K Number
K063375

Validate with FDA (Live)

Date Cleared
2007-01-11

(64 days)

Product Code
Regulation Number
870.5310
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

AED Therapy: To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless.

Manual Defibrillation: Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronous defibrillation is indicated for termination of atrial fibrillation.

Non-invasive External Pacing Therapy: The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.

Pulse Oximetry: The SpO2 option is intended for use when it is beneficial to assess a patient's oxygen saturation level.

Non-invasive Blood Pressure Monitoring: The NBP option is intended for non-invasive measurement of a patient's arterial blood pressure.

End-tidal CO2: The EtCO2 option is intended for non-invasive monitoring of a patient's exhaled carbon dioxide and to provide a respiration rate.

12-Lead ECG: The 12-Lead ECG function is to provide a conventional diagnostic 12-Lead ECG report, which may include measurements and interpretative statements.

Q-CPR: The Q-CPR™ option provides feedback designed to encourage rescuers to perform resuscitation in accordance with AHA/ERC guidelines for chest compression rate, depth, and duty cycle and ventilation rate, volume and flow rate (inflation time). The Q-CPR option is contraindicated for use on neonatal and pediatric patients (under 8 years of age or weighing less that 25 kg). The Q-CPR option is not for use when CPR is contraindicated.

Invasive Pressure: The Invasive Pressure option is indicated for measuring arterial, venous, intracranial and other physiological pressures on patients.

Temperature: The Temperature option is indicated for measuring temperature in patients.

Device Description

The modification is a change that provides an optional Bluetooth™ wireless technology option to transfer data from the MRx to a Bluetooth enabled device. This wireless connection is an alternative to the current RS232 cable connection capability.

AI/ML Overview

The provided text is a 510(k) summary for the Philips HeartStart MRx Monitor/Defibrillator Software Release 7.01. It describes a modification to an existing device, primarily the addition of an optional Bluetooth wireless technology-based data transfer.

This submission is for a modification to a previously cleared device. The summary generally focuses on asserting that the new features (Bluetooth connectivity) do not negatively impact the device's existing performance, rather than re-proving the core functionalities of the monitor/defibrillator. Therefore, many of the typical elements of an AI/algorithm-based device study (like sample sizes, ground truth, expert adjudication, etc.) are not applicable or not explicitly detailed in this type of submission for a physical device with a software update.

Here’s an analysis based on the provided text, addressing your questions where possible:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not specify numerical acceptance criteria or performance metrics for the device's core functions (ECG, NBP, SpO2, defibrillation, etc.) for this particular 510(k). Instead, it states:

AspectAcceptance Criteria (Implicit)Reported Device Performance
Overall Performance"Pass/Fail criteria were based on the specifications cleared for the predicate device.""Test results showed substantial equivalence. The results demonstrate that the device meets all performance claims."
New Feature (Bluetooth)The Bluetooth wireless technology option should function as intended for data transfer and not compromise existing functionalities, safety, or effectiveness.Tested as part of "system level tests, integration tests, environmental tests, and safety testing from hazard analysis." The conclusion is that the device "meets all performance claims."
Safety and EffectivenessShould be substantially equivalent to the predicate device and meet all relevant safety and effectiveness requirements."The new device is substantially equivalent to the previously cleared HeartStart MRx Monitor/Defibrillator software... and has the same Indications for Use as the legally marketed predicate device."

2. Sample Sizes Used for Test Set and Data Provenance

  • Test Set Sample Size: Not specified. The testing described ("system level tests, integration tests, environmental tests, and safety testing from hazard analysis") refers to device validation and verification testing, not typically patient-based clinical trials with a "test set" in the context of an algorithm.
  • Data Provenance: Not applicable in the context of an algorithm-based "test set" here. This is a hardware/software update to a medical device.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Number of Experts: Not applicable. Ground truth, in the sense of expert consensus for an algorithm, is not relevant for this type of device modification submission. The "ground truth" for the device's functionality would be its adherence to engineering specifications and regulatory standards.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. This refers to consensus among experts, which is not relevant for the type of testing described (system, integration, environmental, safety).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No. This type of study assesses how an AI system impacts human reader performance, which is not applicable to a non-AI device modification.
  • Effect size: Not applicable.

6. Standalone (Algorithm Only) Performance Study

  • Was a standalone study done? No. This device is a monitor/defibrillator; its performance is not "algorithm only" in the sense of a diagnostic AI tool. The software update is for connectivity.

7. Type of Ground Truth Used

  • Ground Truth Type: For the device's core functionalities, the "ground truth" would be established by reference standards, engineering specifications, and established medical diagnostics (e.g., a calibrated pressure sensor for NBP, a known electrical signal for ECG, etc.). For the Bluetooth functionality, the ground truth would be successful and reliable data transmission as per engineering specifications. The documentation does not specify the exact methods for establishing this but implies it was based on existing specifications for the predicate device.

8. Sample Size for the Training Set

  • Training Set Sample Size: Not applicable. This is not an AI/machine learning algorithm that requires a "training set" of data in the typical sense. The software development follows traditional engineering and validation processes.

9. How the Ground Truth for the Training Set was Established

  • Ground Truth for Training Set Establishment: Not applicable, as there is no "training set" in the AI sense.

In summary:

This 510(k) submission (K063375) describes a software update (Release 7.01) to an existing Philips HeartStart MRx Monitor/Defibrillator, adding Bluetooth connectivity. The regulatory submission primarily relies on demonstrating substantial equivalence to predicate devices and proving that the new feature does not adversely affect the device's existing safety or effectiveness. The "study" mentioned consists of verification and validation activities (system, integration, environmental, and safety tests) against existing specifications, rather than clinical studies with patient data or AI algorithm performance evaluations. As such, many of the questions related to AI/algorithm performance (e.g., sample sizes for training/test sets, expert adjudication, MRMC studies) are not pertinent to this specific regulatory filing.

{0}------------------------------------------------

K063375

8.0 510(k) Summary

JAN 1 1 2007

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. 8807.92.

    1. The submitter of this premarket notification is Philips Medical Systems
      This summary was prepared on 24 October 2006
    1. The name of this device is the HeartStart MRx Monitor/Defibrillator Software Release 7.01. Classification names are as follows:
ClassificationProCodeDescription
870.2340, II74 DPSElectrocardiograph device
870.1130, II74 DXNNon-invasive blood pressure
870.2700, II74 DQAPulse oximeter
870.2300, II74 MWIMonitor, Physiological, Patient
870.2300, II74 MSXSystem, Network and Communication, PhysiologicalMonitors
868.1400, II74 CCKEnd-tidal Carbon Dioxide
870.5550, II74 DROExternal Trascutaneous Pacemaker (Non-invasive)
870.5300, II74 LDDLow-energy defibrillator
870.1025, III74 MKJDefibrillators, Automatic, External
870.5200, III74 LIXCardiopulmonary Resuscitation Aid
870.1110, II74 DSKBlood Pressure Computer
880.2910, II80 FLLClinical Electronic Thermometer
    1. The new device is substantially equivalent to the previously cleared HeartStart MRx Monitor/Defibrillator software cleared under K031187 and K051134, as well as K061707).
    1. The modification is a change that provides an optional Bluetooth™ wireless technology option to transfer data from the MRx to a Bluetooth enabled device. This wireless connection is an alternative to the current RS232 cable connection capability.
    1. The new device has the same Indications for Use as the legally marketed predicate device.
    1. The new device has the same technological characteristics as the legally marketed predicate device.
    1. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new device with respect to the predicate. Testing involved system level tests, integration tests, environmental tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the device meets all performance claims.

{1}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 1 2007

Philips Medical Systems Ms. Denise Halev Sr. Quality and Regulatory Engineer 3000 Minuteman Road Andover, MA 01810

Re: K063375

Trade Name: HeartStart MRx Models M3535A/M3536A Regulation Number: 21 CFR 870.5310 Regulation Name: Automatic External Defibrillator Regulatory Class: Class III Product Code: MKJ Dated: December 12, 2006 Received: December 12, 2006

Dear Ms. Haley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Ms. Denise Haley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

. A. Hillelenne

for
E
Γ

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

2.0 Indications for Use

K063375 510(k) Number (if known):

Device Name: HeartStart MRx Monitor/Defibrillator

Indications for Use:

The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

AED Therapy

To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless.

Manual Defibrillation

Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronous defibrillation is indicated for termination of atrial fibrillation.

Non-invasive External Pacing Therapy

The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.

Pulse Oximetry

The SpO2 option is intended for use when it is beneficial to assess a patient's oxygen saturation level.

Non-invasive Blood Pressure Monitoring

The NBP option is intended for non-invasive measurement of a patient's arterial blood pressure.

End-tidal CO2

The EtCO2 option is intended for non-invasive monitoring of a patient's exhaled carbon dioxide and to provide a respiration rate.

12-Lead ECG

The 12-Lead ECG function is to provide a conventional diagnostic 12-Lead ECG report, which may include measurements and interpretative statements.

[continued....]

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular DevicesPage 1 of 2
510(k) NumberK063375

PHILIPS MEDICAL SYSTEMS

{4}------------------------------------------------

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

HeartStart MRx Monitor/Defibrillator Device Name:

Indications for Use: [continued ... ]

Q-CPR

The Q-CPR™ option provides feedback designed to encourage rescuers to perform resuscitation in accordance with AHA/ERC guidelines for chest compression rate, depth, and duty cycle and ventilation rate, volume and flow rate (inflation time).

The Q-CPR option is contraindicated as follows:

  • The Q-CPR option is contraindicated for use on neonatal and pediatric patients (under 8 years of age or weighing less that 25 kg).
  • The Q-CPR option is not for use when CPR is contraindicated. -

Invasive Pressure

The Invasive Pressure option is indicated for measuring arterial, venous, intracranial and other physiological pressures on patients.

Temperature

The Temperature option is indicated for measuring temperature in patients.

Prescription Use Yes Over-The-Counter Use No AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 2 of 2

MeleMay
(Division Sign-Off)

510(k) Number K063375

§ 870.5310 Automated external defibrillator system.

(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.