K Number
K063591

Validate with FDA (Live)

Date Cleared
2006-12-15

(14 days)

Product Code
Regulation Number
862.1225
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use only.
VITROS CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and urine.
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
For in vitro diagnostic use only.
VITROS Calibrator Kit 1 is used to calibrate VITROS Chemistry Systems for the quantitative measurement of BUN/UREA, Ca, CREA, GLU, LAC, LI, Mg, PHOS, SALI, THEO, and URIC.

Device Description

The VITROS CREA assay is performed using the VITROS Chemistry Products CREA Slides, and the VITROS Chemistry Products Calibrator Kit 1 on VITROS Chemistry Systems.
The VITROS CREA Slide is a multilayered, analytical element coated on a polyester support.
A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Creatinine diffuses to the reagent layer, where it is hydrolyzed to creatine in the rate-determining step. The creatine is converted to sarcosine and urea by creatine amidinohydrolase. The sarcosine, in the presence of sarcosine oxidase, is oxidized to glycine, formaldehyde, and hydrogen peroxide. The final reaction involves the peroxidase-catalyzed oxidation of a leuco dye to produce a colored product.
Following addition of the sample, the slide is incubated. During the initial reaction phase, endogenous creatine in the sample is oxidized. The resulting change in reflection density is measured at 2 time points.
Once a calibration has been performed for each slide lot. creatinine concentration in unknown samples can be determined using the softwareresident two-point rate math model and the change in reflectance calculated for each unknown test slide.
VITROS Calibrator Kit 1 contains three levels of lyophilized standards with corresponding diluents. The standards are prepared from processed bovine serum to which organic analytes, electrolytes, stabilizers, and preservatives have been added. The diluents are prepared from processed water to which inorganic salts have been added. Once reconstituted, the standards are used to calibrate VITROS Chemistry Systems for the quantitative measurement of creatinine in serum, plasma and urine.
Calibration of the VITROS CREA assay requires the use of all three of the calibrator levels (bottles 1, 2 and 3).

AI/ML Overview

The provided text describes the 510(k) summary for the VITROS Chemistry Products CREA Slides and VITROS Chemistry Products Calibrator Kit 1. It compares a modified device (VITROS CREA assay, modified) to a predicate device (VITROS CREA assay, original). The document focuses on demonstrating substantial equivalence rather than presenting an exhaustive study proving acceptance criteria.

  1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" for the modified device in terms of statistical thresholds (e.g., minimum sensitivity, specificity, or agreement percentages). Instead, it relies on demonstrating substantial equivalence to a cleared predicate device.

    The comparison table provided in the 510(k) summary (page 2, "Table 1 List of Assay Characteristics: Comparison to Predicate Device") highlights similarities and minor differences in device characteristics. The "reported device performance" is essentially that these characteristics are sufficiently similar to the predicate device.

    Device CharacteristicAcceptance Criteria (Predicate)Reported Device Performance (Modified)
    Intended UseFor in vitro diagnostic use only. VITROS CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and urine. VITROS Calibrator Kit 1 calibrate VITROS Chemistry Systems for quantitative measurement of BUN/UREA, Ca, CREA, GLU, LAC, Li, Mg, PHOS, SALI, THEO, and URIC.No change (Identical to predicate)
    Calibration traceabilityTraceable to Certified NIST (National Institute of Standards and Technology) Reference Material SRM® (Standard Reference Material) 914a.The values assigned to the VITROS Chemistry Products Calibrator Kit 1 for Creatinine are traceable to a Gas Chromatography Isotope Dilution Mass Spectrometry (GC/IDMS) method and NIST SRM®914, creatinine standard reference material. (Slight update in method description but still traceable to NIST)
    Reference Interval (Serum)Males: 0.8 – 1.5 mg/dL; Females: 0.7 – 1.2 mg/dLMales: 0.7 – 1.3 mg/dL; Females: 0.6 – 1.0 mg/dL (Slightly different ranges, but within clinical acceptability)
    Reference Interval (Urine, 24 hr)800 – 2800 mg/day (Male and Female)Males: 1000 - 2000 mg/day; Females: 800 - 1800 mg/day (More specific breakdown, still within clinical acceptability)
    Reportable range (Serum)0.05 – 14.0 mg/dLNo Change (Identical to predicate)
    Reportable range (Urine)1.05 – 346.5 mg/dL (after multiplying by a dilution factor of 21)1.2 – 346.5 mg/dL (after multiplying by a dilution factor of 21) (Slight difference in lower limit, but likely within acceptable limits for the clinical context)
    Sample typeSerum, Plasma, UrineNo change (Identical to predicate)
    Basic principleTwo-point colorimetric rateNo change (Identical to predicate)
    InstrumentationVITROS Chemistry SystemsNo change (Identical to predicate)

    The "study" described is a performance evaluation demonstrating substantial equivalence, rather than meeting pre-defined quantitative acceptance criteria in isolation. The implicit acceptance criterion is that the modified device performs similarly enough to the predicate device that it can be considered equally safe and effective.

  2. Sample size used for the test set and data provenance:
    The document states: "Equivalence was demonstrated using manufactured reagents along with quality control fluids, proficiency samples and human serum, plasma, and urine samples with measured creatinine values spanning the assay range."

    • Sample Size: The exact number of human serum, plasma, and urine samples used is not specified in the provided text.
    • Data Provenance: The document does not specify the country of origin. It indicates that the samples were "human serum, plasma, and urine samples," implying real-world clinical samples, but does not state if they were retrospective or prospective.
  3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:
    This information is not provided in the document. For an in vitro diagnostic device measuring a quantitative analyte like creatinine, the "ground truth" would typically be established by a reference method or a highly accurate laboratory assay rather than expert human interpretation of images.

  4. Adjudication method for the test set:
    This information is not applicable/provided. For a quantitative chemical assay, ground truth is typically established by laboratory reference methods, not human adjudication between experts.

  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This is not applicable. This document describes an in vitro diagnostic device for quantitative creatinine measurement, not an AI-assisted diagnostic imaging device or a device involving human readers/interpreters.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    This is not applicable in the context of an "algorithm" as typically referred to for AI/imaging. The device itself (the VITROS CREA assay on VITROS Chemistry Systems) operates in a standalone manner to measure creatinine concentrations in samples. Its performance is evaluated directly against predicate devices or reference methods, without human-in-the-loop interaction in the measurement process itself.

  7. The type of ground truth used:
    The ground truth for the calibration traceability section is stated as:

    • Gas Chromatography Isotope Dilution Mass Spectrometry (GC/IDMS) method.
    • NIST SRM®914, creatinine standard reference material.
      These are highly accurate analytical methods/standards used to establish the true concentration of creatinine values.
  8. The sample size for the training set:
    This information is not provided. The concept of a "training set" as understood in machine learning is not directly applicable here. The device is a chemical assay, and its performance is validated through analytical studies on known samples (standards, quality controls, and patient samples spanning the assay range), not by training a machine learning model.

  9. How the ground truth for the training set was established:
    As explained above, the concept of a "training set" for a chemical assay device is different from that in machine learning. The device's calibration curve is established using known calibrator values, which are traceable to GC/IDMS and NIST SRM®914 (as mentioned in point 7). This essentially acts as the "ground truth" for the device's quantitative measurements.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Ortho-Clinical Diagnostics. The logo consists of a symbol on the left, followed by the text "Ortho-Clinical Diagnostics" on the right. The text is underlined. The symbol on the left appears to be a stylized caduceus.

Johnson-fohnson company

100 Indigo Creek Drive Rochester, New York 14626-5101

DEC 1 5 2006

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: Kol.359|

  • Submitter 1. Ortho-Clinical Diagnostics, Inc. name, ad-100 Indigo Creek Drive MC00881 dress, con-Rochester, New York 14626-5101 tact Phone: (585) 453-3143 FAX: (585) 453-3368 Email: mbyrne3@ocdus.jnj.com
    Contact Person: Michael M. Byrne

    1. Preparation November 29, 2006 date
    1. Device Trade or Proprietary Name: name VITROS Chemistry Products CREA Slides VITROS Chemistry Products Calibrator Kit 1

Common Name: CREA test Classification Name: Creatinine test system (21 CFR 862.1225).

Common Name: Calibrator Kit 1 Classification Name: Calibrator (21 CFR 862.1150)

  • The VITROS CREA assay (modified device) is substantially equivalent to the 4. Predicate device VITROS CREA assay (original). This assay was originally cleared under the VITROS Chemistry Products CREA Slides and VITROS Chemistry Products Calibrator Kit 1 Premarket Notification K001310.
    Continued on next page

{1}------------------------------------------------

510(k) Summary, Continued

The VITROS CREA assay is performed using the VITROS Chemistry 5. Device description Products CREA Slides, and the VITROS Chemistry Products Calibrator Kit 1 on VITROS Chemistry Systems.

The VITROS CREA Slide is a multilayered, analytical element coated on a polyester support.

A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Creatinine diffuses to the reagent layer, where it is hydrolyzed to creatine in the rate-determining step. The creatine is converted to sarcosine and urea by creatine amidinohydrolase. The sarcosine, in the presence of sarcosine oxidase, is oxidized to glycine, formaldehyde, and hydrogen peroxide. The final reaction involves the peroxidase-catalyzed oxidation of a leuco dye to produce a colored product.

Following addition of the sample, the slide is incubated. During the initial reaction phase, endogenous creatine in the sample is oxidized. The resulting change in reflection density is measured at 2 time points.

Once a calibration has been performed for each slide lot. creatinine concentration in unknown samples can be determined using the softwareresident two-point rate math model and the change in reflectance calculated for each unknown test slide.

VITROS Calibrator Kit 1 contains three levels of lyophilized standards with corresponding diluents. The standards are prepared from processed bovine serum to which organic analytes, electrolytes, stabilizers, and preservatives have been added. The diluents are prepared from processed water to which inorganic salts have been added. Once reconstituted, the standards are used to calibrate VITROS Chemistry Systems for the quantitative measurement of creatinine in serum, plasma and urine.

Calibration of the VITROS CREA assay requires the use of all three of the calibrator levels (bottles 1, 2 and 3).

  1. Device in-For in vitro diagnostic use only. tended use

VITROS CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and urine.

For in vitro diagnostic use only.

VITROS Calibrator Kit 1 is used to calibrate VITROS Chemistry Systems for the quantitative measurement of BUN/UREA, Ca, CREA, GLU, LAC, Li, Mg, PHOS, SALI, THEO, and URIC.

Continued on next page

{2}------------------------------------------------

510(k) Summary, Continued

  • The VITROS CREA assay (modified device) is substantially equivalent to the 7. Comparison to predicate predicate, VITROS CREA assay (original), which was cleared by the FDA device (K001310)) for in vitro diagnostic use.
DeviceCharacteristicModified DeviceVITROS CREA assay(modified)Predicate DeviceVITROS CREA assay (original)
Intended UseNo changeFor in vitro diagnostic use only.VITROS CREA Slides quantitativelymeasure creatinine (CREA) concentra-tion in serum, plasma, and urine.For in vitro diagnostic use only.VITROS Calibrator Kit 1 is used tocalibrate VITROS Chemistry Systemsfor the quantitative measurement ofBUN/UREA, Ca, CREA, GLU, LAC,Li, Mg, PHOS, SALI, THEO, andURIC.
Calibration trace-abilityThe values assigned to theVITROS Chemistry Products Cali-brator Kit 1 for Creatinine aretraceable to a Gas ChromatographyIsotope Dilution Mass Spectrome-try (GC/IDMS) method1 and NISTSRM®914, creatinine standard ref-erence material.Traceable to Certified NIST (NationalInstitute of Standards and Technology)Reference Material SRM® (StandardReference Material) 914a.
Reference Interval(Serum)Males: 0.7 – 1.3 mg/dLFemales: 0.6 – 1.0 mg/dL0.8 –1.5 mg/dL0.7–1.2 mg/dL
Reference Interval(Urine, 24 hour)Males: 1000 - 2000 mg/day2Females: 800 - 1800 mg/day2800 – 2800 mg/day (Male and Female)
Reportable range
SerumNo Change0.05 – 14.0 mg/dL
Urine1.2 – 346.5 mg/dL (after multiply-ing by a dilution factor of 21)1.05 – 346.5 mg/dL (after multiplyingby a dilution factor of 21)
Sample typeNo changeSerum, Plasma, Urine
Basic principleNo changeTwo point colorimetric rate
InstrumentationNo changeVITROS Chemistry Systems
Table 1 List of Assay Characteristics: Comparison to Predicate Device
-----------------------------------------------------------------------------------

Continued on next page

{3}------------------------------------------------

510(k) Summary, Continued

The information presented in the pre-market notification demonstrates that the 8. Conclusions performance of the VITROS CREA assay (modified device) for use with human serum, plasma and urine is substantially equivalent to the cleared predicate device.

Equivalence was demonstrated using manufactured reagents along with quality control fluids, proficiency samples and human serum, plasma, and urine samples with measured creatinine values spanning the assay range.

The information presented in the premarket notification provides a reasonable assurance that the VITROS CREA assay (modified device) for use with human serum, plasma, and urine is safe and effective for the stated intended use.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Michael M. Byrne Regulatory Affairs Associate Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive MC00881 Rochester, NY 14626-5101

DEC 1 5 2006

Re: K063591

Trade/Device Name: VITROS Chemistry Products CREA Slides and VITROS Chemistry Products Calibrator Kit 1 Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: Class II Product Code: JFY, JIX Dated: November 30, 2006 Received: December 1, 2006

Dear Mr. Byrne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2 –

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Sean M. Cooper MS, Dim.

Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

{6}------------------------------------------------

Indications for Use

510(k) Number (if known):KU63591
Device Name:VITROS Chemistry Products CREA SlidesVITROS Chemistry Products Calibrator Kit 1
Indications for Use:For in vitro diagnostic use only.VITROS CREA Slides quantitatively measure creatinine(CREA) concentration in serum, plasma, and urine.
A creatinine test system is a device intended to measurecreatinine levels in plasma and urine. Creatinine measurementsare used in the diagnosis and treatment of renal diseases, inmonitoring renal dialysis, and as a calculation basis formeasuring other urine analytes.
For in vitro diagnostic use only.VITROS Calibrator Kit 1 is used to calibrate VITROSChemistry Systems for the quantitative measurement ofBUN/UREA, Ca, CREA, GLU, LAC, LI, Mg, PHOS, SALI,THEO, and URIC.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

063591

Page 1 of 1

§ 862.1225 Creatinine test system.

(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.