of precision and analytical bias in the determination of C3 Complement (C3), C4 Complement (C4), ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB) on the Dimension Vista® System.

Prescription Use × (Per 21 CFR 801 Subpart D) Over-The-Counter-Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

ia m chan

Division Sign-Off

Page 1 of

Office of In Vitro Diagnostic Device Evaluation and Safety KOB3655

§ 866.5210 Ceruloplasmin immunological test system.

(a)
Identification. A ceruloplasmin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the ceruloplasmin (copper-transporting serum protein) in serum, other body fluids, or tissues. Measurements of ceruloplasmin aid in the diagnosis of copper metabolism disorders.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9.