K Number
K063444

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-03-20

(125 days)

Product Code
Regulation Number
872.3200
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Total-Etch Dualbond is indicated for:

  • Bonding resin-based materials (especially dual-cure composite / compomer . materials) to tooth structure (dentin and enamel)
  • Treatment of hypersensitive teeth .

Total-Etch Bond is indicated for:

  • Bonding resin-based materials (especially light-cure composite / compomer . materials) to tooth structure (dentin and enamel)
  • Treatment of hypersensitive teeth .
Device Description

The Total Etch Bonding Agent consists of two dental bonding agent formulations for use in dual- and light-cured bonding procedures:

Total-Etch Bonding Agent for Dual-Cured Resin-Based Materials (Total-Etch Dualbond) Total-Etch Bonding Agent for Light-Cured Resin-Based Materials (Total-Etch Bond)

AI/ML Overview

This document is a 510(k) summary for a dental bonding agent, not a study report. Therefore, it does not contain the detailed information requested regarding acceptance criteria and a study proving device performance in the context of AI/device performance metrics.

Here's why and what information can be extracted:

The document is a submission to the FDA for market clearance, asserting "substantial equivalence" to existing predicate devices. It focuses on chemical composition and physical performance data to establish this equivalence, rather than a clinical trial demonstrating specific performance metrics against pre-defined acceptance criteria in the way a diagnostic algorithm or device might.

However, I can provide a structure based on what you would expect to see in such a study, and indicate what is missing from this document.


Acceptance Criteria and Study for Total Etch Bonding Agent

Missing Information: The provided 510(k) summary does not contain the detailed acceptance criteria or a comprehensive study plan/results in the format requested for evaluating device performance against such criteria. The document states "This submission contains safety and performance information (chemical composition and physical performance data) sufficient to establish substantial equivalence in comparison to the devices listed above," but it does not elaborate on specific performance metrics, and certainly not in the context of AI or human reader improvement.

Therefore, the following tables and sections will indicate where the requested information would typically be if this were a performance study report, and highlight that this specific document does not provide it.


1. Table of Acceptance Criteria and Reported Device Performance

As this document is a 510(k) submission focused on "substantial equivalence" rather than a clinical or performance study report with explicit acceptance criteria for this specific device's novel performance, it does not provide a table like the one requested below. Performance is implied through comparison to predicates regarding chemical composition and physical properties.

Performance Metric (Hypothetical, for a dental bonding agent)Acceptance Criteria (Hypothetical)Reported Device Performance (Not provided in this document)
Bond Strength (e.g., Shear Bond Strength to Enamel)≥ X MPa (e.g., comparable to predicate devices)(Not reported)
Bond Strength (e.g., Shear Bond Strength to Dentin)≥ Y MPa (e.g., comparable to predicate devices)(Not reported)
Microleakage (e.g., Dye penetration score)≤ Z (e.g., comparable to predicate devices)(Not reported)
Working TimeA ± B seconds(Not reported)
Curing TimeC ± D seconds(Not reported)
Cytotoxicity (e.g., ISO 10993)Non-cytotoxic(Implied to be non-cytotoxic based on substantial equivalence)

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size for Test Set: Not specified. The document refers to "physical performance data" but does not detail the size of any test cohorts for specific performance metrics.
  • Data Provenance: Not specified. Given it's a dental product, any physical performance data would likely involve in-vitro studies conducted in a lab setting, possibly with extracted human teeth or artificial substrates. The country of origin for such data is not mentioned.
  • Retrospective or Prospective: Not applicable in the context of this 510(k) document, as it doesn't describe a clinical trial or a dataset analysis in the way these terms are typically used for AI or diagnostic device studies. Any "performance data" would likely be from lab testing (in-vitro), not human subject trials.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Not Applicable. This document describes a dental bonding agent, not a diagnostic device that requires expert interpretation for ground truth establishment. "Ground truth" in this context would refer to objective physical or chemical measurements (e.g., bond strength measured by a machine, microleakage visualized by microscopy).

4. Adjudication Method for the Test Set

  • Not Applicable. As there are no human experts establishing a "ground truth" through interpretation, no adjudication method would be required.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI vs. Without AI Assistance

  • Not Applicable. This device is a dental bonding agent, not an AI-powered diagnostic tool. Therefore, an MRMC study and analysis of AI assistance is completely irrelevant to this submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not Applicable. This device is a dental bonding agent, not an algorithm.

7. The Type of Ground Truth Used

  • For physical and chemical properties: The "ground truth" would be established through objective, standardized laboratory measurements (e.g., spectrophotometry for chemical composition, tensile/shear testing machines for bond strength, microscopy for microleakage).
  • For biocompatibility: "Ground truth" would be established through in-vitro cytotoxicity assays and potentially in-vivo animal studies if full biocompatibility testing was required beyond similarity to predicate devices.
  • The document implies that these types of tests were conducted to demonstrate substantial equivalence, but specific reports are not provided.

8. The Sample Size for the Training Set

  • Not Applicable. This device is not an AI algorithm, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. There is no training set for an AI algorithm in this context.

{0}------------------------------------------------

510(k) Summary

Trade Name:Total Etch Bonding Agent (Total-Etch Bond & Total-Etch Dual-Bond)
Sponsor:DMG USA, Inc.23 Frank Mossberg DriveAttleboro, MA 02703Registration # not yet assignedOwner/Operator No. 9005969
Device Generic Name:Dental Bonding Agent
Classification:According to Section 513 of the Federal Food, Drug, andCosmetic Act, the device classification is Class II.

Product Description:

The Total Etch Bonding Agent consists of two dental bonding agent formulations for use in dual- and light-cured bonding procedures:

Total-Etch Bonding Agent for Dual-Cured Resin-Based Materials (Total-Etch Dualbond) Total-Etch Bonding Agent for Light-Cured Resin-Based Materials (Total-Etch Bond)

Indications for Use:

Total-Etch Dualbond is indicated for

  • Bonding resin-based materials (especially dual cure composite, compomer . materials) to tooth structure (dentin and enamel)
  • . Treatment of hypersensitive teeth

Total-Etch Bond is indicated for

  • Bonding resin-based materials (especially light cure composite, compomer . materials) to tooth structure (dentin and enamel)
  • . Treatment of hypersensitive teeth

Predicate Devices:

The components of the proposed Post and Core Kit materials are substantially equivalent to several currently marketed dental restorative materials including the following:

Product NamePredicates
Excite DSCK003293 (Ivoclar North America Inc.)
Optibond Solo Plus (Dual-cure)Unknown (Kerr Mfg. Co.)
All Bond 2K910628 (Bisco)
Scotchbond MP PlusK942493 (3M Espe)

Total-Etch Dualbond:

{1}------------------------------------------------

Total-Etch Bond:

Product NamePredicate
Contax bonding agent contained inthe DMG Composite / CompomerRepair KitK041960 (DMG - USA)

Safety and Performance:

This submission contains safety and performance information (chemical composition and physical performance data) sufficient to establish substantial equivalence in comparison to the devices listed above.

Conclusion:

Based on the indications for use, technological characteristics, and comparison to the predicate devices, the Total Etch Bond has been shown to be safe and effective for its intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DMG USA, Incorporated C/O Ms. Pamela Papineau Regulatory Affairs Consultant Delphi Medical Device Consulting 5 Whitcomb Avenue Ayer, Massachusetts 01432

MAR 2 0 2007

Re: K063444

Trade/Device Name: Total Etch Bonding Agent (Total Etch Bond & Total Etch Dual Bond) Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: March 01, 2007 Received: March 07, 2007

Dear Ms. Papineau:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 – Ms. Pamela Papineau

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Clair S. Lizio, Ph.D.

Chiu S. Lin. PhD Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page _ 1 of

510(k) Number (if known): K063444

Device Name: Total Etch Bonding Agent (Total Etch Bond & Total Etch Dual Bond)

Indications for Use:

Total-Etch Dualbond is indicated for:

  • Bonding resin-based materials (especially dual-cure composite / compomer . materials) to tooth structure (dentin and enamel)
  • Treatment of hypersensitive teeth .

Total-Etch Bond is indicated for:

  • Bonding resin-based materials (especially light-cure composite / compomer . materials) to tooth structure (dentin and enamel)
  • Treatment of hypersensitive teeth .

Prescription Use × (Per 21 CFR 801 Subpart D) OR

Over-the -Counter Use (Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sunor Teare

f Arkansas Building, General Hospital,
Infection Control, Clean Devices

July 26

00013

§ 872.3200 Resin tooth bonding agent.

(a)
Identification. A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.(b)
Classification. Class II.