K Number
K063487

Validate with FDA (Live)

Date Cleared
2007-01-22

(66 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Merete BLP™ Small Fragment Locking Bone Plate System is used for adult and pediatric patients as indicated for small bone fracture fixation. Indications for use include fixation of fractures, osteotomies, non unions of the clavicle, scapula. olecranon, radius, ulner, fibula, metacarpals, metatarsals, Hallux Valgus osteotomy corrections, middle hand and middle foot bones, particular in osteopenic bone.

Device Description

The Merete BLP™ Small Fragment Locking Bone Plate System consists of anatomically shaped U-oblique plates, right or left in the length of 33, 35 and 37 mm and 3.0 mm locking screws. The system is available in titanium (ASTM F-136), Locking plates/screws incorporate a screw-to-plate locking feature which creates a locked, fixed angle construction to hold fracture or osteotomy reduction.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (Merete BLP™ Small Fragment Locking Bone Plate System), not an AI/ML device. Therefore, the requested information regarding acceptance criteria, study details, ground truth, and AI/ML specific evaluations (like MRMC studies, standalone performance, training sets) is not applicable or present in the provided document.

A 510(k) submission demonstrates substantial equivalence to a legally marketed predicate device, rather than proving performance against specific acceptance criteria for AI algorithms. The document confirms that the device is "substantially equivalent" to predicate devices, meaning it has similar indications for use, materials, and technological characteristics.

Therefore, I cannot populate the requested tables and sections as the information is not relevant to this type of regulatory submission.

{0}------------------------------------------------

Summary of Safety and Effectiveness 3

K063487

Submitted by:Merete Medical GmbHAlt Lankwitz 102,12247 Berlin, Germany
----------------------------------------------------------------------------------

JAN 2 2 7007

FDA Registration Number: 3002949614

  • Contact Person: Donna Coleman Merete Medical, Inc. 49 Purchase Street Rye, New York 10580 Phone: 914 967 1532
    Merete BLP™ Small Fragment Locking Bone Plate Device Name:

CFR 888.3030 Single/multiple component Device Classification: 21 Metallic bone fixation appliances and accessories and 888.3040 Smooth /threaded metallic bone fixation fastener.

  • Product Code: KTT
    Proposed Regulatory Class: Class II

Predicate Device:

  • Synthes Small Fragment Locking Compression Plate (LCP) K000684 -
  • Smith & Nephew Locking Bone Plate System K033669 -
  • Merete MetaFix™ Small Fragment Locking Bone Plate K050457 -

Device Description:

The Merete BLP™ Small Fragment Locking Bone Plate System consists of anatomically shaped U-oblique plates, right or left in the length of 33, 35 and 37 mm and 3.0 mm locking screws. The system is available in titanium (ASTM F-136), Locking plates/screws incorporate a screw-to-plate locking feature which creates a locked, fixed angle construction to hold fracture or osteotomy reduction.

Intended use:

The Merete BLP™ Small Fragment Locking Bone Plate System is used for adult and pediatric patients as indicated for small bone fracture fixation. Indications for use include fixation of fractures, osteotomies, non unions of the clavicle, scapula. olecranon, radius, ulner, fibula, metacarpals, metatarsals, Hallux Valgus osteotomy corrections, middle hand and middle foot bones, particular in osteopenic bone.

{1}------------------------------------------------

Technological Characteristics:

The components of the Merete BLP™ Small Fragment Locking Bone Plate System are similar to legally marketed predicate devices listed above in that they share similar indications for use, are manufactured from similar materials and incorporate similar technological characteristics.

Potential Risks:

The risks associated with this device are the same as with any metallic internal fixation device. These include but not limited to the following: Delayed or nonunion which may lead to breakage the implant. Bending or fracture of the implant. Metal sensitivity, or allergic reaction to a foreign body. Pain, discomfort, or abnormal sensation due to the presence of the device.

র্ব Standards

The Merete BLP™ Small Fragment Locking Bone Plate system is produced from titanium alloy Ti-6Al-4V according to ASTM F-136 and ISO 5832/3.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, clutching a caduceus in its talons. The eagle is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Merete Medical, Inc. % Ms. Donna Coleman 49 Purchase Street Rye, New York 10580

JAN 2 2 2007

Re: K063487

Trade/Device Name: Mecrete BLP"" Small Fragment Locking Bone Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: November 15, 2006 Received: November 17, 2006

Dear Ms. Coleman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 -- Ms. Donna Coleman

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0210 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

for

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

cc: HFZ-401 DMC

{4}------------------------------------------------

Indications for Use 2

Indications for Use

510(k) Number (if known):

Device Name: Merete BLP™ Small Fragment Locking Bone Plate System

Indications for Use:

The Merete BLP™ Small Fragment Locking Bone Plate System is used for adult and pediatric patients as indicated for small bone fracture fixation. Indications for use include fixation of fractures, osteotomies, non unions of the clavicle, scapula, olecranon, radius, ulner, fibula, metacarpals, metatarsals, Hallux Valgus osteotomy corrections, middle hand and middle foot bones, particular in osteopenic bone.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) n

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
Merete Medical GmbH510(November 2006)
Page 5 of 22

N/A