K Number
K063533

Validate with FDA (Live)

Device Name
BIPAP SYNCHRONY
Manufacturer
Date Cleared
2007-02-20

(90 days)

Product Code
Regulation Number
868.5895
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Respironics BiPAP Synchrony 2 is intended to provide non-invasive ventilation for pediatric patients 7 years or older (> 40 lbs) and adult patients (> 66 Ibs) with respiratory insufficiency or obstructive sleep apnea, in the hospital or home.

Device Description

The Respironics BiPAP Synchrony 2 Ventilatory Support System is a microprocessor controlled blower based Bi-level positive pressure system that delivers two positive pressure levels (IPAP/EPAP). The dual pressure levels provide a more natural means of delivering pressure support therapy to the patient resulting in improved patient comfort. A flow sensor and redundant pressure sensors in the patient airway feed data on measured flow and pressure into a microprocessor controller, which in turn regulates the blower assembly. A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.

The BiPAP Synchrony 2 also has a CPAP mode in which a fixed pressure is delivered and four bilevel operating modes which determine how the changes between IPAP and EPAP pressures are made, Spontaneous, Spontaneous/Timed, Timed and Pressure Control.

The BiPAP Synchrony 2 is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.

The BiPAP Synchrony 2 Ventilatory Support System is intended for use with a patient circuit that is used to connect the device to the patient interface device (mask). A typical patient circuit consists of a six-foot disposable or reusable smooth lumen 22mm tubing, a method of venting exhaled gases, and the Mirage Kidsta mask (K060105).

AI/ML Overview

Acceptance Criteria and Performance Study for BiPAP Synchrony 2

The provided document describes the predicate device comparison and bench testing performed for the BiPAP Synchrony 2 Ventilatory Support System, particularly for its expanded indication for pediatric patients. Since this is an abbreviated 510(k) submission, the full details of the specific acceptance criteria tables and complete study reports are not always provided in this summary. However, based on the text, we can infer the type of criteria and the nature of the study.

Note: The document explicitly states that the device is a "Non-Continuous ventilator" and "non-life supporting." This context is important as the rigor of testing and acceptance criteria can differ significantly from life-supporting devices.


1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of the submission focuses on demonstrating equivalence to predicate devices and expanding a pediatric indication primarily through bench testing, the acceptance criteria are generally established against the performance of the predicate devices or relevant industry standards for ventilator performance.

Acceptance Criteria CategoryDescription of Acceptance Criteria (Inferred from text)Reported Device Performance (Inferred from text)
Equivalence to Predicate DevicesPerformance of BiPAP Synchrony 2 should be equivalent to or better than the predicate devices (BiPAP Synchrony HC, S/T, Harmony, and Resmed VPAP III ST-A / Kidsta system) for the intended use, operating principle, and technology.Confirmed: "The BiPAP Synchrony 2/ Kidsta system has the following similarities to the previously cleared predicate device: Same intended use. Same operating principle. Same technology." "Bench testing has confirmed that the BiPAP Synchrony 2 performs equivalently to the device predicate VPAP III ST-A (K060105)."
Pediatric Use (Safety & Efficacy)Demonstrated safety and efficacy for pediatric patients (7 years and older > 40 lbs) without introducing new risks compared to adult use.Confirmed: "Respironics has performed a risk analysis to identify the additional considerations of using the BiPAP Synchrony 2 / Kidsta system on pediatric patients (> 7 years and > 40 lbs). Additional testing... was performed to ensure the safety and efficacy..." "We conclude that the existing and cleared adult indications for use can be safety and effectively applied to pediatric patients (> 7 years and > 40 lbs)."
Pressure PerformanceDelivery of specified and consistent pressure levels (IPAP/EPAP), and CPAP modes within acceptable limits.Confirmed: "This testing included, pressure performance... All tests were verified to meet the required acceptance criteria."
Waveform PerformanceAccurate and stable delivery of pressure waveforms.Confirmed: "This testing included... waveform performance... All tests were verified to meet the required acceptance criteria."
Triggering & CyclingCorrect and responsive triggering (initiation of breath) and cycling (termination of breath) mechanisms.Confirmed: "This testing included... triggering, cycling... All tests were verified to meet the required acceptance criteria."
Alarm FunctionalityProper function of both pre-set and user-adjustable alarms.Confirmed: "This testing included... alarm functionality testing. All tests were verified to meet the required acceptance criteria."
Kidsta Mask CompatibilityDemonstrated compatibility and performance when used with the Mirage Kidsta mask.Confirmed: "Additionally bench testing was performed to demonstrate compatibility of the Kidsta mask (K060105) with the BiPAP Synchrony."

2. Sample Size for the Test Set and Data Provenance

The study described primarily involves bench testing.

  • Sample Size: The document does not specify the exact number of units or test cycles used in the bench testing. It refers to "additional testing" and "bench testing" which implies a sufficient number of tests were conducted to cover the various performance parameters.
  • Data Provenance: The testing was conducted by Respironics, Inc., presumably at their facilities. The data is thus prospective, generated specifically for this submission. The country of origin of the data is the United States.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

N/A. This was not a clinical study involving human patients or ground truth established by medical experts in the context typically seen for diagnostic AI/ML devices. The "ground truth" for this device's performance was established through engineering specifications, regulatory standards, and comparison to predicate device performance, which would have been validated by Respironics' internal engineering and quality assurance teams.


4. Adjudication Method for the Test Set

N/A. As this was primarily bench testing against engineering specifications and predicate device performance, there was no adjudication method in the traditional sense involving multiple human readers/experts. Performance verification was likely based on measured values falling within predefined tolerance bands relative to specifications or predicate device performance.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. An MRMC comparative effectiveness study was not performed. This type of study is typically conducted for diagnostic devices where human readers interpret results, and the AI system aims to assist or improve their performance. The BiPAP Synchrony 2 is a therapeutic device (ventilator) and the submission focuses on its mechanical performance and safety.


6. Standalone (Algorithm only without human-in-the-loop performance) Study

N/A. This device is not an algorithm-only device. It is a physical medical device (ventilator) with integrated software/firmware for controlling its mechanical functions. The performance described is inherently that of the device operating as intended, without a direct "human-in-the-loop" evaluative component in the same way an AI diagnostic tool would be assessed.


7. Type of Ground Truth Used

The ground truth used for this submission's testing was:

  • Engineering Specifications: Performance parameters (pressure, waveform, triggering, cycling, alarms) were likely measured against predefined internal engineering specifications and relevant industry standards for non-invasive ventilators.
  • Predicate Device Performance: Performance data from the legally marketed predicate devices (Respironics BiPAP Synchrony HC, S/T, Harmony, and Resmed VPAP III ST-A / Kidsta system) served as a benchmark for demonstrating equivalence.
  • Risk Analysis Outcomes: The pediatric indication added considerations from a risk analysis, which would have informed specific test requirements to ensure safety and efficacy in that population based on clinical understanding.

8. Sample Size for the Training Set

N/A. This device is not an AI/ML device in the sense that it requires a "training set" for an algorithm. It is a microprocessor-controlled mechanical device. Its control algorithms are developed through engineering design, not machine learning from a data set.


9. How the Ground Truth for the Training Set Was Established

N/A. See point 8.

{0}------------------------------------------------

Abbreviated 510(k) Tab 5 – 510(k) summary

4063533

BiPAP Synchrony 2

TAB 5

510(K) SU

Date of Submission20 November 2006
Official ContactZita A. YurkoManager, Regulatory AffairsRespironics, Inc.1001 Murry Ridge LaneMurrysville, PA 15668Zita.yurko@respironics.com724-387-4120 t724-882-4120 c724-387-4216 f
Classification Reference21 CFR 868.5895
Product CodeMNS - Non-Continuous ventilator
Common/Usual NameVentilator, continuous, non-life supporting
Proprietary NameRespironics BiPAP Synchrony 2 Ventilatory Support System
Predicate Device(s)Respironics BiPAP Synchrony HC (K992530)Respironics BiPAP Synchrony S/T (K012323)Respironics BiPAP Harmony (K031656)Resmed VPAP III ST-A / Kidsta system (K060105)
Reason for submissionInclude pediatric (7 years and older > 40 lbs) indication

Substantial Equivalence

The BiPAP Synchrony 2/ Kidsta system has the following similarities to the previously cleared predicate device:

  • ロ Same intended use.
  • Same operating principle.
  • Same technology.

{1}------------------------------------------------

Tab 5 – 510(k) summary

D Same manufacturing process.

The BiPAP Synchrony was cleared in K992530/K012323. The BIPAP Harmony was cleared in K031656 using the BiPAP Synchrony as the device predicate. The Mirage Kidsta was cleared in K060105. Respironics has performed a risk analysis to identify the additional considerations of using the BiPAP Synchrony 2 / Kidsta system on pediatric patients (> 7 years and > 40 lbs). Additional testing based on the BiPAP Synchrony 2 / Kidsta risk analysis was performed to ensure the safety and efficacy of the BiPAP Synchrony 2/ Kidsta system when used on the pediatric patient (> 7 years and > 40 lbs). Bench testing has confirmed that the BiPAP Synchrony 2 performs equivalently to the device predicate VPAP III ST-A (K060105). Additionally bench testing was performed to demonstrate compatibility of the Kidsta mask (K060105) with the BiPAP Synchrony. This testing included, pressure performance, waveform performance, triggering, cycling and alarm functionality testing. All tests were verified to meet the required acceptance criteria.

Respironics has followed the FDA's Guidance for Industry and FDA Staff document "pre-market assessment of pediatric medical devices" and applied the principle of FDA's Least Burdensome Approach to demonstrate the Substantial Equivalence of the BiPAP Synchrony 2/ Kidsta mask system to its predicate devices (Respironics BiPAP Synchrony, BiPAP Harmony and the VPAP ST-A/Kidsta mask system). We conclude that the existing and cleared adult indications for use can be safety and effectively applied to pediatric patients (> 7 years and > 40 lbs).

Intended Use

The Respironics BiPAP Synchrony 2 is intended to provide non-invasive ventilation for pediatric patients 7 years or older (> 40 lbs) with respiratory insufficiency or obstructive sleep apnea, in the hospital or home.

Device Description

The Respironics BiPAP Synchrony 2 Ventilatory Support System is a microprocessor controlled blower based Bi-level positive pressure system that delivers two positive pressure levels (IPAP/EPAP). The dual pressure levels provide a more natural means of delivering pressure support therapy to the patient resulting in improved patient comfort. A flow sensor and redundant pressure sensors in the patient airway feed data on measured flow and pressure into a microprocessor controller, which in turn regulates the blower assembly. A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.

{2}------------------------------------------------

The BiPAP Synchrony 2 also has a CPAP mode in which a fixed pressure is delivered and four bilevel operating modes which determine how the changes between IPAP and EPAP pressures are made, Spontaneous, Spontaneous/Timed, Timed and Pressure Control.

The BiPAP Synchrony 2 is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.

The BiPAP Synchrony 2 Ventilatory Support System is intended for use with a patient circuit that is used to connect the device to the patient interface device (mask). A typical patient circuit consists of a six-foot disposable or reusable smooth lumen 22mm tubing, a method of venting exhaled gases, and the Mirage Kidsta mask (K060105).

(End of Tab.)

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Respironics, Incorporated Ms. Zita A. Yurko Manager, Regulatory Affairs Sleep & Home Respiratory Group 1001 Murry Ridge Lane Murrysville, Pennsylvania 15668

FEB 2 0 2007

Re: K063533

Trade/Device Name: BiPAP Synchrony 2 Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: MNS Dated: November 20, 2006 Received: November 22, 2006

Dear Ms. Yurko:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Yurko

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Chris

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): __ ______

Device Name: _________________________________________________________________________________________________________________________________________________________________

The Respironics BiPAP Synchrony 2 is intended to provide non-invasive ventilation for pediatric patients 7 years or older (> 40 lbs) and adult patients (> 66 Ibs) with respiratory insufficiency or obstructive sleep apnea, in the hospital or home.

Prescription Use ____________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mh. A. Anker

sthesiology, General Hos

S0001 Number K063533

040019

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).