K Number
K063486

Validate with FDA (Live)

Date Cleared
2007-03-07

(110 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic gram-negative and gram-positive bacteria of human origin.

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.

This premarket notification is for the addition of the antimicrobial agent Minocycline at concentrations of 1-32 µg/mL to Gram-negative ID/AST or AST only Phoenix panels. Minocycline has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Acinetobacter species
Enterobacter aerogenes
Escherichia coli
Klebsiella species
Shigella species

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software. .
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for . AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
  • BD Phoenix AST Broth used for performing AST tests only. .
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of twofold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

AI/ML Overview

Here's an analysis of the provided text, outlining the acceptance criteria and study details for the BD Phoenix™ Automated Microbiology System with Minocycline:

Acceptance Criteria and Device Performance

The core performance metrics for this device are Essential Agreement (EA) and Category Agreement (CA) when compared to a CLSI reference broth microdilution method.

Acceptance Criteria (Not explicitly stated as percentage, but implied by typical FDA guidance)Reported Device Performance (Table 1 - Minocycline for gram-negative organisms)
Essential Agreement (EA): Agree exactly or within ± one two-fold dilution to the reference result. (Typically >90% for new AST devices)The provided Table 1 is incomplete and corrupted, making it impossible to extract the exact Essential Agreement percentage for Minocycline.
Category Agreement (CA): Agree with the reference method with respect to FDA categorical interpretive criteria (susceptible, intermediate, and resistant). (Typically >90% for new AST devices)The provided Table 1 is incomplete and corrupted, making it impossible to extract the exact Category Agreement percentage for Minocycline.

Note: The provided "Table 1" is severely corrupted and unreadable. Therefore, the specific reported device performance percentages for Essential Agreement and Category Agreement for Minocycline cannot be extracted from this document. The document states, "Table 1 summarizes the performance for the isolates tested in this study." and implies this table would contain the numerical results.

Study Details

  1. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: Not explicitly stated as a single number. The study tested "Clinical, stock and challenge isolates."
    • Data Provenance: "multiple geographically diverse sites across the United States." The study included both "Clinical, stock and challenge isolates." Clinical isolates were compared to the CLSI reference method, and challenge isolates were compared to expected results.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    • Not specified. The ground truth method is the CLSI reference broth microdilution method, which is a standardized laboratory procedure, not typically established by individual "experts" in the same way as, for example, image interpretation. For challenge isolates, "expected results" were used, but the method of establishing these expected results or by whom is not detailed.
  3. Adjudication Method for the Test Set:

    • Not applicable in the context of this type of study. The comparison is between the automated system's results and the standardized CLSI reference method. Discrepancies would be analyzed, but a multiple-expert adjudication process isn't described for determining the "true" MIC or category.
  4. If a Multi Reader Multi Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is an automated system for antimicrobial susceptibility testing, not an AI-assisted diagnostic imaging tool for human readers. It automates a laboratory process.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, this was a standalone performance study. The BD Phoenix™ system is an "Automated Microbiology System" that performs ID and AST. Its performance was compared directly to a "CLSI reference broth microdilution method," indicating an algorithm-only evaluation against a gold standard.
  6. The type of ground truth used:

    • CLSI reference broth microdilution method for clinical isolates.
    • Expected results for challenge isolates. This implies a predetermined "true" susceptibility profile for the challenge strains, likely established through extensive prior testing.
  7. The sample size for the training set:

    • Not applicable/mentioned. As this is an older device (510(k) in 2007) for an automated microbiology system, the term "training set" in the context of machine learning isn't directly relevant to its development and validation as described. The system's rules for interpreting growth and MICs are likely based on established microbiological principles, not trained on a distinct "training set" of data in the modern sense of AI algorithms.
  8. How the ground truth for the training set was established:

    • Not applicable. The primary "ground truth" reference method for validation is the CLSI reference broth microdilution, which serves as the comparative standard, not a "training set" ground truth. The system's internal logic and algorithms would have been developed based on scientific understanding of microbial growth and susceptibility, rather than a data-driven training process with externally established ground truth labels.

{0}------------------------------------------------

510(K) SUMMARY

SUBMITTED BY:Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: (410)-316-4287Fax: (410)-316-4499
CONTACT NAME:Monica GiguereRegulatory Affairs SpecialistMAR 07 2007
DATE PREPARED:February 12, 2007
DEVICE TRADE NAME:BD Phoenix™ Automated Microbiology System –Minocycline 1-32 µg/mL
DEVICE COMMON NAME:Antimicrobial susceptibility test system-short incubation
DEVICE CLASSIFICATION:Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645
PREDICATE DEVICES:VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin (K020321,May 23, 2002, and K060324, May 25, 2006), Ofloxacin (K020323,April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
INTENDED USE:The BD Phoenix™ Automated Microbiology System is intendedfor the rapid identification and in vitro antimicrobial susceptibilitytesting of isolates from pure culture of most aerobic and facultativeanaerobic gram-negative and gram-positive bacteria of humanorigin.

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software. .
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for . AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
  • BD Phoenix AST Broth used for performing AST tests only. .
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.

{1}------------------------------------------------

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of twofold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document. "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.

Site Reproducibility

Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of gram-negative Phoenix panels containing this antimicrobial agent and associated reagents.

The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the gram-negative isolates tested.

BD Diagnostic Systems Becton, Dickinson and Company

Page 6

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).

Table 1 summarizes the performance for the isolates tested in this study.

Performance of BD Phoenix System for Gram-Negative Organisms by Drug Table 1:

THE REPORTERIAL FOR IN AN ANDTHE LA LA CONSUL BOOK CHANAL CO WEEK SHOW SHOW SHOW FLAND LEASEAntimicrobialand and the first for the first from the many of the first for the first for the first andFree of Canadian In and CalleryConcentrationSeptember 1980 - And ArticleComments of Children a more the couple the couple of concess concessionיהוnPrint10/11.00Comments of the comments of the survey a response of the later in the larger of the larges of the larges of the larges of the larges of the larges of the larges of the lands-A0114C
-I . I . LEBERT SENSION OF A NO AND IN WANTED OF A NOT A WARRENT OND ANY AND THE WATERMinocycline------------------------------------------------------------------------------------------------------------------------------------------------------------------------------(ug/mCarrely Controller Charles And All A000200 -01 0Carrelation Collection Collection Collection Comments and1C/ Hol

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA', February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002, and K060324, May 25, 2006), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Monica Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152

MAR 0 7 2007

K063486 Trade/Device Name: BD Phoenix TM Automated Microbiology System Minocycline (1-32 ug/mL) - Gram-Negative ID/AST or AST only Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: November 15, 2006 Received: November 17, 2006

Dear Ms. Giguere:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sally, attorn

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number: KO63486

Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Minocycline (1-32 ug/mL) - Gram-negative ID/AST or AST only Phoenix panels.

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.

This premarket notification is for the addition of the antimicrobial agent Minocycline at concentrations of 1-32 µg/mL to Gram-negative ID/AST or AST only Phoenix panels. Minocycline has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Acinetobacter species Enterobacter aerogenes Escherichia coli

Klebsiella species Shigella species

Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Luddie Toole
Division Sign-Off

Sion Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)_K063486

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”