(33 days)
IMCO-STATTM is a software device that receives digital images and data from existing imaging equipment using DICOM 3.0 communication protocols. Images and data are stored on the IMCO-STATTM server in DICOM 3.0 Part 10 and JPEG format.
IMCO-STAT™ is designed to send reports. images, audio and video data to other workstations, Personal Digital Assistants (PDA) or Tablet PCs in wired or wireless environments. This is accomplished using an executable client application on a receiving entity with the appropriate hardware.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
IMCO-STAT™ is a software device that receives digital images and data from existing imaging equipment using DICOM 3.0 communication protocols. Images and data are stored on the IMCO-STAT™ server in DICOM 3 0 Part 10 and JPEG format.
IMCO-STAT™ is designed to send reports, images, audio and video data to other workstations, Personal Digital Assistants (PDA) or Tablet PCs in wired or wireless environments. This is accomplished using an executable client application on a receiving entity with the appropriate hardware.
The images may be embedded for reference in a DICOM image comprised report data, for distribution across a network and storage in a Picture Archive Communication System (PACS) with the original exam series data. Wavelet files can also be created and stored utilizing the same process. The algorithms used to create JPEG and wavelet images follow known and accepted protocols.
IMCO-STAT™ uses standard off-the-shelf hardware and commercially available IMOO-07/11-1650 and operating systems. The software communicates using computer plationne and operthe network used to support the TCP/IP stack is superfluous to IMCO-STAT™
The provided text describes the IMCO-STAT™ software, its intended use, and its substantial equivalence to other legally marketed devices, but it does not contain information about explicit acceptance criteria, performance metrics, or a study proving the device meets specific criteria.
Instead, the document focuses on:
- Substantial Equivalence: The primary method used to demonstrate safety and effectiveness is by showing that IMCO-STAT™ has the same intended uses and technical characteristics as two predicate devices: Medical Insight EasyViz system (K051809) and Marotech, Inc. Marosis PACS System (K012844).
- General Safety and Effectiveness Concerns: It mentions that the device labeling contains instructions for use and indications for use, and that device components are off-the-shelf.
- Validation and Effectiveness: It states that "Testing of the software and related hardware has been performed by programmers, non-programmers, quality control individuals and potential customers." However, it does not provide details on the specific tests conducted, their results, or any quantifiable performance metrics.
Therefore, many of the requested items cannot be directly answered from the provided text.
Here's a breakdown of what can and cannot be answered:
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not explicitly defined in the document. The primary "acceptance" mechanism is demonstrating substantial equivalence to predicate devices, implying that if it functions similarly and poses no new safety risks, it is acceptable. | Not explicitly defined in the document. The document states "Testing of the software and related hardware has been performed by programmers, non-programmers, quality control individuals and potential customers" but does not report specific performance metrics (e.g., accuracy, speed, uptime). |
2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
- Sample Size for Test Set: Not specified.
- Data Provenance: Not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Number of Experts: Not applicable, as detailed information about a "test set" and "ground truth" establishment is not provided. The document mentions "programmers, non-programmers, quality control individuals and potential customers" performed testing, but does not detail their role in establishing ground truth for a formal test set for performance evaluation.
- Qualifications of Experts: Not specified.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Adjudication method: Not applicable, as no formal test set with adjudication is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No, an MRMC comparative effectiveness study is not mentioned or described. The IMCO-STAT™ is presented as a software device for image management and viewing, not explicitly as an AI assistant intended to improve human reader performance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: Not explicitly stated or detailed. The document focuses on the software's functionality in handling and transmitting images and data.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Type of Ground Truth: Not applicable, as the document does not describe the establishment of ground truth for performance evaluation.
8. The sample size for the training set:
- Sample Size for Training Set: Not applicable. The IMCO-STAT™ is described as an image management and viewing software, not a machine learning or AI algorithm that requires a "training set" in the conventional sense.
9. How the ground truth for the training set was established:
- Ground Truth for Training Set: Not applicable, as there is no mention of a training set or ground truth establishment for such a purpose.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo for IMCO Technologies. The logo features the letters "IMCO" in a bold, sans-serif font, with the letters arranged horizontally. The word "TECHNOLOGIES" is written in a smaller font below the letters "IMCO". The letters are in a dark color, and the background is white.
DEC 1 2 2006
The Image Management Company
510(k) Summary
In accordance with the provisions of the Safe Medical Device Act of 1990. IMCO Technologies is providing a summary of safety and effectiveness information regarding the IMCO-STAT™ software
-
1 1 Company Identification
IMCO Technologies N27W23957 Paul Road Pewaukee WI 53072 Contact. Mark Schwartz, President and CEO Telephone, 262-523-4445 Fax: 262-523-1141 Email: mschwartz@jmco-tech.com -
1.2 Official Correspondent
Mark Schwartz. President and CEO IMCO Technologies N27W23927 Paul Rd Pewaukee WI 53072 Telephone: 262-523-4445 Fax: 262-523-1141 Email: mschwartz@imco-tech.com -
1.3 Date of Submission
November 7, 2006 -
1 4 Device Name
Classification Name Common/Usual Name Proprietary Name:
System. Image Processing, Radiological Soft-copy reading system IMCO-STAT TH
- 1.5 Substantial Equivalence
The IMCO-STAT™ system has the same intended uses and technical characteristics as the Medical Insight EasyViz system (K051809) and Marotech, Inc. Marosis PACS System (K012844).
IMCO Technologies 510(k) Summary Page | 01 |
{1}------------------------------------------------
| Product Name | IMCO Stat | MedicalInsightEasyViz | MarosisPACS |
|---|---|---|---|
| Graphical UI | Yes | Yes | Yes |
| Windows O.S. - Client | Yes | Yes | Yes |
| Uses Standard Monitor | Yes | Yes | Yes |
| Scales Image to Display | Yes | Yes | Yes |
| Image Input | DICOM 3.0 | DICOM 3.0 | DICOM 3.0 |
| Images stored on remote Window server | Yes | Yes | Yes |
| Network Protocol | TCP-IP | TCP-IP | TCP-IP |
| Compression | JPEG 2K | Proprietary | JPEG 2K |
| Wireless Capability | Yes | Yes | No |
| Support Tablet PC, PDA, etc | Yes | Yes | No |
| Annotation | Yes | Yes | Yes |
| Image Measurement | Yes | Yes | Yes |
| Cine tool | Yes | Yes | Yes |
| Comparison Mode | Yes | Yes | Yes |
| Review Report from RIS | No | Yes | Yes |
| Designed for Use Inside and Outside Radiology | Yes | Yes | Yes |
| Flip / Rotate of Images | Yes | Yes | Yes |
| User Log In | Yes | Yes | Yes |
| Multiple Layout Options | Yes | Yes | Yes |
| WWL control & Pre-sets | Yes | Yes | Yes |
| Patient & Study Browser | Yes | Yes | Yes |
| Print to Paper Capability | No | Yes | Yes |
1.6
Device Description and Intended Use
IMCO-STAT™ is a software device that receives digital images and data from existing imaging equipment using DICOM 3.0 communication protocols. Images and data are stored on the IMCO-STAT™ server in DICOM 3 0 Part 10 and JPEG format.
IMCO-STAT™ is designed to send reports, images, audio and video data to other workstations, Personal Digital Assistants (PDA) or Tablet PCs in wired or wireless environments. This is accomplished using an executable client application on a receiving entity with the appropriate hardware.
The images may be embedded for reference in a DICOM image comprised report data, for distribution across a network and storage in a Picture Archive Communication System (PACS) with the original exam series data. Wavelet files can also be created and stored utilizing the same process. The algorithms used to create JPEG and wavelet images follow known and accepted protocols.
{2}------------------------------------------------
IMCO-STAT™ uses standard off-the-shelf hardware and commercially available IMOO-07/11-1650 and operating systems. The software communicates using computer plationne and operthe network used to support the TCP/IP stack is superfluous to IMCO-STAT™
General Safety and Effectiveness Concerns. 1 . ?
The device labeling contains instructions for use and indications for use. The The "device" laboling" components specified are off-the-shelf computer components
Validation and Effectiveness:
Testing of the software and related hardware has been performed by programmers, non-programmers, quality control individuals and potential customers
Lossy compressed mammographic images and digitized film screen images must Eosby comproboa nor primary image interpretations. Mammographic images may not be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
Substantial Equivalence
As stated previously, IMCO-STAT™ is substantially equivalent to EasyViz software package and Marosis PACS.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Mr. Mark L. Schwartz President / CEO IMCO Technologies N27W23957 Paul Road #101 PEWAUKEE WI 53072
DEC 1 2 2006
Re: K063392
Trade/Device Name: IMCO-STATTM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 7, 2006 Received: November 9, 2006
Dear Mr. Schwartz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a circular logo with the text "1906-2006" at the top. The letters "FDA" are in large, bold font in the center of the logo. Below the letters, the word "Centennial" is written in a cursive font. Three stars are arranged in a horizontal line below the word "Centennial".
Protecting and Promoting Public Health
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours.
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2 -
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the text "IMCO" in a stylized font. The top half of the letters "MC" are black, while the rest of the letters are in a lighter shade. Below the logo, the word "TECHNOLOGIES" is written in a simple, sans-serif font.
The Image Management Company
INDICATIONS FOR USE
KOL 3392 510(k) Number (if known):
IMCO-STATTM Device Name:
IMCO Technologies Sponsor Name:
Indications for Use:
IMCO-STATTM is a software device that receives digital images and data from existing imaging equipment using DICOM 3.0 communication protocols. Images and data are stored on the IMCO-STATTM server in DICOM 3.0 Part 10 and JPEG format.
IMCO-STAT™ is designed to send reports. images, audio and video data to other workstations, Personal Digital Assistants (PDA) or Tablet PCs in wired or wireless environments. This is accomplished using an executable client application on a receiving entity with the appropriate hardware.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
Prescription Use (21 CFR 801 Subpart D) And/Or
Over-The-Counter Use | | (21 CFR 807 Subpart C)
Do Not Write Below This Line - Continue on Another Page if Needed
Concurrence of CDRH, Office of Device Evaluation
David ln Ingram
(Division Sign-Off) Division of Reproductive, A and Radiological Devices 510(k) Number
IMCO Technologies N27 W23957 Paul Road, Pewaukee, WI 53072 PH: 800/300-7734 262/523-4445 FX: 262/523-1141
N/A