K Number
K063600

Validate with FDA (Live)

Device Name
VISTA AVS
Date Cleared
2006-12-19

(15 days)

Product Code
Regulation Number
892.1540
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is intended for detection of blood flow in veins and arteries and as an aid for the diagnosis of peripheral vascular disease.

Device Description

The Vista AVSTM is a diagnostic ultrasound device used to aid the clinician in obtaining systolic pressure values at the arms, legs and feet. The device will provide a bi-directional Doppler for detecting presence or absence of blood flow, and will control the inflation and deflation of a pressure cuff under direction of the user. Two additional modalities, Pulse Cuff Recording (PCR) and Photoplethysmography (PPG), will provide additional information for the clinician. PCR will be used as a plethysmograph to obtain alternative blood flow waveforms, and the PPG will be used primarily to obtain toe pressures. The unit will calculate Ankle Brachial Index (ABI), Toe Brachial Index (TBI) or segmental values once the clinician has accepted or entered the appropriate pressure values. Waveforms, pressure values, and index results can be printed directly from the device or uploaded using AVS ReportTM report generating software and an external user supplied PC via a USB port. The unit can run from an internal power source or from an external power supply, and mounts to a rolling stand.

AI/ML Overview

The Summit Doppler Systems, Inc. Vista AVS is a diagnostic ultrasound device intended for the detection of blood flow in veins and arteries and as an aid for the diagnosis of peripheral vascular disease. It measures systolic pressure values, provides bi-directional Doppler for blood flow, and controls pressure cuff inflation/deflation. It also incorporates Pulse Cuff Recording (PCR) and Photoplethysmography (PPG) and calculates Ankle Brachial Index (ABI) or Toe Brachial Index (TBI).

Here's an analysis of the provided information regarding its acceptance criteria and supporting studies:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document (K063600) is a 510(k) premarket notification for substantial equivalence, not a detailed clinical study report with specific acceptance criteria and performance metrics for the device itself. The submission claims substantial equivalence to a predicate device (IMEXLAB 9100 Diagnostic Ultrasound Device, K973562).

Therefore, there are no explicit "acceptance criteria" or "reported device performance" metrics directly stated for the Vista AVS in this document in the typical sense of a clinical trial. Instead, the "acceptance criteria" can be inferred as demonstration of equivalence in:

Acceptance Criteria (Inferred from Substantial Equivalence Claim)Reported Device Performance (as inferred from the document)
Intended Use Equivalence: Detect blood flow and aid in peripheral vascular disease diagnosis.The Vista AVS is intended for "detection of blood flow in veins and arteries and as an aid for the diagnosis of peripheral vascular disease." This exactly matches the intended use of its predicate device, as implied by the substantial equivalence claim.
Technological Equivalence: Utilize similar ultrasound technology.The "Technologies Summary" states: "Doppler ultrasound technology is the same as substantially equivalent device shown above." This indicates the core technology, specifically Doppler ultrasound, is considered equivalent to the predicate.
Performance Equivalence: Provide comparable diagnostic information and functionality.The "Description" section details the Vista AVS's capabilities: "obtain systolic pressure values at the arms, legs and feet," "provide a bi-directional Doppler," "control the inflation and deflation of a pressure cuff," include "Pulse Cuff Recording (PCR) and Photoplethysmography (PPG)," and "calculate Ankle Brachial Index (ABI), Toe Brachial Index (TBI) or segmental values." This implies that the device performs functions comparable to established vascular ultrasound monitors.
Safety and Effectiveness: Demonstrate safety and effectiveness comparable to the predicate.The "Conclusion" states: "Based on comparisons of device features, materials, intended use and performance, and user instructions, the Vista AVS is shown to be substantially equivalent to the commercially available and legally marketed device indicated above." This is the primary claim of safety and effectiveness based on equivalence.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not describe a test set or clinical study specific to the Vista AVS for demonstrating performance against defined criteria. The submission is a 510(k) for substantial equivalence, which primarily relies on comparing the new device to a legally marketed predicate device. There is no mention of patient data, sample sizes, or data provenance (country of origin, retrospective/prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Since no specific test set or clinical study is described for the Vista AVS in this submission, there is no information about experts used to establish ground truth.

4. Adjudication Method for the Test Set

As there is no described test set, no adjudication method is mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No MRMC comparative effectiveness study is mentioned in the provided document. The submission focuses on demonstrating substantial equivalence to a predicate device, not on directly comparing human reader performance with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The Vista AVS is described as a diagnostic ultrasound device that aids clinicians. It is not an AI algorithm. Therefore, a standalone performance study in the context of an algorithm's performance without human-in-the-loop is not applicable here, and no such study is mentioned.

7. The Type of Ground Truth Used

The concept of "ground truth" as typically used in AI/diagnostic studies (e.g., pathology, outcomes data) is not applicable to this 510(k) submission. The submission is based on demonstrating substantial equivalence to a predicate device, meaning its "ground truth" for regulatory clearance is that it functions safely and effectively like the predicate device for its intended use.

8. The Sample Size for the Training Set

The Vista AVS is a physical medical device utilizing established Doppler ultrasound technology. It is not an AI system that requires a "training set" in the machine learning sense. Therefore, there is no information about a training set size.

9. How the Ground Truth for the Training Set Was Established

Since the Vista AVS is not an AI system developed with a machine learning training set, this question is not applicable.

{0}------------------------------------------------

K063600

SAFETY AND EFFECTIVENESS SUMMARY Summit Doppler Systems, Inc. Vista AVS

Name and Address:Summit Doppler Systems, Inc.4620 Technology Dr. #100Golden, CO 80403Phone: (303) 423-7572Fax: (303) 431-5994DEC 19 2006
Contact:Ken Jarrell -- President
Preparation Date:November 3, 2006
Device Name:Vista AVS
Common Name:Vascular Ultrasound Monitor
Classification:Class II per:Non-fetal, UltraSound MonitorFR Number892.1540Product CodeJAF
Indications for Use:This device is intended for detection of blood flow in veins and arteries and asan aid for the diagnosis of peripheral vascular disease.
Description:The Vista AVSTM is a diagnostic ultrasound device used to aid the clinician inobtaining systolic pressure values at the arms, legs and feet. The device willprovide a bi-directional Doppler for detecting presence or absence of bloodflow, and will control the inflation and deflation of a pressure cuff underdirection of the user. Two additional modalities, Pulse Cuff Recording (PCR)and Photoplethysmography (PPG), will provide additional information for theclinician. PCR will be used as a plethysmograph to obtain alternative bloodflow waveforms, and the PPG will be used primarily to obtain toe pressures.The unit will calculate Ankle Brachial Index (ABI), Toe Brachial Index (TBI)or segmental values once the clinician has accepted or entered the appropriatepressure values. Waveforms, pressure values, and index results can be printeddirectly from the device or uploaded using AVS ReportTM report generatingsoftware and an external user supplied PC via a USB port. The unit can runfrom an internal power source or from an external power supply, and mounts toa rolling stand.
Substantial Equivalence:The Vista AVS is substantially equivalent to the cleared device shown below.
IMEXLAB 9100 Diagnostic Ultrasound DeviceK973562, Cleared 4/15/98
Technologies Summary:Doppler ultrasound technology is the same as substantially equivalent deviceshown above.
Conclusion:Based on comparisons of device features, materials, intended use andperformance, and user instructions, the Vista AVS is shown to be substantiallyequivalent to the commercially available and legally marketed device indicatedabove.

:

A2

{1}------------------------------------------------

Attachment C – Indication for Use

: : :

()

System Indications for Use................................................C2
5.0BD Probe Indications for Use........................................C3
8.0BD Probe Indications for Use........................................C4

: :

and the comments of the comments of

100 100 100

. . . . . . .

.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Summit Doppler Systems, Inc. c/o Ms. Laura Danielson Program Manager TÜV America, Inc. 1775 Old Highway 8 NEW BRIGHTON MN 55112

DEC 1 9 2006

Re: K063600 Trade Name: Vista AVS Regulation Number: 21 CFR §892.1540 Regulation Name: Nonfetal ultrasonic monitor Regulatory Class: Class II Product Code: JAF Dated: November 22, 2006 Received: December 4, 2006

Dear Ms. Danielson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Vista AVS, as described in your premarket notification:

Transducer Model Number

5.0BD and 8.0BD MHz CW Peripheral Vascular Probes

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Laura Danielson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Ewa Czerska, M.D., at (240) 276-3666.

Sincerely yours,

David A. Lyman
for N. G. Pond

Nancy C. Brogd Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{4}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

Vista AVS with 5.0BD and 8.0BD Ultrasound Probes

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(specify)Other(specify)
Ophthalmic
Fetal
Abdominal
Intraoperative (specify)
Intraoperative Neurological
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Transesophageal
Transrectal
Transvaginal
Transurethal
Intravascular
Peripheral VascularE
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletal Superficial
Other (specify)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Previously cleared indication is K024197, issued 1/3/03

The system consists of main unit plus an 5.0BD or 8.0BD probe Additional Comments: _________________________________________________________________________________________________________________________________________________________

Only one transducer can be used with the main unit at a time.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CRF 801.109)

David A. Seppam

(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devices 510(k) Number _

C2

{5}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

5.0BD MHz CW Peripheral Vascular Probe

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(specify)Other(specify)
Ophthalmic
Fetal
Abdominal
Intraoperative (specify)
Intraoperative Neurological
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Transesophageal
Transrectal
Transvaginal
Transurethal
Intravascular
Peripheral VascularE
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletal Superficial
Other (specify)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

The above is for 5.0 MHz Bi-Directional Peripheral Vascular Probe Additional Comments: _

Previously submitted appendix E on K060410 cleared 4/7/06

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEQED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Lyman

(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number

Prescription Use (Per 21 CRF 801.109)

C3

{6}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

8.0BD MHz CW Peripheral Vascular Probe

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical Application .ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(specify)Other(specify)
Ophthalmic
Fetal
Abdominal
Intraoperative (specify)
Intraoperative Neurological
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Cardiac
Transesophageal
Transrectal
Transvaginal
Transurethal
Intravascular
Peripheral VascularE
Laparoscopic
Musculo-skeletalConventional
Musculo-skeletal Superficial
Other (specify)
N= new indication; P= previously cleared by FDA; E= added under Appendix E

The above is for 8.0 MHz Bi-Directional Peripheral Vascular Probe Additional Comments: _

Previously submitted appendix E on K060410 cleared 4/7/06

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Prescription Use (Per 21 CRF 801.109)

Denis le Sequer

(Division Sign-Off) Division of Reproductive, Abdon and Radiological Devices 510(k) Number _ C4

. 9. 9

§ 892.1540 Nonfetal ultrasonic monitor.

(a)
Identification. A nonfetal ultrasonic monitor is a device that projects a continuous high-frequency sound wave into body tissue other than a fetus to determine frequency changes (doppler shift) in the reflected wave and is intended for use in the investigation of nonfetal blood flow and other nonfetal body tissues in motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.